You are on page 1of 7

675

ARTICLE

Evaluation of total corneal power


measurements with a new
Downloaded from https://journals.lww.com/jcrs by IJFFeEaHn1fAmUjQu206yff+8ovkWFQvnpnPNd9og3pPkoW3tT9tX/tRPn3V6aa7wAulI58LSnRfQEIksSCtuXVn3YbotFMXHmKOu3UKtRZQrzgUA7VCQQWUCNiTAsNWmuQ/nMhjfBKwFlVdV5gh3njvmC2BCiJw82h8bDjeeqM3R6mHu/BWmnLI5Hco1HDV on 05/03/2020

optical biometer
Mehdi Shajari, MD, FEBO, Ruven Sonntag, Michaela Ramsauer, MD, Thomas Kreutzer, MD, FEBO,
Efstathios Vounotrypidis, MD, FEBO, Thomas Kohnen, MD, PhD, FEBO, Siegfried Priglinger, MD, PhD, FEBO,
Wolfgang J. Mayer, MD, PhD, FEBO

Purpose: To evaluate the repeatability and validity of total corneal 0.43 for true net power (TNP), and 0.39 for simulated K. Bland–
power measurements (total keratometry [TK]) obtained with a re- Altman analysis showed no significant difference between the
cently introduced optical biometer (IOLMaster 700, Carl Zeiss optical biometer’s TK compared with standard K and the
Meditec AG) to a Scheimpflug device (Pentacam, Oculus). Scheimpflug device’s simulated K, TNP, and TCRP, the P value
exceeding .05 in each case. A difference of >0.50 diopters
Setting: Department of Ophthalmology, Ludwig-Maximilians- between astigmatism measurements between TK and TCRP
University, Munich, Germany. was found in 10 cases compared with standard K and TNP
with 0 and 2 cases.
Design: Prospective randomized controlled trial.

Methods: The inclusion criteria were a corneal astigmatism of Conclusions: Repeatability of both devices was high, although
3.00 diopters (D) or less, no previous ocular surgery, no known measurements of meridian showed a great variability, suggesting
corneal irregularities, and no known dry eyes. All eyes were that numerous measurements are needed to enhance accuracy.
measured 3 times using the optical biometer and the Scheimp- TCRP measurements between recently introduced TK and TCRP
flug device. The results were statistically compared using cannot be used interchangeably. Future studies are necessary to
Bland–Altman, within-subject SD, and astigmatism vector evaluate which measurement will result in a better outcome when
analysis. respective measurements are used for toric intraocular lens
calculations.
Results: Ninety-three eyes of 93 subjects were included. Over-
all, the repeatability of all eyes measured was 0.42 for standard J Cataract Refract Surg 2020; 46:675–681 Copyright © 2020 Published by
K, 0.40 for TK, 0.45 for total corneal refractive power (TCRP), Wolters Kluwer on behalf of ASCRS and ESCRS

I
ntraocular lens (IOL) power calculation plays a vital manufacturer states that TCRP should realistically reflect
role in recovering a patient’s visual acuity by cataract the actual refractive power. The Scheimpflug device also
surgery. Different formulas can be used for IOL power measures the true net power (TNP), incorporating the
calculation, all depending strongly on the axial length and curvature of the anterior and posterior corneal surface,
the corneal power. Hence, a device that accurately and and the refractive indices of the different media.1,2
reliably measures these variables will generate a superior Simulated keratometry (K) is an estimate of the total
outcome for the patient. corneal power based on the assumption that the cornea is
One optical biometer capable of measuring the dis- a single refracting surface. The anterior corneal curvature
cussed variables is the Pentacam (Oculus), using and refractive index are used for its calculation.3
Scheimpflug technology. It measures the total corneal Although older optical biometers by Zeiss, such as the
refractive power (TCRP) using the ray-tracing method, IOLMaster 500, only measured the anterior corneal curva-
applying Snell’s law of refraction to calculate the re- ture, the Pentacam also takes the posterior corneal curvature
fractive power at any given point on the cornea; the into consideration. Studies have shown that the

