You are on page 1of 14

GE Healthcare

Chalfont St. Giles


Buckinghamshire, UK
www.gehealthcare.com

LOGIQ S8

Product Description
The LOGIQ* S8 is our premium multi-purpose ultrasound imaging system designed for abdominal, vascular, breast,
cardiac, small parts, obstetrics, gynecology, neonatal, pediatrics, urology and transcranial applications.

DOC1487369: LOGIQ S8 R2 Specification Sheet


General Specifications LCD Monitor • Micro convex Array
• Linear Array
• 19” high-resolution LCD
Dimensions and • Matrix Array
• LCD translation (independent of
• Volume Probes (4D)
Weight console):
• Convex Array
• 660 mm horizontal (end to end)
• Height • Micro convex Array
• 135 mm vertical (end to end)
• Maximum: 1750mm, 68.9in • Split Crystal
• 90° swivel
• Minimum: 1150mm, 45.3in
• Fold-down and lock mechanism for
• Width
transportation
System Standard
• Keyboard: 500mm, 19.7 in
• Caster: 620mm, 24.4 in
• Brightness and contrast adjustment Features
• Resolution: 1280 X 1024 • Advanced user interface with high
• Depth
• Horizontal/Vertical viewing angle of resolution wide 9” LCD touch panel
• Maximum: 880mm, 34.6 in
+/- 170° • Automatic Optimization
• Caster: 790mm, 31.1 in
• Weight • CrossXBeam*
• 85 kg, 187.4 lbs
System Overview • Speckle Reduction Imaging (SRI-HD)
• Fine Angle Steer
Electrical Power Applications • Coded Harmonic Imaging
• Abdominal • Virtual Convex
• Voltage: 100-120 Vac or 220-240
• Obstetrical • Patient information database
Vac
• Gynecological • Image Archive on integrated
• Frequency: 50/60 Hz
• Breast CD/DVD and hard drive
• Power consumption maximum of
900VA with peripherals • Small parts • 3D
• Vascular /Peripheral • Raw Data Analysis
Console Design • Transcranial • Real-time automatic Doppler calcs
• 4 Active probe ports and 1 parking • Pediatrics and neonatal • OB Calcs
• Musculoskeletal • Fetal Trending
• Integrated HDD (at least 500 GB)
• Urological • Multigestational Calcs
• Integrated DVD multi drive
• Cardiac • Hip Dysplasia Calcs
• On-board storage of thermal
printer • Gynecological Calcs
• Integrated speakers
Operating Modes • Vascular Calcs
• Locking mechanism that provides • B-Mode • Urological Calcs
rolling lock and caster swivel lock • M-Mode • Renal Calcs
• Integrated cable management • Color Flow Mode (CFM) • Cardiac Calcs
• Front and rear handles • TVI (Option) • InSite*ExC capability
• Easily removable air filters • B-Flow*/B-Flow Color (Option) • On-board electronic documentation
• Extended Field of View (LOGIQ
User Interface View, Option) Peripheral Options
• Power Doppler Imaging (PDI) • Integrated options for:
Operator Keyboard • PW Doppler • Digital BW thermal printer
• Operating keyboard adjustable in • CW Doppler (Option) • DVD video recorder
two dimensions: • Volume Modes (3D/4D) • Digital color thermal printer
• Height • Static 3D • Digital A6 color thermal printer
• Rotation • Real Time 4D (option) • External USB printer connection
• Backlit alphanumeric keyboard • Anatomical M-Mode • DVI-D output available for
• Ergonomic hard key layout • Curved Anatomical M-Mode compatible devices
• Interactive back-lighting • B Steer+ (Option) • Foot Switch, with programmable
• Integrated recording keys for • Coded Contrast Imaging (Option) functionality
remote control of up to 6 • Elastography (Option) • Video Converter
peripheral or DICOM** devices
• Integrated gel warmer
Scanning Methods Display Modes
• Electronic Sector • Live and Stored Display Format: Full
Touch Screen • Electronic Convex size and split screen - both w/
• Wide 9” high-resolution, color, • Electronic Linear thumbnails. for still and Cine
touch, LCD screen • Mechanical Volume Sweep • Review Image Format: 4x4 and
• Interactive dynamic software "thumbnails". for still and Cine
menu
Transducer Types • Simultaneous Capability
• Brightness adjustment • Sector Phased Array • B or CrossXBeam /PW
• User-configurable layout • Convex Array • B or CrossXBeam /CFM or PDI

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 2 of 14


• B/M • Gain • Body Patterns
• B/CrossXBeam • Angle • Customized Comment Home
• Real-time Triplex Mode • Sample Volume Depth and Width Position
(B or CrossXBeam + CFM or • Wall Filter
PDI/PW or CW(Option)) • Velocity and/or Frequency Scale User Manual available
Selectable alternating Modes • Spectrum Inversion

• B or CrossXBeam /PW • Time Scale
on board through Help
• B or CrossXBeam + CFM • PRF (F1)
(PDI)/PW(CW(Option)) • Doppler Frequency
User Manual and Service Manual are
• B/CW (Option) • Color Flow Mode
included on CD with each system. A
• Multi-image (split/quad screen) • Line Density
printed manual is available upon
• Live and/or frozen • Frame Averaging
request.
• B or CrossXBeam + B or • Packet Size
CrossXBeam /CFM or PDI • Color Scale: 3 types
• PW/M Power, Directional PDI, and
Cine Memory/Image
• Independent Cine playback Symmetrical Velocity Imaging Memory
• Time line display • Color Velocity Range and Baseline • 776 MB of Cine Memory
• Independent Dual B or • Color Threshold Marker • Selectable Cine Sequence for Cine
CrossXBeam /PW Display • Color Gain Review
• CW • PDI • Prospective Cine Mark
• Display Formats • Inversion • Measurements/ Calculations and
• Top/ Bottom selectable format • Doppler Frequency Annotations on Cine Playback
• Side/Side selectable format • TGC Curve • Scrolling timeline memory
• Timeline only • Acoustic Frame Rate • Dual Image Cine Display
• Virtual Convex • Cine Gage, Image Number / Frame • Quad Image Cine Display
Number • Cine Gauge and Cine Image
Display Annotation • Body Pattern: Multiple human and Number Display
• Patient Name: First, Last and animal types • Cine Review Loop
Middle • Application Name • Cine Review Speed: 11 steps (11, 13,
• Patient ID • Measurement Results 14, 17, 22, 25, 31, 48, 100, 200,
• 2nd Patient ID • Operator Message 400%)
• Age, Sex and Birth Date • Displayed Acoustic Output
• Hospital Name • TIS: Thermal Index Soft Tissue Image Storage
• Date format: 3 types selectable • TIC: Thermal Index Cranial (Bone) • On-board database of patient
MM/DD/YY, DD/MM/YY, YY/MM/DD • TIB: Thermal Index Bone information from past exams
• Time format: 2 types selectable: • MI: Mechanical Index • Storage Formats:
24 hours, 12 hours • % of Maximum Power output • DICOM – compressed/
• Gestational Age from • Biopsy Guide Line and Zone uncompressed, single/ multi
LMP/EDD/GA/BBT • Heart Rate frame, with/ without Raw Data
• Probe Name • Export JPEG, JPEG2000, WMV
• Map names General System (MPEG 4) and AVI formats
• Probe Orientation Parameters • Storage Devices:
• Depth Scale Marker • USB Memory Stick: 64MB to 4GB
• Lateral Scale Marker System Setup (for exporting individual
• Focal Zone Markers images/clips)
• Pre-programmable Categories
• Image Depth • CD-RW storage: 700MB
• User Programmable Preset
• Zoom Depth • DVD storage: -R (4.7GB)
Capability
• B-Mode • Hard Drive Image Storage:
• Factory Default Preset Data
• Gain ~112GB
• Languages: English, French,
• Dynamic Range • Compare previous exam images
German, Spanish, Italian,
• Imaging Frequency with current exam images
Portuguese, Russian, Greek,
• Frame Averaging • Reload of archived data sets
Swedish, Danish, Dutch, Finnish,
• Gray Map • Network Storage support for Import,
Norwegian, Japanese (message
• SRI-HD Export, DICOM Read, SaveAs,
only), Chinese (message only)
• M-Mode SaveAs Image, Report SaveAs,
• OB Report Formats including Tokyo
• Gain MPEGVue
Univ., Osaka Univ., USA, Europe, and
• Dynamic Range
ASUM
• Time Scale
• User Defined Annotations
• Doppler Mode

