Professional Documents
Culture Documents
SUMMARY Despite the widespread clinical appliciktion of cardiopulmonary resuscitation (CPR), the
mechanism responsible for blood flow during this maneuver remains undefined, although it has been assumed
that blood is squeezed from the heart by direct compression of the sternum. We studied the hemodynamics of
CPR in 15 arrested dogs. During chest compression, pressures in the left ventricle, aorta, right atrium and
pulmonary artery were essentially identical. These pressures were also equal to the intrathoracic. pressure as
estimated by an esophageal balloon catheter. Unequal transmission of pressures to the extrathoracic arterial
and venous system resulted from collapse of the great veins at the thoracic outlet as intrathoracic pressures
rose. This phenomenon gave rise to a peripheral arteriovenous pressure gradient and antegrade flow. When in-
trathoracic pressure was increased by maintaining the lungs fully inflated during chest compression, aortic
systolic pressure rose from 27.3 ± 4.0 mm Hg to 58.4 7.9 mm Hg (p < 0.001) and carotid blood flow in-
creased from 9.0 ± 2.2 ml/min to 28.6 ± 5.9 ml/min (p < 0.001). Increasing the intrathoracic pressure by
tightly binding the abdomen to prevent paradoxical diaphragmatic motion during chest compression also
resulted in a rise in aortic systolic pressure, from 29.4 ± 3.2 to 57.7 ± 7.7 mm Hg (p < 0.001), and an in-
crease in carotid blood flow, from 14.5 ± 8.1 ml/min to 32.3 9.7 ml/min (p < 0.005). It appears that
pressure generation and blood flow during CPR in the dog result from a generalized rise in intrathoracic
pressure, not from direct cardiac compression. Maneuvers that raise the intrathoracic pressure can
dramatically increase carotid blood flow during CPR.
THE CONCEPT that blood moves during car- motion of the rest of the chest wall was restricted by
diopulmonary resuscitation (CPR) as a result of direct placing a belt around the thoracic cage that sternal
compression of the heart between the sternum and the compression resulted in a detectable rise in arterial
vertebral column was originally proposed by pressure. All of these observations suggested to us that
Kouwenhoven, Jude and Knickerbocker' in 1960. the common denominator for movement of blood dur-
Although contradictory views have been expressed,2' ing CPR in man was an increase in intrathoracic
Downloaded from http://ahajournals.org by on April 11, 2020
this concept remains widely accepted despite the lack pressure, not direct cardiac compression.
of supporting scientific data. Recently, Taylor and This notion was further supported by the observa-
associates4 reported that prolonged compression is tion of Criley and associates,5 who demonstrated that
vital in optimizing carotid flow in man during CPR. repeated coughing alone without chest compression
At a constant compression force, we repeatedly can produce effective arterial pressure and sufficient
observed that the compression initiated at end- cerebral blood flow to maintain the conscious state in
inspiration was characterized by augmented arterial patients during ventricular fibrillation.
pressure and blood flow. Because the lung is inflated, In this study, we explored the hypothesis that a rise
this compression is one in which the sternum is farther in the intrathoracic pressure results in movement of
from the vertebral column and there should be less blood during chest compression and tried to clarify
direct compression of the heart. In addition, how intrathoracic pressure might be transmitted un-
emphysematous patients with increased antero- equally to the peripheral arterial and venous tree to
posterior chest dimension and relatively small heart produce the peripheral arteriovenous pressure
size are readily resuscitated despite a thoracic gradient necessary for blood flow. A preliminary
anatomy that makes direct cardiac compression un- report has been presented.6
likely. We also observed two patients with flail ster-
nums as a result of previous chest trauma who suffered Methods
cardiac arrest and noted that their arterial blood
pressure did not rise during chest compression, Studies were performed in 15 mongrel dogs that
although a flail sternum should allow ready compres- weighed 20-45 kg. After sodium pentobarbital
sion of the heart. It was only when the paradoxical anesthesia (25 mg/kg i.v.), vessels were exposed in the
neck and inguinal areas. All dogs were intubated with
a cuffed endotracheal tube. Millar micromanometer
catheters were placed via the right femoral artery and
From the Peter Belfer Laboratory for Myocardial Research, Car- vein into the descending thoracic aorta and right
diology Division and Department of Medicine, The Johns Hopkins
Medical Institutions, Baltimore, Maryland. atrium, respectively. Fluid-filled catheters were placed
Supported by grants P-50 HL 17655-03 and 5 T32 HL 07227 in the left ventricle and pulmonary artery via the left
from the NHLBI, NIH. femoral vessels and in the left external jugular vein
Address for correspondence: Myron L. Weisfeldt, M.D., The and carotid artery via side branches. In addition, a
Johns Hopkins Medical Institutions, Baltimore, Maryland 21205.
