Professional Documents
Culture Documents
OP100
Orthoceph®
OC100
Replacement Instructions
03/2003 63444
Copyright © 2003 by Instrumentarium Imaging
Documentation, trademark and the software are copyrighted with all rights
reserved. Under the copyright laws the documentation may not be copied,
photocopied, reproduced, translated, or reduced to any electronic medium or
machine readable form in whole or part, without the prior written permission
of Instrumentarium Imaging.
Eliminate electric charges from your body by touching a grounding point, eg.
screw head, before you touch the components or boards. Always switch OP
100 power off prior to disconnecting of cables or removal of boards. If
possible hold the boards at their edge, only.
POWER SUPPLY
INTERFACE
HEAD SUPPORT LOCK
AEC
CPU
SOFT TISSUE DISPLAY
INVERTER
FILAMENT
TUBE HEAD
OPTO SENSORS
OC100PCB
1 Software 1.2.07 or higher Select user program Pr68 INS: FRE . Soft-
wares 1.2.01 - 1.2.06. Select service program Sr 78 FRE.
2 Check the AEC base frequency: the display on the control panel should
read 5.00 ± 1.0 kHz (without radiation). If needed, adjust with BASE
FREQUENCY trimmer on the AEC board.
NOTE
If the control panel exposure button is disabled with the service program Sr
89 rEo / Sr 89 COP 1 rEo, the button must be enabled before the previous
test. After the test, disable the button again.
NOTE
If the CPU is totally out of order, the previous steps may not be possible.
7 If the board was supplied without the EPROM chips, carefully remove
the EPROM’s from the old CPU board. Insert the EPROM’s on the new
board, white one (User program) onto the right socket (ICD 29) and
yellow one (Service program) onto the left socket (ICD 28).
CAUTION
Be sure that the notch on the EPROM chips is pointing downwards, incorrect
installation will destroy the circuits. Applicable software versions, see
Service Manual for software versions. If the imaging parameters want to be
retained exchange the parameter memory ICD8 from the old board to the new
one. PAL circuit ICD25 on the CPU board:
If Inverter board is type code 60123, ICD25 circuit must be version 1.0.1 or
1.1.0.
If Inverter board is type code 60166, ICD25 circuit must be version 2.0.0 or
2.1.0.
1 Set the jumper X 10 to the left pins, Service mode is selected. Set the
OPTION jumper X 11 to OFF position (in board versions 1.0 to 1.2 set
the switch S 2 to OFF position).
2 Turn power on. If error code Sy 24 FIL appears, reset it by pressing the
OK key.
3 Do the preheat adjustment by using service program Sr 77 Prh. See
Service Program Manual Section 2.8, Preheat automatic adjustment for
details.
4 Program the recorded values (or obtain new correct values) of lateral
cassette adjustment (Sr 72 LCA), AEC offset (Sr 73 AEC) and special
features (Sr 89 rEo / Sr 89 COP). See Service Program Manual
Section 2.20, Special programs for details.
5 Turn power off. Set the jumper X 10 to the right, User mode is selected.
6 Software 1.2.07 or higher
Press and keep pressing the OK key. First a beeb is heared and the
display shows exposure counter value. After a while more ‘’beeps’’ are
heard and the display shows user program information, eg.Pr50LAY. At
this point, release the OK button. Note that if the button is released too
early, program resumes to normal operation. Start again. Softwares
1.2.01 - 1.2.06.Turn power on and keep the OK key pressed during
power up. The unit enters User programming mode.
7 Program the recorded values to the different user programs. After
programming, press the RESET button on the CPU board to return to
normal mode.
8 Check that all the buttons on the control panel and patient positioning
panel operate.
9 Drive the unit up and down, check that the vertical movement stops at
the limit switches. With OC 100 use also ceph up/down keys. Drive the
cassette holder assembly up and down to check the operation.
