You are on page 1of 35

DOCUMENT NO.

KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 1 of 35

성명 (NAME) 서명 (SIGN) 일자 (DATE) 부서 (Dep’t)

작성자
KIM, CHAN SIK 2006. 06. 27 QA Eng’r
(PREPARED BY)

검토자
SONG, CHANG HO 2006. 06. 27 LEVEL Ⅲ
(REVIEWED BY)

승인자
KIM, PYONG GOO 2006. 06. 30 QAD MGR
(APPROVED BY)

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 2 of 35

TABLE OF CONTENDS

PARA & TITLE

1. 0 PURPOSE

2. 0 REFERENCES

3. 0 PROCEDURE QUALIFICATION

4. 0 PERSONNEL QUALIFICATION

5. 0 GENERAL REQUIREMENTS

6. 0 EQUIPMENT AND MATERIALS

7. 0 EXAMINATION

8. 0 PROCEDURE

9. 0 ACCEPTANCE STANDARD

10. 0 RADIATION SAFETY

11. 0 REPAIR REQUIREMENTS

12. 0 EXAMINATION RESULTS

13. 0 EXHIBITS

ATTACHMENT 1

ATTACHMENT 2

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 3 of 35

1.0 PURPOSE
This procedure is to describe the minimum requirements for methods, techniques and acceptance
standards to be used when performing the radiographic examination of materials and weldments
for use in Code construction and welder and welding operator performance qualification.

2.0 REFERENCES
2.1 ASME Sec. V 1998 Ed. with 1998 through 2000 Add.
2.2 ASME Sec. VIII Div. 1 1998 Ed. with 1998 through 2000 Add.
2.3 ASME B 31.1 1998 Ed. with 1998 through 2000 Add.
2.4 ASME Sec. I 1995 Ed. with 1995 through 2000 Add.
2.5 ASME Sec. IX 1998 Ed. with 1998 through 2000 Add.
2.6 ASNT SNT-TC-1A 1996 Ed.

3.0 PROCEDURE QUALIFICATION


QUALIFICATION
This procedure shall be demonstrated to the satisfaction of the A.I and QC Dep't Manager of JIMC,
before use.

4.0 PERSONNEL QUALIFICATION


Personnel performing nondestructive examination shall be qualified in accordance with the KIE
procedure "Personnel Certification Program" which meets the requirements of ASNT Recommended
practice, SNT-TC-1A, and Section V of ASME Code.

5.0 GENERAL REQUIREMENTS


5.1 Surface Preparation
5.1.1 Materials
Surfaces shall satisfy the requirements of the applicable materials specifications, with
additional conditioning, if necessary, by any suitable process to a degree that surface
irregularities cannot mask or be confused with discontinuities.
5.1.2 Welds
The weld ripples or weld surface irregularities on both the inside (where accessible)
and outside shall be removed by any suitable process to such a degree that the resulting
radiographic image due to any irregularities cannot mask or be confused with the image
of any discontinuity.
5.1.3 Surface Finish
The finished surface of all buttwelded joints may be flush with the base material or may
have reasonably uniform crown, with reinforcement not to exceed that specified in Table1.
5.1.4 All surfaces requiring radiography shall be free of loose rust, scale, dirt, and oil /

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 4 of 35

grease in the area of interest and extending approximately 4~6 inch on both sides of the
weld edge or area of interest, both I.D and O.D.
Table 1. Limitation on Reinforcement
Maximum Reinforcement, in.
Material Nominal
Circumferential Joints in Other Welds
Thickness, in.
Pipe and Tubing
Less than 3/32 3/32 1/32
3/32 to 3/16, in. 1/8 1/16
Over 3/16 to 1/2, in. 5/32 3/32
Over 1/2 to 1, in 3/16 3/32
Over 1 to 2, in. 1/4 1/8
Over 2 to 3, in. 1/4 5/32
Over 3 to 4, in. 1/4 7/32
Over 4 to 5, in. 1/4 1/4
Over 5 5/16 5/16
a) For pressure vessel
Reinforcement of Girth and Longitudinal butt welds
Maximum Thickness of Reinforcement for Design Temperature
Thickness of Base Metal , in. (mm) > 750℉(400)℃ > 350℉-750℉(175℃-400℃) < 350℉(175℃)
in. mm. in. mm. in. mm.
Up to 1/8(3.0) incl. 1/16 2.0 3/32 2.5 3/16 5.0
Over 1/8 to 3/16(3.0 to 5.0),incl. 1/16 2.0 1/8 3.0 3/16 5.0
Over 3/16 to 1/2(5.0 to 13.0),incl. 1/16 2.0 5/32 4.0 3/16 5.0
Over 1/2 to 1(13.0 to 25.0),incl. 3/32 2.5 3/16 5.0 3/16 4.0
Over 1 to 2 (25.0 to 50.0),incl. 1/8 3.0 1/4 6.0 1/4 6.0
The greater of 1/4 in.(6mm) or 1/8 times the
Over 2 (50.0) 5/32 4.0
width of the weld in inches(millimeters).
NOTE : 1) For double welded butt joints, this limitation on reinforcement given above shall apply
separately to both inside and outside surfaces of the joint.
2) For single welded butt joints, the reinforcement limits given above shall apply to the
outside surface of the joint only.
3) The thickness of weld reinforcement shall be based on the thickness of the thinner of the
materials being joined.
4) The weld reinforcement thicknesses shall be determined from the higher of the abutting
surfaces involved.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 5 of 35

5) Weld reinforcement may be removed if so desired.


