-Once aligned, tighten the screws and repeat the following commands to confirm a good alignment:
PLTH RWELA01729 IOFF
If alignment is not ok, realign and confirm. If alignment is OK, do the following:
-Now check your alignment at H12 by typing:
PLTH RWELH12722 IOFF
Check alignment on E12 and H12 following the same steps as when checking A-04 and D-04.
-Realign if necessary.
-If the alignment is good at one end but not the other, and after realinging several times, the plate
may be overshooting/undershooting. Refer to section “4 Inspect Plate Y Overshoot/Undershoot”.
Paremeter Confirmation
When the alignment is done, perform the following through the ChemWell Communications window:
RWELA01722 IOFF POSI
Visually confirm that the alignment is ok, and record the positioning values (obtained with !POSI) for
the plate X and Y coordinates. Now record the new values in the parameter list by typing:
PARM4Xxxxx PARM4Yyyyy
where xxxx = new POSI coordinate for plate X, and yyyy = new POSI coordinate for Plate Y