Professional Documents
Culture Documents
1. Introduction
This manual describes detailed operation of a dedicated GC configuration based on the
TRACE Ultra gas chromatograph and manufactured by Thermo Electron, equipped with FID
(Flame Ionization Detector) and methanizer able to perform the analysis at low level of
concentration of CO and CO2 in high purity hydrogen and other gas-phase samples.
The lower limit of detection for either CO and CO2 in hydrogen is 0.1 ppm.
Higher concentrations of methane (above 5%) will interfere with the determination of carbon
dioxide.
A methanizer placed before the FID is used to convert CO and CO2 in methane.
Sample is injected by Valve #1 which is also responsible for placing the packed column in
backflush mode. The system is equipped with hydrogen sensor to allow safety use of
Hydrogen in the equipment.
NV: is a needle valve which can be used to adjust the carrier gas flow rate. Regulation of
these valves can be accomplished from the front part of the valve oven.
- On the TRACE GC keypad press CONFIG button and scroll to “Right carrier”
- Press MODE/TYPE to display a submenu of carrier gases
- Scroll to the gas to be used and press ENTER to confirm the selection. An asterisk
appears on the left of the gas selected
NOTE: Never close completely the needle valve otherwise the valve needle could be
irreversible damage.
The Valve Oven temperature can be programmed between 40°C up to 175°C in the aux page
of the TRACE GC. In order to heat the valve oven use the following steps:
- On the TRACE GC keypad press AUX button
- Scroll to Temperature Zones and press ENTER button
- Enter the target temperature and press ENTER button.
The Valve switching is controlled through the Valve interface page on the Trace GC keypad
and may be manually actuated by pressing ON/OFF buttons on the keypad or time
programmed through the RUN TIME menu on the Trace GC keyboard.
Manual switching of the valve is useful for set-up the analyzer step by step, or for
troubleshooting.
In order to switch the valve manually use the following steps (for instance for valve #1):
- Press VALVES button on the TRACE GC keypad
- Select gas sample valve #1
- Press ON to place the valve in the inject position or OFF to place the valve in the load
position.
- The “Inject at” parameter is the time into the run that the sample is to be injected. The
“Inject for” parameter is the time that the valve will be returned to the Load position.
If the Inject for time exceeds the GC run time, the GC will automatically reset the
valve to the Load position at the end of the run.
The sample is injected directly onto the column where the separation of the CO, CO2 and
methane is made from the remaining components. After conversion of CO and CO2 in
methane all these components are detected by the FID.
Once these components have been eluted, the Haysep column is placed in reverse flow and
the remaining components are backflushed to the vent.
2. Cap the right base body using the blind jet p/n 40401901 provided in the standard
outfit
3. Set the Right DPFC pressure at 200 kPa and maintain the system in this condition
for two minutes.
4. Set the Right carrier OFF and observe for any pressure decrease in the display, in
the next two minutes. If the pressure does not decrease this means that the system is
leak free.
5. In case of leak try to identify the leak point repeating the above operations and
checking all connections starting from the valve. A leak tester will also help in this
operation.
6.Cap the vent outlet number 2. Then set the Left DPFC pressure at value of 100 kPa
and maintain the system in this condition for one minute
7. Set the Left carrier OFF and observe for any pressure decrease in the display, in the
next two minutes. If the pressure does not decrease this means that the system is leak
free.
If the system is leak free place the Valves #1 in the load positions and repeat steps 3
to 7.
Carbon monoxide 1 ppm, Carbon dioxide 1 ppm and Methane 1 ppm in Helium
NOTE: The methanizer set-up, operation and parameters are described in the
relevant manual included with the TRACE GC Ultra set of documentation.
Pls verify accurately the activation of the methanizer catalyst before starting
injecting the sample
NOTE: The valve #1 switching time has been selected to permits complete
elution to the methanizer/FID of CO, CH4 and CO2 but to avoid heavier
hydrocarbons, if present, to be eluted to the methanizer.
With this setting all components eluted after CO2 are back flushed to the vent.
4. Troubleshooting
The following solutions can fix the most common problems generated during the
operation of the analyzer.