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 3 of 14


Connectivity and • Continuous Dynamic Receive Focus Digital Spectral
/ Continuous Dynamic Receive
DICOM Aperture Doppler Mode
• Ethernet network connection • 274dB Dynamic Range • Adjustable:
• DICOM 3.0 (Option) • System frequency range: 1-13MHz • Acoustic Power: 0-100%, 25 steps
• Verify • 256 shades of gray • Gain: 0 – 85dB range, 1dB steps
• Print • Adjustable Field of View (FOV) • Gray Scale Map: 8 maps
• Store • Image Reverse: Right/ Left • Transmit Frequency: Up to 5 steps
• Modality Worklist • Image Rotation of 0°, 180° • Wall Filter: 18 settings, Min: 5 Hz,
• Storage Commitment Max: 3,339Hz
• Modality Performed Procedure Digital B-Mode • PW Colorization: 6 maps
Step (MPPS) • Adjustable: • Velocity Scale Range: 6.9 –
• Media Exchange • Acoustic Power: 0 - 100%, 25 1032.8cm/s
• Off network / mobile storage steps • Sweep Speed: 0 – 7, 8 steps
queue • Gain: 0 – 90 dB range, 1dB steps • Sample Volume Length: 1 – 16mm,
• Query / Retrieve • Dynamic Range: 36 – 96 dB, 16 12 steps
• Public SR Template steps • Angle Correction: -90 - 90 degrees,
• Structured Reporting – • Frame Averaging: up to 8 steps 1 degree steps
compatible with vascular and • Gray Scale Map: 10 maps • Steered Linear: 0 – 30 degrees
OB standard • Frequency: up to 6 steps • Spectrum Inversion
• InSite ExC capability • Speed of Sound (probe, • Trace Method
application dependent) • Baseline Shift: 5 – 95%, 11 steps
Physiological Input • Line Density: up to 5 steps • Doppler Auto Trace
• Compression: 0.5 – 2.4, 9 steps
Panel (Option) • Scanning Size (FOV or Angle -
• Trace Direction
depending on the probe, see
• Physiological Input
probe specifications) • Trace Sensitivity
• ECG, 2 lead
• B Colorization: 9 maps
• Dual R-Trigger
• Reject: 6 steps
• Pre-settable ECG R Delay Time Digital Color Flow
• Suppression: up to 6 steps
• Pre-settable ECG Position
• Adjustable ECG Gain Control
• SRI-HD: 0 – 5, 6 steps Mode
• Edge Enhance: 7 steps • Adjustable:
• Automatic Heart Rate Display
• Acoustic Power: 0 - 100%, 25
Report Writer (Option) Digital M-Mode steps
• Adjustable: • Color Maps, including velocity-
• On-board reporting package
• Acoustic Power: 0 - 100%, 25 variance maps: 15 maps
automates report writing
steps • Gain: 0 – 40dB, 81 steps
• Formats various exam results into
• Gain: -20 – 20 dB (delta from B) • Velocity Scale Range: 2 – 370cm/s
a report suitable for printing or
• Dynamic Range : 36 – 96 dB, 16 • Wall Filter: 0 – 3, 4 steps
reviewing on a standard PC
steps • Packet Size: 5 – 24, 9 steps
• Exam result reports can include
• Gray Scale Map: 10 maps • Line Density: 5 steps
patient info, exam info,
• Frequency: up to 6 steps • Spatial Filter: 6 steps
measurements, calculations,
• Sweep Speed: 0–7, 8 steps • Steering Angle: 0 – 20 degrees
images, comments and physician
• M Colorization: 9 maps • Baseline Shift: 0 – 100%, 11 steps
diagnosis
• M Display Format: V-1/3B, V-1/2B, • Frame Average: 0 – 6, 7 steps
• Standard templates provided
V-2/3B, H-1/2B, H-1/4B, TL Only • Threshold: 0 – 100%, 11 steps
• Customizable templates
• Rejection: 6 steps • Accumulation mode: 7 levels
Scanning Parameters Anatomical M-Mode • Sample Volume Control
• Digital Beamformer • Flash Suppression
• M-mode cursor adjustable at any
• 478,405 Channels
• Frame Rate: 2399Hz Maximum
plane Digital Power Doppler
• Can be activated from a Cine loop
• Displayed Imaging Depth: 0 – 33
from a live or stored image Imaging
cm • Adjustable:
• M and A capability
• Minimum Depth of Field: 0 – 2 cm • Acoustic Power: 0 - 100%, 25
• Available with Color Flow Mode
(Zoom) (probe dependent) steps
• Curved Anatomical M-Mode
• Maximum Depth of Field: 0 – 33 • Color Map: 16 maps
cm (probe dependent) • Gain: 0 – 40 dB, 81 steps
• Wall Filter: 0 – 3, 4 steps
• Packet Size: 5 – 24, 9 steps