Received December 27, 1978; revision accepted July 26, 1979. balloon-tipped, fluid-filled catheter was passed via the
Circulation 61, No. 2, 1980. oral cavity into the thoracic portion of the esophagus.
345
346 ClIRCULATION VOL 61, No 2, FEBRUARY 1980
Pressures in the fluid-filled catheters were measured and after each additional protocol differed by less than
with Statham P23Db transducers. Catheter positions 15%, falling slightly in most dogs. Arterial Po2 was
were confirmed by characteristic pressure recordings >300 mm Hg in all dogs and Pco2 was 22-35 mm Hg.
and position at the termination of the study. Mean
systemic pressure was determined periodically by Results
stopping the chest compression and ventilation and Pressure and Flow Pattern During CPR
allowing 20 seconds for equilibration of intravascular
pressures. Millar catheters were calibrated against Each chest compression produced positive pressure
mercury in vitro and adjusted for baseline drift in vivo pulses of similar magnitude in the right atrium, left
by comparison with pressure recorded through the ventricle, aorta and carotid artery (fig. 1) and
lumen of the catheter. After administration of heparin antegrade carotid blood flow (12.3 ± 2.7 ml/min)
(300 units/kg i.v.), cannulating electromagnetic flow (mean ± SEM for five-beat cycle in 13 dogs). When
probes were placed in the right carotid artery and ex- compression was released, there was significant
ternal jugular vein and flow was measured with a retrograde carotid blood flow (3.3 ± 1.0 ml/min), as
Biotronix BL 613 sine-wave flow meter. All data were well as antegrade jugular flow (fig. 2). There was
recorded on a Brush Mark 600 recorder. Minute flows negligible retrograde jugular flow during chest com-
were subsequently calculated by integration of the in- pression despite a large pressure gradient between the
stantaneous flow trace using a Hewlett Packard right atrium and the jugular vein outside the chest
9810A calculator and 9468A digitizer. In expressing (table 1). Jugular pressure rose slowly during systole,
these results, antegrade flow refers to flow in the probably as a result of antegrade filling.
jugular vein toward the chest and in the carotid artery The systolic aortic pressure was dependent on the
away from the chest. state of lung inflation. Beat 1, the first chest compres-
After recording of control pressures and flows, ven- sion after lung inflation, produced higher systolic aor-
tricular fibrillation was induced with a transthoracic tic pressures than the subsequent beats of the com-
electrical current. Anteroposterior external chest com- pression cycle that occurred with the lungs deflated
pression was initiated after 15 seconds by means of a (table 1). This augmentation of aortic pressure was
pneumatic chest compression device set to compress produced by lung inflation rather than by the
the chest once per second with a compression duration associated prolonged diastolic pause, because the
of 0.5 seconds and a compression force of 140 pounds.
These procedures were not changed during the course CONTROL
of the study. Every fifth compression, diastole was
Downloaded from http://ahajournals.org by on April 11, 2020
0-
kr i-: 1-.
v
11-
4-
m
-l.'
CONTROL
-1
i;
.i
.11tA1t11 AR .A ill.,11
1.