b) For Power Piping
5.2 Back-Scatter Radiation Check
A lead symbol "B", with minimum dimensions of 1/2 in. height and 1/16 in. in thickness, shall
be attached to the back of each film holder during each exposure to determine if backscatter
radiation is exposing the film.
If a light image of the "B", appears on a darker background of the radiograph, protection from
backscatter is insufficient and the radiograph shall be considered unacceptable.
A dark image of the "B" on a lighter background is not cause for rejection.
5.3 System of Identification
A system shall be used to produce permanent identification on the radiograph traceable to the
contract, component, weld or weld seam, or part numbers, and repairs (R1,R2,etc) as appropriate.
In addition, the manufacturer's symbol or name and the date of the radiograph shall be plainly
and permanently included on the radiograph. In any case, this information shall not obscure
the area of interest.
5.4 Monitoring Density Limitations of Radiographs
Either a densitometer or step wedge comparison film shall be used for judging film density.
5.5 Radiographic Density
5.5.1 Density Limitations
The transmitted film density through the radiographic image of the body of the appropriate
hole penetrameter and the area of interest shall be 1.8 minimum for single film viewing
for radiographs made with an X-ray source and 2.0 minimum for radiographs made with a
gammaray source.
The maximum density shall be 4.0. A tolerance of 0.05 in density is allowed for variations
between densitometer readings.
5.5.2 Density Variation
a) General
If the density of the radiograph anywhere through the area of interest varies by more
than minus 15% or plus 30% from the density through the body of the hole penetrameter
within the minimum/maximum allowable density ranges specified in 5.5.1, then an
additional penetrameter shall be used for each exceptional area or areas and the
radiograph retaken. When calculating the allowable variation in density, the
calculation may be rounded to the nearest 0.1 within the range specified in 5.5.1.
b) With Shims
When shims are used the plus 30% density restriction of (a) above may be exceeded,
provided the required penetrameter sensitivity is displayed and the density
limitations of 5.5.1 are not exceeded.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 6 of 35

5.6 Geometric Unsharpness


Geometric unsharpness of the radiograph shall be determined in accordance with :
Ug = F.d/D
where
Ug = geometric unsharpness
F = source size : the maximum projected dimension of the radiating source (or effective
focal spot) in the plane perpendicular to the distance D from the weld or object
being radiographed, in.
D = distance from source of radiation to weld or object being radiographed, in.
d = Distance from source side of weld or object being radiographed to the film, in.
Geometric unsharpness of the radiograph shall not exceed the following.
Table 2] Geometric unsharpness
Material Ug
Thickness, in. Maximum, in.
Under 2 0.020
2 through 3 0.030
Over 3 through 4 0.040
Greater than 4 0.070
Note : 1. For ASME Sec. I, When it is impracticable to use a combination of radiography
parameters such that a geometric unsharpness of 0.07in. will not be exceed.
2. For ASME B 31.1, The above requirements are to be used as a guide but not
for the rejection of radiographs unless the geometrical unsharpness exceeds
0.070 in.
5.7 I.Q.I Selection
1) Image Quality Indicator Design
Penetrameters shall be either the hole type or the wire type and shall be manufactured and
identified in accordance with the requirements or alternates allowed in SE-142 or SE-1025
(for hole type) and SE-747 (for wire type).and Appendices.
2) Material
IQIs shall be selected from either the same alloy material group or grade as identified
in SE-1025 or from an alloy material group or grade with less radiation absorption than
the material being radiographed.
3) Size
The designated hole IQI or essential wire shall be as specified in Table 4. A thinner or
thicker hole type IQI may be substituted for any section thickness listed in Table 5, provided
equivalent IQI sensitivity is maintained. See para.5.8 2).
a) Welds with Reinforcements

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 7 of 35

The thickness on which the IQI is based is the nominal single wall thickness plus the
estimated weld reinforcement not to exceed the maximum permitted by the referencing
Code Section. Backing ring or strip is not to be considered as part of the thickness
in IQI selection. The actual measurement of the weld reinforcement is not required.
b) Welds without Reinforcements
The thickness on which the IQI is based is the nominal single wall thickness. Backing
rings or strips are not to be considered as part of the weld thickness in IQI selection.
5.8 I.Q.I Sensitivity
1) Required sensitivity
Radiographic shall be performed with a technique of sufficent sensitivity to display the
designated hole IQI image and the 2T hole, or the essential wire of a wire IQI. The radiographs
shall also display the IQI identifying numbers and letters. If the designated hole IQI image
and 2T hole, or essential wire, do not show on any film in a multiple film technique, but
do show in composite film viewing.
2) Equivalent Hole-Type Sensitivity
If a thinner or thicker hole-type IQI than listed in table 4 was substituted, an equivalent
IQI sensitivity, as sepecified in table 5, shall have been mailtained as well as all other
requirements for radiographiy having been met.

Table 3] Wire IQI Designation


Set A Set B Set C Set D
Wire Wire Wire Wire Wire Wire Wire Wire
Diameter, in. Identity Diameter, in. Identity Diameter, in. Identity Diameter, in. Identity
0.0032 1 0.010 6 0.032 11 0.100 16
0.004 2 0.013 7 0.040 12 0.126 17
0.005 3 0.016 8 0.050 13 0.160 18
0.0063 4 0.020 9 0.063 14 0.200 19
0.008 5 0.025 10 0.080 15 0.250 20
0.010 6 0.032 11 0.100 16 0.320 21

TABLE 4] MATERIAL THICKNESS, PENETRAMETER DESIGNATIONS, AND ESSENTIAL HOLES


Penetrameter
Nominal single-wall
Source Side Film Side
Material Thickness
Range, in. Hole Type Wire-Type Hole Type Wire-Type
Designation Essential Wire Designation Essential Wire
Up to 0.25, incl 12 5 10 4

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 8 of 35

Over 0.25 through 0.375 15 6 12 5


Over 0.375 through 0.50 17 7 15 6
Over 0.5 through 0.75 20 8 17 7
Over 0.75 through 1.00 25 9 20 8
Over 1.00 through 1.50 30 10 25 9
Over 1.50 through 2.00 35 11 30 10
Over 2.00 through 2.50 40 12 35 11
Over 2.50 through 4.00 50 13 40 12
Over 4.00 through 6.00 60 14 50 13
Over 6.00 through 8.00 80 16 60 14
Over 8.00 through 10.00 100 17 80 16
Over 10.00 through 12.00 120 18 100 17
Over 12.00 through 16.00 160 20 120 18
Over 16.00 through 20.00 200 21 160 20

TABLE 5] DIAMETER OF WIRE PENETRAMETER CORRESPONDING TO HOLE TYPE(1T, 2T, 4T)