Submitted: June 24, 2019 | Final revision submitted: December 9, 2019 | Accepted: January 11, 2020
From the Department of Ophthalmology, Ludwig-Maximilians University (Shajari, Sonntag, Ramsauer, Kreutzer, Vounotrypidis, Priglinger, Mayer), Munich, Department of
Ophthalmology (Shajari, Kohnen), Goethe University, Frankfurt, Germany.
M. Shajari and R. Sonntag contributed equally to this work.
Corresponding author: Wolfgang J. Mayer, MD, PHD, FEBO, Department of Ophthalmology, Ludwig-Maximilians University, Mathildenstrasse 8, Munich 80336,
Germany. Email: wolfgang.j.mayer@med.uni-muenchen.de.

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS 0886-3350/$ - see frontmatter
Published by Wolters Kluwer Health, Inc. https://doi.org/10.1097/j.jcrs.0000000000000136

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
676 EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS

measurement of the posterior corneal curvature reduces The total refractive power TK (IOLMaster 700) was compared
systematic measurement error and improves the refractive with standard K (IOLMaster), simulated K (Scheimpflug), TCRP
(Scheimpflug), and TNP (Scheimpflug). In addition, total corneal
accuracy, suggesting that patients could benefit from taking
curvature TR (IOLMaster) was compared with standard R (IOL-
the posterior corneal curvature into consideration when Master), simulated R (Scheimpflug), and TNP (Scheimpflug). For
measuring the total corneal astigmatism.4,5 The newer op- each comparison, the validity was evaluated with the Bland–Altman
tical biometer IOLMaster 700 includes the posterior corneal analysis.12 The difference between 2 measurements is plotted over the
curvature, and can incorporate the influence of the posterior mean of those 2 measurements to obtain a Bland–Altman diagram.
The graph indicated any systematic error, showing how far 2 readings
corneal curvature with the measurement of the total corneal
are separated from each other. The coefficient of repeatability (CoR)
power (total K [TK]).6 and the relative CoR (CoR/average measurement) were calculated.
The recently introduced IOLMaster 700 measures axial Furthermore, limits of agreement (LOA = mean ± CoR) were re-
length and corneal power using swept-source optical co- corded, which define the variability of the values measured. The
herence tomography (SS-OCT) technology. SS-OCT scans analysis of the refractive power also included an independent 2-
sample t-test. A P value of less than 0.05 was considered statistically
the eye with a rapid-cycle, tunable wavelength laser, which
significant.
is known to give a better signal-to-noise ratio and has Another analysis of validity for the comparison mentioned above
improved tissue penetration and image quality.7 Several was performed, emphasizing clinical significance rather than sta-
published studies have shown good results for agreement tistical significance. The difference between the first measurement of
between the IOLMaster 700 and other devices, as well as each of the 2 variables compared was calculated, and all values were
highlighted, exceeding a defined limit. For the comparison with TK,
high repeatability; however, comparison so far has only
any difference in the total corneal astigmatism greater than 0.50
looked at the anterior corneal surface and not the total diopters (D) was highlighted, and for the comparison of the axis, any
corneal power. In addition, the IOLMaster 700 showed difference greater than 20 degrees, respectively. The sum of cases
a higher success rate in obtaining measurements in eyes where a value was highlighted would therefore show which variables
with posterior subscapular and dense nuclear cataracts.8–10 when compared had a difference greater than the defined clinically
significant limit.
The purpose of this study was to evaluate the re-
Vector analysis was performed to evaluate the corneal astig-
peatability and validity of the corneal power measurements matism, according to Abulafia et al.13 Double-angle plots were
compared with the corresponding variables of both optical generated with the Astigmatism Double Angle Plot Tool.14
biometers. The main focus was on the comparison of TK