Systolic
80- Pressure
(mmHg)
JUGULAR ,,J
(rmrHg) 'A 11 h P\ f'J
0-
80-
PA --------- K
(nmmHg)
80-
n=9
RA
_-j :- t :k f,' 1k c-
(mn*4g)
FIGURE 3. Aortic and esophageal systolic pressure during
80-
Ao
(mmHg)
cardiopulmonary resuscitation. Systolic aortic and es-
\, /
4A- -
X ophageal pressures are nearly identical. Corresponding aor-
0-
tic and esophageal pressure are 69.9 ± 6.4 and 70.1 + 5.8
FIGURE 2. Further pressures and jugular flow during car- mm Hg for beat 1, 32.4 ± 3.7 and 33.4 ± 3.3 mm Hg for
diopulmonary resuscitation. During chest compression, beat 3, and 31.3 ± 3.5 and 32.4 ± 3.2 mm Hg for beat 5.
jugular pressure is less than other pressures shown. When
chest compression is released there is jugular flow into the pression cycle, but higher than jugular venous pressure
chest. With chest compression, however, there is only a (table 2).
brief, small, retrograde jugularflow followed by essentially The near-equality of systolic pressures in each of the
no jugular flow despite the presence of a pressure gradient cardiac chambers, the intrathoracic great vessels and
from right atrium to jugular vein. Jugular venous, the intrathoracic pressure as reflected in the esoph-
pulmonary artery (PA), right atrial (RA) and aortic (Ao) agus, and the variation of all these pressures with
pressures are shown. the state of lung inflation, strongly suggests that the
Downloaded from http://ahajournals.org by on April 11, 2020
EXPIRATION INSPIRATtON
CAROTID +
FOW
0-
Occlude
Airway
80 : eniato
3 FIGURE 4. Airway occlusion at end-
Ao
(mmrHq)
inspiration. All pressures and carotid blood
0- flow increased during subsequent chest com-
80- pressions. A ortic (A o), right atrial (RA), left
RA ventricular (L V) and esophageal (ESOPH)
(mmHg) pressures are shown.
0-
80-
LV
(mmHg) O-
ESOPH80-
fmmhlg) -i
Of. 1 1
0
,\ 11-1-.'-- -111
o------
tilation immediately before beat 1. The increased remaining change in mean systemic pressure reflected
systolic pressures were accompanied by an increase in increased transmural filling pressure probably related
carotid blood flow in 12 of the 13 dogs studied (fig. 6). to an increase in effective intravascular volume. The
Chest compression with the lung held inflated again second major effect was a rise in systolic pressure in
produced nearly identical systolic aortic and es- excess of that accounted for by the diastolic rise
ophageal pressures of 69.1 ± 8.4 and 66.9 ± 7.2 mm produced nearly identical systolic aortic and esoph-
Hg for beat 1 and 58.4 ± 7.9 and 60.3 ± 6.2 mm Hg ageal pressures of 69.1 ± 8.4 and 66.9 i 7.2 mm
for beat 3. Pressures in the right atrium, ventricle and tic pressure. The difference between systolic aortic and
pulmonary artery were also nearly identical to aortic esophageal pressure was only 2.0 ± 1.8 mm Hg.
Downloaded from http://ahajournals.org by on April 11, 2020
pressure. These data confirm the dependence of in- Systolic carotid pressure increased for each beat (fig.
travascular pressures on intrathoracic pressure and in- 7). Carotid blood flow increased in 13 of 15 dogs
dicate the ability to effect changes in blood flow and studied (fig. 8).
pressure through manipulation of intrathoracic The relative importance of the volume shift vs the
pressure. increased systolic pressure in the augmentation of
Abdominal Binding carotid blood flow was assessed by intravenous infu-
sion 1000-2000 ml of normal saline in six dogs. Mean
Tightly binding the abdomen in 15 dogs had two systemic pressure increased 6.7 ± 2.0 mm Hg (com-
major effects (fig. 7). First, mean systemic pressure pared with 9.5 mm Hg with abdominal binding).
rose from 10.7 to 20.0 mm Hg (p < 0.001). Only However, carotid blood flow did not increase
2.6 ± 1.1 mm Hg of this rise could be accounted for significantly. These data suggest that the rise in
by a rise in diastolic intrathoracic pressure. The systolic pressure accounts for most of the increased
p<.OOI * p<.O5
Aortic Carotid Alortic Carotid Aortic Carotid FIGURE 5. Aortic and carotid arterial
I
m 80 systolic pressure during cardiopulmonary
E resuscitation before and after airway occlu-
E
60 sion at peak inspiration. After airway occlu-
a) sion, aortic pressures for beats 3 and 5 and
tn carotid pressure for beats 1, 3 and 5 were
A) 40 higher.