Hole Type Designation Equivalent Hole Type Designation
2T Hole 1T Hole 4T Hole
10 15 5
12 17 7
15 20 10
17 25 12
20 30 15
25 35 17
30 40 20
35 50 25
40 60 30
50 70 35
60 80 40
80 120 60
100 140 70
120 160 80
160 240 120
200 280 140

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 9 of 35

5.9 Use of Penetrameters to Monitor Radiographic Examination.


5.9.1 Placement of Penetrameter
a) Pemetrameter Location for Welds - Hole penetrameters.
The penetrameter(s) may be placed adjacent to or on the weld. The identification
number(s) and, when used, the lead letter "F" shall not be in the area of interest,
except for the conditions described in (1) or (2) below:
1) When geometric configuration makes it impractical to place the penetrameter(s) as
outlined in 5.9.1 (a) ; or
2) When the weld metal is not radiographically similar to the base material (as
categorized in SE-142, Annex A1)
b) Penetrameter Location for Welds - Wire Penetrameters.
The penetrameter(s) shall be placed on the weld so that the length of the wires
is perpendicular to the length of the weld. The identification numbers and, when
used, the lead letter "F" shall not be in the area of interest, except for the
condition described in (a)(1) or (2) above.
c) Penetrameter Location for Materials other than Welds.
The penetrameter(s) with the penetrameter identication number(s), and, when used,
the lead letter "F", may by placed in the area of interest.
5.9.2 Number of Penetrameters
For components where one or more film holders are used for an exposure, at least one
penetrameter image shall appear on each radiograph except as outlined in 5.9.2(b) below.
a) Multiple Penetrameters
If the requirements of 5.5 are met by using more than one penetrameter, one shall
be representative of the lightest area of interest and the other the darkest area
of interest ; the intervening densities on the radiograph shall be considered as
having acceptable density.
b) Special Cases
1) For cylindrical vessels where the source is placed on the axis of the object and
one or more film holders are used for a single exposure of a complete circumference,
at least three penetrameters shall be spaced approximately 120 deg. apart. Where
sections of longitudinal welds adjoining the circumferential weld are radiographed
simultaneously with the circumferential weld, an additional penetrameter shall be
placed on each longitudinal weld at the end of each section most remote from the
junction with the circumferential weld being radiographed.
2) For cylindrical vessels where the source is placed on the axis of the object and
four or more film holders are used for a single exposure of a section of the
circumference, at least three penetrameters shall be used. One penetrameter shall

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 10 of 35

be in the approximate center of the section exposed and one at each end. When the
section of the circumference exposed exceeds 240deg., the rules of 5.9.2 (b) (1)
apply. Additional film locations may be required to obtain necessary penetrameter
spacing ; otherwise at least one penetrameter image shall appear on each radiograph.
3) For spherical vessels, where the source is located at the center of the vessel and
one or more film holders are used for a single exposure of a complete circumference,
at least three penetrameters shall be spaced approximately 120 deg. apart. For other
welds radiographed simultaneously, one additional penetrameter shall be placed on
each other weld.
4) For segments of spherical vessels where the source is located at the center of the
vessel and four or more film holders are used for an exposure of a circumferential
weld, at least three penetrameters shall be used. One penetrameter shall be in the
approximate center of the portion exposed and one at each end. When the portion
exposed exceeds 240 deg., the rules of 5.9.2(b)(3) above apply. Additional film
locations may be required to obtain necessary penetrameter image shall appear on
each radiograph.
5) When an array of objects in a circle is radiographed, at least one penetrameter
shall show on each object image.
6) In order to maintain the continuity of records involving subsequent exposures, all
radiographs exhibiting penetrameters which quality the techniques permitted in
accordance with 5.9.2 (b)(1), (2), (3) or (4) above must be retained.
5.9.3 Shims Under Hole Penetrameters
A shim of material radiographically similar to the weld metal shall be placed between
the part and the penetramerter, if needed, so that the radiographic density throughout
the area of interest is no more than minus 15% from (lighter than) the radiographic density
through the penetrameter. The shim dimensions shall exceed the penetrameter dimensions
such that the outline of at least three sides of the penetrameter image shall be visible
in the radiograph and shimed penetrameter shall be placed so as not to overlap the backing
strip or ring.
5.10 Location Markers
Location markers (see Fig 1), which are to appear as radiographic images on the film, shall
be placed on the part, not on the exposure holder/cassette. Their locations shall be
permanently marked on the surface of the part being radiographed when permitted, or on a map,
in a manner permitting the area of interest on a radiograph to be accurately traceable to
its location on the part, for the required retention priod of the radiograph. Evidence shall
also be provided on the radiograph that the required coverage of the region being examined
has been obtained. Location markers shall be placed as follows.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 11 of 35

5.10.1 Single-Wall Viewing


a) Source Side Markers
Location markers shall be placed on the source side when radiographing the following :
1) Flat components or longitudinal joints in cylindrical or conical components.
2) Curved or spherical components whose concave side is toward the source and when the
"source-to-material" distance is less than the inside radius of the component;
3) Curved or spherical components whose convex side is toward the source.
b) Film Side Markers
1) Location markers shall be placed on the film side when radiographing either curved
or spherical components whose concave side is toward the source and when the
"source-to-material" distance is greater than the inside radius.
2) As an alternative to source side placement in 5.10.1(a) (1)above, location markers
may be placed on the film side when the radiograph shows coverage beyond the location
markers to the extent demonstrated by Fig. 1(e).
c) Either Side Markers
Location markers may be placed on either the source side or film side when radiographing
either curved or spherical components whose concave side is toward the source and the
"source-to-material" distance equals the inside radius of the component.
5.10.2 Double-Wall Viewing
For double-wall viewing, at least one location marker shall be placed on the source side
surface adjacent to the weld (or on the material in the area of interest) for each
radiograph.
5.10.3 Mapping the Placement of Location Markers
When inaccessibility or other limitations prevent the placement of markers as stipulated
in 5.10.1 and 5.10.2, a dimensioned map of the actual marker placement shall accompany
the radiographs to show that full coverage has been obtained.
5.11 Verification of Source Size
The equipment manufacturer's or supplier's publications, such as technical manuals, decay
curves, or written statements documenting the actual or maximum source size or focal spot,
shall be acceptable as source size verification.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 12 of 35

