Reference TOTL-WI-F15

Office Egypt

ISO/IEC 17025 Form Issue n° / Date 00 / 01 July 20


Working Instruction Forms Author / Approver M. Saleh / M. Abd El Razek

Working Instructions Page n° Page 1 of 1

CHNS verification by Petroleum coke CRM


Equipment Name: CHNS analyzer Serial No.: 502-684
Manufacturer: Leco
Nominal Value: Max. Possible Error
Carbon 88.5 ± 1.30
Hydrogen 3.91 ± 0.29
Nitrogen 1.40 ± 0.17
Sulfur 4.77 ± 0.19

Date Result Conclusion Verifier Signature


C= Pass Fail
H= Pass Fail
N= Pass Fail
S= Pass Fail
C= Pass Fail
H= Pass Fail
N= Pass Fail
S= Pass Fail
C= Pass Fail
H= Pass Fail
N= Pass Fail
S= Pass Fail

Instructions:
• The CHNS Verification shall be done Every working day.