(IOLMaster 700) with TCRP and TNP (Pentacam).
RESULTS
The study evaluated 93 healthy volunteers (43 men and 50
METHODS women). The mean age was 52.35 ± 19.1 (SD) years (range
This prospective controlled study included 93 eyes of 93 volun- 18 to 85 years).
teers, acquiring proper informed consent for participation of every
patient enrolled. All patients were examined in the Department of
Ophthalmology, Ludwig-Maximilians-University, Munich, Ger- Repeatability
many. The study was performed in conduct with the principles of Table 1 shows the within-subject SD (Sw) and repeatability
the Declaration of Helsinki and was approved by the local in- (2.77 × Sw) for each variable measured, where a smaller
stitutional review board. value indicates a better repeatability.
Exclusion criteria included previous ocular surgery, known
corneal irregularities, previous corneal pathology, dry eyes, and
a corneal astigmatism above 3.00 diopters (D). In addition, only Comparability
measurements were included, which achieved the image quality Table 2 shows the P value, CoR, relative CoR, and the limits
recommended by the manufacturer (for the IOLMaster 700, of agreement of the comparison between the true values
measurements marked with an exclamation mark were excluded; (TR and TK measured with the IOLMaster 700) and the
for the Pentacam, it needed to be marked “OK” by the automated
quality check). corresponding variables. Figure 1 and Figure 2 show the
One of 3 trained optometrists performed 3 complete auto- corresponding Bland-Altman plots for each comparison
mated measurements with each of the 2 devices. The order of mentioned in Table 2.
measurement was chosen randomly. In between each finished The 2-sample t-test for the comparison of astigmatism
measurement, the patient was asked to stand up, and the po- measured using TK and the astigmatism measured with
sition of the device’s joystick and chin rest was changed to
ensure that the measurements were independent of one standard K, simulated K, TNP, and TCRP showed no
another. significant difference between any of the mentioned vari-
For the Pentacam, the patients were asked to fixate on the ables (P = .125 for standard K, P = .153 for simulated K, P =
light, and Scheimpflug pictures were taken as soon as the correct .167 for TNP, and P = .557 for TCRP).
alignment was achieved. For the IOLMaster 700, the patients The test we conducted to analyze in how many cases the
positioned their chin on the headrest and looked at the fixation
light. difference between 2 variables exceeded a clinically signifi-
The repeatability of the devices used to collect 3 measurements cant limit we defined beforehand showed a certain trend. For
of each eye was determined by calculating the within-subject SD the difference in astigmatism, 0.50 D was exceeded in 0 cases
(Sw) according to Bland and Altman.11 In addition, pffiffiffithe re- for the comparison with standard K and in 2 cases for TNP
peatability was calculated by multiplying the Sw with 2 × 1.96 out of the 93 eyes examined. Simulated K exceeded this limit
(=2.77). This value indicates that the difference between 2
measurements for the same patient is expected to be less than in 4 cases, TCRP in 10. For the difference in meridian, 20
2.77 × Sw for 95% of pairs of measurements.11 degrees was exceeded in 8 and in 9 cases for standard K and

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS 677

Table 1. Repeatability of the variables measured.


Pentacam IOLMaster 700

Repeatability Repeatability
Variable SW (SW × 2.77) Variable SW (SW × 2.77)
CCT (µm) 2.47 6.83 CCT (µm) 3.45 9.55
ACD (mm) 0.08 0.21 ACD (mm) 0.03 0.09
Simulated K Standard K
R1 (mm) 0.03 0.07 R1 (mm) 0.06 0.17
R2 (mm) 0.02 0.06 R2 (mm) 0.02 0.07
Asti (D) 0.14 0.39 Asti (D) 0.15 0.42
Meridian (°) 7.45 20.63 Meridian (°) 8.65 23.95
TCRP TK
K1 (D) 0.16 0.44 R1 (mm) 0.02 0.06
K2 (D) 0.57 1.59 R2 (mm) 0.06 0.18
Asti (D) 0.16 0.45 Asti (D) 0.14 0.40
Meridian (°) 9.37 25.95 Meridian (°) 12.44 34.47
Posterior cornea
K1 (D) 0.04 0.11 WzW (mm) 0.23 0.65
K2 (D) 0.04 0.11 AL (mm) 0.11 0.30
Asti (D) 0.04 0.12 LD (mm) 0.02 0.05
Meridian (°) 5.73 15.86
TNP
R1 (mm) 0.03 0.08
R2 (mm) 0.02 0.06
Asti (D) 0.16 0.44
Meridian (°) 10.68 29.58

