You are on page 1of 7

THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI 7

FASCICLE VIII, 2006 (XII), ISSN 1221-4590


TRIBOLOGY

FAILURE OF A TOTAL KNEE PROSTHESIS TIBIAL TRAY


DUE TO THE FATIGUE WEAR
Lucian CAPITANU, Aron IAROVICI, Justin ONIŞORU
Institute of Solid Mechanics, Bucharest, Romanian Academy, Romania
luciancapitanu@yahoo.com

ABSTRACT
Some studies report the failure of a tibial insert of knee prostheses due to the
subsurface cracking and delamination wear [1]. The present study starts by
analysing retrieved total knee prosthesis revealing the loss of a functionality of the
prosthesis due to the fact that a larger part of the polyethylene tibial insert was
dislocated. Some qualitatively remarks on the depreciative phenomenon are done
based on optical and microscopic analysis of the affected areas of the polyethylene
component. The failure mechanism was postulated as a combination of the high level
of loading during normal activities and a non-conforming contact mechanism
between the metallic femoral condyles and the tibial insert. The question that arises
is: Could be this phenomenon evaluated quantitatively a priori, e.g. could be the
failure due to delamination wear predicted? In order to do that some finite element
simulations were performed to dynamically determinate the area of the contact and
the contact pressure for three different activities – normal walking, stair ascending
and stairs descending – that could be considered as defining the level of the activity
of the patient. The finite element model uses a modern contact mechanism, including
friction, between the metallic femoral condyles (considered rigid) and the tibial
polyethylene insert. The results reveal the disadvantages induced by the non-
conformity of the contact, it means the fact that the loads are transferred by relatively
small areas of higher stresses. A special summation technique was adopted in order
to consider the cumulative effect of all activities, based on the frequency of each
activity and on the magnitude of the determined damage function. The computed
surface of cumulative maximum damage traversing the layers of the polyethylene
insert could be identified as the cracking surface in the retrieved prosthesis. The
conclusion is that a partial answer to the question enunciated before could be
achieved by the method presented here, that is able to determinate the path of the
subsequent failure of the polyethylene inserts even that for an accurate and complete
answer some extensions of the simulations needed to be done.

KEYWORDS: Total knee prosthesis, wear, experimental methods.

1. INTRODUCTION Examining an early retrieved prosthesis, Ries et


al. [9] noticed gross wear pits and delamination of the
Recent clinical studies (Blunn et al. [1], tibial insert, accompanied by an increase of
Wasielevski et al. [2]) identify the massive wear of percentage crystallinity at a plane coincident with the
the polyethylene tibial insert due to the mechanics of plane of delamination and the presence of subsurface
non-conforming contact as the main cause to oxidation peak typically observed in UHMWPE
prosthetic failure. In [3] Knight et al. report that from inserts, gamma-irradiated in air and aged. Still
209 cases of primary TKR, 18 prostheses failed, all controversial, the idea that increasing the thickness of
tibial inserts showing wear pits and delamination. The tibial inserts will reduce wear was rather false.
average time to failure was around 80 months, close Indeed, it is the contact non-conformity which
to that reported by Heck et al. in [4] (i.e. 72 months), conducts to high contact pressure localized on small
but other studies (Engh et al. [5], Jones et al. [6], areas and subsequently to the fatigue wear and
Kilgus et al. [7], Mintz et al. [8]) show that gross delamination of polyethylene part [3, 4].
delamination could occur even earlier.
8 THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

a) b)

c) d)

Fig. 1. Pictures of retrieved TKP after 21 years operation: (a) CoCr alloy femoral-component; (b) UHMWPE
insert and 16 years operation; (c) CoCr alloy femoral-component; (d) UHMWPE insert [15].

insert in figure 3a is from an early retrieved tibial part


(the revision was due to the loosening of the implant)
and one could observe the wear pits of the contact
surface, that are specific for adhesive wear, and also
delamination of the peripheral area of the medial part
as effect of fatigue wear.
The insert in figure 3b was from a total knee
prosthesis retrieved at 67 months from implantation.
In this case, the fatigue wear has more catastrophic
effects – massive loss of polyethylene in the medial
part, and severe delamination in the lateral part.
From the qualitative evaluation of the two
retrieved implants we could conclude that the fatigue
Fig. 2. Optical image of a UHMWPE insert estimated wear debuted with some wear pits of the contact
by using a shape tracer. Lines indicate surface surface and delamination (especially in the peripheral
profiles. (+) marks indicate the maximum medial part of the tibial insert) but, in time, severe
deformation point at each area. [15]. damages occur (gross delaminations, massive losses
of polyethylene).
Analysing two retrieved TKP (after 21 and 16
years from implantation), Oonishi et al. [15] noticed 2. METHODOLOGY AND FE MODEL
the presence of parallel scratches (oriented antero-
posterior) on femoral component (fig. 1), and large The methodology for predicting the fatigue wear
delamination of polyethylene tibial insert (fig. 2). phenomena combines FE analyses of active loading
Our study, based on examination of two cycles of relevant routine activities with a summation
retrievals (see fig. 3 and fig. 4) identifies also some technique that is based on computation of a
characteristic marks of fatigue wear phenomena. The cumulative estimator of damage.
THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI 9
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