O 20
(P
o
Beat Beat 3 Beat 5
BLOOD FLOW DURING CPR/Rudikoff et al. 349
1H1
Carotid Carotid
Blood Flow Blood Flow
(mI/min) (mI/min)
n=13 0
~~~~~~~~~~~n=15
0
Before After Before After
FIGURE 6. Airway occlusion at end-inspiration. In 12 of FIGURE 8. Abdominal binding. Carotid blood flow in-
13 dogs studied, carotid blood flow increasedfrom 9.0 ± 2.2 creased in 13 of 15 dogs, from 14.5 ± 8.1 ml/min to
ml/min to 28.6 ± 5.9 ml/min (p < 0.001). 32.3 ± 9.7 ml/min (p < 0.005).
blood flow produced by abdominal binding. In addi- rose for each beat (table 3). In seven dogs there was an
tion, abdominal binding may decrease the size of the increase in carotid flow, from 18.3 ± 9.1 to
perfused arterial bed. 33.9 ± 14.1 ml/min. However, carotid collapse oc-
curred in the remaining seven dogs and there was a fall
Airway Occlusion After Abdominal Binding
Downloaded from http://ahajournals.org by on April 11, 2020
CAROTID o- -
FLOW
I
FIGURE 9. Airway occlusion at end-
.. .......
.. .... ... a.rwa ....s inspiration after abdominal binding. Carotid
RA
(rnr ,^ ,:
'i3-- 4't blood flow fell markedly and a large
RK pres.vsure gradient developed between the
-77-
.-
t --
-- aorta and carotid artery, indicating carotid
---:....... -f -------7
collapse. RA = right atrial pressure; A o
160- aortic pressure; ESOPH - esophageal
ESOPH K
(mmHgI
TI->
.,......
pressure.
OAROT
,mHg)
This rise in intrathoracic pressure is reflected as an cur just inside the thorax. The high resistance of the
equal rise in pressure in each of the cardiac chambers collapsed segment of vessel and negligible driving
and intrathoracic great vessels. The relatively small gradient result in little or no flow.8' 9
gradient between the aorta and carotid artery in- Our data indicate that this collapse phenomenon is
dicates that there is relatively direct transmission of less marked on the arterial side. The arterial system is
the pressure from the arterial system within the chest intrinsically somewhat stiffer than the venous system
to the arterial system outside the chest. Despite a large and may thereby be less subject to collapse from the
apparent pressure gradient, there is minimal flow of high surrounding intrathoracie pressure. In addition,
blood from the intrathoracic venous system to the ex- because the capacity of the extrathoracic arterial
trathoracic veins. This results in a smaller systolic system is less than that of the venous system, any
pressure rise in the extrathoracic veins with chest com- blood leaving the thorax will produce a relatively
pression (fig. 2). This unequal transmission of in- greater extrathoracie intravascular pressure, which
trathoracic pressure to the extrathoracic arterial and will tend to prevent vascular collapse with rapid move-
venous systems results in a peripheral systolic ment of blood from the thoracic arteries at the onset
arteriovenous pressure gradient with antegrade blood of compression. Clearly, complete collapse of arterial
flow. When chest compression is released, in- structures does occur under conditions of extremely
Downloaded from http://ahajournals.org by on April 11, 2020
trathoracic pressure falls below venous pressure and high intrathoracic pressure (fig. 9).
