You are on page 1of 1

Hydrogen Sulphide H S 2 No.

4HH

n=1
4HH
0.1
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
%
detecting layer (pale blue)

Performance
Measuring range 0.1 to 2 % 2 to 4 %
Number of pump strokes 1(100 ml) 1/2(50 ml)
Correction factor 1 2
Sampling time 45 sec 30 sec
Detecting limit : 0.01 %(1 pump stroke)
Colour change : Pale blue → Blackish brown
Corrections for temperature & humidity : Unnecessary
Relative standard deviation : 10 %(for 0.1 to 0.6 %), 5 %(for 0.6 to 2 %)
Shelf life : 3 years

Reaction principle
H2S + CuSO4 → CuS + H2SO4

Possible coexisting substances and their interferences


Substance Concentration Interference Changes colour by itself to
Ammonia Two layers Dark blue
(Blackish brown
& dark blue)
Mercaptans ≧1% +
No
Sulphur dioxide ≧ 0.5 % −