Professional Documents
Culture Documents
Total Procedure NDT Procedure (Ut For Steel Structure) - 1
Total Procedure NDT Procedure (Ut For Steel Structure) - 1
NDT PROCEDURE
(UT FOR STEEL STRUCTURE)
22 66
23 67
24 68
25 69
26 70
27 71
28 72
29 73
30 74
31 75
32 76
33 77
34 78
35 79
36 80
37 81
38 82
39 83
40 84
41 85
42 86
43 87
44 88
CONTENTS
1.0 SCOPE
1.1 This procedure describes the requirements of ultrasonic examination straight or angle beam,
pulse echo, ultrasonic examination of welds in accordance with the requirements of AWS
D1.1 Structural Welding Code-Steel for TOTAL SOUTH PARS PROJECT. (Phases two and
tree).
1.2 This procedure will govern the ultrasonic examination of full penetration welds between the
thickness of 5/16in.(8.0mm) and 8in. (203mm) inclusive. These procedures and standards
1.3 For the requirements not specified in this procedure, codes and standards in para.2.0 or/and
2.0 REFERENCES
2.1 Unless otherwise specified by TSP's Specifications, the edition in para.2.2 through 2.8 shall
be used.
2.3 ASNT SNT-TC-1A : Recommended Practice for NDT Personnel Qualification and
RP 5561V 999 1882 001 Rev. 43 Structural Steel Design, Fabrication and Erection
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 6/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
3.0 RESPONSIBILITY
3.1 The welding engineer is responsible for corporate with sub-contractor for NDT work
3.3 The QA/QC Manager is responsible for control of sub-contractor and documents related to
the NDT.
4.1 Personnel who are performing ultrasonic examination to this procedure shall be qualified and
4.2 Personnel qualified Level II or III shall perform the test and interpret test results.
5.1.1 The search unit contact surfaces on each side of the weld shall be free of weld spatter,
roughness, or any irregularities that would interfere with the free movement of the search
unit or with the transmission of ultrasound into the test part. The fabricator shall provide
for the search unit contact surfaces to be sufficiently finished so as to prevent non
discontinuities.
5.1.2 Where the weld surface interferes with the examination, the weld shall be prepared as
The volume shall be examined by moving the search unit over the examination surface so as
to scan the entire examination volume. Each pass of the search unit shall overlap a minimum
The rate of search unit movement for examination shall not exceed 6in./sec (152mm/sec)
5.5 Application
Information furnished to the NDE contractor shall clearly identify the materials, thickness,
6.1 Instrument
(1)The ultrasonic instrument shall be the pulse echo type suitable for use with transducers
oscillating at frequencies between 1 and 5MHz. The display shall be an "A" scan rectified
video trace.
(2) The horizontal linearity of the test instrument shall be qualified over the full sound path
distance to be used in examination.
(3) Test instruments shall include internal stabilization so that after warm-up, no variation in
response greater than ±1dB occurs with supply voltage change of 15% nominal or, in the
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 8/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
case of a battery, through out the charge operating life. There shall be an alarm or meter
to signal a drop in battery voltage prior to instrument shutoff due to battery exhaustion.
(4) The test instrument shall have a calibrated gain control(attenuator) adjustable in discrete
1 or 2dB steps over a range of at least 60dB. The accuracy of the gain control setting
shall be within plus or minus 1dB.
(5) The dynamic range of the instrument's display shall be such that a difference of 1dB of
than 645 mm2. The transducer shall be round or square. Transducers shall be capable of
resolving the three reflections, when set the transducer in position F on the IIW Block shown in
Fig. 3.
25mm in width and from 9mm to 21mm in height. The maximum width to height ratio shall
be 1.2 to 1.0, and the minimum width-to-height ratio shall be 1.0 to 1.0, (except that
para. 7.3.
(6) The dimensions of the search unit shall be such that the distance form the leading edge
of the search unit to the index point shall not exceed 25mm.