Filmside Source side


Source side
unacceptable acceptable
acceptable

Filmside Source side Filmside


unacceptable acceptable unacceptable

Flat component or longitudinal Curved components with radiation source to Curved component with convex
seam[See para. 5.10.1 a) 1)] filmdistance less than radius of component surface towards radiation source
[See sketch (e) for alternate] [See para. 5.10.1 a) 2)] [See para. 5.10.1 a) 1)]
(a) (b) (c)

Either side
location marker
is acceptable
D
t

Source side Filmside x Mf x


unacceptable acceptable

Curved components with radiation Source side marker alternate Flat Curved component with radiation
source to filmdistance greater than component or longitudinal seam source at center curvature
radius of curvature x = (t / D) (Mf / 2) [See para.5.10.1 c)]
[See para.5.10.1 b) 1)] x = additional required coverage,in. (f)
(d) beyond filmside location marker
LEGEND t = component thickness, in.
Radiation source - Mf = filmside location marker interval, in.
Location marker - D = source to component distance,in.
Component center - [See para.5.10.1 b) 2)]
(e)

Fig. 1 Location maker sketches

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 13 of 35

5.12 Step Wedge Film and Densitometer


5.12.1 Densitometer
Densitometer shall be calibrated at least every 90 days during use as follows:
a) A national standard step tablet or a step wedge calibration film, traceable national
standard step tablet and having at least 5 steps with neutral densities from at least
1.0 through 4.0, shall be used. The step wedge calibration film shall have been
verified within the last year by comparison with a national standard step tablet.
b) The densitometer manufacturer's step-by-step instructions for the operation of the
densitometer shall be followed.
c) The density step closest to 1.0, 2.0, 3.0 and 4.0 on the national standard step tablet
or step wedge calibration film shall be read.
d) The densitometer is acceptable if the density readings do not vary by more than ±0.05
density units from actual density stated on the national standard step tablet or step
wedge calibration film
5.12.2 Step Wedge Comparison Films
Step wedge comparison films shall be verified prior to first use, unless performed by
the manufacturer, as follows:
a) The density of the steps on a step wedge comparison film shall be verified by a
calibrated densitometer.
b) The step wedge comparison film is acceptable if the density readings do not vary by
more than ±0.1 density units from the density stated on the step wedge comparison
film.
5.12.3 Periodic Verification
a) Densitometer
Periodic calibration verification checks shall be performed as described in 5.12.1
at the beginning of each shift, after 8 hr of continuous use, or after change of
apertures, whichever comes first. The densitometer is acceptable if the density
readings are within ±0.05 of the calibration readings determined in 5.12.1 c).
b) Step Comparison Films
Verification check shall be performed annually per 5.12.2.
5.12.4 Documentation
a) Densitometer calibration readings required by 5.12.1 c) shall be recorded in an
appropriate calibration log.
b) Periodic verification reading required by 5.12.3 do not have to be recorded.
5.13 Direction of Radiation
The direction of the central beam of radiation should be centered on the area of interest
whenever practical.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 14 of 35

5.14 Quality of Radiographs


All radiographs shall be free from mechanical, chemical or other blemishes to the extent that
they cannot mask or be confused with the image of any discontinuity in the area of interest
of the object being radiographed.
a) fogging
b) processing defects such as streaks, water marks, or chemical stains
c) scratches, finger marks, crimps, dirtiness, static marks, smudges, or tears
d) loss of detail due to poor screen-to-film contact
e) false indications due to defective screens
5.15 Radiographic sensitivity
All radiographs shall be obtained the requirements of 5.5, 5.6, 5.8. To obtained the
requirements of 5.5, 5.6, 5.8, vary the following factor.
5.15.1 Source size
The maximum source size shall be determined to obtain 5.6, 5.8. To obtain the requirements
of Ug, more than smaller source shall be used.
5.15.2 Source to object distance
Minimum source to object distance shall be determined to obtain 5.5, 5.6, 5.8. To obtain
the 5.5, 5.6, 5.8, take the longer source to object distance.
5.15.3 Source side of object to the film distance.
The distance from source side of object to the film shall be decrease to obtain the
radiographic sensitivity.
5.15.4 Film
The type I or II film shall be used as described in 6.1.

6.0 EQUIPMENT AND MATERIALS


6.1 Film
6.1.1 Selection
Radiographs shall be made using industrial radiographic film. The film brand and
designated used are as follows.
Type of industrial Radiographic Film
FILM DESIGNATION
TYPE
KODAK AGFA FUJI
Ⅰ M, T, R D-4,D-5 #50, #80
Ⅱ AA D-7 #100
6.2 Screen
Lead intensifying screen may be used in direct contact with the film. The front and back screens

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 15 of 35

shall be 0.005 inches or 0.01 inches thick.


6.3 Facilities for Viewing of Radiographs
Viewing facilities shall provide subdued background lighting of an intensity that will not
cause troublesome reflections, shadows, or glare on the radiograph. Equipment used to view
radiograph for interpretation shall provide a variable light source sufficent for the essential
penetrameter hole or designated wire to be visible for the specified density range. The viewing
conditions shall be such that light from around the outer edge of the radiograph or coming
through low-density portions of the radiograph does not interfere with interpretation.
6.4 Selection of Energy of Radiation
6.4.1 Radiation
Gamma ray equipment such as Ir-192 or Co-60 may be used. The minimum thickness range used
in gamma ray radiography shall be as specified in Table 6.
Table 6] Minimum Thickness
Material Iridium 192 Cobalt 60
Steel 6mm 38mm
Copper or High Nickel 16.5mm 33mm
Aluminum 63.5mm ...
Note) Overal radigraphic sensitivity is primary influenced by factors such as:
1) film selection
2) intensifying screen selection
3) geometric unsharpness
4) film density
The maximum thickness for the use of radioactive isotopes is primarily dictated by exposure
time; therefore, upper limit are not shown. The minimum recommended thickness limitation
may be reduced when the radiographic techniques used demonstrate that the required
radiographic sensitivity has been obtained.
The minimum recommended thickness limitation of above table 6 (limitation for steel by
Ir-192) was verified by demonstrate to obtaine the required radiographic sensitivity.

