Professional Documents
Culture Documents
Outline
3
Objectives
2
1
11/12/2020
5 7
6 8
There are many different types of x-ray imaging
systems, usually identified according to the
energy of the x-rays they produce or the purpose
for which the x-rays are intended.
2
11/12/2020
9 11
The general purpose x-ray examination room
contains a radiographic imaging system and a
fluoroscopic imaging system.
The fluoroscopic x-ray tube is usually located under the
examining table; the radiographic x-ray tube is attached to an
overhead movable crane assembly that permits easy
positioning of the tube and aiming of the x-ray beam.
12
3
11/12/2020
13 15
14 16
Plan drawing of a general-purpose x-ray examination room,
showing locations of the various x-ray apparatus items.
4
11/12/2020
Operating Console
17 19
Operating Console
▸ The part of x-ray imaging system most familiar to the Provides for control of ff:
radiologic technologists. 1. Autotransformer
▸ The operating console allows radiologic technologists to 2. Line compensator
control the x-ray tube current and voltage so that the
useful x-ray beam is of proper quantity and quality.
3. kVp
4. mA
5. Exposure timer
Note: All of the electric circuits that connect the
meters and controls on the operating console
low voltage to minimize the
are at
possibility of hazardous shock.
18 20
5
11/12/2020
21 23
22 24
6
11/12/2020
25 27
26 28
▸ The line compensator measures the ▸ On most operating consoles, the kVp meter
voltage provided to the x-ray imaging system registers, even though no exposure is being
made and the circuit has no current. This type
and adjusts that voltage to precisely 220 V.
of meter is known as a pre-reading kVp meter. It
▸ Older units required technologists to adjust allows the voltage to be monitored before an
the supply voltage while observing a line exposure.
voltage meter. ▸ The kVp meter is placed across the output
▸ Today’s x-ray imaging systems have terminals of the autotransformer.
automatic line compensation and hence have
no meter.
7
11/12/2020
29 31
30 32
8
11/12/2020
33 35
34
Exposure timers
36
High-voltage generator
4. Automatic Exposure Control (AEC)
▸ AEC requires a special understanding on the
part of the radiologic technologist.
▸ AEC is a device that measures the quantity of FIGURE 5-12:
radiation that reaches the image receptor. Cutaway view of a
▸ It automatically terminates the exposure typical high-voltage
when the image receptor has received the generator showing
required radiation intensity. oil-immersed diodes
and transformers.
▸ It must be calibrated regularly.
9
11/12/2020
38 40
1. High-voltage transformer
It is a step-up transformer, that is, the secondary voltage
is higher than the primary voltage because the number of Practice question:
secondary windings is greater than the number of primary
windings. 1. The secondary side of a transformer has 300,000
turns; the primary side has 600 turns. What is the
The ratio of the number of secondary windings to the number of turns ratio?
primary windings is called the turns ratio.
10
11/12/2020
41 43
Practice question:
Practice question:
2. The turns ratio of a high-voltage transformer is
700:1, and the supply voltage is peaked at 120 V. 1. The turns ratio of a filament transformer is 0.125.
What is the secondary voltage supplied to the x- What is the filament current if the current through
ray tube? the primary winding is 0.8 A?
42 44
The secondary windings are thick and at approximately one direction is a rectifier.
12 V electric potential and carry a current of 5 to 8 A. Voltage rectification is required to ensure that
electrons flow from x-ray tube cathode to anode only.
11
11/12/2020
45 47
Voltage rectification
Rectification is accomplished Voltage rectification
with diodes, an electronic
device that contains two Rectification is essential
electrodes. for the safe and efficient
Originally, all diode rectifiers operation of the x-ray tube.
were vacuum tubes called
valve tubes; these have (3) Types of rectification:
been replaced by solid-state 1. Unrectified voltage
rectifiers made of silicon. 2. Half-wave rectification
Semiconductors are classed 3. Full-wave rectification
FIGURE 5-14: Rectifiers in most modern x-ray
into two types: n-type and generators are the silicon, semiconductor type.
The multiple black components on this 75-kVp
p-type. FIGURE 5-15: A p-n junction semiconductor
high-voltage multiplier board are rectifiers.
shown as a solid-state diode.
1. Unrectified voltage
46 48
Voltage rectification
N-type semiconductors
have loosely bound The current that passes
electrons that are through the x-ray tube,
relatively free to move.
however, exists only
P-type semiconductors during the positive half
have spaces, called holes of the cycle when the
(as mobile as electrons), anode is positive and the
where there are no cathode is negative. FIGURE 5-17: Unrectified voltage
electrons. and current waveforms on the
secondary side.
FIGURE 5-15: A p-n junction semiconductor
shown as a solid-state diode.
12
11/12/2020
3. Full-wave rectification
50 52
Types of Generator
The negative half-
cycle corresponding to 1. Single-phase
the inverse voltage is power
2. Three-phase power
reversed so that the
anode is always 6-pulse
positive. FIGURE 5-21: Voltage 12-pulse
across a full-wave–rectified 3. High-frequency
circuit is always positive. generator
13
11/12/2020
All of the voltage waveforms discussed so far are produced Produce a nearly constant potential voltage
by single-phase electric power. Single-phase waveform thus improving image quality at
power results in a pulsating x-ray beam. This is caused by lower patient radiation dose.
the alternate swing in voltage from zero to maximum
potential 120 times each second under full-wave
rectification.
Increasing application for many x-ray
imaging systems.
The x-rays produced when the single-phase voltage
waveform has a value near zero are of little Full-wave rectification or high-frequency
diagnostic value because of their low voltage generation is used in almost all
energy; such x-rays have low penetrability. stationary x-ray imaging systems.
14
11/12/2020
57 59
Voltage Ripple
3. Three-phase, 12-pulse
power results in only 4%
ripple; therefore, the voltage
supplied to the x-ray tube
does not fall to below 96%
of the maximum value.
Practice question:
1. When a system with low-voltage ripple is energized at 100 kVp,
4. High-frequency generators 100 ms, the maximum possible tube current is 800 mA. What is
have approximately 1% ripple the power rating?
and therefore greater x-ray
quantity and quality. 2. An interventional radiology system is capable of 1200 mA when
operated in 100 kVp, 100 ms. What is the power rating?
Less voltage ripple results in greater radiation quantity and quality.
58
Power Rating
60
X-ray circuit
Transformers and high-voltage generators
usually are identified by their power rating
in kilowatts (kW).
FIGURE 5-32:
The schematic
circuit of an
x-ray imaging
system.
15