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 4 of 14


• Line Density: 5 steps • Line Density: 5 steps • Auto fit on monitor
• Spatial Filter: 6 steps • Edge Enhance: 7 steps • Measurements in B-Mode
• Steering Angle: 0 – 20 degrees • Frame Average: 0 – 7, 8 steps • Up to 60cm scan length
• Frame Average: 0 – 6, 7 steps • Gray Scale Map: 8 maps
• Threshold: 0 – 100%, 11 steps • Tint Map :9maps 3D
• Accumulation mode:7 levels • Dynamic Range: 36 – 96 dB, 16 • Allows unlimited rotation and
• Sample Volume Control steps planar translations
• Flash Suppression: 0 – 4, 5 steps • Rejection: 6 steps • 3D reconstruction from Cine sweep
• Gain: 0 – 90 dB range, 1dB steps
Continuous Wave • B-Flow Color Advanced 3D
Doppler (Option) • Accumulation: 7 levels • Acquisition of Color data
• Available on M5S-D, 3Sp-D, 6S-D, • Automatic rendering
S4-10-D, 6Tc-RS, P2D, P6D
B Steer+ (Option) • 3D Landscape technology
• Steerable CW mode includes • Available on 9L-D, 11L-D, ML6-15-D, • 3D Movie
Adjustable: L8-18i-D
• Acoustic Power: 0 - 100%, 25
Real Time 4D (Option)
steps
Coded Contrast • Acquisition Modes:
• Gain: 0 – 85dB range, 1dB steps Imaging (Option) • Real Time 4D
• Gray Scale Map: 8 maps • Static 3D
• Available on 3CRF-D, C1-5D, C2-9-D,
• Transmit Frequency: 1 or 2 steps • Visualization Modes:
C2-6b-D, IC5-9-D, 9L-D, 11L-D, ML6-
• Wall Filter: 18 settings, Min:5.5Hz, • 3D Rendering (diverse surface
15-D, M5S-D, L8-18i-D, RAB6-D
Max:5095Hz and intensity projection modes)
• 2 Contrast Timers
• CW Colorization: 6 maps • Sectional Planes (3 Section planes
• Timed Updates: 0.05 – 10 seconds
• Velocity Scale Range: 20 – 1116 perpendicular to each other)
• Accumulation mode, six levels
cm/s (2000cm/s :P2D) • Volume Contrast Imaging-Static
• Maximum Enhance Mode
• Sweep Speed: 0 – 7, 8 steps (option)
• Flash
• Angle Correction: -90 - 90 • Tomographic Ultrasound Imaging
• Time Intensity Curve (TIC) Analysis
degrees, 1 degree steps (option)
• Auto MI control
• Spectrum Inversion • Render Mode:
• The LOGIQ S8 is designed for
• Trace Method • Surface Texture, Surface Smooth,
compatibility with commercially
• Baseline Shift: 5 – 95%, 11 steps max-, min- and X-ray (average
available ultrasound contrast
• Doppler Auto Trace intensity projection), mix mode of
agents. Because the availability of
• Compression: 0.5 – 2.4, 9 steps two render modes
these agents is subject to
• Trace Direction • Curved 3 point Render start
government regulation and
• Trace Sensitivity • 3D Movie
approval, product features intended
• Scalpel: 3D Cut tool
for use with these agents may not
Automatic be commercially marketed nor
• Display Format:
• Quad: A-/B-/C-Plane/3D
Optimization made available before the contrast
• Dual: A-Plane/3D
• Optimize B-Mode image to help agent is cleared for use. Contrast
• Single: 3D or A- or B- or C-Plane
improve contrast resolution. related product features are
• Automated Volume Calculation -
• Selectable amount of contrast enabled only on systems for
VOCAL II (option)
resolution enhancement (low, delivery to an authorized country or
• Betaview
medium, high) region of use.
• Auto Sweep
• Auto-Spectral Optimize – adjusts
baseline, invert, PRF (on live image), LOGIQ View (Option) Volume Navigation
and angle correction • Extended Field of View Imaging
• Available on the following probes: (Option)
Coded Harmonic 9L-D, 11L-D, ML6-15-D, L8-18i-D, • Available on C1-5-D, C2-9-D, C2-6b-
3CRF-D, C1-5D, C2-9-D, C2-6b-D, D, 9L-D, ML6-15-D, 3CRF-D, IC5-9-D,
Imaging 10C-D, IC5-9-D, S1-5-D, M5S-D, S1-5-D, S4-10-D, M5S-D, L8-18i-D
• Available on all 2D and 4D probes 3Sp-D, 6S-D, S4-10-D, RAB6-D, • Sensor-based acquisition
RIC5-9-D, 6Tc-RS • Position Markers
B-Flow (Option) • For use in B-Mode • Needle tip tracking
• Available on C1-5-D, C2-9-D, 9L-D, • CrossXBeam is available on linear • Virtual tracking
11L-D, ML6-15-D, M5S-D, S1-5-D, probes • Tru3D feature includes:
L8-18i-D, 10C-D • Auto detection of scan direction • Display of data in: Main-, Parallel-,
• Background: On/Off • Pre or post-process zoom Angular-Mode
• Sensitivity/PRI : 1 – 50, 17 steps • Rotation

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 5 of 14


• Render Modes: Gray Surface, Elastography • Compatible with CrossXBeam
Texture, Min-, Max-, Average- • Available on all linear and sector
Intensity Quantification (Option transducers 9L-D, 11L-D, ML6-15-D,
• Measurements: distance, angle, – not available in the L8-18i-D, S1-5-D, M5S-D, 3Sp-D, 6S-
area, volume D, S4-10-D, 6Tc-RS
• 3D Movie United States)
• Relative quantification tool SRI-HD
Scan Assistant • Available on ML6-15-D, 11L-D, 9L-D, • Speckle Reduction Imaging
IC5-9-D, C1-5-D, C2-9-D • Provides multiple levels of speckle
(Option)
reduction
• Factory Programs Quantitative Flow • Compatible with Side by Side
• User –defined programs
• Steps include image annotations, Analysis (Option) DualView Display
• Available in Color and Power • Compatible with ALL linear, convex
mode transitions, basic imaging
Doppler and sector transducers
controls and measurement
• Compatible w/ B-Mode, Color,
initiation
TVI (Option) Contrast Agent and 3D imaging
Compare Assistant • Myocardial Doppler Imaging with
CrossXBeam
color overlay on tissue image
(Option) • Available on the sector probes • Provides 3,5,7, or 9 angles of spatial
• Allows side-by-side comparison of compounding
• Tissue color overlay can be
previous ultrasound and other • Live Side by Side DualView Display
removed to show just the 2D image,
modality exams during live • Compatible with:
still retaining the tissue velocity
scanning • Color Mode
information
• Curved Anatomical M-mode: free • PW
Power Assistant (curved) drawing of M-mode • SRI-HD
(Option) generated from the cursor • Coded Harmonic Imaging
independent from the axial plane • Virtual Convex
• Allows moving the system without
• Q-Analysis: Multiple Time -Motion • Available on 9L-D, 11L-D, ML6-15-D,
a complete system shutdown and
trace display from selected points L8-18i-D, 3CRF-D, C1-5D, C2-9-D,
boot-up power cycle.
in the myocardium C2-6b-D, 10C-D, IC5-9-D, RAB6-D,
Breast Productivity RIC5-9-D

Package (Option) Stress Echo (Option) Controls Available


• Advanced and flexible stress-echo
• Allows automatic contour and While “Live”
examination capabilities
measurement of breast lesions
• Provides exercise and • Write Zoom
• Worksheet summary
pharmacological protocol • B/M/CrossXBeam-Mode
• Feature Assessment
templates • Gain
• BI_RADS Assessment
• 8 default templates • TGC
• User editable
• Template editor for user • Dynamic Range
configuration of existing templates • Acoustic Output
Thyroid Productivity or creation of new templates • Transmission Focus Position
Package (Option) • Reference scan display during • Transmission Focus Number
• Worksheet summary includes acquisition for stress level • Line Density Control
measurements and locations for comparison (dual screen) • Sweep Speed for M-Mode
nodule, parathyroid and lymph • Baseline level/Previous level • Number of Angles for
node selectable CrossXBeam
• Feature Assessment • Raw data continuous capture • PW-Mode
• User editable • Over 100 sec available • Gain
• Wall motion scoring (bulls-eye and • Dynamic Range
Elastography (Option) segmental) • Acoustic Output
• Available on ML6-15-D, 11L-D, 9L- • Smart stress: Automatically set up • Transmission Frequency
D, C2-9-D, C1-5-D, IC5-9-D various scanning parameters (for • PRF
instance geometry, frequency, gain, • Wall Filter
etc.) according to same projection • Spectral Averaging
on previous level • Sample Volume Gate
• Length
Virtual Convex • Depth
• Provides a convex field of view • Velocity Scale