7.0 EXAMINATION
7.1 Radiographic Technique
A single-wall exposure technique shall be used for radiography whenever practical. When it
is not practical to use a single-wall technique, a double-wall technique shall be used. An
adequate number of exposures shall be made to demonstrate that the required coverage has been
obtained.
7.1.1 Single-Wall Technique
In the single-wall technique, the radiation passes through only one wall of the weld

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 16 of 35

(material), which is viewed for acceptance on the radiograph.


7.1.2 Double-Wall-Technique
When it is not practical to use a single-wall technique, one of the following double-wall
techniques shall be used.
a) Single-Wall Viewing
For materials and for welds in components, a technique may be used in which the radiation
passes through two walls and only the weld (material) on the film side wall is viewed
for acceptance on the radiographic. When complete coverage is required for
circumferential welds(materials), a minimum of three exposures taken 120 deg. to each
other shall be made.
b) Double-Wall Viewing
For materials and for welds in components 3 1/2 in. or less in nominal outside diameter,
a technique may be used in which the radiation passes through two walls and the weld
(material) in both walls is viewed for acceptance on the same radiograph. For double-
wall viewing, only a source side penetramenter shall be used. Care should be exercised
to ensure that the required geometric unsharpness is not exceeded. If the geometric
unsharpness requirement cannot be met, then single-wall viewing shall be used.
1) For welds, the radiation beam may be offset from the plane of the weld at an angle
sufficient to separate the images of the source side and film side portions of the
weld so that there is no overlap of the areas to be interpreted. When complete
coverage is required, a minimum of two exposures taken 90 deg. to each other shall
be made for each joint.
2) As an alternative, the weld may be radiographed with the radiation beam positioned
so that the images of both walls are superimposed. When complete coverage is
required, a minimum of three exposures taken at either 60 deg. or 120deg. to each
other shall be made for each joint.
3) Additional exposures shall be made if the required radiographic coverage cannot
be obtained using the minimum number of exposures indicated in (b) (1) or (b)
(2) above.

8.0 PROCEDURE
8.1 Preparation Prior to Examination
1) Obtain procedure, drawing and other information.
2) Confirm the area to be examined with the drawings or any other informations.
8.2 Material and Equipment
1) Use following materials and equipment listed below, and record type or KIECO registered
No. on the Examination Report.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 17 of 35

a) X-Ray (or Gamma Ray) b) Film


c) Screen d) Penetrameter
e) Shim f) Other information (if necessary)
8.3 Examined Areas
1) Confirm examined areas on the part, exposure number, and film number.
2) Mark film location line, film No. on inner and outer surfaces of weld examined.
3) Confirm film direction and film top side on inner and outer surface of weld examined.
8.4 Surface Preparation of Weld
1) Check on that the weld ripples or weld surface irregularities which mask or be confused
with the images of any discontinuities had been remove.
2) Check on that weld reinforcements on both inner and outer vessel surface shall not over
the specified in this procedure.
8.5 Identification Markers
1) Check all identification marker on each film.
8.6 Exposure Condition
1) The film density on examined areas shall be covered in 2.0 to 3.5 (min 1.8 to max 4.0 for
X-Ray and max 2.0 to max 4.0 for gamma-Ray)
2) Determine exposure technique.
3) Determine SFD (F.F.D)
4) Determine exposure time using exposure chart (or calculator).
8.7 Placement of Penetrameter
1) Determine essential shim thickness so that the film density on penetrameter shall be equal
to the examined area adjoining it.
2) One penetrameter shall be placed on the nearly center of the examined area and the other
penetrameter with shim shall be placed on the opposite side of the penetrameter.
8.8 "B" Markers
Attach "B" marker for checking the back scattered radiation on the back of the each film holder.
8.9 Processing
8.9.1 Mannual processing
1) The developer shall be maintained at a temperature of 68℉. Development time shall
be adjusted if the temperature is changed more than 2℉. See manufactuler's
recommendations.
2) Solutions shall be stirred prior to the start of processing.
3) Hanger shall be separated by at least 1/2 in. during processing.
4) Film shall be agitated at the start of developing to obtain complete even wetting of
the film and remove any air bubbles. It shall also be agitated periodically during
the development stage increasing to agitation every minute when the developer becomes

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 18 of 35

old.
5) Manufacturer's recommendations for development time shall be followed. (This is usually
5 minutes. It is better to expose the film for the shorter development time.)
6) After the development is complete, rinse the film in water for a few seconds and plunge
into the stop bath to halt the action of the developer. Agitate the film in the stop
bath for the period of time recommended by the manufacturer.
7) Rinse the film in water for a few seconds and plunge it into the fixer. Agitate it
for about 10 seconds. After about one minute agitate again. Normally 10-15 minutes
is the fixing time.
8) The washing efficiency decreases rapidly with decreasing temperatures below 68℉.
Washing time at 68℉ shall be 30 minutes increasing to 40 minutes at 68℉. For
temperatures above 68℉ the washing time shall be decreased to about 20 minutes at
78℉. When the water temperature is above 68℉ the film shall be removed from the wash
immediately following the wash cycle since the film gelatin softens in warm water.
The water flow shall be sufficient to change the volume four times in one hour.
9) When the washing cycle is completed the film shall be agitated in a wetting agent.
10) Film shall be left on their hangers for drying.
8.9.2 Auto processing
The essence of the automatic processing system is control. The processor maintains the
chemical solutions at the proper temperature, agitates and replenishes the solutions
automatically and transports the films mechanically at a carefully controlled speed
throughout the processing cycle. Film characteristics must be compatible with processing
conditions. It is, therefore, essential that the recommendations of the film processor,
and chemical manufactures be followed.
8.10 Checking of Finished Film
1) Check on that the finished film are free of excessive chemical and processing defects.
2) Check on that the film density of the examined area are within 2.0 to 3.5, using densitometer.
3) Check on that the film density through the weld examined do not varies more than –15% or
+30% from the density through the each penetrameter.
4) Check on that the image of "B" mark is not appeared on the each film.
5) Check on that the image of the identification markers were clealy appeared on the each
film.
Note : If the results of each description mentionded above could not be accepted, results
shall be taken appropriately.
8.11 Film Checking Prior to Judgement
1) Check the film quality, image of identification with film cover sheet, image of "B" mark,
film density limitation and variation through the each penetrameter.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 19 of 35

