You are on page 1of 15

11/12/2020

Outline
3

Operating Console High-Voltage Generator


Autotransformer High-Voltage Transformer
Adjustment of Kilovolt Peak (kVp) Filament Transformer
Control of Milliamperage (mA) Voltage Rectification
Exposure Timers Single-Phase Power
Synchronous Timers Three-Phase Power
Electronic Timers High-Frequency Generator
mAs Timers Voltage Ripple
Source: Radiologic Science for Technologists, Stewart C. Bushong, 10e-Mosby (2012) Automatic Exposure Control Power Rating
X-ray Circuit

Objectives
2

X-ray Imaging System


At the completion of this chapter, you should be The function of the x-ray imaging
able to do the following:
1. Identify the components of the x-ray imaging system system is to provide a controlled
operating console.
2. Explain the operation of the high-voltage generator.
flow of electrons intense enough to
3. Relate the differences among single-phase, three- produce an x-ray beam appropriate
phase, and high-frequency power.
4. Discuss the importance of voltage ripple to x-ray for imaging.
quantity and quality.
5. Define the power rating of an x-ray imaging system.

1
11/12/2020

5 7

The three main components of an x-


ray imaging system are
1. Operatingconsole
2. High-voltagegenerator
3. X-raytube

6 8
There are many different types of x-ray imaging
systems, usually identified according to the
energy of the x-rays they produce or the purpose
for which the x-rays are intended.

 Diagnostic x-ray imaging systems come in many


different shapes and sizes, some of which are shown in
Figure 5-1. These systems are usually operated at
voltages of 25 to 150 kVp and at tube currents of 100 to
1200 mA.

2
11/12/2020

9 11
The general purpose x-ray examination room
contains a radiographic imaging system and a
fluoroscopic imaging system.
The fluoroscopic x-ray tube is usually located under the
examining table; the radiographic x-ray tube is attached to an
overhead movable crane assembly that permits easy
positioning of the tube and aiming of the x-ray beam.

12

Electric energy is supplied to the x-ray imaging system


in the form of well-controlled electric current. Patient couch (Examination table)
 This examination couch may be flat or
A conversion takes curved but must be uniform in
place in the x-ray tube, thickness and as transparent to x-rays
where most of this as possible.
electric energy is
transformed into heat,
some of it into x-rays.

3
11/12/2020

13 15

Carbon fiber couches are strong The protective barrier


and absorb little x-radiation. This must have a window for
contributes to reduced patient viewing the patient during the
radiation dose. examination.
Most patient couches are floating—
easily unlocked and moved by the
radiologic technologist—or motor
driven.

14 16
Plan drawing of a general-purpose x-ray examination room,
showing locations of the various x-ray apparatus items.

Just under the couch is an


opening to hold a thin tray
(Bucky tray) for a
cassette and grid.

4
11/12/2020

Operating Console
17 19

Operating Console
▸ The part of x-ray imaging system most familiar to the Provides for control of ff:
radiologic technologists. 1. Autotransformer
▸ The operating console allows radiologic technologists to 2. Line compensator
control the x-ray tube current and voltage so that the
useful x-ray beam is of proper quantity and quality.
3. kVp
4. mA
5. Exposure timer
Note: All of the electric circuits that connect the
meters and controls on the operating console
low voltage to minimize the
are at
possibility of hazardous shock.

18 20

Operating Console Operating Console


▸ Radiation quality refers to the FIGURE 5-6:
number of x-rays or the intensity of Circuit diagram of
the x-ray beam. the operating
▸ Radiation quality refers to the console, with
penetrability of the x-ray beam. controls and
meters identified.

5
11/12/2020

21 23

Operating Console Autotransformer Law


▸ Operating consoles are based on computer technology.
Controls and meters are digital, and techniques are selected
with a touch screen. Numeric technique selection is often
replaced by icons indicating the body part, size, and shape.
▸ Many of the features are automatic, but the radiologic
technologist must know their purpose and proper use.

22 24

1. Autotransformer Autotransformer Law


▸ The power supplied to the x-ray Practice question:
imaging system is delivered first to the 1. If the autotransformer in
autotransformer. Figure 5-7 is supplied with
▸ It has a single winding and is 220 V to the primary
designed to supply a precise voltage connections AA′, which enclose
to the filament circuit and to the high- 500 windings, what is the
voltage circuit of the x-ray imaging secondary voltage across BB′
system. (500 windings), CB′ (700
windings), and DE (200
windings)?

6
11/12/2020

25 27

Autotransformer Law 3. Adjustment of kilovolt peak (kVp)


Practice question: Equation: ▸ Some older x-ray operating consoles
2. An autotransformer connected to a have adjustment controls labeled major kVp
440-V supply contains 4000 turns, and minor kVp.
all of which are enclosed by the ▸ Selecting a combination of these controls, radiologic
primary connections. If 2300 turns technologists can provide precisely the required
are enclosed by secondary kilovolt peak.
connections, what voltage is
supplied to the high-voltage kVp determines the quality of the x-ray beam.
generator?