10 Take a normal panoramic image (program 1) with the test phantom. No
error codes should appear. Process the film and verify the correct image.
11 Refit the vertical carriage center cover and the main support top cover.
Refit the FH-projector knob.
CAUTION
Do not open the fixing screws of the filament control support plate. Opening
may cause misalignment of the tube head assembly.
1 Set the unit into service mode by placing the jumper X 10 on the CPU
board to the left. Turn on power. After power up self check, no error
codes should appear.
2 Verify that the green power supply LED’s on the filament control board
are lit. None of the red LED’s should be lit.
3 Verify that the Inverter board +15V green LED (H5) is lit. Inverter board
+25V green LED (H7) and the red LED’s should not be lit. H1 (+310V)
may glow slightly.
4 Select service program Sr 75 EPS. Select low exposure values and a
time of few seconds. Make an exposure and verify that all of the LED’s
except H7 (preh) and H12 (TMPFAIL) on the filament control boars are
lit. H7 should be lit during the preheat time. Verify that all LED’s on the
inverter board are lit during the exposure.
5 Do the preheat calibration using Sr 77 Prh. See Service Program Manual
chapter 2.20.
6 If a voltmeter is available, mA measurement is recommended. Connect
the voltmeter to test points TP1 (GND) and TP2 (mAfb). Make exposure
with several mA settings and verify that the feedback voltages are
within the tolerances specified in Service Manual Volume IV Section
1.8.1.
7 Turn power off. Set the unit to normal mode (CPU jumper X 10 to
right). Refit the inverter board and close the tube head rear cover.
8 Turn on power. Take an image with a test phantom to verify the correct
operation of the unit. Take a QA film, mark it with the serial number of
the unit and a date and store it for further reference.
1 Fit the new board on the mounting studs. Fix the board with the mount-
ing nuts.
2 Refit the connectors. Verify that the connectors are properly locked.
1 Turn power on. Verify that all of the green LED’s on the interface board
are lit.
2 Verify that the jumper X 21 is installed (board version 1.3 ⇒ )
3 Drive the vertical carriage up and down. The carriage should first move
slowly and after few seconds the speed should increase.
4 Drive the cassette holder assembly up and down.
5 Select the test (T) mode from the control panel. Press the exposure
button and verify the movements. Hold the button until the imaging
program is finished. No error codes should appear.
6 Take an exposure in manual (M) mode. Verify that the warning lamps on
the sides of the unit are lit during the exposure.
7 Refit the main support top cover.
1 Fit the new board on the mounting screws on the inverter board support
bracket. Verify that the insulators are under the board and that the spacer
bushings are between the insulator and the board. Fix the board with the
mounting nuts.
2 Set the lower grooves of the inverter board support bracket onto the
lower mounting screws so that the board is hanging at a horizontal
(service) position. Refit the inverter board connectors. Verify that the
connectors are properly locked.
CAUTION
Be sure that you have the right selection of line voltage.
R7 R6
1 Open the cephalostat top cover.
2 Open the connector X 43.
3 Open the fixing screws of the soft tissue display mounting
brackets.
4 Remove the board with the mounting brackets.
5 Remove the mounting brackets from the board. Note the orientation of
the brackets.
1 Calibrate the display. Check that the display has ‘’0’’ when the nose
pupport is pushed towards ear holders. Adjust to ‘’0’’ if necessary by
turning the trimpot R6 at the Soft Tissue Display Board. A negative
value is indicated by ‘’c0’’.Check that the display has ‘’60’’ when the
nose support is away from ear holders. Adjust to ‘’60’’ if necessary by
turning the trimpot R7 at the Soft Tissue Display Board.
2 Refit the cephalostat top cover.
2 Motors
2.1 Z-motor M1
Fig 2.1. Secure the counterweight to column with the screw and bushing (A) provided with
OP100. Switch off the unit.
2 Lift the vertical carriage system about 20 cm and support the system
with eg. two wooden studs (B) or equal at both sides of the column,
under the carriage lower plate. See that the counterweight cables do
not jam.
