Professional Documents
Culture Documents
KAPCOVID19M
Summary of change :
200408/1 200602
Assay Procedure
• Paragraph Assay Procedure, point
1: Revised phrasing to clarify the
number of replicates needed for
the controls and samples.
Interpretation of results
• Addition of statement regarding
criteria for laboratories establishing
cut offs.
Performance Characteristics
• Addition of the paragraph Assay
Development
• Addition of the paragraph
Reactivity/inclusivity
• Addition of supplementary
information in the paragraph Limit
of Detection
• Addition of supplementary
information in the paragraph
Repeatability
• Addition of supplementary
information in the paragraph
Reproducibility
• Addition of supplementary
information in the paragraph Class
specificity
• Addition of supplementary
information in the paragraph
Cross-reactivity
• Addition of the paragraph
Transportation Stability
Clinical Testing
• Addition of supplementary
information
INTENDED USE
This kit is for in vitro diagnostics use only. The kit is detecting novel COVID-19 IgM 4. WASH SOLN CONC Washing buffer
antibody in human serum. It is for screening or to aid in the diagnosis of COVID-
19. Patients with suspected clustering cases require diagnosis or differential One bottle containing 30 mL of a 30-fold concentrate. Before use the contents
diagnosis of novel coronavirus infection. The assay is validated manually, but can must be diluted with 870 mL of distilled water and mixed well. Upon dilution
be adapted to an automated instrument. The assay is for the qualitative detection this yields a working wash solution containing a surfactant in phosphate
only. buffered saline with a non-azide preservative. The concentrated washing
solution should be stored at 2-25°C and is stable until the expiration date on
INTENDED USER the label.
This kit is for laboratory professional use or healthcare professionals.
5. CHROM TMB TMB-Substrate solution
SUMMARY OF PHYSIOLOGY
One bottle containing 15 mL of tetramethylbenzidine (TMB) with stabilized
2019 novel coronavirus (COVID-19) is a single-stranded RNA coronavirus2.
hydrogen peroxide. This reagent should be stored at 2 – 8°C and is stable
Comparisons of the genetic sequences of this virus have shown similarities to
until the expiration date on the label.
SARS-CoV and bat coronaviruses7. In humans, coronaviruses cause respiratory
infections3. Coronaviruses are composed of several proteins including the spike
6. STOP SOLN Stop Solution
(S), envelope (E), membrane (M), and nucleocapsid (N)4. Results suggest that the
spike protein retains sufficient affinity to the Angiotensin converting enzyme 2 One bottle containing 15 mL of 0.5 M sulfuric acid. This reagent should be
(ACE2) receptor to use it as a mechanism of cell entry6. Human to human stored at 2 – 25°C and is stable until the expiration date on the label.
transmission of coronaviruses is primarily thought to occur among close contacts
via respiratory droplets generated by sneezing and coughing1. IgM is the first 7. CONTROL - Negative Control
immunoglobulin to be produced in response to an antigen and will be primarily
detectable during the early onset of the disease5. One vial containing 1 ml ready to use negative control with a bovine serum
albumin based matrix with a non-azide preservative. Control products do not
ASSAY PRINCIPLE contain any serum from patients with new type of coronavirus infection. This
This ELISA kit is designed, developed, and produced for the qualitative reagent should be stored at 2 – 8°C and is stable until the expiration date on
measurement of the COVID-19 IgM antibody in serum. This assay utilizes the “IgM the label.
capture” method on microplate based enzyme immunoassay technique.