a) b)
Fig. 3. Two retrieved polyethylene inserts showing (a) pitting and (b) severe delamination [6].

Fig. 4. Pictures of fatigue wear of a total hip prostheses components.

A FE model of the artificial joint contact was Elastic linear constitutive laws are considered
used (fig. 5) for all analyses. The model includes one for metallic as for the plastic parts. The materials
femoral condyle and one half (the medial one) of the properties are listed in Table 1. The routine activities
polyethylene insert, and of the metallic tibial tray. For considered relevant for this study are the active cycles
deformable parts (the tibial parts) solid brick elements of the normal walking, stair descending, and stair
with 8 nodes and 3 DOF’s per node (all three ascending activities.
translations) are used. The kinematics of all three movements are
illustrated in figure 6 – in (a), (b), and (c) the
segmental movement of the lower member are
sketched (Bergmann et al. [10]), in (d) the
compressive force of contact are listed comparative
for all activities (Taylor and Walker, [11]), in (e) and
(f) two determinants of the cinematic conditions (the
flexion angle and the internal-external rotation of
tibia) are plotted (Bergmann et al. [10]).
The compressive force is applied to the tibial
tray (as distributed pressure on the lower basis of it).
The two rotations (flexion and internal-external
rotation) are dynamically constraining the femoral
Fig. 5. Finite element model used in analyses [16].
part. We considered that the actual level of flexion
(not higher than 70 degrees) does not induce an
The femoral condyle having a thoroidal shape
antero-posterior translation (usually appearing when
with a radius of 22 mm in the sagital plane (the
the ligaments are overstrained) as for extended
flexion plane), and a radius of 30 mm in the
flexions.
transversal plane, was considered rigid.
10 THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

300
normal walking
stairs descending
250 stairs ascending

Joint contact force [%BW]


200

150

100

50

0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Normalized time [s/s]
a) d)
70

60 normal walking
stairs descending
stairs ascending
50
Flexion angle [deg]

40

30

20

10

0
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Normalized time [s/s]
b) e)

25

normal walking
20
stairs descending
stairs ascending
Internal-external rotation angle [deg]

15

10

-5

-10
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Normalized time [s/s]
c) f)

Fig. 6. Dynamic conditions during all activities: kinematics of lower member [10] (a – normal walking, b – stairs
descending, c – stairs ascending) and loading [11] and kinematics [10] of the joint surfaces (d – contact force, e –
flexion angle, f – internal-external rotation angle).

Table 1. The mechanical properties of materials used in FE analyses


Material Young Modulus E [GPa] Poisson ratio ν Remarks
Co-Cr alloy 200 0.3 from ISO 5832-4
UHMWPE 1.06 0.36 from Lewis [12]
THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI 11
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