ACD = anterior chamber depth; AL = axial length; Asti = astigmatism; CCT = central corneal thickness; IOL = intraocular lens; K = refractive power; LD = lens
thickness; R = corneal curvature; SW = within-subject SD; TCRP = total corneal refractive power; TK = total keratometry; TNP = true net power; WzW =
horizontal corneal diameter

TNP, respectively. In contrast, simulated K exceeded this DISCUSSION


limit 14 times and TCRP 19 times. In this study, we tested the repeatability of variables
Table 3 shows the astigmatism vector mean of each measured by both the Pentacam and the IOLMaster 700.
variable measured with SD and its meridian in degrees. This analysis was performed, calculating the within-subject
Table 4 shows the difference in the vectors obtained when SD (Sw). Looking at astigmatism measured with both de-
comparing the different variables, similar to the Bland– vices, the repeatability (Sw × 2.77) is very similar
Altman analysis. Figure 3 shows the vector analysis with throughout, with 0.42 for standard K, 0.40 for TK, 0.45 for
double-angle plots, representing astigmatism measure- TCRP, 0.43 for TNP, and 0.39 for simulated K.
ments obtained using the different corneal power param- Srivannabonn et al. showed that the repeatability of the
eters evaluated. Looking at the 95% confidence ellipse of the IOLMaster 700 was excellent, and compared with the
dataset, standard K and TK look very much alike. This is IOLMaster 500, agreement was high. Furthermore, the lens
also the case for the Pentacam’s simulated K, TNP, and penetration of the IOLMaster 700 was superior compared
TCRP. with its older predecessor.7 Other studies have come to

Table 2. Comparison of true corneal curvature and TK astigmatism with the corresponding variables measured.
Mean Upper Lower LOA
Variable Eyes (n) P Value Difference CoR rCoR LOA LOA Range R2
CCD vs CCD 93 .003 3.495 16.035 2.853 19.529 12.540 32.069 0.099
ACD vs ACD 93 .005 0.004 0.209 6.506 0.213 0.205 0.418 0.083
TR vs standard R 93 .839 0.006 0.048 0.626 0.042 0.054 0.096 0.001
TR vs SimR 93 .895 0.056 0.111 1.439 0.055 0.166 0.222 0.0002
TR vs TNP R 93 .174 0.289 0.104 1.335 0.185 0.394 0.209 0.202
TK Asti vs stan Asti 93 .080 0.481 0.330 39.516 0.378 0.282 0.660 0.033
TK Asti vs SimK Asti 93 .466 0.047 0.629 65.155 0.676 0.581 1.258 0.006
TK Asti vs TNP Asti 93 .089 0.430 0.584 60.561 0.541 0.627 1.169 0.0002
TK Asti vs TCRP Asti 93 .896 0.226 0.725 71.469 0.748 0.703 1.451 0.031

ACD = anterior chamber depth; Asti = astigmatism; CCT = central corneal thickness; CoR = coefficient of repeatability; LOA = limit of agreement; R = corneal
curvature; rCoR = relative CoR; stan = standard; TCRP = total corneal refractive power; TK = total keratometry; TNP = true net power

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
678 EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS

Figure 1. Bland–Altman plots of comparison between astigmatism measured using TK with astigmatism measured using (a) standard K,
(b) simulated K, (c) TNP, and (d) TCRP (K = keratometry; TCRP = total corneal refractive power; TK = total keratometry; TNP = true net power).