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 9/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
6.4 Shoes
Transducers may be fitted with shoes which maintain adequate sound penetration in the test
materials. Shoes for curved surfaces shall have the same nominal radius of curvature as the
6.5 Couplant
(1) A couplant shall be used between the search unit and the test material, shall be either
glycerin or a cellulose gum and water mixture of a suitable consistency. Light machine
longer required.
blocks may be used, provided the reference level sensitivity for instrument search unit
combination is adjusted to be the equivalent of that achieved with the IIW Block.
(3) The combination of search unit and instrument shall resolve three holes in the RC
(4) Calibration block shall be of similar metallurgical structure and have the same or
The horizontal linearity of the examination instrument shall be requalified after each 40
hours of instrument use in each of the distance ranges that the instrument will be used in
(1) Couple a straight beam search unit meeting the requirements of para.6.2 to the IIW or DS
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 10/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
block in Position G, T or U (See Fig. 3) as necessary to attain five back reflections in the
(2) Adjust the first and fifth back reflections to their proper locations with use of the distance
(3) Each indication shall be adjusted to reference level with the gain or attenuation control for
(4) Each intermediate trace deflection location shall be correct within ±2% of the screen width.
The instrument's gain control or attenuator shall meet the requirements of para. 6.1.(4) and
shall be checked for correct calibration at two month intervals in accordance with followings.
(1) Procedure
In order to attain the required accuracy (±1%) in reading the indication height, the display
shall be placed on the display between 60% and 100% of screen height. This may be
accomplished with use of a graduated transparent part of the ultrasonic unit, care
should be taken that the overlay does not obscure normal examination displays.
a) Couple a straight beam search unit, meeting the requirements of para. 6.2 to the DS block
b) Adjust the distance calibration so that the first 2 in.(50mm) back reflection indication
c) Adjust the calibrated gain or attenuation control so that the indication is exactly at or
d) Move the search unit toward position U, see Fig. 3, until the indication is at exactly 40%
screen height.
e) Increase the sound amplitude 6dB with the calibrated gain or attenuation control. The
f) Record the dB reading under "a" and actual % screen height under "b" from step e) on
g) Move the search unit further toward position U, Fig. 3, until the indication is at exactly
i) Repeat Step f) ; except, information should be applied to the next consecutive line on the
certification sheet.
j) Repeat step g), h) and i) consecutively until the full range of the gain control or
k) Apply the information from columns "a" and "b" to equation of para. 7.2.(2) to calculate the
corrected dB.
m) Subtract column "c" value from Column "a" value and the difference in column "d", dB
error.
o) Evaluation form provides a relatively simple means of evaluating data from item(n).
p) Apply the dB information from column "e" vertically and dB reading from column "a"
q) The longest horizontal length, as represented by the dB reading difference, which can
be inscribed in a rectangle representing 2dB in height, denotes the dB range in which the
equipment meets the code requirement. The minimum allowable range is 60dB.
r) Equipment than does not meet this minimum requirement may be used, provided
correction factors are developed and used for flaw evaluation outside the instrument
acceptable linearity range, or the weld examination and flaw evaluation is kept within
dB1 = Column a
dB2 = Column c
%1 = Column b
Maximum internal reflections from each search unit shall be verified at a maximum time
(2) Remove the search unit from the calibration block without changing any other equipment
adjustments.
(3) Increase the calibrated gain or attenuation 20dB more sensitive than reference level.
(4) The display area beyond 1/2in.(13mm) sound path and above reference level height shall
checked after each eight hours of use to determine that the contact face is flat, that the
sound entry point is correct, and that the beam angle is within the permitted ±2deg.
tolerance. Search units which do not meet these requirements shall be corrected or replaced.
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 16/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
8.1 All calibrations and examinations shall be made with the reject control turned off. Use of the
reject control may alter the amplitude linearity of the instrument and invalidate examination
results.
8.2 Calibration for sensitivity and horizontal sweep (distance) shall be made by the ultrasonic
interval, or when the electrical circuitry is disturbed in any way which includes the following:
(1) Transducer change
(2) Battery change
(3) Electrical Outlet change
(4) Coaxial cable change
(5) Power outage (failure)
8.4 Straight Beam Examination of Base Metal
Calibration for straight beam examination of base metal shall be made with the search unit
Set the transducer in position G on the IIW block and adjust instrument to produce
indications at 1in., 2in., 3in., 4in., etc., on the display. The horizontal sweep shall be
adjusted for distance calibration to present the equivalent of the at least two plate
The sensitivity shall be adjusted at a location free of indications so that the first back
reflection from the far side of the plate (Set the transducer in position G on the IIW
Set the transducer in position D on the IIW block shown in Fig. 3, and move the
transducer until the signal from the radius maximized. The point on the transducer
which aligns with the radius line on the calibration block is the point of sound entry.
Set the transducer in position B for angles 40deg. through 60deg. or in position C for angle
60deg. through 70deg. on IIW block shown in Fig. 3, and move the transducers back and
forth over the line indicative of the transducer angle until the signal from the radius is
maximized. Compare the sound entry point on the transducer with the angle mark on the
The horizontal sweep shall be adjusted to represent to actual sound path distance by
using the IIW block or alternative blocks. The distance calibration shall be made using
either the 5in.(130mm) scale or 10in.(255mm) scale on the display. The transducer in
position D on the IIW Block shown in Fig. 3 (any angle), and adjust the instrument to
The zero reference level sensitivity used for flaw evaluation is attained by adjusting the
calibrated gain control or attenuator of the flaw detector. The transducer in position A on
the IIW block (any angle) shown in Fig. 3, and adjust the maximized signal from the
0.06in.(1.5mm) hole to attain a horizontal reference line height indication. This basic
sensitivity then becomes the zero reference level for discontinuity evaluation. The
maximum decibel reading obtained in above shall be used as the "Reference Level" "b"
(5) Resolution
The resolution shall be evaluated with the instrument controls set at normal examination
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 18/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
settings and with indications from the hole brought to mid-screen height. Set the
transducer on resolution block shown in Fig. 3, position Q for 70deg. angle, position R or
60deg. angle, or position S for 45deg. angle, transducer and instrument shall resolve the
three test holes, at least to the extent of distinguishing the peaks of the indications from
8.6 Any of the above checks found to be out of tolerance shall be cause for system repair
followed by full re-calibration of affected points.
Note1: The result of such calibration should be signed by contractor and certifying
authority and should be maintained for reference in the NDT department and available for
Company inspection as required.
All surfaces to which a search unit is applied shall be free of weld spatter, dirt, grease, oil,
paint, and loose scale and shall have a contour permitting intimate coupling.
This procedure is not intended to be employed for the procurement testing of base metals.
However, welding related discontinuities (cracking, lamellar tearing, delaminations, etc.) in the
adjacent base metal which would not be acceptable under provisions of AWS D1.1 shall be
9.3 An "X" line for flaw location shall be marked on the test face of the weldment in a direction
parallel to the weld axis. The location distance perpendicular to the weld axis is based on the
dimensional figures on the detail drawing and usually falls on the centerline of the butt joint
welds, and always falls on the near face on the connecting member of T and corner joint
9.4 A "Y" accompanied with a weld identification number shall be clearly marked on the base
metal adjacent to the weld that is ultrasonically tested. This marking is used for the
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 19/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
following purposes:
(3) Distance measurements and direction (+ or -) from the " X" line
(1) The entire base metal through which ultrasound must travel to test the weld shall be
tested from laminar reflectors using a straight beam search unit. If any area of base
metal exhibits total loss of back reflection or an indication equal to or greater than the
original back reflection height is located in a position that will interfere with the normal
weld scanning procedure (See Fig. 4) , its size, location, and depth from the A face shall
be determined and reported on the ultrasonic examination report, and an alternate weld
Appendix 1 due to laminar content recorded in accordance with para. 9.5.(1), the exami-
nation shall be conducted using one or more of the following alternative procedures as
(1) Welds shall be examined using an angle beam search unit with the instrument calibrated
using the angle as shown in Appendix 1. Following calibration and during examination, the
only instrument adjustment permitted is the sensitivity level adjustment with the calibrated
Sensitivity shall be increased from the reference level "b" for used scanning in accordance
Notes :
1. Testing patterns are all symmetrical around the weld axis with the exception of
2. Testing from both sides of the weld axis is to be made wherever mechanically
possible.