2) If the results of each description mentioned above could not be accepted, reshoot shall
be taken.
8.12 Acceptance Standard
1) Evaluation shall be performed by Level II or III.
2) Acceptance standard shall be in accordance with Para 9.0 of this Procedure.
3) When any images that could interfere with proper interpretation of radiography appeared
on, reshoots for performing proper interpretation shall be taken.
8.13 Report and Record
1) Report results and sign and date on the format with black ball-pen (when accepted)
2) Describe detail of Nonconformance and submit it to Q.C personnel. (when not accepted)

9.0 ACCEPTANCE STANDARD


Welds that are shown by radiography to have any of the following types of discontinuities are
unacceptable ;
9.1 ASME Sec. I
1) Any indication characterized as a crack or zone of incomplete fusion or penetration;
2) Any other elongated indication on the radiograph that has length greater than :
(1) 1/4 in.(6mm) for t up to 3/4 in.(19mm)
(2) 1/3 t for t from 3/4 in.(19mm) to 2 1/4 in.(57mm)
(3) 3/4 in.(19mm) for t over 2 1/4 in.(57mm)
3) Any group of slag inclusions in line that have an aggregate length greater than t in a length
of 12 t, except when the distance between the successive imperfections exceeds 6 L where
L is the length of the longest imperfection in the group.
4) Rounded indications in excess of that specified by the acceptance standards attached (See
Attachment 1).
9.2 ASME Sec. Ⅷ
9.2.1 ASME Sec. Ⅷ Div. 1(Full Radiographic Examination)
1) Any type of crack, or zone of incomplete fusion or penetration.
2) Any elongated slag inclusion which has length greater than :
a) 1/4 in. for t up to 3/4 in.
b) 1/3 t for t from 3/4 in. to 2 1/4 in.
c) 3/4 in. for t over 2 1/4 in.
where ; t = the thickness of the weld
3) Any group of slag inclusions in line that have an aggregate length greater than t in
a length of 12 t, except when the distance between the successive imperfections exceeds
6 L where L is the length of the longest imperfection in the group.
4) Rounded indications in excess of that specified by the acceptance standards attached

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 20 of 35

(See ATTACHENT 1).


9.2.2 ASME Sec. Ⅷ Div 1(Spot Radiographic Examination)
Spot radiographing of a welded joint is recognized as an effective inspection tool. The
spot radiography rules are also considered to be an aid to quality control. Spot radiographs
made directly after a welder or an operator has completed a unit of weld proves that the
work is or is not being done in accordance with a satisfactory procedure. If the work is
unsatisfactory, corrective steps can then be taken to improve the welding in the subsequent
units, which unquestionably will improve the weld quality. Spot radiography in accordance
with these rules will not ensure a fabrication product of predetermined quality level
throughout. It must be realized that an accepted vessel under these spot radiography rules
may still contain defects which might be disclosed on further examination. If all
radiographically disclosed weld defects must be eliminated from a vessel, then 100%
radiography must be employed.
(a) Butt welded joints which are to be spot radiographed shall be examined locally as provided
herein.
(b) Minimum Extent of Spot Radiographic Examination
(1) One spot shall be examined on each vessel for each 50 ft increment of weld or fraction
thereof for which a joint efficiency. However, for identical vessels, each with less
than 50 ft of weld for which a joint efficiency, 50ft increments of weld may be
represented by one spot examination.
(2) For each increment of weld to be examined, a sufficient number of spot radiographs
shall be taken to examine the welding of each welder or welding operator. Under
conditions where two or more welders or welding operator. Under conditions where two
or more welders or welding operators make weld layers in a joint, or on the two sides
of a double-welded butt joint, one spot may represent the work all welders or welding
operators.
(3) Each sot examination shall be made as soon as practicable after completion of the
increment of weld to be examined. The location of the spot shall be chosen by the
Inspector after completion of the increment of welding to be examined, except that
when the Inspector has been notified in advance and cannot be present or otherwise
make the selection, the fabricator may exercise his own judgment in selecting the spots.
(c) Standards for Spot Radiographic Examination.
Spot examination by radiography shall be made in accordance with the technique prescribed
in this procedure. The minimum length of spot radiograph shall be 6in. Spot radiographs
may be retained or be discarded by the Manufacturer after acceptance of the vessel by the
Inspector. The acceptability of welds examined by spot radiography shall be judged by
the following standards.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 21 of 35