a similar conclusion, demonstrating the high repeatability achieved by TCRP measurements with the Pentacam in
of both the IOLMaster 700 and the Pentacam.7,8,15 The comparison with various topographers, including the
within-subject SD for standard K found by Kunert et al.16 IOLMaster 500. The IOLMaster 700 seems to have closed
was slightly larger, where the Sw was 0.20 D compared with the gap because our results suggested a slightly better
0.15 D in this study. Another study from 2017 also reported repeatability using TK compared with TCRP.18 However,
a higher within-subject SD for standard K with 0.24 D.17 the results shown in Table 1 also suggest that the re-
Reasons for this might be a higher number of patients peatability is not given when measuring the meridian of
tested, in addition to 3 measurements of the same eye the astigmatism for any of the variables evaluated.
conducted in comparison to 2 measurements per eye in the Therefore, before calculating a patient’s toric IOL cyl-
study from 2017. inder power, repeated measurements of the astigmatism
A study from 2016 showed that the highest re- meridian should be taken into consideration to ensure an
peatability of corneal astigmatism measurements was acceptable outcome.

Figure 2. Bland–Altman plots of comparison between true corneal curvature (TR) with corneal curvature measured using (a) standard R,
(b) simulated R and (c) TNP (TNP = true net power).

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS 679

Table 3. Comparison of the astigmatism vector mean of


compared with TK in a clinical setting. The differences in
each variable measured. meridian show more cases where the limit of 20 degrees
Comparability was exceeded, no matter the variable compared with TK.
This is probably due to the fact that the repeatability of
Astigmatism Vector
Device Variable (Mean ± SD at Degrees)
meridian measurements was rather poor, as suggested by
the analysis of the within-subject SD. The bigger range in
IOLMaster 700 TK 0.43 ± 0.75 at 177
K 0.62 ± 0.74 at 179
measurements would therefore lead to larger differences
Pentacam SimK 0.65 ± 0.77 at 180 in the test conducted. Once again, this would imply that
TCRP 0.57 ± 0.79 at 001 the meridian of the astigmatism should be measured
TNP 0.45 ± 0.76 at 177 more than once in a clinical setting to enhance the ac-
IOL = intraocular lens; K = keratometry; TCRP = total corneal refractive curacy of treatment.
power; TK = total keratometry; TNP = true net power Analysis of the vectors, as shown in Tables 3 and 4,
showed great agreement between the astigmatism vector
This study also evaluated the validity of the IOLMaster mean of TK and TNP, with a mean difference of 0.01 ±
corneal power measurements (TK and TR) compared 0.32 D. Comparison between TK and TCRP (difference of
with simulated K, TNP, and TCRP. The Bland–Altman 0.15 ± 0.36 D) and between TK and standard K (difference
analysis has shown no significant difference between the of 0.19 ± 0.09 D) showed weaker agreement. Together
IOLMaster 700’s TR and TK compared with the corre- with the results summarized above, this would suggest that
sponding Pentacam measurements of simulated K, TNP, TK and TNP are most comparable.