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 6 of 14


• Time Resolution • Post Gain • TAMAX (Time Averaged Maximum
• Color Flow Mode • Change display – single, dual, Velocity)
• CFM Gain quad sectional or rendered • PVAL (Peak Velocity Value)
• CFM Velocity Range • Volume Flow (TAMEAN and Vessel
• Acoustic Output Measurements / Area)
• Wall Echo Filter Calculations
• Packet Size OB Measurements /
• Frame Rate Control General B-Mode
• CFM Spatial Filter
Calculations
• Depth and Distance • Gestational Age by:
• CFM Frame Averaging
• Circumference (Ellipse / Trace) • GS (Gestational Sac)
• CFM Line Resolution
• Area (Ellipse / Trace) • CRL (Crown Rump Length)
• Frequency / Velocity Base Line
• Volume (Ellipsoid) • FL (Femur Length)
Shift
• % Stenosis (Area or Diameter) • BPD (Biparietal Diameter)
Controls Available on • Angle between two lines • AC (Abdominal Circumference)
• HC (Head Circumference)
“Freeze” or Recall General M-Mode • APTD x TTD (Anterior/Posterior
• Automatic Optimization • M-Depth Trunk Diameter by Transverse
• SRI-HD • Distance Trunk Diameter)
• CrossXBeam – Display non- • Time • FTA (Fetal Trunk Cross-sectional
compounded and compounded • Slope Area)
image simultaneously in split • Heart Rate • HL (Humerus Length)
screen • BD (Binocular Distance)
• 3D reconstruction from a stored General Doppler • FT (Foot Length)
Cine loop • OFD (Occipital Frontal Diameter)
• B/M/CrossXBeam Mode
Measurements/ • TAD (Transverse Abdominal
• Gray Map Optimization Calculations Diameter)
• TGC • Velocity • TCD (Transverse Cerebellum
• Colorized B and M • Time Diameter)
• Frame Average (loops only) • A/B Ratio (Velocities / Frequency • THD (Thorax Transverse Diameter)
• Dynamic Range Ratio • TIB (Tibia Length)
• Anatomical M Mode • PS (Peak Systole) • ULNA (Ulna Length)
• Max Read Zoom to 8x • ED (End Diastole) • Estimated Fetal Weight (EFW) by:
• Base Line Shift • PS/ED (PS/ED Ratio) • AC, BPD
• Sweep Speed • ED/PS (ED/PS Ratio) • AC, BPD, FL
• PW Mode • AT (Acceleration Time) • AC, BPD, FL, HC
• Gray Map • ACCEL (Acceleration) • AC, FL
• Post Gain • TAMAX (Time Averaged Maximum • AC, FL, HC
• Baseline shift Velocity • AC, HC
• Sweep Speed • Volume Flow (TAMEAN and Vessel • BPD, APTD, TTD, FL
• Invert Spectral wave form Area) • BPD, APTD, TTD, SL
• Compression • Heart Rate • Calculations and Ratios
• Rejection • PI (Pulsatility Index) • FL/BPD
• Colorized Spectrum • RI (Resistivity Index) • FL/AC
• Display Format • FL/HC
• Doppler Audio Real-time Doppler • HC/AC
• Angle Correct • CI (Cephalic Index)
• Quick Angle Correct Auto Measurements / • AFI (Amniotic Fluid Index)
• Auto Angle Correct Calculations • CTAR(Cardio-Thoracic Area Ratio)
• Color Flow • PS (Peak Systole) • Measurements / Calculations by:
• Overall Gain (loops and stills) • ED (End Diastole) ASUM, ASUM 2001, Berkowitz,
• Color Map • MD (Minimum Diastole) Bertagnoli, Brenner, Campbell, CFEF,
• Transparency Map • PI (Pulsatility Index) Chitty, Eik-Nes, Ericksen, Goldstein,
• Frame Averaging (loops only) • RI (Resistivity Index) Hadlock, Hansmann, Hellman, Hill,
• Flash Suppression Hohler, Jeanty, JSUM, Kurtz,
• AT (Acceleration Time)
• CFM Display Threshold Mayden, Mercer, Merz, Moore,
• ACC (Acceleration)
• Spectral Invert for Color/Doppler Nelson, Osaka University, Paris,
• PS/ED (PS/ED Ratio)
• Anatomical M-Mode on cine loop Rempen, Robinson, Shepard,
• ED/PS (ED/PS Ratio)
• 4D Shepard/Warsoff, Tokyo University,
• HR (Heart Rate)
• Gray Map, Colorize Tokyo/Shinozuka, Yarkoni

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 7 of 14


• Fetal Graphical Trending • SYS MICA (Systolic Mid Internal • Band Width: 1 – 6MHz
• Growth Percentiles Carotid Artery) • Number of Elements: 192
• Multi-Gestational Calculations (4) • DIAS MICA (Diastolic Mid Internal • Field of View (Max): 70 degree
• Fetal Qualitative Description Carotid Artery) • Physical Foot Print: 66 x 18mm
(Anatomical survey) • SYS PICA (Systolic Proximal Internal • B-mode Frequency: 2, 3, 4, 5MHz
• Fetal Environmental Description Carotid Artery) • Harmonic Frequency: 3, 4, 5MHz
(Biophysical profile) • DIAS PICA (Diastolic Proximal • PW Doppler Frequency: 1.8, 2.1, 2.5,
• Programmable OB Tables Internal Carotid Artery) 3.6MHz
• Over 20 selectable OB Calcs • SYS DECA (Systolic Distal External • Color Doppler Frequency: 1.8, 1.9,
• Expanded Worksheets Carotid Artery) 2.5, 3.1, 3.6MHz
• DIAS DECA (Diastolic Distal External
OB Measure Assistant Carotid Artery) C2-9-D Convex Probe
(Option) • SYS PECA (Systolic Proximal External • Applications: Abdomen, Pediatrics,
Carotid Artery) OB, Gynecology, Neonatal,
• Allows automatic contour and
• DIAS PECA (Diastolic Proximal Vascular
measurement of BPD, HC, FL and
External Carotid Artery) • Biopsy Guide: Multi-Angle,
AC
• VERT (Systolic Vertebral Velocity) disposable with a reusable bracket
• User editable
• SUBCLAV (Systolic Subclavian (H4913BA)
Velocity) • Band Width: 2 – 9 MHz
Breast Measure • Automatic IMT (Option) • Number of Elements: 192
Assistant (Option) • Summary Reports • Field of View (Max): 65 degree
• Allows automatic contour and • Physical Foot Print: 52 x 9 mm
measurement of breast lesions Auto EF (Option) • B-mode Frequency: 2.5, 4.5, 6, 7.5,
• User editable • Allows semi-automatic 9MHz
measurement of the global EF • Harmonic Frequency: 5, 6, 7, 8,
GYN Measurements/ (Ejection fraction) 9MHz
Calculations • User editable • PW Doppler Frequency: 2.8, 3.1, 3.6,
4.2, 5MHz
• Right Ovary Length, Width, Height Urological Calcs • Color Doppler Frequency: 2.8, 3.1,
• Left Ovary Length, Width, Height
• Bladder Volume 3.6, 4.2, 5MHz
• Uterus Length, Width, Height
• Prostate Volume
• Cervix Length, Trace
• Lt/Rt Renal Volume C2-6b-D Biopsy Convex Probe
• Ovarian Volume
• Generic Volume • Applications: Abdomen, Intervention
• ENDO (Endometrial thickness)
• Post-Void Bladder Volume • Biopsy Guide: Multi-Angle,
• Ovarian RI
disposable (H46332LY)
• Uterine RI Probes (All Optional) • Band Width: 2 – 6MHz
• Follicular measurements
3CRF-D Micro Convex Biopsy Probe • Number of Elements: 128
• Summary Reports
• Applications: Abdomen • Field of View (Max): 61degree
• Biopsy Guide: Single-Angle, • Physical Foot Print: 65 (Normal) x
Vascular 13mm
disposable with a reusable bracket
Measurements/ (H40442LR), Multi-Angle with a • B-mode Frequency: 2, 3, 4, 5MHz
Calculations reusable bracket (H40452LP) • Harmonic Frequency: 3.5, 4, 5MHz
• Band Width: 1 - 6MHz • PW Doppler Frequency: 2.5, 3, 3.6,
• SYS DCCA (Systolic Distal Common
• Number of Elements: 128 4.2MHz
Carotid Artery)
• Field of View (Max): 80 degree • Color Doppler Frequency: 2.5, 3.1,
• DIAS DCCA (Diastolic Distal
• Physical Foot Print: 29 x 12 mm 3.6, 4.2MHz
Common Carotid Artery)
• B-mode Frequency: 3, 3.5, 4MHz
• SYS MCCA (Systolic Mid Common
• Harmonic Frequency: 3.6, 4, 4.4MHz 10C-D Micro convex Probe
Carotid Artery)
• PW Doppler Frequency: 2.1, 2.5, • Applications: Neonatal, Pediatrics,
• DIAS MCCA (Diastolic Mid Common
3.6MHz Vascular
Carotid Artery)
• Color Doppler Frequency: 2.1, 2.5, • Band Width: 4 – 12MHz
• SYS PCCA (Systolic Proximal
3.6MHz • Number of Elements: 128
Common Carotid Artery)
• Field of View (Max): 102degree
• DIAS PCCA (Diastolic Proximal
C1-5-D Convex Probe • Physical Foot Print: 28 x 13mm
Common Carotid Artery)
• Applications: Abdomen, OB, • B-mode Frequency: 4, 6, 8, 10MHz
• SYS DICA (Systolic Distal Internal
Gynecology, Urology, Vascular • Harmonic Frequency: 7, 8, 9, 10MHz
Carotid Artery)
• Biopsy Guide: Multi-Angle, • PW Doppler Frequency: 4.2, 5, 6.3,
• DIAS DICA (Systolic Distal Internal
disposable with a reusable bracket 8.3MHz
Carotid Artery)
(H40432LE)

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 8 of 14


• Color Doppler Frequency: 4.2, 5, • Number of Elements: 288 • PW Doppler Frequency: 3.1, 3.6, 4.2,
6.3, 8.3MHz • Field of View (Max): 120degree 5MHz
• Physical Foot Print: 28 x 17mm • Color Doppler Frequency: 3.1, 3.6,
IC5-9-D Micro convex Probe • B-mode Frequency: 2, 2.5, 3.5, 4.2, 5, 6.3MHz
• Applications: OB, Gynecology, 4.5MHz
Urological • Harmonic Frequency: 2.5, 3, 3.2, 3.3, 11L-D Linear Probe
• Biopsy Guide: Single Angle, 3.5, 3.7, 4, 4.5, 4.6MHz • Applications: Vascular, Small Parts,
Disposable with a disposable • PW Doppler Frequency: 1.8, 2.1, 2.5, Pediatrics, Neonatal
bracket (E8385MJ) or reusable 3.1, 3.6MHz • Biopsy Guide Available: Multi- Angle,
bracket (H40412LN) • Color Doppler Frequency: 1.8, 2.1, disposable with a reusable bracket
• Band Width: 3 – 10MHz 2.5, 3.1, 4.2MHz (H40432LC)
• Number of Elements: 192 • CWD Doppler Frequency: 1.9, • Band Width: 3 – 11MHz
• Field of View (Max): 145degree 2.5MHz • Number of Elements: 192
• Physical Foot Print: 26 x 6mm • Field of View (Max): 38mm
• B-mode Frequency: 4, 5, 6, 7, 8, 3Sp-D Sector Probe • Physical Foot Print: 44 x 7mm
9MHz • Applications: Cardiac, Transcranial • B-mode Frequency: 7, 8, 9, 11MHz
• Harmonic Frequency: 7, 8, 9MHz • Band Width: 1 – 6MHz • Harmonic Frequency: 9, 11, 12MHz
• PW Doppler Frequency: 3.6, 4.2, 5, • Number of Elements: 64 • PW Doppler Frequency: 4.2, 5, 6.3,
6.3MHz • Field of View (Max): 120degree 8.3MHz
• Color Doppler Frequency: 4.2, 6.3, • Physical Foot Print: 15 x 14mm • Color Doppler Frequency: 4.2, 5, 6.3,
8.3MHz • B-mode Frequency: 2, 3, 4, 5MHz 8.3MHz
• Harmonic Frequency: 3, 3.2, 3.3, 3.5,
S1-5-D Sector Probe 3.7, 4, 4.5, 4.6MHz ML6-15-D Matrix Array Linear Probe
• Applications: Abdomen, OB, • PW Doppler Frequency: 1.8, 2.1, 2.5, • Applications: Small Parts, Vascular,
Gynecology, Vascular 3.1, 4.2MHz Neonatal, Pediatrics
• Biopsy Guide: Multi-angle, • Color Doppler Frequency: 1.6, 1.8, • Biopsy Guide Available: Multi-Angle,
disposable with a reusable 2.1, 2.5, 3.1, 4.2MHz disposable with a reusable bracket
bracket (H4908SD) • CWD Doppler Frequency: 1.9, (H40432LJ)
• Band Width: 1 – 6 MHz 2.5MHz • Band Width: 4 – 15MHz
• Number of Elements: 128 • Number of Elements: >1008
• Field of View (Max): 90degree 6S-D Sector Probe • Field of View (Max): 50mm
• Physical Foot Print: 28 x 14mm • Applications: Cardiac, Pediatrics, • Physical Foot Print: 50 x 6mm
• B-mode Frequency: 2, 3, 4MHz Neonatal • B-mode Frequency: 9, 11, 13,
• Harmonic Frequency: 2.5, 3, 4, 4.5, • Band Width: 2 – 8 MHz 15MHz
5MHz • Number of Elements: 128 • Harmonic Frequency: 8, 9, 12,
• PW Doppler Frequency: 1.8, 2.1, • Field of View (Max): 90degree 15MHz
2.5, 3.4MHz • Physical Foot Print: 20 x 15mm • PW Doppler Frequency: 5, 6.3,
• Color Doppler Frequency: 1.8, 1.9, • B-mode Frequency: 4, 5, 6.5, 8MHz 8.3MHz
2.1, 2.5MHz • Harmonic Frequency: 4.8, 5.4, • Color Doppler Frequency: 5, 7.7, 10,
6.2MHz 12.5MHz
S4-10-D Sector Probe • PW Doppler Frequency: 2.8, 3.1, 3.6,
• Applications: Pediatrics, Neonatal 5MHz L8-18i-D Linear Probe
• Band Width: 3 - 9MHz • Color Doppler Frequency: 2.8, 3.1, • Applications: Small Parts, Vascular,
• Number of Elements: 128 3.6, 5MHz Intraoperative
• Field of View (Max): 90degree • CWD Doppler Frequency: 2.8, 3.6, • Band Width: 4 – 15MHz
• Physical Foot Print: 20 x 15mm 5MHz • Number of Elements: 168
• B-mode Frequency: 4, 6, 8, 10MHz • Field of View (Max): 25mm
• Harmonic Frequency: 5, 6, 7, 8.5, 9L-D Linear Probe • Physical Foot Print: 35 x 10mm
10MHz • Applications: Vascular, Small Parts, • B-mode Frequency: 8, 9, 12, 15,
• PW Doppler Frequency: 3.6, 4.2, 5, Pediatrics, Abdomen 18MHz
6.3MHz • Biopsy Guide: Multi-angle, • Harmonic Frequency: 9, 15, 18MHz
• Color Doppler Frequency: 5, 6.3, disposable with a reusable bracket • PW Doppler Frequency: 5, 6.3,
8.3MHz (H4906BK) 8.3MHz
• CWD Doppler Frequency: 4.2, • Band Width: 2 – 8MHz • Color Doppler Frequency: 5, 7.7, 10,
5MHz • Number of Elements: 192 12.5MHz
• Field of View (Max): 44mm
M5S-D Sector Probe • Physical Foot Print: 49 x 9mm RAB6-D Convex Volume Probe
• Applications: Cardiac, Transcranial • B-mode Frequency: 5, 6, 7, 9MHz • Applications: Abdomen, OB,
• Band Width: 1 – 5MHz • Harmonic Frequency: 8.5, 10MHz Gynecology, Pediatrics