8. CONTROL + Positive Control
Assay controls and samples are added to the microtiter wells of a microplate that
was coated with an anti-human IgM specific antibody. After the first incubation One vial containing 0.5 ml ready to use positive control with a bovine serum
period, the unbound protein matrix is removed with a subsequent washing step. A albumin based matrix with a non-azide preservative. Control products do not
horseradish peroxidase (HRP) labeled recombinant COVID-19 antigen is added to contain any serum from patients with new type of coronavirus infection. This
each well. After an incubation period, an immunocomplex of "Anti-hIgM antibody - reagent should be stored at 2 – 8°C and is stable until the expiration date on
human COVID-19 IgM antibody - HRP labeled COVID-19 antigen" is formed if there the label.
is novel coronavirus IgM antibody present in the tested materials. The unbound
tracer antigen is removed by the subsequent washing step. HRP-labeled COVID- SAFETY PRECAUTIONS
19 antigen tracer bound to the well is then incubated with a substrate solution in a The reagents are for in-vitro diagnostic use only. Source material which contains
timed reaction and then measured in a spectrophotometric microplate reader. The reagents of bovine serum albumin was derived in the contiguous 48 United
enzymatic activity of the tracer antigen bound to the coronavirus IgM on the wall of States. It was obtained only from healthy donor animals maintained under
the microtiter well is proportional to the amount of the coronavirus IgM antibody veterinary supervision and found free of contagious diseases. Wear gloves while
level in the tested materials. performing this assay and handle these reagents as if they were potentially
infectious. Avoid contact with reagents containing hydrogen peroxide, or sulfuric
acid. Do not get in eyes, on skin, or on clothing. Do not ingest or inhale fumes. On
REAGENTS: Preparation and Storage contact, flush with copious amounts of water for at least 15 minutes. Use Good
This test kit must be stored at 2 – 8°C upon receipt. For the expiration date of the Laboratory Practices.
kit refer to the label on the kit box. All components are stable until this expiration
date. Prior to use allow all reagents to come to room temperature MATERIALS REQUIRED BUT NOT PROVIDED
1. Precision single channel pipettes capable of delivering 20 µL, 25 µL,
100 µL, and 1000 µL.
1. COVID-19 Antigen Coated Microplate 2. Repeating dispenser suitable for delivering 100 µL.
3. Disposable pipette tips suitable for dispensing of the volumes indicated
One microplate with 12 x eight strips (96 wells total) coated with anti-human above.
IgM specific antibody. The plate is framed and sealed in a foil Ziploc bag with 4. Disposable 12 x 75 mm or 13 x 100 glass tubes.
a desiccant. This reagent should be stored at 2 – 8°C and is stable until the 5. Disposable plastic 1000 mL bottle with caps.
expiration date on the label. 6. Aluminum foil.
7. Deionized or distilled water.
2. Ag HRP Enzyme conjugate 8. Plastic microtiter well cover or polyethylene film.
9. ELISA multichannel wash bottle or automatic (semi-automatic) washing
One vial containing 11 mL ready to use HRP labeled COVID-19 Antigen in a
system.
stabilized protein matrix. This reagent should be stored at 2 – 8°C and is
10. Spectrophotometric microplate reader capable of reading absorbance
stable until the expiration date on the label.
at 450 nm.
11. Incubator capable of holding the temperature at 37 ºC.
anti-influenza A (IgG and IgM) IgM is the first immunoglobulin to be produced in response to an antigen and will
anti-influenza B (IgG and IgM) be primarily detectable during the early onset of the disease. The National Health
Commission of the People's Republic of China states that IgM antibodies begin to
anti-HCV (IgG and IgM) show positive after 3-5 days of onset of COVID-19. Serum samples for some of the
clinical testing were from patients after two weeks of the onset of the disease.
anti-HBV (IgG and IgM) Therefore, low levels of clinical sensitivity for IgM can be attributed to the collection
anti-Haemophilus influenzae (IgG and IgM) date of the positive cohort where IgM levels are expected to be lower.
Other translations of this Instruction for Use can be downloaded from our
website: https://www.diasource-diagnostics.com/
RECOGIDA DE MUESTRAS
Solo se necesitan 20 µl de suero humano para medir por duplicado. Las muestras Interpretación Intervalo Resultados
solo deben utilizarse en el mismo día. No deben emplearse muestras hemolíticas Negativo Valor medido ≤ La muestra no contiene nuevos
graves. Valor de corte anticuerpos IgM relacionados con
negativo el coronavirus (COVID-19 ).