and TCRP, the P value exceeding .05 in each case. Visser et al.19 compared various topographers using
The Bland–Altman graphs in Figures 1 to 2 show vector analysis and found a mean difference in corneal
a mean difference near zero, implying that the meas- astigmatism of 0.03 ± 0.29 between standard K with
urements are somewhat comparable. The mean difference the IOLMaster 500 and simulated K. When we com-
was lowest for the comparison between the IOLMaster pared standard K with simulated K, we found the same
700’s TK and the Pentacam’s simulated K with 0.047. The mean difference of 0.03 ± 0.29, although we used the
comparison between TK and standard K, TNP, and TCRP newer IOLMaster 700 and had different mean astig-
showed a higher mean difference with 0.481, 0.430, and
matism vectors. This would suggest a similar compa-
0.226, respectively. This suggests that the IOLMaster 700
rability and agreement of the measurements made with
measurements are most comparable to the Pentacam’s
simulated K followed by TCRP. both devices.
The test to evaluate the differences in total corneal As the concept of a measured TK is fairly new, there
astigmatism exceeding 0.50 D showed a clear trend. are a few things to consider. Zeiss claims that TK can be
Differences between astigmatism measured with TK incorporated in classic IOL power calculations, using
compared with standard K (0 cases) and TNP (2 cases) optimized IOL constants, such as User Group for Laser
did not exceed the limit as often as it was the case with Interference Biometry (ULIB) constants, which was
simulated K and TCRP (4 and 10 cases). Differences in verified by Haigis et al.20 So far, ULIB constants are
astigmatism measurements exceeding a clinically sig- based on keratometry measurements of the anterior
nificant limit are concealed in a calculation for the mean corneal curvature and fixed ratios, rather than the total
difference, as in the Bland–Altman analysis. This test corneal power, considering the actually measured pos-
reveals these cases and suggests a different conclusion terior corneal curvature. Therefore, the question arises
that standard K and TNP show a higher validity how TK can still be used with standard ULIB lens
constants and whether TK has been adjusted to do so.
Because standard K and TK show no significant dif-
ference in astigmatism measured (P = 0.125), it can be
Table 4. Difference between the astigmatism vector
mean of the variables compared. assumed that Zeiss has adjusted TK to some degree. The
Mean Difference
growing use of optical biometers capable of measuring
the posterior corneal curvature could affect the ULIB
Astigmatism Vector
constants in the future, shifting them toward meas-
Variables Compared (Mean ± SD at Degrees)
urements considering both the anterior and posterior
TK vs TCRP 0.15 ± 0.36 at 12
corneal curvature. Although the repeatability of both
TK vs TNP 0.01 ± 0.32 at 03
TK vs K 0.19 ± 0.09 at 03 devices has been discussed in several studies, the
K vs SimK 0.03 ± 0.29 at 21 comparability of total corneal power measurements,
TCRP = total corneal refractive power; TK = total keratometry; TNP = true such as TK, needs to be explored further, especially in
net power a clinical setting.