The examination angle and scanning procedure shall be in accordance with those
shown in Appendix 1. All butt joint welds shall be examined from each side of the weld
axis. Corner & T-joint welds shall be primarily examined from one side of the weld axis
only. All welds shall be examined using the applicable scanning pattern or patterns
intended that, as a minimum, all welds be examined by passing sound through the entire
volume of the weld and the heat affected zone in two crossing directions, wherever
practical.
When a discontinuity indication appears on the screen, the maximum attainable indication
trace deflection on the display. This adjustment shall be made with the calibrated gain
control. or attenuator and the instrument reading in decibels shall be recorded on the
ultrasonic examination report under the reading "Indication Level, a", for calculating the
The attenuation Factor, "c", on the examination report is attained by subtracting 1in.
(25mm) from the sound path distance and multiplying the remainder by 2. This factor shall
be rounded out to the nearest dB value. Factional values less than 1/2dB shall be reduced
to the lower dB level and those of 1/2 or greater increased to the higher level.
The "indication rating, d," in the UT report, represents the algebraic difference in decibels
between the indication level and the reference level with correction for attenuation as
a-b-c = d
b-a-c = d
10.1 Each weld discontinuity shall be accepted or rejected on the basic of its indication rating and
its length, in accordance with para. 11.0 for statically loaded structures.
Only those discontinuities which are rejectable need be recorded on the examination report.
10.2 Each rejectable discontinuity shall be indicated on the weld by a mark directly over the
discontinuity for its entire length. The depth from the surface and indication rating shall be
10.3 Welds found unacceptable by ultrasonic examination shall be repaired and shall be reexamined
ultrasonically.
smaller than the transducer size. When the discontinuity is larger than the transducer, a full
loss of back reflection will occur and a 6dB loss of amplitude and measurement to the
centerline of the transducer is usually reliable for determining flaw edges. However, the
approximate size evaluation of those reflectors, which are smaller than the transducer, must
be made by beginning outside of the discontinuity with equipment and moving the transducer
toward the area of discontinuity until an indication on the display begins to form. The leading
edge of the search unit at this point is indicative of the edge of the discontinuity.
relatively constant sound amplitude response, are longer than the search unit element
width, and are separated longitudinally from other indications by a distance of at least one
search unit element width. The length of such a discontinuity as entered under "Discontinuity
Length" on the examination report shall be determined by measuring the distance between
the center line of transducer locations where a 6dB reduction of sound amplitude below the
10.5.2 The rejectable length of discontinuties that have variable dB ratings more serious than
Class D Level, and separation less than the search unit width, shall be determined as follows :
(1) Find the location where the most serious indication occurs and note the indication level.
(2) Classify this level as being a Class A, B, C or D flaw.
(3) Adjust the calibrated gain control or attenuator to the most serious Class D Level.
(4) Move the search unit from the point of most serious indication level (both directions)
level(s).
(7) If this distance exceeds the allowable length of a Class C flaw, the indication is rejectable at
this length.
(8) If this length is equal to or less than the allowable Class C flaw length, further evaluation is
necessary as follows:
a) Adjust the calibrated gain control or attenuator to the most serious Class C Level.
b) Move the search unit toward the point of most serious indication level (Step (3)) until
level.
e) If this distance exceeds the allowable length of a Class B flaw, the indication is
f) If this length is equal to or less than the allowable Class B flaw length, further
evaluation is necessary.
i)Adjust the calibrated gain control or attenuator to the most serious Class B level.
ⅱ) Move the search unit further toward to most serious indication found in Step(1)
until the indication height reaches a reference level trace deflection height.
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 24/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
ⅲ) Mark the locations of the leading edge of the search unit on the test piece.
ⅳ) The distance between these points is the rejectable length of indication as a Class
A Flaw.
