You are on page 1of 31

1

Cathode Ray Oscilloscope(CRO)

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Cathode Ray Oscilloscope(CRO)
2

 Very useful and versatile laboratory instrument (diagnostic


lab equipment).
 Used for display, measurements and analysis of waveforms
(signals) in electrical and electronic circuits at frequencies
ranging from dc to GHz.
 Basic instrument for the study of all types of signals.
 Very fast in displaying/plotting (X-Y plotters).
 Capable of generating a graph of an input signal versus time
(voltage-time mode) or a second variable (X-Y mode).
 Can be employed to measure quantities such as peak
voltage, frequency, period, phase difference, pulse width etc.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Parts of a CRO:
3

1) Cathode Ray Tube (CRT) and its associated controls.


An electron gunI.
II. Deflection plates
III. Phosphorescent screen
2) Input circuitry (a time-base generator, vertical and
horizontal amplifiers).

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Simple CRO construction:
4

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO construction:
5

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Block diagram of a basic CRO :
6

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Operation: Vertical Deflection Plates(VDP)
7

 Electron beam generated from electron gun will be


accelerated towards phosphorous screen.
 Beam passes through a set of parallel plates (vertical
deflection plates) oriented horizontally.
 Voltage to be displayed is amplified by a vertical amplifier
and produce an electric field which deflects the path of the
electron vertically.
 The polarity of the signal determines whether the
deflections will be up or down.
 Magnitude of the signal determines the vertical
displacement of the electron.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Operation: Horizontal Deflection Plates(HDP)
8

 After passing through vertical deflection plates, the beam


passes through a set of similar plates (horizontal
deflection plates) which are oriented vertically.
 The potential difference applied to these plates produce an
electric field which deflects left or right.
 Under normal configuration these HDP are connected to a
time-base circuitry.
 This circuit can control how fast the electron beam
sweeps from the left to right

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Operation:
9

 If two deflection voltages were held constant, the electron beam would
strike a point on the phosphorescent film and a stationary point would
visible on the screen.
 Most voltages of interested are time-varying and so the voltage applied
to the HDP is varied with time in such a way that the spot moves from
left to right on the screen as time passes.
 The phosphorescent material has the property of emitting light for
several milliseconds after the electrons have passed, the total effect is
for the electrons to leave behind a visible trail –a time-varying signal.
 The horizontal deflection voltage is also varied in such a way that
when the beam reaches the right-hand edge of the screen, it starts over
at the left-right side.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Displaying a signal on the screen:
10

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Displaying a signal on the screen:
11

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO in X-Y Mode:
12

 Beside showing a time-varying signal, the oscilloscope


lets you display an input signal rather than the time base
on the horizontal axis. This is called X-Y mode as both
the X and Y axis are tracing input voltage.
 The waveform resulting from XY arrangement of two
periodic signals of different periods is called Lissajous
Pattern.
 From the shape of Lissajous pattern we can determine
relative phases of the signals as well as frequency ratio.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO controls:
13

1. Y-Gain: Amplifies Y-deflection (small voltages are


amplified by built-in amplifiers before applying Y-plate)
Y-Gain=0.5V/Div
2. Time-Base: A saw-tooth wave is applied internally
across X-plates.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO control: 2.Time-Base
14

 Controls the speed at which the spot sweeps across the


screen horizontally from left to right.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO control: 2.Time-Base
15

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


CRO control: 2.Time-Base
16

 Helps to display the actual waveform applied across Y-


plates.
 Calibrated in s/cm or ms/cm or µs/cm.
 Gives the time required for the spot to sweep 1cm
horizontally across the screen.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Most Common Oscilloscope Controls:
17

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of frequency:
18

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of phase difference:
19

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Lissajous Figures:
20

 Can be displayed by applying two a.c signals


simultaneously to X-plates and Y-plates of a CRO.
 As the frequency, amplitude and phase difference are
altered, different patterns are seen on the screen of the
CRO.
 Obtained when operated in X-Y mode.
 If the ratio of the two frequencies of the voltages is just
equal to a rational number, standing figures appear on the
oscilloscope (Lissajous’ figures); if the frequency ratio,
however, deviates slightly from a rational number, these
figures are moving.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of phase difference using Lissajous Pattern:
21

Obtained by applying two sinusoidal signals to


be compared at the vertical and horizontal plates.
The phase difference then measured from
resulting Lissajous pattern.
The pattern may be either a straight line or a
circle or an ellipse depending on the amount of
phase difference.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of phase difference using Lissajous Pattern:

22

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of phase difference using Lissajous Pattern:

23

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of phase difference using Lissajous Pattern:

24

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of frequency using Lissajous Pattern:
25

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of frequency using Lissajous Pattern:
26

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Measurement of frequency using Lissajous Pattern:
27

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Storage type oscilloscope:
28

 Used for presentation of very slowly swept signals.


 Used in mechanical and biomedical fields.
 In conventional CRO the persistence of the screen is
smaller than the rate at which the signal sweeps across the
screen (the display will disappear within µs to seconds).
 The persistence of the screen can be adjusted to match
with the sweep time. So the display stays for few seconds
or even hours.
 A special CRT called storage tube is used which is same
as conventional CRT but with additional number of
special electrodes.

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Storage type oscilloscope:
29

1) Two special guns called flood guns


2) Collimation electrodes
3) Collector mesh
4) Storage mesh or storage target

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Example1:
30

 Horizontal frequency=50Hz
 Vertical frequency=?

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20


Example2:
31

 Phase difference=?

Mr.P.Krishna, EEE Department, IIITN 19-Feb-20

You might also like