The document summarizes biomedical maintenance work orders for Divine Grace Medical Center from March to August 2021. Key maintenance tasks included preventative maintenance, calibration, and repairs of various medical equipment such as physiologic monitors, aspirators, sphygmomanometers, and electrosurgical units. The work orders describe equipment issues, maintenance performed, and test results, with the goal of ensuring all equipment is functional and safe for patient care.
The document summarizes biomedical maintenance work orders for Divine Grace Medical Center from March to August 2021. Key maintenance tasks included preventative maintenance, calibration, and repairs of various medical equipment such as physiologic monitors, aspirators, sphygmomanometers, and electrosurgical units. The work orders describe equipment issues, maintenance performed, and test results, with the goal of ensuring all equipment is functional and safe for patient care.
The document summarizes biomedical maintenance work orders for Divine Grace Medical Center from March to August 2021. Key maintenance tasks included preventative maintenance, calibration, and repairs of various medical equipment such as physiologic monitors, aspirators, sphygmomanometers, and electrosurgical units. The work orders describe equipment issues, maintenance performed, and test results, with the goal of ensuring all equipment is functional and safe for patient care.
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
DEPARTMENT / AREA : CSR
WORK ORDER NO MGD/CM/000006 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/23/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG ECG Test (Passed) Sp02 Test (Passed) EQUIPMENT NO. MGD1000129 NIBP Test (Passed) MODEL NAME: PRO 1 NIBP Leak Test (Passed) MANUFACTURER NAME OPRIMA Equipment is functional ASSET DESCRIPTION MONITORS, BEDSIDE, PHYSIOLOGIC
WORK ORDER NO MGD/CM/000007 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/22/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) FOR REPLACEMENT OF POWER CORD WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG ECG Test (Passed) Sp02 Test (Passed) EQUIPMENT NO. MGD1000128 NIBP Test (Passed) MODEL NAME: PRO 1 NIBP Leak Test (Passed) MANUFACTURER NAME OPRIMA Equipment is functional ASSET DESCRIPTION MONITORS, BEDSIDE, PHYSIOLOGIC FOR REPLACEMENT OF POWER CORD
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
CSR Page 1 of 1 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
DEPARTMENT / AREA : INTENSIVE CARE UNIT
WORK ORDER NO MGD/CM/000003 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/30/2021 PM AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG BP Manometer Accuracy Test (Passed) BP Leak Test (Passed) EQUIPMENT NO. MGD1000131 Equipment is functional MODEL NAME: ROLL-BY MANUFACTURER NAME WELCH ALLYN ASSET DESCRIPTION SPHYGMOMANOMETERS, ANEROID
WORK ORDER NO MGD/CM/000004 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/30/2021 PMS Performed Preventive Maintenance Equipment is functional WORK ORDER STATUS Completed Physical Inspection (Passed) Functional Testing (Passed) WORK ORDER TYPE OTHERS Suction Test (Passed) ASSIGNED TO LARRY CALUBAG - Max: 0.08 MPa Cleaning EQUIPMENT NO. MGD1000130 Equipment is functional MODEL NAME: T-150 MANUFACTURER NAME TMS ASSET DESCRIPTION ASPIRATORS
WORK ORDER NO MGD/CM/000005 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/30/2021 PMS Performed Preventive Maintenance Equipment is functional WORK ORDER STATUS Completed Physical Inspection (Passed) Functional Testing (Passed) WORK ORDER TYPE OTHERS Suction Test (Passed) ASSIGNED TO LARRY CALUBAG - Max: 0.08 MPa Cleaning EQUIPMENT NO. MGD1000132 Equipment is functional MODEL NAME: T-150 MANUFACTURER NAME TMS ASSET DESCRIPTION ASPIRATORS
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
INTENSIVE CARE UNIT Page 1 of 1 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
DEPARTMENT / AREA : OPERATING ROOM
WORK ORDER NO MGD/CM/000001 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/14/2021 Low Output Power Physical Inspection Equipment is defective WORK ORDER STATUS Completed Functional Testing with Fluke Electrosurgical Equipment is obsolete as of July 2005 (Please see attached Analyzer(Failed) Oblolescence Letter from manufacturer) WORK ORDER TYPE CORRECTIVE MAINTENANCE Set Cut Power=300 watts | Output=38.4 watts ASSIGNED TO LARRY CALUBAG Set Coag Power=120 watts | Output= 19.7 watts EQUIPMENT NO. MGD1000126 Output Energy Calibration (Failed) MODEL NAME: FORCE 2-20PC - no output response during calibration MANUFACTURER NAME VALLEYLAB ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
WORK ORDER NO MGD/CM/000002 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Aug/02/2021 Request for technical evaluation Perform test and evaluation. Equipment is Created equipment decommissioning report, refer to EDR WORK ORDER STATUS Completed subject for decommissioning Control No. EDRMGD000001
WORK ORDER TYPE CORRECTIVE MAINTENANCE
ASSIGNED TO LARRY CALUBAG
EQUIPMENT NO. MGD1000126