WARNING
Ensure that the system is stationary before opening of the suspension plate.
3 Remove the right side plate (C) by undoing the M4 screws (D). Undo
the 4 pcs screws (E) of the suspension plate (F) and remove it. Protect
the surface of the column. Pull out the assembly to get access to pulleys.
4 To remove complete top plate assembly first remove the cables from
pulleys: undo the screw from the other pulley and remove the pulley,
then release the cable. Undo the nut and remove the spring washers
and drive wheel (pulley) then release the cable.
5 Disconnect the motor cable. Lift the suspension plate with motor on
side.
6 For complete top plate replacement, skip the following two steps.
7 Disassemble the clutch system (G) and pulley (H). Undo the motor axle
(I) with a screwdriver. Undo the screws (J) and washers of the motor and
open connector X 104.
10 Adjust the clutch with the nut (K). Hold the axle with a screw driver.
The clutch is correctly adjusted when the cable does not slide over the
pulley when carriage system is moved up and down by hand.
11 Mount the side plate.
2 Open the connector X 111. Undo the motor axle fixing screw (B) inside
the fly wheel with 2 mm Allen wrench provided with OP 100.
3 Undo the 4 pcs nuts (C) with 8 mm wrench. Remove the 4 pcs attenuator
rubbers (D) by undoing the nuts (E) on the motor side. Remove the
motor.
2 Remove the lever(s) (A) of the collimator. Remove the THA front cover
by undoing the 4 pcs M4 screws.
3 Undo the 4 pcs M4 screws (C) of the lover trim cover and remove the
cover.
4 Open connector X 112. Replace the motor/gear assembly.
5 Reassemble in a reverse order. No adjustments are needed, if the spring
loaded clutch assembly is not moved.
6 Run Service Program “Sr 80 ro-” to check the operation.
2 Undo the gear pivot screw (C) and release the spring (D). Remove the
motor assembly and open the connector X 115.
2 Undo the 2 pcs M3 screws (A) of the secondary plate and loosen the 2
pcs allen screws (B). Remove the secondary plate. Check that the screws
(B) are not scraping the main support.
3 Lamps
3.1 X-RAY warning lamps LA 1 / LA 2
1 Switch off the unit. Open side box lid (A) left for LA 1 or right for LA 2.
Open side box cover (B) L or R by undoing the M4 screws (C and D).
2 Pull out the defective lamp (E) and replace it with a new one.
3 Reassemble in a reverse order.
4 After reassembling switch on the unit and check that the lamps work
during exposure.
3 Remove the light shield by undoing the screw. Push down the projector
to its lowest position. Undo the 2 pcs screws of the lamp holder and
remove the holder.
4 Pull out the defective lamp and replace it with a new one.
NOTE
Do not touch lamp with fingers!
6 Pull out the defective lamp and replace it with a new one.
NOTE
Do not touch the lamp with fingers!
3 Pull out the defective lamp and replace it with a new one.
NOTE
Do not touch lamp with fingers!
Open the cover. Fix the strain relief and fix the wires.
1 Fit the beam alignment tool on the secondary slot. Remove the sinus rest
from the chin rest support piece and fit the alignment spike to the sup-
port piece.
2 Remove the collimator if fitted.
3 Align the rotating unit perpendicular to the main support, i.e. when the
rotation microswitch detecting the cassette lifting position “clicks”.
4 With sw. 1.2.07 or higher Select Pr68 INS,EPS or with software 1.2.01-
1.2.06 Sr 75 EPS , maximum exposure values and make the exposure.
Note the position of the shadow of the alignment spike on the
fluorescent screen of the alignment tool. If no shadow is seen, move the
linear movement by turning the linear actuator shaft by hand until the
shadow of the spike is visible.