26 28

2. Line Compensator Adjustment of kilovolt peak (kVp)

▸ The line compensator measures the ▸ On most operating consoles, the kVp meter
voltage provided to the x-ray imaging system registers, even though no exposure is being
made and the circuit has no current. This type
and adjusts that voltage to precisely 220 V.
of meter is known as a pre-reading kVp meter. It
▸ Older units required technologists to adjust allows the voltage to be monitored before an
the supply voltage while observing a line exposure.
voltage meter. ▸ The kVp meter is placed across the output
▸ Today’s x-ray imaging systems have terminals of the autotransformer.
automatic line compensation and hence have
no meter.

7
11/12/2020

29 31

4. Control of milliamperage (mA) Exposure timers


▸ Thermionic emission is the release of electrons 1. Synchronous timer
from a heated filament. ▸ A special type of electronic motor.
▸ Thex-raytubecurrent,crossing from cathodetoanode,is measured ▸ It a precision device designed to drive a shaft precisely
in milliamperes (mA). 60 revolutions per second (rps).
▸ Thenumber of electrons emitted by the filamentis determined by ▸ Synchronous timers are recognizable because the
minimum exposure time possible is 1/60 s (17ms).
the temperature of the filament.
▸ Synchronous timers cannot be used for serial
▸ Filaments normally operate at currents of 3 to 6A. exposures because they must be reset after each
exposure.

30 32

5. Exposure timers Exposure timers


▸ For any given radiographic examination, the 2. Electronic timer
number of x-rays that reach the image
receptor is directly related to both the x-ray ▸ The most sophisticated, most complicated, and
tube current and the time that the x-ray most accurate.
tube is energized. ▸ Allow a wide range of time intervals to be
selected and are accurate to intervals as
Types of exposure timers:
1. Synchronous timer
small as 1 ms.
2. Electronic timer ▸ Used for rapid serial exposures, particularly
3. mAs timer suitable for interventional radiology
4. Automatic Exposure Control (AEC)
procedures.

8
11/12/2020

33 35

Exposure timers Automatic Exposure Control (AEC)


3. mAs timer
▸ A special kind of electronic timer, that
monitors the product of mA and time and
terminates the exposure when desired
mAs value is attained.
▸ This timer is designed to provide the highest
safe tube current (mA) for the shortest
FIGURE 5-10 Automatic exposure
exposure for any mAs selected. control terminates the x-ray exposure
FIGURE 5-11 Solid-state radiation
at the desired film optical density. This detectors are used to check timer
accuracy.
is done with an ionization chamber or
a photodiode detector assembly.

34

Exposure timers
36
High-voltage generator
4. Automatic Exposure Control (AEC)
▸ AEC requires a special understanding on the
part of the radiologic technologist.
▸ AEC is a device that measures the quantity of FIGURE 5-12:
radiation that reaches the image receptor. Cutaway view of a
▸ It automatically terminates the exposure typical high-voltage
when the image receptor has received the generator showing
required radiation intensity. oil-immersed diodes
and transformers.
▸ It must be calibrated regularly.

9
11/12/2020

High-voltage generator High-voltage transformer


37 39

 Responsible for increasing the output voltage


from the autotransformer to the kVp
necessary for x-ray production.

(3) Parts of high-voltage generator  The voltage increase is proportional


1.High-voltage transformer to the turns ratio, according to the
2.Filament transformer transformer law.
3.Voltage rectification FIGURE 5-13: Voltage induced in the
secondary winding of a high-voltage
step-up transformer is alternating like
the primary voltage but has a higher
value.

38 40

1. High-voltage transformer
 It is a step-up transformer, that is, the secondary voltage
is higher than the primary voltage because the number of Practice question:
secondary windings is greater than the number of primary
windings. 1. The secondary side of a transformer has 300,000
turns; the primary side has 600 turns. What is the
 The ratio of the number of secondary windings to the number of turns ratio?
primary windings is called the turns ratio.

 The turns ratio of a high-voltage transformer is usually between


500:1 and 1000:1.

10
11/12/2020

41 43

Practice question:
Practice question:
2. The turns ratio of a high-voltage transformer is
700:1, and the supply voltage is peaked at 120 V. 1. The turns ratio of a filament transformer is 0.125.
What is the secondary voltage supplied to the x- What is the filament current if the current through
ray tube? the primary winding is 0.8 A?

