Professional Documents
Culture Documents
11.0 Pressure Transmitter Calibration
11.0 Pressure Transmitter Calibration
No: INST-SOP-021
Rev: 0
CONTROLLED COPY
INST-SOP-021
Doc. No: INST-SOP-021
Rev: 0
CONTROLLED COPY
APPROVALAND REVISION
APPROVAL
This procedure has been reviewed and approved for implementation throughout Roltex
Resources Nigeria Limited.
REVISION HISTORY
Rev: 0
CONTROLLED COPY
8. Apply 0% pressure as per LRV with hand held test pump and check multimeter if it is not 4ma
adjust the zero pot in the transmitter and correct transmitter output to 4ma
9. Apply 100%pressure as per the URV and correct 20ma in multimeter by adjusting span pot in the
transmitter
10. Repeat these steps to rectify error.
Doc. No: INST-SOP-021
Rev: 0
CONTROLLED COPY
1. We have to use HART communicator, connect the communicator with the transmitter select the
HART Communicator Menu for lower range value trim and upper range value trim.
2. Basic Set up – Calibration – Zero Trim/Sensor Trim —Lower/Upper range value trims.
3. HART communicator will automatically calibrate the transmitter.
4. Restore the process connection
5. Take the transmitter on line. Ensure there is no
leak
Rev: 0
CONTROLLED COPY
Calibration Procedure :
1. Set up the differential pressure transmitter, HART communicator, power supply,
hand pump, and the multimeter as below (see below calibration setup Diagram).
2. Make sure the equalizing valve manifold is closed.
3. Apply a pressure to the transmitter equal to a lower range pressure (usually it
correspond to 4 mA in the transmitter output). For example we have 0 to 100
mBar calibrated range, then the lower range pressure is 0, or let’s say we have -2
psig to 5 psig then we have lower range pressure equal to -2 psig.
4. Read the pressure in the transmitter LCD (or in the HART communicator).
Adjust (if any) through the HART communicator so that the output of the
transmitter (on LCD) is the same with the applied pressure.
5. Read the mA output of the transmitter by using a multimeter. Adjust (if any)
through the HART communicator so that the output of the transmitter (on
multimeter) is 4 mA.
6. Apply a pressure to the transmitter equal to an upper range pressure (usually it
correspond to 20 mA in the transmitter output).
Doc. No: INST-SOP-021
Rev: 0
CONTROLLED COPY
7. Read the pressure in the transmitter LCD (or in the HART communicator).
Adjust (if any) through the HART communicator so that the output of the
transmitter (on LCD) is the same with the applied pressure.
8. Read the mA output of the transmitter by using a multimeter. Adjust (if any)
through the HART communicator so that the output of the transmitter (on
multimeter) is 20 mA.
Doc. No: INST-SOP-021
Rev: 0
CONTROLLED COPY