blood from the extrathoracic venous system flows into The dependence of the aortic pressure and thereby,
the chest. carotid arterial pressure on intrathoracie pressure
The absence of larger-volume retrograde flow into suggested that blood pressure and flow would be im-
the jugular vein during chest compression is critical to proved by maneuvers that raise intrathoracic
this proposed mechanism. We are accustomed to con- pressure.'0 When the airway was occluded at the end-
sidering flow as proportional to the gradient between inspiration, the lungs remained fully inflated during
the high-pressure, or inflow, and low-pressure, or out- subsequent chest compressions. Aortic pressure in-
flow, ends of a conduit. In a very collapsible tube, such creased and remained similar to intrathoracie pressure
as a systemic vein, this relationship applies only if the measured in the esophagus. The increase in aortic and
pressure surrounding the tube is less than the outflow carotid pressures was associated with a more than
pressure. When surrounding pressure exceeds outflow doubling of carotid blood flow. Abdominal binding,
pressure, flow will depend instead on the gradient first suggested by Harris et al.,'1 proved to be a
between inflow and surrounding pressures.7 In these strikingly effective technique for augmenting carotid
studies on the venous side, outflow pressure (jugular) blood flow during external chest compression. By
is low and inflow pressure (right atrial) and outside restricting paradoxical diaphragmatic movement dur-
pressure (intrathoracic) are similar during compres- ing chest compression, abdominal binding increased
sion. Thus, there is little gradient for retrograde flow. intrathoracic pressure and thereby, aortic systolic
A small gradient is probably present at the onset of pressure. Also, there may be some benefit in terms of
compression, because filling of the right heart and cen- carotid flow from limitation of abdominal flow with
tral veins causes right atrial pressure to exceed in- vascular compression. Volume infusion alone did not
trathoracie pressure slightly. The large pressure have a consistently beneficial effect. This suggests that
gradient between the right atrium and the ex- the favorable effect of binding is not mainly due to
trathoracic jugular vein during compression probably blood volume shifts. Liver lacerations from the ab-
results from venous collapse. These conclusions are dominal compression were sought routinely but not
supported by the minimal measured retrograde flow in found in any of the dogs studied.
the jugular vein after the initial brief retrograde jet By maintaining the lungs inflated after the abdomen
(fig. 2). In isolated systems, collapse occurs just inside had been bound, we were able to achieve very high
the region with high outside pressure; that is, during systolic aortic pressures, frequently greater than 100
chest compression, collapse would be expected to oc- mm Hg. Despite the high aortic pressure, carotid
.~
BLOOD FLOW DURING CPR/Rudikoff et al. 351
pulse pressure and flow frequently fell. The develop- major role, and 2) a group in which mean radial artery
ment of a large systolic pressure gradient from the pressure (n = 26, or 70%, beat 1 = 40.4 ± 2.0 mm
aorta to the carotid artery indicates arterial collapse. Hg) approximates the mean carotid pressures in the
The mechanism of the sudden complete arterial canine study group (table 2). After completion of the
collapse seen in the first beat after airway occlusion reported studies, a 12-kg dog with a flat sternum was
(fig. 9) probably results from the mechanism of venous resuscitated, and a higher aortic than right atrial
collapse discussed above. systolic pressure was observed. Direct cardiac com-
Recently, Mashiro and associates'2 measured in- pression probably occurred here, as with 30% of the
tracardiac pressures and dimensions after ventricular humans studied. Mean radial pressure was selected
fibrillation induced by coronary embolism or by elec- because vasoconstriction markedly damps radial
trolyte changes. Evidence of left ventricular contrac- artery pressure during CPR in man. It has not escaped
ture was found with left ventricular pressures ex- our attention that clinical CPR13 and, in fact, other
ceeding right ventricular pressures for 15 seconds. To circulatory support systems, might well be substan-
ensure sustained ventricular fibrillation, we did not tially improved by use of the principle of increased
begin chest compression for 15 seconds. At the onset blood flow through increases in intrathoracic pressure
of compression, pressure and flows were as presented. in many patients. Expansion of blood volume and/or
Thirty seconds after the onset of fibrillation, diastolic augmentation of pulmonary blood flow might limit
pressure in the aorta was 13 ± 1 mm Hg in seven the possibility of arterial collapse and the strategy
dogs. Aortic and right atrial systolic pressures were might provide very substantial overall increases in
similar, with a maximal difference of 3 mm Hg and a blood flow to the carotid and other arterial beds.