Welds that are subject to ultrasonic examination, in addition to visual inspection, shall be
12.0 REPORT
12.1The results of ultrasonic examination shall be documented on the formal report form
(Appendix 4). The report made out and signed by the certified Level I personnel and approved
13.0 APPENDICES
Appendix 1
Testing Angle and Testing Procedure
Procedure Chart
Material Thickness, in.(mm)
5/16(8.0) >1½ >1¾ >2½ >3½ >4½ >5 >6½ >7
Weld Type to to to to to to to to to
1½(38.1 1¾(44.5 2½(63.5) 3½(88.9 4½(114.3 5(127.0) 6½(165.1 7(177.8) 8(203)
) ) ) ) )
* * * * * * * * *
1G 1G 6 8 9 12
Butt 1 0 1 F or F or F or F or F or F or F 12 F
4 5 7 10 11 13
F F F F F F F
T 1 0 1 or 4 or 5 or 7 or 10 or 11 or 13 or - -
XF XF XF XF XF XF XF
F 1G F 1G F 6 F 8 F 9 F 13 F
Corner 1 0 1 or or or or or or or or or or or or or - -
XF 4 XF 5 XF 7 XF 10 XF 11 XF 14 XF
Electrogas 1G 1G P1 6 11 11 11 11
& 1 0 1 0 or 1** or or or P3 or P3 or P3 or P3 or P3
Electroslag 4 3 P3 7 15 15 15 15**
Notes :
1. Where possible, all examinations shall be made from Face A and in Leg 1, unless otherwise
2. Root areas of single groove weld joints which have backing not requiring removal by contract,
shall be tested in Leg 1, where possible, with Face A being that opposite the backing. (Grinding
of the weld face or testing from additional weld faces may be necessary to permit complete
3. Examination in Leg Ⅱ or Ⅲ shall be made only to satisfy provisions of this table or when
necessary to test weld areas made inaccessible by an unground weld surface, or interference
4. A maximum of Leg Ⅲ shall be used only where thickness or geometry prevents scanning of
5. On tension welds in cyclically loaded structures, the top quarter of thickness shall be tested
with the final leg of sound progressing from Face B to-ward Face A, the bottom quarter of
thickness shall be tested with the final leg of sound progressing from Face A toward Face B ;
i.e., the top quarter of thickness shall be tested either from Face A in Leg Ⅱ or from Face B in
Leg Ⅰ at the contractor's option, unless otherwise specified in the contract documents.
6. The weld face indicated shall be ground flush before using procedure 1G, 6, 8, 9, 12, 14 or 15.
Legend :
O - Not required.
A Face - the face of the material from which the initial scanning is done (on T- and
C Face - the face opposite the weld on the connecting member or a T- or corner joint.
the weld metal-base metal interface while searching at scanning level with primary
P - Pitch and catch shall be conducted for further discontinuity evaluation in only the
middle half of the material thickness with only 45deg. or 70deg. transducers of
equal specification, both facing the weld. (Transducers must be held in fixture to
control positioning-see sketch.) Amplitude calibration for pitch and catch is normally
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 27/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
made by calibrating a single search unit. When switching to dual search units for
pitch and catch inspection, there should be assurance that this calibration does not
F - Weld metal-base metal interface indications shall be further evaluated with either
Procedure legend
Area of weld thickness
No. Top Middle Bottom
quarter half quarter
1 70° 70° 70°
Appendix 2
Notes :
1. Class B and C discontinuities shall be separated by at least 2L, L being the length of the
longe discontinuity, except that when two or more such discontinuities are not separated
by at least 2L, but the combined length of discontinuities and their separation distance is
equal to or less than the maximum allowable length under the provisions of Class B or C,
2. Class B and C discontinuities shall not begin at a distance less than 2L from weld ends
3. Discontinuities detected at "scanning level" in the root face area of complete joint
penetration double groove weld joints shall be evaluated using an indicating rating 4dB
more sensitive than described in para. 8.6.(6) when such welds are designated as "tension
2in.(51mm) in length shall be suspected as being piping porosity and shall be further
5. For indications that remain on the display as the search unit is moved AWS D1.1 6.13.1
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 29/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
*Weld thickness shall be defined as the nominal thickness of the thinner of the two parts
being joined.
Any indication in this category having a length greater than 3/4in.(19mm) shall be rejected.
Any indication in this category having a length greater than 2in.(51mm) shall be rejected.
Any indication in this category shall be accepted regardless or length or location in the
weld.
