Bod Incubator
Bod Incubator
Objective : The objective of the Calibration Procedure is to calibrate the BOD Incubator.
Scope : The scope of Calibration Procedure is applicable for all technical staff of the laboratory.
Responsibility : Chemist
Procedure :
1. Set the desired temperature on the BOD incubator panel.
With the help of calibrated thermometer or calibrated thermocouple take minimum 5 reading
2.
(one from each corner & one at center) on set temperature.
Standard deviation (Type – A Uncertainty ) from these readings will be calculated by the
3.
following formula :
SD =√[ (x-y)2/(n-1)]
Where, SD = Standard Deviation
X = Each individual value used to calculate mean
Y = Mean of n values
n = Total no. of values
Type – B uncertainty will be taken from the calibration certificate of the thermometer or
4.
thermocouple.