Higher Pa System
Low Pa System
Leak Check: everything o ? Sux CGO, 10 secs? Repeat w/ open vaporizer,10 secs?
Flowmeters: check full range smooth, undamaged ow tubes. Hypoxic mix? (N2O↑O2↑/O2↓N2O↓)
Scavenging System
Calibrate O2 Monitor
O2 Flow Min, Other gas 0 —> Inspiration TV good? Expiration bellows ll complete? (Bellows UP exp)
Vaporizers OFF APL valve OPEN BAG MODE FLOWMETERS OFF Sux Good
fi
ff
fi
fi
fl
fl
fi
Emergency Ventilation Equipment: Ambu bag + Sux
High Pa System
Low Pa System
MACHINE ON
Scavenging System
Calibrate O2 Monitor
BAG MODE
CO2 Absorbent?
Vent Parameters
VENT MODE
O2 FLUSH - Fill Bellows + Breathing Bag. O2 Flows Min Vt good? Expiration lls bellows complete?
Unidirectional Valves
BAG MODE
Final Status
Flowmeters OFF
Vaporizers OFF
BAG MODE
SUX Good