Scanning Levels
Above Zero
Sound path in in.(mm)
Reference, dB
through 2 1/2(64mm) 14
>2 1/2 through 5(64∼127mm) 19
>5 through 10(127∼254mm) 29
>10 through 15(254∼381mm) 39
Notes :
1. Class B and C discontinuities shall be separated by at least 2L, L being the length of the
longer discontinuity, except that when two or more such discontinuities are not separated at
least 2L, but the combined length of discontinuities and their separation distance is equal to
or less than the maximum allowable length under the provisions of Class B or C, the
2. Class B and C discontinuities shall not begin at a distance less than 2L from weld
3. Discontinuities detected at "scanning level" in the root face area of complete joint
penetration double groove weld joints shall be evaluated using an indicating rating 4dB
more sensitive than described in para. 8.6.(6) when such welds are designated as
4. For indications that remain on the display as the search unit is moved AWS D1.1 6.13.1
* Weld thickness shall be defined as the nominal thickness of the thinner of the two parts
being joined.
Any indication in this category having a length greater than 3/4in.(19mm) shall be rejected.
Any indication in this category having a length greater than 2in.(51mm) shall be rejected.
Any indication in this category shall be accepted regardless or length or location in the
weld.
Scanning Levels
Above Zero
Sound path in in.(mm)
Reference, dB
through 2 1/2(64mm) 20
>2 1/2 through 5(64∼127mm) 25
>5 through 10(127∼254mm) 35
>10 through 15(254∼381mm) 45
Appendix 3
SCANNING TECHNIQUES
(UT for Steel Structures)
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 32/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
Technique 1
3. Probes
a) 0, 4MHz
b) 45, 4MHz
4. Range
a) 0 mm to 50 mm
b) 0 mm to 100 mm scanning with full skip distance
5. Scanning Sensitivity
a) Echo of first back reflection adjusted to 75% of full screen height
b) Refer to AWS D1.1, paragraph 6.25.5.2 “Zero Reference Level”
6. Sizing Method
Length : 6dB drop, Width and Height : 20dB drop
7. Scanning Patterns
SCANNING TECHNIQUES
(UT for Steel Structures)
Technique 2
3. Probes
A. 0, 4MHz
B. 45, 4MHz
4. Range
A. 0 mm to 50 mm
B. 0 mm to 100 mm scanning with full skip distance
5. Scanning Sensitivity
A. Echo of first back reflection adjusted to 75% of full screen height
B. Refer to AWS D1.1, paragraph 6.25.5.2 “Zero Reference Level”
6. Sizing Method
Length : 6dB drop, Width and Height : 20dB drop
7. Scanning Patterns
Appendix 3
Report No.
of
Date of Report
Project Name
(Total South Pars Projects Phase 2&3)
Owner/Customer
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 34/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
□ mm
Calibration Block Couplant Water Surface Condition
□ Technique 1 □ Technique 2
Decibels Discontinuity
Discontinuity
Attenuation Factor
Transducer Angle
Distance
Form Face
Remarks
Leg
Length
Attenuation
A b c d From X From Y
Name
Sign
Date
UT-HYUN-002(Rev.0) A4(210X297)
Appendix 5
SOUTH PARS FIELD DEVELOPMENT
(PHASE TWO AND THREE)
EPSCC3-CONTRACT NO. TSP/PJ/013
ONSHORE FACILITIES
Job No. : 99012 Page : 35/35
NDT PROCEDURE Doc No. :
(UT FOR STEEL STRUCTURE) PP 99012 999 0001 0565 Rev. 2
INSTRUMENT TYPE :
CALIBRATION RECORD OF
I.D No. :
ULTRASONIC EQUIPMENT
ID/SERIAL No. :
1. TEST EQUIPMENT’S
Type Maker
Transducer
Frequency Size Sr. No.
4. HORIZONTAL LINEARITY
1) Range – 100 mm
2) Range – 200 mm
Liner scale (a) 25 (b) 50 (c) 75 (d) 100 (e) 125 (f) 150 (g) 175 (h) 200
Echo position (a) (b) (c) (d) (e) (f) (g) (h)
Examined by Approved by