MODEL NAME: FORCE 2-20PC MANUFACTURER NAME VALLEYLAB ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
WORK ORDER NO MGD/CM/000008 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG BP Manometer Accuracy Test (Passed) BP Leak Test (Passed) EQUIPMENT NO. MGD1000121 Equipment is functional MODEL NAME: ROLL-BY MANUFACTURER NAME WELCH ALLYN ASSET DESCRIPTION SPHYGMOMANOMETERS, ANEROID
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
OPERATING ROOM Page 1 of 5 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
WORK ORDER NO MGD/CM/000009 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG BP Manometer Accuracy Test (Passed) BP Leak Test (Passed) EQUIPMENT NO. MGD1000118 Equipment is functional MODEL NAME: ROLL-BY MANUFACTURER NAME WELCH ALLYN ASSET DESCRIPTION SPHYGMOMANOMETERS, ANEROID
WORK ORDER NO MGD/CM/000010 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS Performed Preventive Maintenance Equipment is functional WORK ORDER STATUS Completed Physical Inspection (Passed) Functional Testing (Passed) WORK ORDER TYPE OTHERS O2 Leak Test (Passed) ASSIGNED TO LARRY CALUBAG Checking of O2 Fittings (Passed) Cleaning EQUIPMENT NO. MGD1000117 Equipment is functional MODEL NAME: 110 ANESTHETIC UNIT MANUFACTURER NAME SIMS NEU PAC ASSET DESCRIPTION ANESTHESIA UNITS
WORK ORDER NO MGD/CM/000011 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Breath Rate Test (Passed) I:E Test (Passed) EQUIPMENT NO. MGD1000116 Equipment is functional MODEL NAME: ULCO MEDICAL MANUFACTURER NAME ULCO ASSET DESCRIPTION ANESTHESIA UNITS
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
OPERATING ROOM Page 2 of 5 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
WORK ORDER NO MGD/CM/000012 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Breath Rate Test (Passed) I:E Test (Passed) EQUIPMENT NO. MGD1000120 Equipment is functional MODEL NAME: FABIUS PLUS MANUFACTURER NAME DRAGER ASSET DESCRIPTION ANESTHESIA UNITS
WORK ORDER NO MGD/CM/000013 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Breath Rate Test (Passed) I:E Test (Passed) EQUIPMENT NO. MGD1000122 Equipment is functional MODEL NAME: AEON 7400A MANUFACTURER NAME AEONMED ASSET DESCRIPTION ANESTHESIA UNITS
WORK ORDER NO MGD/CM/000014 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG ECG Test (Passed) Sp02 Test (Passed) EQUIPMENT NO. MGD1000119 NIBP Test (Passed) MODEL NAME: PRO 1 NIBP Leak Test (Passed) MANUFACTURER NAME OPRIMA Equipment is functional ASSET DESCRIPTION MONITORS, BEDSIDE, PHYSIOLOGIC
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
OPERATING ROOM Page 3 of 5 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
WORK ORDER NO MGD/CM/000015 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/02/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG ECG Test (Passed) Sp02 Test (Passed) EQUIPMENT NO. MGD1000115 NIBP Test (Passed) MODEL NAME: SVM-7621 NIBP Leak Test (Passed) MANUFACTURER NAME NIHON KOHDEN Equipment is functional ASSET DESCRIPTION MONITORS, BEDSIDE, PHYSIOLOGIC
WORK ORDER NO MGD/CM/000016 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/14/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Cut Power Output Calibration Coag Power Output Calibration EQUIPMENT NO. MGD1000123 Cleaning MODEL NAME: FORCE 2-20PC Equipment is functional MANUFACTURER NAME VALLEYLAB ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
WORK ORDER NO MGD/CM/000017 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/14/2021 PMS AMD CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Cut Power Output Calibration Coag Power Output Calibration EQUIPMENT NO. MGD1000124 Cleaning MODEL NAME: FORCE 40S-20 Equipment is functional MANUFACTURER NAME VALLEYLAB ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
OPERATING ROOM Page 4 of 5 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
WORK ORDER NO MGD/CM/000018 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/14/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Cut Power Output Calibration Coag Power Output Calibration EQUIPMENT NO. MGD1000125 Cleaning MODEL NAME: SYSTEM 5000 Equipment is functional MANUFACTURER NAME CONMED ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
WORK ORDER NO MGD/CM/000019 NATURE OF PROBLEM ASSESSMENT/WORK DONE STATUS/RECOMMENDATIONS
WO DATE Jul/14/2021 PMS AND CALIBRATION Performed Preventive Maintenance and Equipment is functional WORK ORDER STATUS Completed Calibration Physical Inspection (Passed) WORK ORDER TYPE OTHERS Functional Testing (Passed) ASSIGNED TO LARRY CALUBAG Cut Power Output Calibration Coag Power Output Calibration EQUIPMENT NO. MGD1000127 Cleaning MODEL NAME: FORCE 2-20PC Equipment is functional MANUFACTURER NAME VALLEYLAB ASSET DESCRIPTION ELECTROSURGICAL UNITS, MONOPOLAR
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY:
Reports generated by CEMS
OPERATING ROOM Page 5 of 5 MT. GRACE RESOURCES MANAGEMENT, INC. BIOMEDICAL MAINTENANCE SERVICES GROUP 132 PIONEER ST., MANDALUYONG CITY
CM WORK ORDER SUMMARY
DIVINE GRACE MEDICAL CENTER Hospital Name PERIOD OF Mar/01/2021 - Aug/31/2021
PREPARED BY: REVIEWED BY: VERIFIED AND ACKNOWLEDGED BY: