Professional Documents
Culture Documents
Deflecting Torque
The deflecting or operating torque (Td) is produced by utilizing effects of electric current such as
magnetic, electrostatic, electrodynamic, thermal or inductive etc. The actual method of torque
production depends on the type of instrument. This deflecting torque causes the moving system
(and hence the pointer attached to it) to move from its ‘zero’ position i.e. its position when the
instrument is disconnected from the supply.
Controlling Torque
The deflection of the moving system would be indefinite if there were no controlling or restoring
ot
torque. This torque oppose the deflecting torque and increases with the deflection of the moving
system. The pointer is brought to rest at a position where the two opposing torques are equal.
The deflecting torque ensures that currents of different magnitudes shall produce deflections of
the moving system in proportion to their size. Without such at torque, the pointer would swing
ajy
over to the maximum deflected position irrespective of the magnitude of the current to be
measured. Moreover, in the absence of a restoring torque, the pointer once deflected, would not
return to its zero position on removing the current. The controlling or restoring or balancing
torque in indicating instruments is obtained either by a spring or by gravity.
Pr
Damping Torque:
A damping force is one which acts on the moving system of the instrument only when it is
moving and always opposes its motion. Such stabilizing or damping force is necessary to bring
the pointer to rest quickly, otherwise due to inertia of the moving system, the pointer will
oscillate about its final deflected position for quite some time before coming to rest in the steady
position. The degree of damping should be adjusted to a value which is sufficient to enable the
pointer to rise quickly to its deflected position without overshooting. In that case, the instrument
is said to be dead-beat. Any increase of damping above this limit i.e. overdamping will make the
instruments slow and lethargic. The damping force can be produced by (i) air frictions (ii) eddy
currents and (iii) fluid friction.
shown in
Fig.(a).
It is seen from Fig. (b) that the controlling or restoring torque is proportional to the sine of the
angle of deflection i.e.
Tc ∝ sin θ
GGSP
Eddy Current Damping:
ot
ajy
Permanent Magnet Type Instruments
The operation of a permanent-magnet moving-coil type instrument is be based upon the
principle that when a current-carrying conductor is placed in a magnetic field, it is acted upon by
a force
which tends to move it to one side and out of the field.
Pr
Construction
As its name indicates, the instrument consists of a permanent magnet and a rectangular coil of
many turns wound on a light aluminium or copper former inside which is an iron core as shown
in fig below.
GGSP
ot
ajy
The powerfull U-shaped permanent magnet is made of Alnico and has soft-iron end-pole pieces
which are bored out cylindrically. Between the magnetic poles is fixed a soft iron cylinder whose
Pr
function is
(i) to make the field radial and uniform.
(ii) to decrease the reluctance of the air path between the poles and hence increase the magnetic
flux. Surrounding the core is a rectangular coil of many turns wound on a light aluminium frame
which is supported by delicate bearings and to which is attached a light pointer. The aluminium
frame not only provides support for the coil but also provides damping by eddy currents induced
in it.
The sides of the coil are free to move in the two airgaps between the poles and core.
Control of the coil movement is affected by two phosphor-bronze hair springs, one above and
one below, which additionally serve the purpose of lending the current in and out of the coil. The
two springs are spiralled in opposite directions in order to neutralize the effects of temperature
changes
Deflecting Torque
When current is passed through the coil, force acts upon its both sides which produce a
deflecting torque as shown in Fig. below:
GGSP
b = breadth of coil in metre
N = number of turns in the coil
If 'i' ampere is the current passing through the coil, then the magnitude of the force experienced
by each of its sides is = Bil newton
For N turns, the force on each side of the coil is = NBil newton
∴ deflecting torque Td = force × perpendicular distance
= NBil × b = NBi(I × b) = NBiA N-m
where A is the face area of the coil.
It is seen that if B is constant, then Td is proportional to the current passing through the coil i.e.
Td ∝ Current (I)
ot
6. Error due to stray magnetic field is negligible.
7. Eddy current damping is used without any extra damping mechanism.
8. Strong operating magnetic field due to permanent magnet and iron core arrangement.
ajy
Disadvantages of Permanent Magnet Moving Coil Instruments:
1. With time the strength of permanent magnets reduces, causing error in reading.
2. Thermoelectric e.m.f. may cause errors when it is used with shunts.
3. Costly.
4. Can be used on D.C. only and not on A.C. This is because the direction of magnetic field
Pr
due to permanent magnet is fixed. And if the direction of current in the coil is A.C. then
the force produced will be in positive half cycle in one direction and due to negative half cycle
in the other direction. Hence the pointer will oscillate and will not deflect.
5. Delicate construction, so to be used with due care.
GGSP
Moving Iron Instrument
There are two basic forms of these instruments i.e. the attraction type and the repulsion type.
The operation of the attraction type depends on the attraction of a single piece of soft iron into a
magnetic field and that of repulsion type depends on the repulsion of two adjacent pieces of
iron magnetised by the same magnetic field. For both types of these instruments, the necessary
magnetic field is produced by the ampere-turns of a current-carrying coil. In case the instrument
is to be used as an ammeter, the coil has comparatively fewer turns of thick wire so that the
ammeter has low resistance because it is connected in series with the circuit. In case it is to be
used as a voltmeter, the coil has high impedance so as to draw as small a current as possible
since it is connected in parallel with the circuit. As the current through the coil is small, it has
large number of turns in order to produce sufficient ampere-turns.
ot
into the core of the coil when the current flows. Whatever be the direction of current in the coil,
the magnetisation of the moving iron is always such that it will get attracted. Thus the
instruments are unpolarized instruments i.e. they are independent of the direction in which the
current passes, hence it can be used on AC or DC. The principle of moving iron attraction type
ajy
instruments is represented in Fig below:
Pr
In the instrument shown, the gravity control is used which was previously used, but nowa-days
spring control is used. The damping method used is air friction damping. In this type, the
moving iron, which is made of high permeability steel (µ metal) is essentially pivoted and
consists of thin disc. The shape of the disc is such that a suitably divided scale is obtained. The
stray field (field due to neighbouring currents) usually influence the readings, hence magnetic
shielding is required.
GGSP
piece is fixed on the wall of bobbin which is called as fixed iron and the other is fixed to the
spindle which is free to move and called as moving iron.
When a current is passed through the coil, a magnetic field is produced within the hollow
space. The two iron pieces are within the field. They are magnetised in the same direction of
polarity, irrespective of direction of current in the coil. Hence, the two iron pieces will repel each
other. Hence, a deflecting torque is produced and moving iron is deflected.
The deflection is indicated by the pointer attached to the spindle. Generally, a spring is used
to produce controlling torque. Air friction damping is adopted to produce damping torque in
such type of instruments.
ot
The deflecting torque for MI instruments is given by:
Td ∝ Current (I)^2
ajy
The controlling torque for MI instruments is given by:
Tc ∝ sin θ (For Gravity Control)
Tc ∝ θ (For Spring Control)
Pr
Thus
sin θ ∝ I^2 (For Gravity Control)
θ ∝ I^2 (For Spring Control)
Disadvantages:
1. The readings shown by the meter get affected due to:
(i) Increase in resistance of coil.
(ii) Decrease in permeability and
(iii) Decrease in stiffness of spring, which are the effects due to increase in temperature.
2. Error is caused due to hysteresis in the iron.
3. The readings are affected due to stray fields.
4. The scale is not uniform because the deflection θ ∝ I^2 (crowded at the beginning).
GGSP
5. The instrument readings are true for only one frequency for which it has been designed.
ot
ajy
Ans: PMMC Instrument is Not Suitable to Measure AC: PMMC type instrument uses two
permanent magnets in order to create stationary magnetic field. These types of instruments are
used only for measurement of the DC quantities. If we use this instrument on AC, whenever the
direction of current is reversed during negative half cycle, the direction of torque will also be
reversed. This results in average value of torque as zero. The pointer will not deflect due to high
frequency from its mean position showing zero reading. However it can measure the direct
current very accurately.
PMMC Vs MI
GGSP
ot
ajy
Pr
GGSP
Fig. below shows the connections of a dynamometer, ammeter
ot
Low Current Meter High Current Meter
ajy
Fig. below shows the connections of a dynamometer, voltmeter.
Pr
• The torque of the instrument is dependent upon the strengths of the magnetic fields of
both fixed and moving coils i.e. in an ammeter the torque is roughly proportional to the
current squared and in voltmeter to the voltage squared.
Dynamometer instruments can thus be used in alternating current circuits for which square
law is essential. Two hair springs are used for the control and as leads to the moving coil.
Damping is often by air friction or enclosed vane. Although in some cases eddy current damping
by an aluminium disc rotating in a permanent magnet field is used.
Advantages:
GGSP
1. It is free from hysteresis and eddy current errors, but is used only as voltmeter or ammeter for
precision measurements. It is generally used as a wattmeter.
2. It can be used for D.C. as well as A.C. measurements.
3. It has a fairly high degree of accuracy.
Disadvantages: •
As compared to permanent magnet instruments, it has the following disadvantages:
1. As there is no iron core in the fixed coils, the field intensity is very weak. Therefore, to obtain
necessary deflecting torque, large number of ampere turns must be used on moving coil. This
makes the moving element heavy and the power loss is high. The torque to weight ratio is small
and frictional error is high.
2. Deflecting torque being proportional to square of voltage or current, the scale is
not uniform.
3. It is comparatively costly.
4. Operates with weak magnetic field because the coils are air cored.
Basic Ammeter
Basic ammeter has thick copper winding of few turns and there is a shunt of low
resistance connected across this coil. The basic circuit of ammeter is as shown below.
Ra: Resistance of coil of ammeter
Rsh: Resistance shunted across the ammeter coil
Ammeter is connected in series with the circuit of which current is to be measured as shown in
Fig. below
GGSP
Pr
GGSP
ajy
ot
ot
ajy
Pr
2. Universal Shunts:
In such case, Rs is the shunt resistance having tappings and desired tapping can be selected
by moving the selector as shown in the following figure. Tappings 1, 2, 3, 4, 5 …… Selecting
GGSP
tapping desired range of meter is selected.
ot
ajy
Pr
GGSP
Pr
GGSP
ajy
ot
Loading Effect:
ot
In many measurement techniques the sensor or detector utilizes the energy of the medium, thus
altering slightly the parameters to be measured.
Voltmeter has loading effect on the circuit. A well calibrated voltmeter may give a
misleading voltage reading when connected across a high resistance circuit. But gives a
ajy
dependable reading when connected across a low resistance circuit.
Corrective Measures:
Loading effect can be avoided by using voltmeter of high value resistance when measuring
a low resistance by ammeter voltmeter method.
Pr
Loading effect should be considered while planning any measurement, and corrections
should be made.
Voltmeter Sensitivity
GGSP
Input is AC. A diode which serves as half-wave rectifier is connected to the supply through
a multiplier resistance (Rs) and DC is available to the PMMC voltmeter.
Rm - meter resistance and Rs - multiplier resistance.
When DC voltage V is applied to the meter the current through the meter is
Im = Vm / (Rm + Rs) which produces full-scale deflection.
Because of the inertia of moving parts, PMMC indicates a deflection corresponding to
average value of current which is dependent upon the average value of the applied voltage.
The sensitivity of a half-wave rectifier instrument is 0.45 times its sensitivity at d.c.
ot
ajy
Pr
The sensitivity of a full-wave rectifier type of instruments with a sinusoidal a.c. as an input is 90%
of that with d.c. voltage of the same magnitude. The sensitivity of a full-wave rectifier type of
instrument is twice that of a half-wave rectifier type.
Ohm-meter
This meter directly measures the resistance with low accuracy. There is a wide field
of applications for this instrument.
An ohmmeter is useful for determining the approximate resistance of circuit components
such as heater elements or machine field coils, measuring and sorting of resistors used in
electronic circuits.
Ohm-meter can be classified into 2 types:
1. Series Ohm-meter
2. Shunt Ohm-meter
1. Series Ohm-Meter
GGSP
ot
ajy
Pr
1. Shunt Ohm-Meter
GGSP
Clamp-on Type or Portable Type CTs
By the use of a construction with a suitably split and hinged core, upon which the secondary
winding is wound, it is possible to measure the current in a heavy-current conductor or bus-bar
ot
without breaking the current circuit. The split core of the CT along with the secondary winding is
simply clamped around the main conductor, which acts as the primary winding of the CT. When
used with range selectable shunts and a calibrated ammeter, clamp on type CTs can be very
ajy
conveniently used for direct and quick measurement of current. Fig below shows photograph of
such a clamp-on type CT.
Pr
GGSP
The CT consists of a split core. The secondary winding of which is connected to a rectifier
type moving coil instrument.
The primary becomes the conductor through which current flowing is to be measured.
The slitted core gets aligned (jointed) by the force of a spring tension.
ot
It consists of a selector switch so that secondary turns are changed which ultimately changes
the range of the current on the dial. For measuring the current, the core is opened by pressing
trigger shown and then clipped over the conductor carrying current.
ajy
The flow of current in the wire will produce an alternating magnetic flux. This flux thus flows
through the CT core and based on transformer action, the secondary current is obtained.
The dial is however calibrated to read the current in the main wire directly.
Pr
GGSP