Bio-Rad Laboratories Instrumentation

PR3100 TSC Quick Reference Guide


Refer to the PR3100 TSC User Manual and PR3100 TSC Microplate Assay Testing Reference Manual (MATRM)
for more detailed instructions.

Daily Operation
1. Daily Start-up (Power on) g. Place plate in the plate support tray. i. Touch [Finish] box.
(Stand clear minimum of 8 inches h. Touch . j. Result can be viewed by touching [View
before powering on reader). Data] box on or printing by touching
i. If assay requires any other input such
a. Power on Reader with switch on the as Min/Max control value or verification [Print] box.
rear right. Log in using User Name of valid BR Anti-HBs (0-400), an [Edit k. Touch to save the .wsp file. Remarks
and Password. Touch . Prompt] box will appear. Otherwise, can be entered in Remarks box.
b. Allow one minute to load assay files it will start shaking the plate for l. Touch [Finish] to get back to Wizard
from the APF USB stick. 20 seconds outside the reader List screen.
before taking in the plate to read.
2. Create Sample ID List (applies 5. Daily Shutdown – Do not leave
only for methods with v2.0 – see Table 3
j. Result will print automatically if the
plate in Reader
in Section 9 of the MATRM) plate meets all the validation criteria.
k. Verify that all validations have passed. a. Take plate off the Plate Support.
a. Touch [Create/Edit a Sample ID List]
l. Verify patient results are within b. Inspect reader for any spill and clean
b. Highlight the assay under Filename as needed if spill occurs.
and touch defined specifications.
m. If any validation is not met, c. Touch [Exit Magellan] box.
c. Touch [Manual enter sample ID list…]
the result report can be printed d. When the Popup message “It is now
d. Touch “sm1” well then [Edit] box. ok to turn off Instrument” comes up,
without the Interpretation.
e. Scan first bar code label. It will advance Power down Reader with switch
1. Touch [Yes] button when the QC
to next available well. Scan in remaining on the rear right.
Validation Criteria Failed box appears
bar code labels.
with failed validation criteria asking
f. Touch and review plate map. to proceed or to stop evaluation.
If correct, touch and to proceed
2. Result will print automatically.
to Save as screen.
g. Save using MMDDYY_run#_method 4. Read “Raw Absorbance of Plate”
convention. a. Touch [Start Measurement] box.
3. Read “Preprogrammed Assay Plate” b. Touch radial button for Obtain Raw
Data and Touch .
a. Touch [Start Measurement] box.
c. Touch [Measurement Parameters]
b. Touch radio button for User Predefined
to access the [Meas Params] folder. 6. Power off the PR3100 TSC before
Method or Start Favorite.
d. Select single or dual wavelength(s) for taking out the Export USB stick
c. Select Assay File Name and touch .
reading by selecting the wavelength(s) to access the ASCII or ASTM file
d. Enter Workspace Name incorporating to be read. a. Insert the Export USB stick into a USB
Date, Lot # and Kit Exp Date.
e. Touch [Shaking] box to input port on a PC and open using appropriate
(i.e., 071207,BB725,122607)
shake options. software. It can be opened with standard
For plate that does not utilize Sample ID List f. Touch [OK] and . text file program, EXCEL or ASTM
created in step 2, skip to step g. interface program.
g. At Start Measurement Screen,
e. Touch [Insert…] to input the Sample ID enter workspace name. b. When inserting back to PR3100 Reader,
List created for this particular run. make sure the reader is powered off.
h. Place plate in the plate support tray.
(Saved in step 2,g)
f. If Quantitative or Semi-quantitative Assay
requires change in standard value, touch
[Modify Layout] box.
Instrumentation

PR3100 TSC Quick Reference Guide

QC Pac2 – Reader Verification Software Navigation for Magellan Operators


1. Installing the QC Pac2 USB stick Wizard List Screen
(Stand clear minimum of 8 inches before Start Measurement
powering on reader).
Obtain Raw Data PC Remote
a. Power down Reader with switch on the rear right. Measurement Parameters box (Not applicable to Bio-Rad
(Required to be entered) Configuration)
b. Take the APF USB stick out from the back
of Touchscreen compartment. There is a sliding • General Tab: General
Change User
door to the rear right of the Touchscreen. Information
• Meas Param…: Allows User change without
c. Insert the QC Pac2 USB stick into the back of
Measurement rebooting PR3100 TSC
Touchscreen into the USB port compartment. (Make
sure the small light hole on the USB stick is facing down.) and Reference
Options
wavelength options
d. Power on Reader with switch on the rear right. Touchscreen
– Read Mode options
(Normal, Accuracy Date, Time, Contrast and
2. Running the QC Pac2
or Center) Keyboard (choice of regular
a. Touch [Q2] box on the top toolbar. • Shaking: Shaking options or large keyboard)
b. Follow instruction on screen. Touch .
Use Predefined Method Printer
c. Enter User Name. Printer driver selection
All predefined methods
d. Enter QC Plate #. Font size
• Enter Modify Layout
• Enter #, first time the plate is used. for Quant and Semi USB Port selection
• Touch [ ] to show previous Plate # to be selected. Quant assays (in Start Letter Paper Size selection
Measurement screen)
e. Place the Verification Plate on the plate support with Portrait Orientation selection
marking A1 on the plate in the top left corner. Start Favorite
Plate View
f. Touch . Reader will start reading the plate. 15 Selected methods (Not applicable to Bio-Rad
g. When the plate comes out and stays out, turn the • Enter Modify Layout Configuration)
for Quant and Semi
plate 180º with marking A1 at the bottom right corner.
Quant assays (in Start Misc
h. Touch . Reader will start reading the plate. Measurement screen) Wizard screen options
i. When the plate comes out and stays out for the Start-up screen options
Evaluate Results
second time, take the plate off the plate support and Audit Trail
place it back into the plastic bag before putting back Select a file
Change password…
into the case. View Data
Language
j. Touch [Move…] box to access plate movement. Print (Reprint)
(Do Not Change. Conflicts
k. Retract Plate Support by touching [In]. with Preprogrammed
Create/Edit a Sample-ID List Bio-Rad Assays)
l. Touch [Finish] to go to save result option – default (Applies only to the 22 methods
Add HUID
filename contains the Date in the filename. with v2.0 in Remarks section
(Do Not Change. Conflicts
when selecting method to run.)
m. Touch [OK] and Save. with Preprogrammed
Bio-Rad Assays)
n. Review screen for any number other than “0” Plate In/Out
next to FAIL. “0” means all passed. Moves plate In or Out About Magellan
o. Touch to print the complete report. Information on this
Select the printer driver. Default is PCL Instrument Control Instrument Software
Compatible Printer (Color). Movements (Plate and Filter)
p. Touch [OK] to print complete report. Moves plate In or Out Exit Magellan

q. Touch File on Toolbar followed with Exit. Moves Absorbance Filter Allows to power down
slide In or Out
r. Power down Reader with switch on the rear right.
s. Take out the QC Pac2 USB stick and replace with Handheld Bar Code Scanner
the APF USB stick. Allows to run Handheld
Bar code scanner test
t. Ready for Daily Start-up.

For more information, call us toll free at 1-800-2BIO-RAD (1-800-224-6723); outside the US,
contact your local sales office. Visit us on the web at www.bio-rad.com/diagnostics