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 9 of 14


• Biopsy Guide: Multi Angle, • CWD Doppler Frequency: 6.3MHz acoustic output indices on
disposable with a reusable diagnostic ultrasound equipment
bracket (H48681ML) External Inputs and (MI, TIS, TIB, TIC)
• Band Width: 2 – 8 MHz Outputs (not including • EMC Emissions Group 1 Class B
• Number of Elements: 192 device requirements as per Sub
• Field of View (Max): 80degree on-board peripherals) clause 4.2 of CISPR 11 (Group 1
• Physical Foot Print: 62 x 34mm • DVI-D signal with HDMI connector Class A when 6S-D is connected)
• B-mode Frequency: 2, 4, 6, 8MHz • Ethernet
• Harmonic Frequency: 5, 6.5, 8MHz • Multiple USB 2.0 ports
• PW Doppler Frequency: 2.5, 3, 3.6,
4.2MHz Safety Conformance
• Color Doppler Frequency: 2.5, 3,
3.6, 4.2, 5MHz The LOGIQ S8 is:
• Classified to UL 60601-1 by a
RIC5-9-D Convex Volume Probe Nationally Recognized Test Lab
• Applications: OB, Gynecology, • Certified to CAN/CSA-C22.2
Urology No.601.1-M90 by an SCC accredited
• Biopsy Guide: Single Angle, Test Lab
Reusable (H46721R) • CE Marked to Council Directive
• Band Width: 3 – 10MHz 93/42/EEC on Medical Devices
• Number of Elements: 192 • Conforms to the following
• Field of View (Max): 145degree standards for safety:
• Physical Foot Print: 32 x 27mm • EN 60601-1 Medical electrical
• B-mode Frequency: 4, 5, 6, 7, 8, equipment –Part 1: General
9MHz requirements for safety
• Harmonic Frequency: 7, 8, 9MHz • EN 60601-1-1 Medical electrical
• PW Doppler Frequency: 3.6, 4.2, 5, equipment – Part 1-1 General
6.3MHz requirements for safety – Collateral
• Color Doppler Frequency: 4.2, 6.3, Standard: Safety requirements for
8.3MHz medical electrical systems
• EN 60601-1-2 Medial electrical
6Tc-RS Trans-esophageal Probe equipment –Part 1-2 General
• Applications: Cardiac requirements for safety – Collateral
• TEE RS-DLP Adapter (H46352LK) Standard: Electromagnetic
required compatibility – requirements and
• Band Width: 3 – 8MHz tests
• Number of Elements: 64 • EN 60601-1-4 Medical electrical
• Field of View (Max): 90degree equipment Part 1-4 General
• Physical Foot Print: 37 x 13 x requirements for safety – Collateral
10mm Standard: programmable electrical
• B-mode Frequency: 6, 7, 8MHz medical systems
• Harmonic Frequency: 6MHz • EN 60601-1-6 Medical electrical
• PW Doppler Frequency: 2.8, 3.1, equipment Part 1-6 General
3.6, 4.2, 5MHz requirements for basic safety and
• Color Doppler Frequency: 3.1, 4.2, essential performance – Collateral
5MHz Standard: Usability
• CWD Doppler Frequency: 3.1, 3.6, • EN 60601-2-37 Medical electrical
4.2MHz equipment – Part 2-37: Particular
requirements for the safety of
P2D CW Split Crystal Probe ultrasonic medical diagnostic and
• Applications: Cardiac, Vascular monitoring equipment
• Number of Elements: 2 • ISO 10993-1 Biological evaluation
• Physical Foot Print: 17mm dia of medical devices – Part 1
• CWD Doppler Frequency: 2.1MHz Evaluation and testing
• NEMA UD2 Acoustic output
P6D CW Split Crystal Probe measurement standard for
• Applications: Cardiac, Vascular diagnostic ultrasound equipment
• Number of Elements: 2 • NEMA UD3 Standard for real time
• Physical Foot Print: 8mm dia display of thermal and mechanical

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 10 of 14


• Left Ventricle Posterior Wall • Right Ventricle Wall Thickness
Supplement: Fractional Shortening (LVPWd, (RVAWd, RVAWs)
Cardiac LVPWs) • Right Ventricle Outflow Tract
• Left Ventricle Stroke Index, Diameter (RVOT Diam)
Measurements/Calculations Teichholz/Cube (LVIDd, LVIDs and • Left Pulmonary Artery (LPA)
Body Surface Area) • Main Pulmonary Artery (MPA)
B-Mode Measurements • Left Ventricle Fractional • Right Pulmonary Artery (RPA)
• Aorta Shortening (LVIDd, LVIDs) • System
• Aortic Root Diameter (Ao Root • Left Ventricle Stroke Volume, • Inferior Vena Cava
Diam) Teichholz/Cubic (LVIDd, LVIDs) • Systemic Vein Diameter (Systemic
• Aortic Arch Diameter (Ao Arch • Left Ventricle Stroke Index, Single Diam)
Diam) Plane, Two Chamber, Method of • Patent Ductus Arterosis Diameter
• Ascending Aortic Diameter (Ao Disk (LVI Dd, LVIDs, LVSd, LVSs) (PDA Diam)
Asc) • Left Ventricle Stroke Index, Single • Pericard Effusion (PEs)
• Descending Aortic Diameter (Ao Plane, Four Chamber, Method of • Patent Foramen Ovale Diameter
Desc Diam) Disk (LVI Dd, LVIDs, LVSd, LVSs) (PFO Diam)
• Aorta Isthmus (Ao Isthmus) • Left Ventricle Stroke Index, Bi- • Ventricular Septal Defect
• Aorta (Ao st junct) Plane, Bullet, Method of Disk Diameter (VSD Diam)
• Aortic Valve (LVAd, LVAs) • Interventricular Septum (IVS)
• Aortic Valve Cusp Separation (AV • Interventricular Septum (IVS) Fractional Shortening (IVSd, IVSs)
Cusp) • Left Ventricle Internal Diameter • Tricuspid Valve
• Aortic Valve Area Planimetry (LVI D) • Tricuspid Valve Area (TV
(AVA Planimetry) • Left Ventricle Posterior Wall Panimetry)
• (Trans AVA) Thickness (LVPW) • Tricuspid Valve Annulus Diameter
• Left Atrium • Mitral Valve (TV Annulus Diam)
• Left Atrium Diameter (LA Diam) • Mitral Valve Annulus Diameter
• LA Length (LA Major) (MV Ann Diam) M-Mode Measurements
• LA Width (LA Minor) • E-Point-to-Septum Separation • Aorta
• Left Atrium Diameter to AoRoot (EPSS) • Aortic Root Diameter (Ao Root
Diameter Ratio (LA/Ao Ratio) • Mitral Valve Area Planimetry (MVA Diam)
• Left Atrium Area (LAA(d), LAA(s)) Planimetry) • Aortic Valve
• Left Atrium Volume, Single Plane, • Pulmonic Valve • Aortic Valve Diameter (AV Diam)
Method of Disk (LAEDV A2C, • Pulmonic Valve Area (PV • Aortic Valve Cusp Separation (AV
LAESV A2C) (LAEDV A4C, LAESV Planimetry) Cusp)
A4C) • Pulmonic Valve Annulus Diameter • Aortic Valve Ejection Time (LVET)
• Left Ventricle (PV Annulus Diam) • Left Atrium
• Left Ventricle Mass (LVPWd, • Pulmonic Diameter (Pulmonic • Left Atrium Diameter to AoRoot
LVPWs) Diam) Diameter Ratio (LA/Ao Ratio)
• Left Ventricle Volume, • Right Atrium • Left Atrium Diameter (LA Diam)
Teichholz/Cubic (LVIDd, LVI Ds) • Right Atrium Diameter, Length • Left Ventricle
• Left Ventricle Internal Diameter (RAD Ma) • Left Ventricle Volume,
(LVIDd, LVI Ds) • Right Atrium Diameter, Width Teichholz/Cubic (LVIDd, LVI Ds)
• Left Ventricle Length (LVLd, LVLs) (RAD Mi) • Left Ventricle Internal Diameter
• Left Ventricle Outflow Tract • Right Atrium Area (RAA) (LVIDd, LVI Ds)
Diameter (LVOT Diam) • Right Atrium Volume, Single Plane, • Left Ventricle Posterior Wall
• Left Ventricle Posterior Wall Method of Disk (RAAd) Thickness (LVPWd, LVPWs)
Thickness (LVPWd, LVPWs) • Right Atrium Volume, Systolic, • Left Ventricle Ejection Time (LVET)
• Left Ventricle Length (LV Major) Single Plane, Method of Disk • Left Ventricle Pre-Ejection Period
• Left Ventricle Width (LV Minor) (RAAs) (LVPEP)
• Left Ventricle Outflow Tract Area • Right Ventricle • Interventricular Septum (IVS)
(LVOT) • Right Ventricle Outflow Tract Area • Left Ventricle Internal Diameter
• Left Ventricle Area, Two (RVOT Planimetry) (LVI D)
Chamber/Four Chamber/Short • Left Pulmonary Artery Area (LPA • Left Ventricle Posterior Wall
Axis (LVA (d), LVA (s)) Area) Thickness (LVPW)
• Left Ventricle Endocardial Area, • Right Pulmonary Artery Area (RPA
Width (LVA (d), LVA(s)) • Mitral Valve
Area) • E-Point-to-Septum Separation
• Left Ventricle Epicardial Area, • Right Ventricle Internal Diameter
Length (LVAepi (d), LVAepi (s)) (EPSS)
(RVIDd, RVIDs) • Mitral Valve Leaflet Separation (D-
• Left Ventricle Mass Index • Right Ventricle Diameter, Length
(LVPWd, LVPWs) E Excursion)
(RVD Ma) • Mitral Valve Anterior Leaflet
• Ejection Fraction, • Right Ventricle Diameter, Width
Teichholz/Cube (LVIDd, LVIDs) Excursion (D-E Excursion)
(RVD Mi)