42 44

2. Filament transformer 3. Voltage rectification


 The full title for this transformer is the filament  The current in a common wall plug is 60 Hz
heating isolation step-down transformer. alternating current (AC) that changes direction 120
 In a step-down transformer, the secondary times each second.
 Transformers operate with alternating current, x-ray
current is larger than the primary current.
 The primary windings are of thin copper and carry a tubes must be provided with direct current.
current of 0.5 to 1 A and approximately 150 V.  The electronic device that allows current flow in only

 The secondary windings are thick and at approximately one direction is a rectifier.
12 V electric potential and carry a current of 5 to 8 A.  Voltage rectification is required to ensure that
electrons flow from x-ray tube cathode to anode only.

11
11/12/2020

45 47
Voltage rectification
 Rectification is accomplished Voltage rectification
with diodes, an electronic
device that contains two Rectification is essential
electrodes. for the safe and efficient
 Originally, all diode rectifiers operation of the x-ray tube.
were vacuum tubes called
valve tubes; these have (3) Types of rectification:
been replaced by solid-state 1. Unrectified voltage
rectifiers made of silicon. 2. Half-wave rectification
 Semiconductors are classed 3. Full-wave rectification
FIGURE 5-14: Rectifiers in most modern x-ray
into two types: n-type and generators are the silicon, semiconductor type.
The multiple black components on this 75-kVp
p-type. FIGURE 5-15: A p-n junction semiconductor
high-voltage multiplier board are rectifiers.
shown as a solid-state diode.

1. Unrectified voltage
46 48
Voltage rectification

 N-type semiconductors
have loosely bound  The current that passes
electrons that are through the x-ray tube,
relatively free to move.
however, exists only
 P-type semiconductors during the positive half
have spaces, called holes of the cycle when the
(as mobile as electrons), anode is positive and the
where there are no cathode is negative. FIGURE 5-17: Unrectified voltage
electrons. and current waveforms on the
secondary side.
FIGURE 5-15: A p-n junction semiconductor
shown as a solid-state diode.

12
11/12/2020

2. Half-wave rectification 3. Full-wave rectification


49 51

 A condition in which the


voltage is not allowed to
swing negatively during the
negative half of its cycle.

FIGURE 5-19: A half-wave–


rectified circuit contains one or FIGURE 5-20: A full-wave–rectified circuit contains at least four
FIGURE 5-16: The
electronic symbol for a more diodes. diodes. Current is passed through the tube at 120 pulses per
solid-state diode. second.

3. Full-wave rectification
50 52

Types of Generator
 The negative half-
cycle corresponding to 1. Single-phase
the inverse voltage is power
2. Three-phase power
reversed so that the
anode is always  6-pulse
positive. FIGURE 5-21: Voltage  12-pulse
across a full-wave–rectified 3. High-frequency
circuit is always positive. generator

13
11/12/2020

1. Single-Phase Power 3. High-Frequency Generator


53 55

 All of the voltage waveforms discussed so far are produced  Produce a nearly constant potential voltage
by single-phase electric power. Single-phase waveform thus improving image quality at
power results in a pulsating x-ray beam. This is caused by lower patient radiation dose.
the alternate swing in voltage from zero to maximum
potential 120 times each second under full-wave
rectification.
 Increasing application for many x-ray
imaging systems.
 The x-rays produced when the single-phase voltage
waveform has a value near zero are of little  Full-wave rectification or high-frequency
diagnostic value because of their low voltage generation is used in almost all
energy; such x-rays have low penetrability. stationary x-ray imaging systems.

2. Three-Phase Power Voltage Ripple


54 56

 Another way to characterize voltage waveforms.


 With three-phase power, multiple voltage 1. Single-phase power has
waveforms are superimposed on one 100% voltage ripple; the
another, resulting in a waveform that voltage varies from zero to
its maximum value.
maintains a nearly constant high voltage.
2. Three-phase, six-pulse
 Voltage applied across the x-ray tube is nearly power produces voltage with
constant, never dropping to zero during only approximately 14%
ripple; the voltage supplied to
exposure. the x-ray tube never falls to
below 86% of the maximum
value.

14
11/12/2020

57 59
Voltage Ripple
3. Three-phase, 12-pulse
power results in only 4%
ripple; therefore, the voltage
supplied to the x-ray tube
does not fall to below 96%
of the maximum value.
Practice question:
1. When a system with low-voltage ripple is energized at 100 kVp,
4. High-frequency generators 100 ms, the maximum possible tube current is 800 mA. What is
have approximately 1% ripple the power rating?
and therefore greater x-ray
quantity and quality. 2. An interventional radiology system is capable of 1200 mA when
operated in 100 kVp, 100 ms. What is the power rating?
Less voltage ripple results in greater radiation quantity and quality.

58
Power Rating
60
X-ray circuit
 Transformers and high-voltage generators
usually are identified by their power rating
in kilowatts (kW).
FIGURE 5-32:
The schematic
circuit of an
x-ray imaging
system.

15

You might also like