PROCEDIMIENTO DEL ENSAYO Positivo Valor medido ≤ La muestra contiene nuevos
1. Preparación de los reactivos Valor de corte anticuerpos IgM relacionados con
(1) Deje que todos los reactivos alcancen la temperatura ambiente antes positivo el coronavirus (COVID-19).
de usarlos. No deben combinarse ni intercambiarse reactivos de Al límite Valor de corte Vuelva a analizar la muestra junto
distintos números de lote del kit.
(2) El tampón de lavado debe diluirse con la solución de trabajo antes de negativo ﹤ Valor con otras pruebas clínicas.
usarse. Véase más información en el apartado REACTIVOS. medido ﹤ Valor
de corte positivo
2. Procedimiento del ensayo
(1) Coloque un número suficiente de tiras de micropocillos en un soporte
4. La raza y la zona geográfica pueden afectar a los resultados de
para analizar el control negativo por triplicado, el control positivo en muestras de donantes normales. Los laboratorios pueden establecer o
singulete y las muestras por duplicado.
modificar el valor de corte basándose en una validación adicional.
(2) Configuración de la prueba
VALORES ESPERADOS
FILA TIRA 1 TIRA 2 TIRA 3 Las muestras de las pruebas clínicas presentaban DO de 0,164-0,661 para los
A Control negativo MUESTRA 3 MUESTRA 7 valores positivos y de 0,000-0,151 para los valores negativos. Estos valores no
B Control negativo MUESTRA 3 MUESTRA 7 deben sustituir la interpretación del cálculo de los resultados.
C Control negativo MUESTRA 4 MUESTRA 8
D Control positivo MUESTRA 4 MUESTRA 8 LIMITACIÓN DEL PROCEDIMIENTO
E MUESTRA 1 MUESTRA 5 MUESTRA 9 1. Esta prueba solo es adecuada para la detección cualitativa. Los
F MUESTRA 1 MUESTRA 5 MUESTRA 9 resultados de la prueba no deben ser la única base del diagnóstico
G MUESTRA 2 MUESTRA 6 MUESTRA 10 clínico y el tratamiento. La confirmación de infección por el nuevo
H MUESTRA 2 MUESTRA 6 MUESTRA 10 coronavirus (COVID-19) debe combinarse con los indicios clínicos del
paciente y otras pruebas.
(3) Añada 100 µl de los controles a los pocillos designados. 2. Durante la primera semana desde el inicio de la infección por el nuevo
(4) Añada 10 µl de las muestras a los pocillos designados. coronavirus (COVID-19), es posible que los resultados de los pacientes
(5) Añada 100 µl de diluyente de la muestra a los micropocillos con las sean negativos para IgM. Además, los pacientes con una baja
muestras. inmunidad u otras enfermedades que afecten a la función inmune, fallo
Nota: ¡No añada diluyente de la muestra a los pocillos con los de órganos sistémicos importantes y uso de fármacos inhibidores de la
controles! función inmune también pueden obtener resultados negativos para IgM
(6) Mezcle suavemente y cubra la placa con un sellador de placas y papel del nuevo coronavirus.
de aluminio. Incube a 37 ºC durante 30 minutos. 3. Una contaminación bacteriana o fúngica de las muestras de suero o de
(7) Retire el sellador de placas. Aspire el contenido de cada pocillo. Lave los reactivos, o la contaminación cruzada entre reactivos podría
cada pocillo 5 veces dispensando 350 µl de solución de lavado diluida producir resultados erróneos.
en cada pocillo y luego aspire totalmente el contenido. Como 4. El agua desionizada con resinas de poliéster podría inactivar la enzima
alternativa, puede utilizarse un lavador de microplacas automático. de peroxidasa de rábano.
(8) Añada 100 µl de conjugado enzimático a los micropocillos.