(1) Welds in which indications are characterized as cracks or zones of incomplete fusion
or penetration shall be unacceptable.
(2) Welds in which indications are characterized as slag inclusions or cavities shall be
unacceptable if the length of any such indication is greater than ⅔t where t is the
thickness of the weld excluding any allowable reinforcement. For a butt weld joining
two members having different thicknesses at the weld, t is the thinner of these
two thicknesses. If a full penetration weld includes a fillet weld, the thickness
of the throat of the fillet shall be included in t. If several indications within the
above limitations exist in line, the welds shall be judged acceptable if the sum of
the longest dimensions of all such indications is not more than t in a length of 6t
(or proportionately for radiographs shorter than 6t) and if the longest indications
considered are separated by at least 3L of acceptable weld metal where L is the length
of the longest indication. The maximum length of acceptable indications shall be 3/4in.
Any such indications shorter than 1/4in. shall be acceptable for any plate thickness.
(3) Rounded indications are not a factor in the acceptability of welds not required to
be fully radiographed.
(d) Evaluation and Retests
(1) When a spot, radiographed as required in (b)(1) or (b)(2) above, is acceptable in
accordance with (c)(1) and (c)(2) above, the entire weld increment represented by this
radiograph is acceptable.
(2) When a spot, radiographed as required in (b)(1) or (b)(2) above, has been examined
and the radiograph discloses welding which does not comply with the minimum quality
requirements of (c)(1) or (c)(2) above, two additional spots shall be radiographically
examined in the same weld increment at locations away from the original spot. The
locations of these additional spots shall be determined by the Inspector or fabricator
as provided for the original spot examination in (b)(3) above.
(a) If the two additional spots examined show welding which meets the minimum quality
requirements of (c)(1) and (c)(2) above, the entire weld increment represented
by the three radiographs is acceptable provided the defects disclosed by the first
of the three radiographs are removed and the area repaired by welding. The weld
repaired area shall be radiographically examined in accordance with the foregoing
requirements of para. 9.2.
(b) If either of the two additional spots examined shows welding which does not comply
with the minimum quality requirements of (c)(1) or (c)(2) bove, the entire
increments of (c)(1) or (c)(2) above, the entire increment of weld represented
shall be rejected. The entire rejected weld shall be removed and the joint shall
be rewelded weld shall be removed and the joint shall be rewelded or, at the

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 22 of 35

fabricator’s option, the entire increment of weld represented shall be completely


radiographed and only defects need be corrected.
(c) Repair welding shall be performed using a qualified procedure and in a manner
acceptable to the Inspector. The rewelded joint, or the weld repaired areas, shall
be spot radiographically examined at one location in accordance with the foregoing
requirements of para. 9.2.
9.3 ASME B 31.1
1) Any type of crack, or zone of incomplete fusion or penetration.
2) Any other elongated indication which has a length greater than :
(1) 1/4 in.(6.0mm) for t up to 3/4 in.(19.0mm), inclusive;
(2) 1/3 t for t from 3/4 in.(19.0mm) to 2 1/4 in.(57.0mm) inclusive;
(3) 3/4 in.(19mm) for t over 2 1/4 in.(57mm)
where ; t = the thickness of the thinner portion of the weld
3) Any group of slag inclusions in line that have an aggregate length greater than t in a length
of 12 t, except when the distance between the successive imperfections exceeds 6 L where L
is the length of the longest imperfection in the group.
4) Porosity in excess of that specified by the acceptance standards attached(See ATTACHMENT 1).
5) root concavity when there is an abrupt change in density, as indicated on the radiograph.
9.4 Welder and Welding Operator Qualification
1) Linear Indications
a) Any type of crack or zone of incomplete fusion or penetration ;
b) Any elongated slag inclusion which has a length greater than ;
(1) 1/8 in. for t up to 3/8 in., inclusive
(2) 1/3 t for t over 3/8 to 2 1/4 in., inclusive
(3) 3/4 in. for t over 2 1/4 in.
c) any group of slag inclusions in line that have an aggregate length greater than t in
a length of 12t, except when the distance between the successive imperfections exceeds
6L where L is the length of the longest imperfection in the group.
2) Rounded Indications
a) The maximum permissible dimension for rounded indications shall be 20% of t or 1/8 in.,
whichever is smaller.
b) For welds in material less than 1/8 in. in thickness, the maximum number of acceptable
rounded indications shall not exceed 12 in a 6 in. length of weld. A proportionately
fewer number of rounded indications shall be permitted in welds less than 6 in. in length.
c) For welds in material 1/8 in. or greater in thickness, the charts in ATTACHMENT 2 represent
the maximum acceptable types of rounded indications illustrated in typically clustered,
assorted and randomly dispersed configurations. Rounded indications less than 1/32 in.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 23 of 35

in maximum diameter shall not be considered in the radiographic acceptance tests of


welders and welding operators in these ranges of material thickness.

10.0 RADIATION SAFETY


10.1 Personnel performing radiographic examination shall be trained by the Radiological Officer's
License Holder.
10.2 Safety practice shall be applied as outlined in KIECO'S regulation approved by the Manistry
of Science and Technology for control of radiation hazards.

11.0 REPAIR REQUIREMENTS


REQUIREMENTS
Repair shall be examined by the same procedure used for detection of the discontinuities.
Acceptability of repairs shall be determined by the same acceptance standards.

12.0 EXAMINATION RESULTS


Radiographic examination results must be reported on the attached form by the certified Level
II or Level III certifying the results of the examination.

13.0 EXHIBITS
1) Radigraphic Examination Report ----- KIE - TR - R - 31 (03. 10. Rev.1)
2) Radigraphic Examination Report ----- KIE - TR - R - 32 (03. 01. Rev.0)

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 24 of 35

ATTACHMEMT 1

ROUNDED INDICATIONS CHARTS


ACCEPTANCE STANDARD FOR
RADIOGRAPHICALLY DETERMINED REUNDED INDICATIONS IN WELDS

1. APPLICABILITY OF THESE STANDARDS


These standards are applicable to ferritic, austenitic, and nonferrous materials.
2. TERMINOLOGY
(a) Rounded Indications.
Indications Indications with a maximum length of three times the width or less
on the radiograph are defined as rounded indications. These indications may be circular,
elliptical, conical, or irregular in shape and may have tails. When evaluating the size of
an indication, the tail shall be included. The indication may be from any imperfection in
the weld, such as porosity, slag, or tungsten.
(b) Aligned
Ali gned Indications.
Indications A sequence of four or more rounded indications shall be considered
to be aligned when they touch a line parallel to the length of the weld drawn through the
center of the two outer rounded indications.
(c) Thickness t . t is the thickness of the weld, excluding any allowable reinforcement. For
a butt weld joining two members having different thickness at the weld, t is the thinner of
these two thicknesses. If a full penetration weld includes a fillet weld, the thickness of
the throat of the fillet shall be included in t.
3 . ACCEPTANCE CRITERIA
(a) Image Density.
Density Density within the image of the indication may vary and is not a criterion
for acceptance or rejection.
(b) Relevant Indications.(See Table 1 for examples) Only those rounded indications which
exceed the following dimensions shall be considered relevant.
1/10t for t less than 1/8in.
1/64 in. for t form 1/8 in. to 1/4 in., include.
1/32 in. for t greater than 1/4 in. to 2 in., include.
1/16 in, for t greater than 2 in.
(c) Maximum Size of Rounded Indication.(See Table 1 for examples) The maximum permissible
size of any indication shall be 1/4t, or 5/32in., whichever is smaller, except that an isolated