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
680 EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS

Figure 3. Double-angle plots visualizing the astigmatism measurements measured with (a) standard K, (b) TK, (c) SimK, (d) TNP, and (e) TCRP
(TCRP = total corneal refractive power; TK = total keratometry; TNP = true net power).

keratomileusis or photorefractive keratectomy for myopia. J Cataract Re-


fract Surg 2016;42:363–369
WHAT WAS KNOWN 7. Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical com-
 Because of variability in the posterior corneal curvature, parison of a new swept-source optical coherence tomography–based
a significant difference exists between measuring the anterior optical biometer and a time-domain optical coherence tomography–based
surface alone and simulating the total corneal power com- optical biometer. J Cataract Refract Surg 2015;41:2224–2232
pared with measuring the total corneal refractive power 8. Kurian M, Negalur N, Das S, Puttaiah NK, Haria D, J TS, Thakkar MM.
Biometry with a new swept-source optical coherence tomography bio-
(TCRP) directly.
meter: repeatability and agreement with an optical low-coherence reflec-
 Scheimpflug devices can measure with a high repeatability
tometry device. J Cataract Refract Surg 2016;42:577–581
the anterior and posterior corneal curvature and TCRP. 9. Akman A, Asena L, Güngör SG. Evaluation and comparison of the new
swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J
WHAT THIS PAPER ADDS Ophthalmol 2016;100:1201–1205
 A recently introduced software update allows a widely used 10. Hirnschall N, Varsits R, Doeller B, Findl O. Enhanced penetration for axial
optical biometer to measure the TCRP with high repeatability. length measurement of eyes with dense cataracts using swept source
 There is a significant difference between total corneal optical coherence tomography: a consecutive observational study. Oph-
thalmol Ther 2018;7:119–124
measurements of the optical biometer and the Scheimpflug 11. Bland JM, Altman DG. Measurement error. BMJ 1996;313:744
device. 12. Martin Bland J, Altman DG. Statistical methods for assessing agreement
between two methods of clinical measurement. Lancet 1986;327:307–310
13. Abulafia A, Koch DD, Holladay JT, Wang L, Hill W. Pursuing perfection in
REFERENCES intraocular lens calculations: IV. Rethinking astigmatism analysis for in-
1. Tonn B, Klaproth OK, Kohnen T. Anterior surface–based keratometry traocular lens-based surgery: suggested terminology, analysis, and stand-
compared with Scheimpflug tomography–based total corneal astigmatism. ards for outcome reports. J Cataract Refract Surg 2018;44:1169–1174
Invest Ophthalmol Vis Sci 2015;56:291–298 14. Astigmatism Double Angle Plot Tool. ASCRS. American Society of Cararact
2. Savini G, Barboni P, Carbonelli M, Hoffer KJ. Comparison of methods to and Refractive Surgery. Available at: https://ascrs.org/astigmatism-double-
measure corneal power for intraocular lens power calculation using a rotat- angle-plot-tool. Accessed May 15, 2019
ing Scheimpflug camera. J Cataract Refract Surg 2013;39:598–604 15. Sel S, Stange J, Kaiser D, Kiraly L. Repeatability and agreement of
3. Herrmann C, Ludwig U, Duncker G. [Corneal topography; analysis of the Scheimpflug-based and swept-source optical biometry measurements.
corneal surface]. Opthalmologe 2008;105:193–206 Contact Lens Anterior Eye 2017;40:318–322
4. Rydström E, Westin O, Koskela T, Behndig A. Posterior corneal astigmatism 16. Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T.
in refractive lens exchange surgery. Acta Ophthalmol (Copenh) 2016;94: Repeatability and agreement in optical biometry of a new swept-source optical
295–300 coherence tomography–based biometer versus partial coherence interferometry
5. Koch DD, Ali SF, Weikert MP, Shirayama M, Jenkins R, Wang L. Contri- and optical low-coherence reflectometry. J Cataract Refract Surg 2016;42:76–83
bution of posterior corneal astigmatism to total corneal astigmatism. 17. Shajari M, Cremonese C, Petermann K, Singh P, Müller M, Kohnen T.
J Cataract Refract Surg 2012;38:2080–2087 Comparison of axial length, corneal curvature, and anterior chamber depth
6. Abulafia A, Hill WE, Koch DD, Wang L, Barrett GD. Accuracy of the Barrett measurements of 2 recently introduced devices to a known biometer—
True-K formula for intraocular lens power prediction after laser in situ ScienceDirect. Am J Ophthalmol 2017;178:58–64

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.
EVALUATION OF TOTAL CORNEAL POWER MEASUREMENTS 681

18. Fityo S, Bühren J, Shajari M, Kohnen T. Keratometry versus total 20. Haigis W, Sekundo W, Kunert KS, Blum M. Total keratometric power
corneal refractive power: analysis of measurement repeatability with (TKP) derived from corneal front and back surfaces using a full eye-length
5 different devices in normal eyes with low astigmatism - journal of SS-OCT scan biometer prototype in comparison to automated keratom-
Cataract & Refractive Surgery. J Cataract Refract Surg 2016;42: etry. Presented at the annual meeting of the XXXII Congress ESCRS,
569–576 London, United Kingdom, September 2014
19. Visser N, Berndschot TosTJM, Verbakel F, de Brabander J, Nujits RMMA.
Comparability and repeatability of corneal astigmatism measurements using
different measurement technologies - ScienceDirect. J Cataract Refract Disclosures: None of the authors has a financial or proprietary
Surg 2012;38:1764–1770 interest in any material or method mentioned.

Volume 46 Issue 5 May 2020

Copyright © 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS. Unauthorized reproduction of this article is prohibited.

You might also like