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 11 of 14


• Mitral Valve D-E Slope (D-E • Aortic Valve Acceleration Time (AV • Mitral Valve Deceleration Time
Slope) Acc Ti me) (MV Dec Time)
• Mitral Valve E-F Slope (E-F Slope) • Aortic Valve Deceleration TIme • Mitral Valve Ejection Time ((MV
• Pulmonic Valve (AV Trace) Trace)
• QRS complex to end of envelope • Aortic Valve Ejection Time (AVET) • Mitral Valve A-Wave Duration (MV
(Q-to-PV close) • Aortic Valve Acceleration to A Dur)
• Right Ventricle Ejection Time Ratio (AV Acc Time, • Mitral Valve Time to Peak (MV
• Right Ventricle Internal Diameter AVET) Trace)
(RVIDd, RVIDs) • Aortic Valve Area according to • Mitral Valve Acceleration
• Right Ventricle Wall Thickness PHT Time/Deceleration Time Ratio
(RVAWd, RVAWs) • Left Ventricle (MVAcc/Dec Time)
• Right Ventricle Outflow Tract • Left Ventricle Outflow Tract Peak • Stroke Volume Index by Mitral
Diameter (RVOT Diam) Pressure Gradient (VLOT Vmax) Flow (MVA Planimetry, MVTrace)
• Right Ventricle Ejection Time • Left Ventricle Outflow Tract Peak • Pulmonic Valve
(RVET) Velocity (LVOT Vmax) • Pulmonic Insufficiency Peak
• Right Ventricle Pre-Ejection • Left Ventricle Outflow Tract Mean Pressure Gradient (PR Vmax)
Period (RVPEP) Pressure Gradient (LVOT Trace) • Pulmonic Insufficiency End-
• System • Left Ventricle Outflow Tract Mean Diastolic Pressure Gradient
• Pericard Effusion (PE (d)) Velocity (LVOT Trace) (PRTrace)
• Tricuspid Valve • Left Ventricle Outflow Tract • Pulmonic Valve Peak Pressure
• QRS complex to end of envelope Velocity Time Integral (LVOT Gradient (PV Vmax)
(Q-to-TV close) Trace) • Pulmonic Insufficiency Peak
• Left Ventricle Ejection Time (LVET) Velocity (PR Vmax)
Doppler Mode Measurements • Mitral Valve • Pulmonic Insufficiency End-
• Aortic Valve • Mitral Valve Regurgitant Flow Diastolic Velocity (Prend Vmax)
• Aortic Insufficiency Mean Acceleration (MR Trace) • Pulmonic Valve Peak Velocity (PV
Pressure Gradient (AR Trace) • Mitral Valve Regurgitant Mean Vmax)
• Aortic Insufficiency Peak Velocity (MR Trace) • Pulmonary Artery Diastolic
Pressure Gradient (AR Vmax) • Mitral Regurgitant Mean Pressure Pressure (PV Trace)
• Aortic Insufficiency End Diastole Gradient (MR Trace) • Pulmonic Insufficiency Mean
Pressure Gradient (AR Trace) • Mitral Regurgitant Velocity Time Pressure Gradient (PR Trace)
• Aortic Insufficiency Mean Integral (MR Trace) • Pulmonic Valve Mean Pressure
Velocity (AR Trace) • Mitral Valve Mean Velocity (MR Gradient (PV Trace)
• Aortic Insufficiency Velocity Time Trace) • Pulmonic Insufficiency Mean
Integral (AR Trace) • Mitral Valve Velocity Time Integral Square Root Velocity (PR Trace)
• Aortic Valve Mean Velocity (AV (MR Trace) • Pulmonic Insufficiency Velocity
Trace) • Mitral Valve Mean Pressure Time Integral (PR Trace)
• Aortic Valve Velocity Time Gradient (MR Trace) • Pulmonic Valve Mean Velocity (PV
Integral (AV Trace) • Mitral Regurgitant Peak Pressure Trace)
• Aortic Valve Mean Pressure Gradient (MR Vmax) • Pulmonic Valve Velocity Time
Gradient (AV Trace) • Mitral Valve Peak Pressure Integral (PV Trace)
• Aortic Valve Peak Pressure Gradient (MR Vmax) • Pulmonic Insufficiency Pressure
Gradient (AR Vmax) • Mitral Regurgitant Peak Velocity Half Time (PR PHT)
• Aortic Insufficiency Peak Velocity (MR Vmax) • Pulmonic Valve Flow Acceleration
(AR Vmax) • Mitral Valve Peak Velocity (MR (PV Acc Time)
• Aortic Insufficiency End-Diastolic Vmax) • Pulmonic Valve Acceleration Time
Velocity (AR Trace) • Mitral Valve Velocity Peak A (MV A (PV Acc Time)
• Aortic Valve Peak Velocity (AV Velocity) • Pulmonic Valve Ejection Time
Vmax) • Mitral Valve Velocity Peak E (MV E (PVET)
• Aortic Valve Peak Velocity at Velocity) • QRS complex to end of envelope
Point E (AV Vmax) • Mitral Valve Area according to (Q-to-PV close)
• Aorta Proximal Coarctation PHT (MV PHT) • Pulmonic Valve Acceleration to
(Coarc Pre-Duct) • Mitral Valve Flow Deceleration Ejection TIme Ratio (PV Acc Time,
• Aorta Distal Coarctation (Coarc (MV Trace) PVET)
Post-Duct) • Mitral Valve Pressure Half Time • Right Ventricle
• Aortic Valve Insufficiency (PV PHT) • Right Ventricle Outflow Tract Peak
Pressure Half Time (AR PHT) • Mitral Valve Flow Acceleration (MV Pressure Gradient (RVOT Vmax)
• Aortic Valve Flow Acceleration Trace) • Right Ventricle Outflow Tract Peak
(AV Trace) • Mitral Valve E-Peak to A-Peak Velocity (RVOT Vmax)
• Aortic Valve Pressure Half Time Ratio (A-C and D-E) (MV E/ARatio) • Right Ventricle Outflow Tract
(AV Trace) • Mitral Valve Acceleration Time Velocity Time Integral (RVOTTrace)
(MV Acc Time)

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 12 of 14


• Right Ventricle Ejection Time (RV • Tricuspid Valve A-Wave Duration Method of Disk(Simpson) (LVAd,
Trace) (TV A Dur) LVAs, HR)
• Stroke Volume by Pulmonic Flow • QRS complex to end of envelope • Cardiac Output Four Chamber,
(RVOT Planimetry, RVOTTrace) (Q-to-TV close) Single Plane, Area-Length/
• Right Ventricle Stroke Volume • Tricuspid Valve Pressure Half Method of Disk (Simpson) (LVAd,
Index by Pulmonic Flow (RVOT Time (TV PHT) LVAs, HR)
Planimetry, RVOT Trace) • Stroke Volume by Tricuspid Flow • Ejection Fraction Two Chamber,
• System (TV Planimetry, TV Trace) Single Plane, Area-Length/
• Pulmonary Artery Peak Velocity • Tricuspid Valve E-Peak to A-Peak Method of Disk (Simpson) (LVAd,
(PV Vmax) Ratio (TV E/A Velocity) LVAs)
• Pulmonary Vein Velocity Peak A • Ejection Fraction Four Chamber,
(reverse) (P Vein A) Color Flow Mode Measurements Single Plane, Area-Length/
• Pulmonary Vein Peak Velocity (P • Aortic Valve Method of Disk (Simpson) (LVAd,
Vein D, P Vein S) • Proximal Isovelocity Surface Area: LVAs)
• Systemic Vein Peak Velocity Regurgitant Orifice Area (AR • Left Ventricle Stroke Volume,
(PDA Diastolic, PDA Systolic) Radius) Single Plane, Two Chamber/Four
• Ventricular Septal Defect Peak • Proximal Isovelocity Surface Area: Chamber, Area-Length (LVAd,
Velocity (VSD Vmax) Radius of Aliased Point (AR LVAs)
• Atrial Septal Defect (ASD Radius) • Left Ventricle Stroke Volume,
Diastolic, ASD Systolic) • Proximal Isovelocity Surface Area: Single Plane, Two Chamber/Four
• Pulmonary Vein A-Wave Regurgitant Flow (AR Trace) Chamber, Method of Disk
Duration (P Vein A Dur) • Proximal Isovelocity Surface Area: (Simpson) (LVIDd, LVIDs, LVAd,
• IsoVolumetric Relaxation Time Regurgitant Volume Flow (AR LVAs)
(IVRT) Trace) • Left Ventricle Volume, Two
• IsoVolumetric Contraction Time • Proximal Isovelocity Surface Area: Chamber/Four Chamber, Area-
(IVCT) Aliased Velocity (AR Vmax) Length (LVAd, LVAs)
• Pulmonary Vein S/D Ratio (P Vein • Mitral Valve • Ejection Fraction, Bi-Plane,
D, P Vein S) • Proximal Isovelocity Surface Area: Method of Disk (LVAd, LVAs, 2CH,
• Ventricular Septal Defect Peak Regurgitant Orifice Area (MR 4CH)
Pressure Gradient (VSD Vmax) Radius) • Left Ventricle Stroke Volume, Bi-
• Pulmonic-to-Systemic Flow • Proximal Isovelocity Surface Area: Plane, Method of Disk (LVAd, LVAs,
Ratio (Qp/Qs) Radius of Aliased Point (MR 2CH, 4CH)
• Tricuspid Valve Radius) • Left Ventricle Volume, Bi-Plane,
• Tricuspid Regurgitant Peak • Proximal Isovelocity Surface Area: Method of Disk (LVAd, LVAs, 2CH,
Pressure Gradient (TR Vmax) Regurgitant Flow (MR Trace) 4CH)
• Tricuspid Valve Peak Pressure • Proximal Isovelocity Surface Area: • Left Ventricle Stroke Index, Single
Gradient (TV Vmax) Regurgitant Volume Flow (MR Plane, Two Chamber/Four
• Tricuspid Regurgitant Peak Trace) Chamber, Area-Length (LVSd,
Velocity (TR Vmax) • Proximal Isovelocity Surface Area: LVSs and BSA)
• Tricuspid Valve Peak Velocity (TV Aliased Velocity (MR Vmax) • Left Ventricle Volume, Single
Vmax) Plane, Two Chamber/Four
• Tricuspid Valve Velocity Peak A Combination Mode Chamber, Method of Disk (LVAd,
(TV A Velocity) Measurements LVAs)
• Tricuspid Valve Velocity Peak E • Left Ventricle Volume, Apical View,
• Aortic Valve
(TV E Velocity) Long Axis, Method of Disk (LVAd,
• Aortic Valve Area (Ao Root Diam,
• Tricuspid Regurgitant Mean LVAs)
LVOT Vmax, AV Vmax)
Pressure Gradient (TR Trace) • Mitral Valve
• Aortic Valve Area by Continuity
• Tricuspid Valve Mean Pressure • Stroke Volume by Mitral Flow
Equation by Peak Velocity (Ao
Gradient (TV Trace) (MVA Planimetry, MV Trace)
Root Diam, LVOT Vmax, AV Vmax)
• Tricuspid Regurgitant Mean • Cardiac Output by Mitral Flow
• Stroke Volume by Aortic Flow
Velocity (TR Trace) (MVA Planimetry, MV Trace, HR)
(AVA Pl ani met ry, AV Trace)
• Tricuspid Regurgitant Velocity • Cardiac Output by Aortic Flow • Pulmonic Valve
Time Integral (TR Trace) (AVA Planimetry, AV Trace, HR) • Stroke Volume by Pulmonic Flow
• Tricuspid Valve Mean Velocity • Aortic Valve Area by Continuity (PV Planimetry, PV Trace)
(TV Trace) Equation VTI (Ao Root Diam, LVOT • Cardiac Output by Pulmonic Flow
• Tricuspid Valve Velocity Time Vmax, AV Trace) (PV Planimetry, PV Trace, HR)
Integral (TV Trace) • Tricuspid Valve
• Left Ventricle
• Tricuspid Valve Time to Peak (TV • Cardiac Output by Tricuspid Flow
• Cardiac Output, Teichholz/Cubic
Acc/Dec Time) (TV Planimetry, TV Trace, HR)
(LVIDd, LVI Ds, HR)
• Tricuspid Valve Ejection Time (TV
• Cardiac Output Two Chamber,
Acc/Dec Time)
Single Plane, Area-Length/

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 13 of 14


Cardiac Worksheet
• Parameter: lists the mode, the
measurement folder and the
specific measurement
• Measured Value: Up to six
measurement values for each item.
Average, maximum, minimum, or
last

Generic Study in Cardiology


• Stroke Volume (SV)
• Cardiac Output (CO)

© 2013 General Electric Company –


All rights reserved.
General Electric Company reserves
the right to make changes in
specifications and features shown
herein, or discontinue the product
described at any time without notice
or obligation. Contact your GE
Representative for the most current
information.

GE and GE Monogramare
trademarks of General Electric
Company.
*Trademark of General Electric
Company
**Third party trademarks are the
property of their respective owners

GE Medical Systems Ultrasound &


Primary Care Diagnostics, LLC, a
General Electric company, doing
business as GE Healthcare.

DOC1487369: LOGIQ S8 R2 Specification Sheet Page 14 of 14