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 25 of 35

indication separated from an adjacent indication by 1 in. or more may be 1/3t, or 1/4in.,
whichever is less. For t greater than 2 in. the maximum permissible size of an isolated
indication shall be increased to 3/8 in.
(d) Aligned Rounded Indications.
Indications Aligned rounded indications are acceptable when the summation
of the diameters of the indications is less than t in a length of 12t. See Fig.1-1. The length
of groups of aligned rounded indications and the spacing between the groups shall meet the
requirements of Fig.1-2.
(e) Spacing. The distance between adjacent rounded indications is not a factor in determining
acceptance or rejection, except as required for isolated indications or groups of aligned
indications.
(f) Rounded Indication Charts.
Charts The rounded indications characterized as imperfections shall
not exceed that shown in the charts. The chart in Figs.1-3 through 1-8 illustrate various
types of assorted, randomly dispersed and clustered rounded indications for different weld
thicknesses greater than 1/8 in. These charts represent the maximum acceptable concentration
limits for rounded indications. The charts for each thickness range represent full scale
6 in. radiographs, and shall not be enlarged or reduced. The distributions shown are not
necessarily the patterns that may appear on the radiograph, but are typical of the
concentration and size of indications permitted.
(g) Weld Thickness t less than 1/8 in. For t less than 1/8 in. the maximum number of rounded
indications shall not exceed 12 in. a 6 in. length of weld. A proportionally fewer number
of indications shall be permitted in welds less than 6 in. in length.
(h) Clustered Indications. The illustrations for clustered indications show up to four times
as many indicaitons in a local area, as that shown in the illustrations for random indications.
The length of an acceptable cluster shall not exceed the lesser of 1 in. or 2t. Where more
than one cluster is present, the sum of the lengths of the clusters shall not exceed 1in.
in a 6in. length weld.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 26 of 35

Maximum Size of Acceptable Rounded Maximum Size of


Thickness t, Indication, in. (mm) Nonrelevatn
in. (mm) Random Isolated Indication, in. (mm)
Less than 1/8 (3.2) 1/4t 1/3t 1/10t
1/8 (3.2) 0.031 (0.79) 0.042 (1.07) 0.015 (0.38)
3/16 (4.8) 0.047 (1.19) 0.063 (1.60) 0.015 (0.38)
1/4 (6.4) 0.063 (1.60) 0.083 (2.11) 0.015 (0.38)

5/16 (7.9) 0.078 (1.98) 0.104 (2.64) 0.031 (0.79)


3/8 (9.5) 0.091 (2.31) 0.125 (3.18) 0.031 (0.79)
7/16 (11.1) 0.109 (2.77) 0.146 (3.71) 0.031 (0.79)
1/2 (12.7) 0.125 (3.18) 0.168 (4.27) 0.031 (0.79)

9/16 (14.3) 0.142 (3.61) 0.188 (4.78) 0.031 (0.79)


5/ 8 (15.9) 0.156 (3.96) 0.210 (5.33) 0.031 (0.79)
0.230 (5.84)
11/16 (17.5) 0.156 (3.96) 0.031 (0.79)

3/4 to 2, incl. 0.156 (3.96) 0.250 (6.35) 0.031 (0.79)


Over 2 0.156 (3.96) 0.375 (9.53) 0.063 (1.60)
NOTE : This table contains examples only.

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


RADIOGRAPHIC

Sum of L1 to LX shall be less than t in a length of 12t.

FIG.4-1 ALIGNED ROUNDED INDICATIONS

KIE - QR - A03 - 2(03. 02. Rev. 0)


EXAMINATION

The Sum of group lengths shall be less than t in a length of 12t.

Maximum Group Length Minimum Group Spacing


L = 1/4 in. for t less than 3/4 in. 3L where L is the
REVISION NO.
DOCUMENT NO.

PROCEDURE

L = 1/3 t for t 3/4in. to 2-1/4in. longest


L = 3/4 in. for greater than 2-1/4 in.

FIG.4-2 GROUPS OF ALIGNED ROUNDED INDICATIONS


27
0

of 35
KIE - TP - R297

TEL:(02)598-2525
DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 28 of 35

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-3 CHARTS FOR t EQUAL TO 1/8 in. to 1/4 in., INCLUSIVE

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-4 CHARTS FOR t OVER 1/4 in. to 3/8in.,

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 29 of 35

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-5 CHARTS FOR t OVER 3/8 in. to 3/4in.,

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 30 of 35

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-6 CHARTS FOR t OVER 3/4 in. to 2in.,

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 31 of 35

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-7 CHARTS FOR t OVER 2 in. to 4in.,

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 32 of 35

RANDOM ROUNDED
Typical concentration and size permitted in any 6 in. length of weld

ISOLATED CLUSTER
Maximum size per Table 5

FIG. 4-8 CHARTS FOR t OVER 4

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 33 of 35

ATTACHMEMT 2

ROUNDED INDICATION CHARTS

Typical quantity and size permitted in 6 in. length of


weld 1/8 in. to 1/4 in. thickness

Typical quantity and size permitted in 6 in. length of


weld over 1/4 in. to 1/2 in. thickness

Typical quantity and size permitted in 6 in. length of


weld over 1/2 in. to 1in. thickness

Typical quantity and size permitted in 6 in. length of


weld over 1in. thickness

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 34 of 35

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525


DOCUMENT NO. KIE - TP - R297

REVISION NO. 0

RADIOGRAPHIC EXAMINATION PROCEDURE 35 of 35

KIE - QR - A03 - 2(03. 02. Rev. 0) TEL:(02)598-2525

You might also like