MEASURE PG I
2-19-6 Yokosuka Reagent for determination of Pepsinogen I
Matsudo, Chiba, Japan Latex Immunoturbidimetric Method

 2 ~ 8 °C IVD In vitro Diagnostics Packages


R1 1  45 mL R2 1  15 mL
 DO NOT freeze  18 months/block from light R1 1  30 mL R2 1  10 mL

1. PURPOSE OF USE and determine after mixing it well (transparent solution) before
In vitro Determination of Pepsinogen I in blood serum or blood use.
plasma.
6. MEASUREMENT PRINCIPLE
2. GENERAL INSTRUCTION When patient sample is reacted with buffer and latex reagents,
1. For in vitro diagnostics use only. specific antigen-antibody reaction is caused by Pepsinogen I in
2. Diagnosis should be made in a comprehensive manner, in patient samples and anti-human Pepsinogen I mouse monoclonal
accordance with other related test results and clinical symptoms antibody sensitized latex, then it gives out turbidity.
by the doctor in attendance. Degree of turbidity is proportional to concentration of Pepsinogen I
3. For guaranteed results, usage of this product must comply with in patient samples, so concentration of Pepsinogen I in patient
the instruction in this manual. samples can be obtained by measuring changes of turbidity.
4. Follow the automatic analyzers instructions carefully.
7. CALCULATION & UNIT CONVERSION
Pepsinogen I and Pepsinogen II are indicators to reflect degree of
Calculation
atrophia of gastric mucosa, and are not specific marker to
- Calculate ∆Abs of specimen & standards vs blank
stomach cancer, therefore high positive ratio is indicated in other
- Plot a calibration curve PGI (ng/mL) = f(∆Abs)
diseases than stomach cancer which accompany atrophia of
- Calculate PGI in specimen using the curve
gastric mucosa.
(doing same procedure for Controls)
This method is not established as an alternative determination
Unit conversion
method to replace radiographic projection in mass-screening.
N/A
3. MATERIALS REQUIRED BUT NOT INCLUDED
8. PERFORMANCE & CORRELATION TEST
- Saline 0.9 % and high grade purified water
Performance
- Micropipet and other basic laboratory equipment.
- Sensitivity: Absorbance was less than 0.02 at 570 nm when
- Calibrators and Controls (separatedly sold)
purified water was used as sample, and change around 0.1 ~ 0.5
4. REAGENT COMPOSITION & PREPARATION in 5 minutes when standard solution of PGI of 100 ng/mL was
- Reagent R-1: Good buffer. used as sample.
- Reagent R-2: Anti-human Pepsinogen I antibody-conjugated - Specificity: The accuracy is within ±10.0%.
latex solution - Reproducibility: CV value < 10.0%.
- Calibrators & Controls (separatedly sold): Calibrators (Level 1 – - Measuring range: 10 ~ 200 ng/mL
Level 6) and controls (L, H) are ready for use wthout dilution. Correlation test
Company A (same principle)
5. SAMPLE PREPARATION & STORAGE
Regression equation: Y = 1.0264X + 5.9878 (n = 69)
- Use serum or plasma as specimens.
Correlation coefficient: R = 0.985
- Collect samples by normal method and determine while they are
Company B (different principle)
fresh.
Regression equation: Y = 1.0020X + 0.2663 (n = 54)
- Analyze sample soon after collection. In case, it could not be
Correlation coefficient: R = 0.997
analyzed soon, store samples at -20 ℃ in freezer and analyze
(Y: Value obtained from using UMA’s reagent)
within 1 month (avoid freeze and thaw).
- Between Serum and Plasma
- When using frozen samples, thaw them in room temperature
Regression equation: Y = 0.9969X - 0.0371 (n=54)

1/2 Revision 11/2014


Correlation coefficient: R = 0.999 Calibration type Spline
(Y: Serum; X: Plasma) Unit ng/mL

Reference Materials for Calibration


- N/A 12. HANDLING, USAGE & DISPOSAL

9. EXPECTED VALUES Handling

Measured Value 1. Specimen can be potentially positive for infectious agents


Judgment
Pepsinogen I and PGI/PGII ratio including hepatitis B virus and HIV. Wear glove and goggle when

Strong Positive below 30.0 and below 2.0 needed.

Medium positive below 50.0 and below 3.0 2. In case reagents got into skin, eye or mouth by mistake, wash it
immediately with plenty of water and consult the doctor if needed.
Positive below 70.0 and below 3.0
3. If reagents are spilled, dilute with water and wipe it out. If
- Correct measurement results cannot be obtained in case non-
specimen is spilled, spray 80% of alcohol over the specimen and
specific reaction materials (heterophile antibody etc.) exist in the
wipe it out.
samples.
Usage
- Reference range should be established at each facility and
1. Store reagents under specified condition. Do not use after
judgement should base on measurement results in a
expiration date.
comprehensive manner together with clinical symptoms and other
2. Do not use the container and auxiliaries included in this kit for
measurement results.
other purposes.
10. INTERFERENCES
3. Do not mix reagents of different lot for use.
No influence of conjugated bilirubin, free bilirubin, hemolysis,
4. Do not add to the reagent being used even if it is the same lot
lipemia and ascorbic acid against determination were observed by
number.
internal experiments.
Disposal
11. INFORMATION FOR AUTOANALYZERS 1. All specimens, as well as all instruments (e.g. test tubes) that
 For Hitachi Analyzers come in contact with the specimens, must be treated by the
Calculation Method 2-end point (fix) following methods:
Temperature 37 C ・ Sterilize with an autoclave, subjecting them to high pressure
Specimen 6.0 saturated steam at 121 °C for more than 20 minutes. Do not
Volume (μL) R1 150
process waste containing sodium hypochlorite solution with an
R2 50
autoclave.
Main 570
Wavelength (nm) ・ Immerse at least one hour in sodium hypochlorite solution
Sub- -
(active chloride concentration of over 1000 ppm).
Point 1 10
Measurement
Point 2 19 2. This reagent contains sodium azide. Sodium azide can react
(cycle)
Point 3 34 with lead pipe and/or steel pipe and can generate explosive metal

Calibration type Spline azide. Make sure to use plenty of water at disposal. Concentration
Unit ng/mL of sodium azide in R-2 is 0.05%.

13. OTHER INSTRUCTIONS AND CAUTION


 For Beckman Coulter Analyzers
- Results may differ depending on the sample/reagent ratio.
Calculation Method FIXED
Adjust parameters for different analyzer.
Temperature 37 C
- Prepare the calibration curve on the day of determination.
Specimen 6.0
Volume (μL) R1 150
R2 50
Main 570
Wavelength (nm)
Sub- -
Point 1 12
Measurement Point 2 27
(cycle) Point 3 -
Point 4 -

2/2 Revision 11/2014