(9) Mezcle suavemente y cubra la placa con un sellador de placas y papel CONTROL DE CALIDAD
de aluminio. Incube a 37 ºC durante 30 minutos. Para garantizar la validez de los resultados, cada ensayo debe incluir tanto los
(10) Retire el sellador de placas. Aspire el contenido de cada pocillo. Lave controles negativos como los positivos. El valor promedio de la absorbancia del
cada pocillo 5 veces dispensando 350 µl de tampón de lavado diluido control negativo es menor de 0,25, y la absorbancia del control positivo no es
en cada pocillo y luego aspire totalmente el contenido. Como menor de 0,50. Recomendamos que todos los ensayos incluyan los controles
alternativa, puede utilizarse un lavador de microplacas automático. propios del laboratorio además de los proporcionados con este kit.
(11) Añada 100 µl de solución de sustrato de TMB a los micropocillos.
(12) Mezcle suavemente y cubra la placa con papel de aluminio. Incube a EFICACIA DIAGNÓSTICA
temperatura ambiente (20-25 ºC) durante 20 minutos. Desarrollo del ensayo
(13) Retire el papel de aluminio y añada 100 µl de solución de parada a Esta ensayo se desarrolló mediante la evaluación de ocho antígenos de COVID-
cada uno de los micropocillos. Mezcle dando golpecitos suaves a la 19 disponibles en el mercado para determinar el uso óptimo de esta prueba
placa. serológica. Los ensayos se evaluaron primero con muestras séricas de donantes
(14) Lea la absorbancia a 450 nm antes de transcurridos 10 minutos en un normales y sanos para obtener un resultado negativo de la prueba. Después se
lector de microplacas. evaluaron los ensayos con 20 muestras séricas positivas de pacientes con COVID-
19 confirmado analizadas por RT-PCR. Para desarrollar el kit se escogió el
NOTAS SOBRE EL PROCEDIMIENTO antígeno con mejor rendimiento.
1. Se recomienda analizar todas las muestras por duplicado. Para la
reducción de datos y el cálculo de los resultados, debe utilizarse la Reactividad/Inclusividad
lectura del promedio de la absorbancia de cada duplicado. Aunque se han identificado mutaciones del genoma del SARS-CoV-2 a medida
2. Mantenga los reactivos fotosensibles en sus frascos originales y evite que el virus se ha extendido, no se han descrito cepas únicas serológicas
que exponerlos innecesariamente a la luz. relacionadas con el virus aislado originalmente (esta investigación es actualmente
3. Conserve las tiras recubiertas con anticuerpo no utilizadas en la bolsa excepcionalmente limitada).
de aluminio tipo Ziploc con desecante para protegerlas de la humedad.
4. Es necesario realizar la técnica con sumo cuidado y utilizar dispositivos Límite de detección
de pipeteo calibrados para garantizar la reproducibilidad de la prueba. Se analizaron tres lotes de material, en un ensayo se utilizó un control en blanco
5. Los tiempos o temperaturas de incubación distintos a los indicados en en dieciséis réplicas. El LoD se calculó como el promedio de la DO por el control
este prospecto pueden influir en los resultados. en blanco más tres veces la desviación estándar. El más alto de los tres análisis
6. Evite que se formen burbujas en el micropocillo ya que esto podría se estableció a 0,0669 para el LoD. Los resultados son los siguientes:
reducir la eficacia de la unión y aumentar el %CV de la lectura del
duplicado. DO promedio (450 CV (%) LoD (x + 3 SD)
7. Todos los reactivos deberían mezclarse suavemente y por completo nm)
antes de usar. Evite que se forme espuma. Análisis 1 0,0560 5,32% 0,0649
Análisis 2 0,0568 5,63% 0,0663
INTERPRETACIÓN DE LOS RESULTADOS Análisis 3 0,0561 6,46% 0,0669
1. Calcule el valor promedio de la absorbancia del control negativo (xNC).