Bench Test Protocols: 1.0 Oxygen Supply Pressure
Bench Test Protocols: 1.0 Oxygen Supply Pressure
0, 21/3/20
3.3 Adjust flow control valve to achieve flow rate of 60 +/- 3 L/min
1
CPAP Bench Test Protocol V1.0, 21/3/20
PERFORMANCE TEST
TABLE 1
Oxygen Available [%]
PEEP Valve
Flowrate [L/min]
5cm H20 10cm H2O 15cm H2O
40 +/- 2 30.8 33.0 34.4
60 +/- 3 29.2 30.6 31.6
80 +/- 4 28.4 29.4 30.0
100 +/- 5 28.2 29.0 29.5
120 +/- 6 28.2 29.0 29.2
2
CPAP Bench Test Protocol V1.0, 21/3/20
TABLE 2
Maximum Flowrate [L/min]
PEEP Valve
5cm H20 10cm H2O 15cm H2O
156 +/- 15 144 +/- 14 132 +/- 13
TABLE 3
Oxygen Supply Consumption at 10cm H20 [L/min]
Flowrate Patient Oxygen Concentration [%]
[L/min] 30 +/- 2 50 +/- 2 70 +/- 2 100 +/- 2
40 +/- 2 6 +/- 0.5 18 +/- 1 24 +/- 2 40 +/- 4
60 +/- 3 8 +/- 0.5 24 +/- 2 36 +/- 3 60 +/- 6
80 +/- 4 10 +/- 1 30 +/- 3 50 +/- 5 78 +/- 7
100 +/- 5 12 +/- 1 40 +/- 4 62 +/- 6 100 +/- 10
120 +/- 6 14 +/- 1 46 +/- 4 74 +/- 7 118 +/- 11