mean systolic aortic pressure of 57 ± 3 mm Hg for Under all conditions studied to date the aortic
beat 1. During the study, the diastolic pressure fell diastolic pressure relative to atmosphere is extremely
slightly. low: 35 mm Hg. The low aortic pressure must indicate
The obvious differences in thoracic anatomy dictate a very small volume of blood in the central arterial
caution in the extrapolation of these data to man. vessels and raises concern regarding the adequacy of
Confirmation of a gradient in the jugular vein at the coronary perfusion in any life-support system using in-
thoracic outlet in man (fig. 10) indicates that venous trathoracic pressure to move blood into the peripheral
collapse occurs during human CPR and suggests that tissues.
the mechanism of flow identified in the dog at least
contributes to flow in man. With differences in chest Addendum
configuration, direct cardiac compression may also
Downloaded from http://ahajournals.org by on April 11, 2020
References
Pressure 1. Kouwenhoven WB, Jude JR, Knickerbocker GG: Closed chest
cardiac massage. JAMA 173: 1064, 1960
2. Weale FE, Rothwell-Jackson RL: The efficiency of cardiac
(mmHg) massage. Lancet 1: 990, 1962
3. MacKenzie G, Taylor S, McDonald A, Donald K:
Hemodynamic effects of external cardiac compression. Lancet
1: 1342, 1964
4. Taylor G, Tucker WM, Greene HL, Rudikoff MT, Weisfeldt
0- ML: Importance of prolonged compression during car-
diopulmonary resuscitation in man. N Engl J Med 296: 1515,
FIGURE 10. During cardiopulmonary resuscitation in a 1977
patient with a central venous catheter inserted per- 5. Criley JM, Blaufuss AH, Kissel GL: Cough-induced cardiac
cutaneously via the internaljugular vein, the catheter was compression. JAMA 236: 1246, 1976
6. Rudikoff MT, Freund P, Weisfeldt ML: Mechanisms of blood
slowly withdrawn. When the catheter tip reached the flow during cardiopulmonary resuscitation. (abstr) Circulation
thoracic outlet, systolic pressure suddenly fell. The paper 56 (suppl III): III-97, 1977
speed,4 is 5 MM/sc,, 7. Permutt 5, Riley RL: Hemodynamics of collapsible vessels
352 CIRCULATION VOL 61, No 2, FEBRUARY 1980
with tone: the vascular waterfall. J Appl Physiol 18: 924, 1963 culation during cardiopulmonary resuscitation. Anesthesia 28:
8. Holt JP: The collapse factor in the measurement of venous 730, 1967
pressure. Am J Physiol 134: 292, 1941 12. Mashiro I, Cohn JN, Heckel R, Nelson RR, Franciosa JA:
9. Brecher GA: Mechanism of venous flow under different degrees Left and right ventricular dimensions during ventricular fibrilla-
of aspiration. Am J Physiol 169: 423, 1952 tion in the dog. Am J Physiol 235: H231, 1978
10. Wilder RJ, Weir D, Rush B, Ravitch M: Methods of coor- 13. Chandra N, Rudikoff MT, Weisfeldt ML: Simultaneous chest
dinating ventilation and closed chest cardiac massage in the compression and ventilation at high airway pressure during car-
dog. Surgery 53: 186, 1963 diopulmonary resuscitation (CPR) in man. (abstr) Circulation
11. Harris L, Kirimli B, Safar P: Augmentation of artificial cir- 58 (suppl II): 11-203, 1978
SUMMARY To determine the role of the pericardium in man, ventricular function was studied in 20
patients who underwent pericardiotomy during coronary artery bypass surgery. Left and right ventricular
function curves were generated before and after pericardiotomy by changing body position to alter venous
pressure. Central venous (0-14 mm Hg) and pulmonary wedge pressures (1-25 mm Hg) ranged over normal
and moderately elevated values.
Pericardiotomy did not alter the relationship between left ventricular stroke work index and pulmonary
wedge pressure, between right ventricular stroke work index and central venous pressure, or between
pulmonary wedge pressure and central venous pressure.
These data suggest that in patients with normal and moderately elevated filling pressures, the pericardium
does not influence right or left ventricular systolic function or the coupling between the right and left ventricles.
Downloaded from http://ahajournals.org by on April 11, 2020