The contact mechanism includes rolling and ground) followed by even a higher value but
sliding of the two joint surfaces – the femoral distributed on a large domain of time (corresponding
condyliar thoroidal surface and the planar surface of to the weight bearing stance).
tibial insert; the friction obeys the Coulomb law with The contact pressure is very important as
a friction coefficient of 0.12 (Villa et al. [13]). estimator of loading magnitude, but even more
For every activity a damage estimator could be important for estimating the fatigue wear will be the
computed from the variation of the shear maximum maximum shear stress, as being the parameter used in
principal stress in every element (Sathasivam and the damage estimator stated by formula (1). From the
Walker [14]): contact mechanism (which involves both rolling and
n
1 frictional sliding) it results that the maximum shear
D kf = ∑2 τ
i =1
i +1 − τ i ⋅ ( τ i +1 + τ i ) (1) stress will be located under the contact surface at a
distance depending of the aspect ratio of the contact
where: elliptic spot.
D kf - damage function for activity k , Using the cumulative estimator defined in for-
mula (2) the areas were the damage is likely to occurs
τ i - shear maximum stress on elements for time ti .
could be identified from the cumulative distributions
The effect of all activities could be cumulated plotted in figure 9 (the L-shaped region of different
by evaluating a weighted sum: colour). One could see that the most affected areas are
Dtot
f = wk D kf∑ (2) the medial and posterior parts of the tibial insert.
One could also notice that the maximum dama-
where the weights are depending on the frequency of
the activity. ge will occur in the subsurface of the insert (where the
shear stresses from normal walking are maximal),
which is usual the starting plane of severe delamina-
3. RESULTS AND DISCUSSION
tions. The magnitude of the damage estimator
Performing the dynamic analyses of the contact (~477MPa2 at the surface and ~535 MPa2 in the
between the femoral metallic condyles and the poly- subsurface) is in concordance with the results of
ethylene tibial insert (in the above-described condi- Sathasivam and Walker [14] which obtained a
tions) one could determinate the characteristics of the damage score of ~230 MPa2 for a constant loading of
contact mechanism in the artificial knee. For example, 1,000 N (approx. 120% BW) which represent one half
the trajectories of the contact spot for all three of the maximum loading from normal walking or
activities are plotted on the medial part of the tibial from stair ascending or stair descending activities.
insert (fig. 7).
Examining the sketch one could see that all 4. CONCLUSIONS
activities do not involve extension of the knee – it
means that only the median and posterior parts of A method combining the Finite Element Method
insert are used. For normal walking the trajectory is a for evaluating the contact mechanism in an artificial
closed curve (like a hysteretic loop). knee joint with a summation technique based on a
The stair ascending and stair descending damage estimator was successfully used in order to
trajectories are quite similar (even opposite as predict the areas were the damage is likely to occur.
direction), both of them having a large internal- Based on the above-described study, one can
external rotation (for stair ascending loading at the formulate several conclusions.
beginning of the cycle, for stair descending at the end First, the fatigue wear was identified as the main
of it). One could see that the normal walking loading phenomenon responsible for the polyethylene tibial
cycle will affect the median area of the medial part of insert failure. The clinical studies show that even
the insert, the stair ascending and descending those tibial parts retrieved for other reasons (implant
activities having their maxima located in the posterior loosening, misalignment etc.) presented markers of
parts of the insert. fatigue wear (cracks in the subsurface, wear pits,
As expected, due to the geometry of the contact
delaminations, loss of large parts of polyethylene).
(the joint surfaces are a toroidal surface for femoral
The cumulative nature of the wear fatigue
condyle and a plane surface for tibial insert) it is clear
phenomena requires a qualitative and quantitative
that the contact spot will have an elliptical form with
evaluation of the transfer of loadings through the joint
the major axis oriented transversally to the saggital
plane of the femoral condyle. The maximum contact and a summation technique for encountering the
pressure is a good estimate of the intensity of the variety of the human activities. The first part was
loading. The variation in time of this contact para- assessed by intermediate of some dynamic finite
meter will follow the variation of loading (see fig. 8). element analyses of the joint contact mechanism. For
One could notice for all activities an initial step the second part, we assumed that normal walking,
maxima (corresponding to the initiation of the stair ascending and stair descending are the regular
movement – the contact between the foot and the activities dominant for the phenomena.
12 THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

Posterior
stair
ascending

stair
descending

Medial

Lateral
normal
walking

Anterior
Fig. 7. The contact paths for all three cases considered.

100

90

80
Contact pressure [MPa]

70

60

50

40 normal walking
stairs descending
stairs ascending
30

20
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Normalized time [s/s]

Fig. 8. The contact pressure variation for all three cases.

a) b)
Fig. 9. The damage score map (a) at the contact surface, (b) in the subsurface.
THE ANNALS OF UNIVERSITY “DUNĂREA DE JOS “ OF GALAŢI 13
FASCICLE VIII, 2006 (XII), ISSN 1221-4590
TRIBOLOGY