5 Turn the rotating unit 180 degrees and align it perpendicular to the main
support. Take care not to move the tubehead sideways on the rotating
unit. Make the exposure and note the position of the spike shadow.
6 If the position of the shadow on step 5 differs from the position on step
4, adjust the linear movement so that the shadow is on the center of
positions on steps 4 and 5. Repeat steps 3 through 5 until the shadow
does not move when the rotating unit is turned 180 degrees.
7 While keeping the rotating unit perpendicular to the main support, adjust
the tubehead sideways until the shadow is on the center of the alignment
tool. Turn rotating unit 180 degrees and check that the shadow is still on
the center. If it is not repeat steps 3 through 6.
When the spike shadow is on the center of the alignment tool on both sides,
tighten the tubehead mounting screws.
8 Turn the rotating unit parallel to the main support, i.e. the center rotation
microswitch “clicks”, taking care not to move the linear movement.
Make the exposure and note the position of the spike shadow on the
alignment tool. If the shadow is not at the center of the alignment tool,
adjust the alignment spike sideways until the shadow is centered.
There are four position indicators for rotating unit. These microswitches are
located inside the main support. Rotation angle information comes from a
code disk, which is located on the rotation unit, under the main support.
Rotational position is indicated by four microswitches (S 11 to S 14) as
follows.
S 14 S 13 S 12 S 11 TUBEHEAD-
ROT4SW ROT3SW ROT2SW ROT1SW POSITION
In5 LED8 In5 LED7 In5 LED6 In5 LED5
LINLIMSW
LINMIDSW
S 15 is for the mid position of the linear movement, also indicating front and
rear segments of the movement. It is used to set OP 100 rotating unit for
patient positioning (Programs 1 to 4 and 6 to 9) and it serves as a reference
point for TMJ pointer movements in TMJ imaging (Programs 6 to 9).
There are two position indicators for cassette carriage movement. These
optical sensors (or microswitches in older models) are located inside the
cassette holder, behind the cassette carriage.
Fig 6.1. Units s/n 7000-70231 Fig 6.2. Units from s/n 70232
CASLIMSW
CASMIDSW
D 2 / S 24 D 3 / S 25 CARRIAGE POSI-
CASLIMSW CASMIDSW TION (viewed from
In0 LED7 In0 LED6 tubehead towards
cassette)
PANCASSW
Indicates the panoramic cassette: when signal is active (opto sensor D1 free
or microswitch S23 closed), cassette is in the cassette holder. Signal is
indicated by In 0 LED8 in service program Sr 74 Ioc, lit LED indicates the
presence of the cassette.
CEPH1CAS
Fig 6.3. Cassette holder limits detected Fig 6.4. Cassette holder limits detected
by sensors: low by sensors: medium
RACKLIMSW
One microswitch (S 22) to indicate either end of the cassette holder vertical
movement.
RACKMIDSW
closed closed up
open closed upper half
open open lower half
closed open down
There are two position indicators for the vertical carriage movement. These
microswitches are located inside the rear support assembly, at the rear of the
column unit. The height of the vertical carriage is detected by two short
detent pieces located in the groove of the column. These pieces are adjusted
at the factory so that they will stop the carriage 10 - 20 mm before the
mechanical limits.
ZLIMSW
ZMIDSW
S5 S4 CARRIAGE
ZLIMSWIn5 ZMIDSW POSITION
LED4 In5 LED3
S 18 COLLIMATOR FUNCTION
COL2SW In 2 LED6
S 31 S 32 S 33 COLLIMATOR POSI-
COL1SW COL2SW COL3SW TION
In2 LED5 In2 LED6 In2 LED7
Open covers:
Sliding chassis
cover holders
Make the movement test in SR80 ro program and adjust the rotation
movement with the spring force.
Check and adjust the head support bracket to the unit midlle line.
Instrumentarium Imaging is constantly improving its products and reserves the right to
change these specifications without notice. Distributor:
www.InstrumentariumImaging.com