Daily Operation
1. Daily Start-up (Power on) g. Place plate in the plate support tray. i. Touch [Finish] box.
(Stand clear minimum of 8 inches h. Touch . j. Result can be viewed by touching [View
before powering on reader). Data] box on or printing by touching
i. If assay requires any other input such
a. Power on Reader with switch on the as Min/Max control value or verification [Print] box.
rear right. Log in using User Name of valid BR Anti-HBs (0-400), an [Edit k. Touch to save the .wsp file. Remarks
and Password. Touch . Prompt] box will appear. Otherwise, can be entered in Remarks box.
b. Allow one minute to load assay files it will start shaking the plate for l. Touch [Finish] to get back to Wizard
from the APF USB stick. 20 seconds outside the reader List screen.
before taking in the plate to read.
2. Create Sample ID List (applies 5. Daily Shutdown – Do not leave
only for methods with v2.0 – see Table 3
j. Result will print automatically if the
plate in Reader
in Section 9 of the MATRM) plate meets all the validation criteria.
k. Verify that all validations have passed. a. Take plate off the Plate Support.
a. Touch [Create/Edit a Sample ID List]
l. Verify patient results are within b. Inspect reader for any spill and clean
b. Highlight the assay under Filename as needed if spill occurs.
and touch defined specifications.
m. If any validation is not met, c. Touch [Exit Magellan] box.
c. Touch [Manual enter sample ID list…]
the result report can be printed d. When the Popup message “It is now
d. Touch “sm1” well then [Edit] box. ok to turn off Instrument” comes up,
without the Interpretation.
e. Scan first bar code label. It will advance Power down Reader with switch
1. Touch [Yes] button when the QC
to next available well. Scan in remaining on the rear right.
Validation Criteria Failed box appears
bar code labels.
with failed validation criteria asking
f. Touch and review plate map. to proceed or to stop evaluation.
If correct, touch and to proceed
2. Result will print automatically.
to Save as screen.
g. Save using MMDDYY_run#_method 4. Read “Raw Absorbance of Plate”
convention. a. Touch [Start Measurement] box.
3. Read “Preprogrammed Assay Plate” b. Touch radial button for Obtain Raw
Data and Touch .
a. Touch [Start Measurement] box.
c. Touch [Measurement Parameters]
b. Touch radio button for User Predefined
to access the [Meas Params] folder. 6. Power off the PR3100 TSC before
Method or Start Favorite.
d. Select single or dual wavelength(s) for taking out the Export USB stick
c. Select Assay File Name and touch .
reading by selecting the wavelength(s) to access the ASCII or ASTM file
d. Enter Workspace Name incorporating to be read. a. Insert the Export USB stick into a USB
Date, Lot # and Kit Exp Date.
e. Touch [Shaking] box to input port on a PC and open using appropriate
(i.e., 071207,BB725,122607)
shake options. software. It can be opened with standard
For plate that does not utilize Sample ID List f. Touch [OK] and . text file program, EXCEL or ASTM
created in step 2, skip to step g. interface program.
g. At Start Measurement Screen,
e. Touch [Insert…] to input the Sample ID enter workspace name. b. When inserting back to PR3100 Reader,
List created for this particular run. make sure the reader is powered off.
h. Place plate in the plate support tray.
(Saved in step 2,g)
f. If Quantitative or Semi-quantitative Assay
requires change in standard value, touch
[Modify Layout] box.
Instrumentation
q. Touch File on Toolbar followed with Exit. Moves Absorbance Filter Allows to power down
slide In or Out
r. Power down Reader with switch on the rear right.
s. Take out the QC Pac2 USB stick and replace with Handheld Bar Code Scanner
the APF USB stick. Allows to run Handheld
Bar code scanner test
t. Ready for Daily Start-up.