Thus, for all three activities the finite element polyethylene tibial inserts and their relationship to technical
considerations during total knee arthroplasty, Clinical
analyses are performed and a damage score was Orthopaedics, Vol. 299, pp. 31–43.
computed. Finally, a cumulative damage score (that 3. Knight J.L., Gorai P.A., Atwater R.D., Grothaus L., 1995,
accounts for all three activities) was determined and Tibial Polyethylene Failure After Primary Porous-coated Anatomic
the areas were the fatigue wear is likely to occur, are Total Knee Arthroplasty, J. of Arthroplasty, Vol. 10, pp. 748-757.
4. Heck D.A., Clingman J.K., Kettelkamp D.G., 1992, Gross
identified. polyethylene failure in total knee arthroplasty, Orthopedics, Vol.
From the distribution of the cumulative damage 15, pp. 23ff.
score (plotted in figure 9) one could see that the areas 5. Engh G.A., Dwyer K.A., Hanes C.K., 1992, Polyethlyene wear
that could be damaged are located medially in the of metal-backed tibial components in total and unicompartmental
knee prostheses, J Bone Joint Surg, Vol. 74B, pp. 9-17.
median and posterior regions of the tibial insert. Even 6. Jones S.M.G., Pinder I.M., Moran C.G., Malcolm A.J., 1992,
the stair ascending and descending are activities that Polyethylene wear in uncemented knee replacements, J Bone Joint
generate higher contact pressure, due to the increased Surg, Vol. 74B, pp. 18-22.
frequency, the normal walking is still dominant. 7. Kilgus D.J., Moreland J.R., Finerman G.A. et al, 1991,
Catastrophic wear of tibial polyethylene inserts, Clin Orthop, Vol.
A closer look to the distribution of damage 273, pp. 223-231.
score reveals that the maximum damage is likely to 8. Mintz L., Tsao A.K., McCrae C.R. et al, 1991, The
occur not at the contact surface but, rather, in the arthroscopic evaluation and characteristics of severe polyethylene
subsurface. The presence of maximum values in an wear in total knee arthroplasty, Clin Orthop, Vol. 273, pp. 215-222.
9. Ries M.D., Bellare A., Livingston B.J., Cohen R.E., Spector,
area were other studies (see Ries et al. [9]) identifies M., 1996, Early Delamination of a Hylamer-M Tibial Insert,
an increase of the percentage cristallinity and the Journal of Arthroplasty, Vol. 11, pp. 974-976.
presence of subsurface oxidation peak induced by the 10. Bergmann G., Deuretzbacher G., Heller M., Graichen F.,
gamma-irradiation and ageing of the prosthesis could Rohlmann A., Strauss J., Duda G.N., 2001, Hip Contact Forces
and Gait Patterns from Routine Activities, Journal of
be an explanation of the early delamination of the Biomechanics, Vol. 34, pp. 859-871.
polyethylene inserts. 11. Taylor S.J.G., Walker P.S., 2001, Forces and moments
We intend to extend the study in two directions: telemetered from two distal femoral replacements during various
firstly, the contact mechanism would be improved by activities, Journal of Biomechanics, Vol. 34, pp. 839-848.
12. Lewis G., 2001, Properties of cross-linked ultra-high-
considering the polyethylene viscoelastoplasticity and molecular-weight polyethylene, Biomaterials, Vol. 22, pp. 371-401.
a better friction mechanism (accounting for the 13. Villa T., Migliavacca F., Gastaldi D., Colombo M.,
dependency of the friction coefficient to the contact Pietrabissa R., 2004, Contact stresses and fatigue life in a knee
pressure and for the velocity of the relative movement prosthesis: comparison between in vitro measurements and compu-
tational simulation, Journal of Biomechanics, Vol. 37, pp. 45-53.
between the joint surfaces); secondary, the summation 14. Sathasivam, S., Walker, P.S., 1999, The conflicting
technique would be extended to a large area of requirements of laxity and conformity in total knee replacement,
activities and the influence of the frequency of the Journal of Biomechanics, Vol. 32, pp. 239-247.
loadings would be better estimated. 15. Oonishi H., Kim S.C., Kyomoto M., Iwamoto M., Ueno M.,
2006, PE wear in ceramic/PE bearing surface in total knee
arthroplasty; Clinical experiences of more than 24 years,
REFERENCES Bioceramics and Alternative Bearings in Joint Arthroplasty, 11-th
BIOLOX Symposium Proceedings, pp.101-110.
1. Blunn G.W., Joshi A.B., Minns R.J. et all, 1997, Wear in 16.Onişoru J., Capitanu L., Iarovici A., 2006, Failure of a tibial
retrieved condylar knee arthroplasties, Journal of Arthroplasty, insert of a total knee prosthesis due to fatigue wear, Buletinul
Vol. 12, pp. 281-290. Institutului Politehnic din Iasi, Tomul LII (LVI), Fasc. 6A, ACME
2. Wasielewski R.C., Galante J.O., Leighty R.M., Natarajan 2006/1, pp.333-340.
R.N., Rosenberg A.G., 1994, Wear patterns on retrieved

You might also like