Professional Documents
Culture Documents
Figure 1 The type and composition of wear er, patient factors are the hardest to
debris particles have since been iden- manage. Active people generally se-
tified as other key factors in the gen- lect a knee replacement so that they
eration of the osteolytic response. can remain active; these individuals
Particles of polyethylene, polymeth- find it hard to give up or modify their
ylmethacrylate, and metal have been activities.
shown to be present in the debris of
failed TKAs and shown to elicit dis-
Implant Factors
tinct inflammatory responses.17 In-
vestigators also have recognized that Polyethylene Structure and
debris from failed total knee prosthe- Thickness
ses consists of larger flake-shaped Factors related to the material
Retrieved polyethylene insert of a failed particles compared to that from properties of an implant also affect
total knee arthroplasty, demonstrating failed total hip arthroplasties, and its wear resistance and the genera-
wear and delamination of the the tissue response is characterized tion of particulate debris.15 Polyeth-
polyethylene bearing surface. by fewer macrophages.25 This sug- ylene powder is generated by a
(Reproduced from Naudie DDR, gests that large polyethylene parti- polymerization process whereby
Rorabeck CH: Sources of osteolysis cles associated with delamination, ethylene monomers are linked to-
around total knee arthroplasty: Wear of
pitting, and fatigue wear do not elicit gether and transferred into a poly-
the bearing surface. Instr Course Lect
the same cellular response as does mer chain of carbon and hydrogen
2004;53:251-259.)
small-particle debris. Some studies groups. The physical properties and
propose that osteolysis around mod- molecular weight of polyethylene
of particulate debris, the composi- ular TKA designs occurs as a result are defined largely by the polymer-
tion of debris, and the location of de- of small-particle debris generated by ization process, which is affected by
bris generation (articular or back- backside wear at the interface be- conditions such as temperature and
side) all must be considered. tween the polyethylene insert and pressure and by chemicals used as
the metal baseplate.5,9,10,26-29 There is, catalysts in polymerization. Process-
however, no evidence to confirm ing flaws can result in subsurface
The Problem of
that small wear particles are pro- cracks, voids, or fusion defects that
Osteolysis
duced only by backside wear or that can act as sites for delamination or
The earliest reports of osteolysis articular-side wear does not contrib- for further crack propagation when
around total knee replacements de- ute to osteolysis. these implants are loaded in vivo.
scribed bone resorption in associa- Authors of retrieval studies have
tion with cementless implants.18-20 examined polyethylene bearings con-
Patient Factors
Holes in the tibial baseplate, tibial taining processing defects and re-
fixation screws, and areas of un- Many factors influence the amount ported higher rates of delamination
bonded or discontinuous porous of wear and osteolysis that occurs and total wear damage than in sim-
coating on the implant were impli- over time. Patient-related factors, ilar bearings that are free of de-
cated as conduits for debris, ulti- such as age, size, and activity level, fects.32,33 Studies also have shown
mately leading to osteolysis.21,22 determine the amount and types that polyethylene materials that are
Soon afterward, investigators recog- of load experienced by an im- free of defects demonstrate virtually
nized that osteolysis also occurred plant.9,30,31 Age alone is not indica- no delamination even after extended
in association with total knee re- tive of the usage of a total knee im- duration in vivo.34 Over the years,
placements in which components plant, however, because younger several attempts have been made to
were inserted with cement.12,23,24 In patients may have undergone knee enhance polyethylene, including by
these cases, debris particles were arthroplasty for inflammatory dis- heat pressing, carbon fiber reinforce-
thought to gain access to bone by ease (eg, rheumatoid arthritis) that ment, and alteration of its crystallin-
way of voids in the cement mantle. itself limits activities. Similarly, a ity, but most of these changes re-
The implant-bone interface was patient’s body weight may have con- sulted in materials that were more
therefore identified as an important tributed to the arthritis in the knee, prone to fatigue modes of wear.35-37
variable in the development of os- but it also may contribute to inactiv- Evidence also suggests that the
teolysis because it provided me- ity after TKA. Therefore, activity method of manufacturing polyethy-
chanical stability to the prosthesis level is likely the most important lene may affect its clinical perfor-
and acted as a barrier to particulate patient factor affecting the loads mance and wear resistance. In gen-
debris. placed on a TKA over time. Howev- eral, components are manufactured
Figure 2 Figure 3
bearing-surface designs aim to min- inserts revealed high failure rates a dished polyethylene insert and a
imize contact loads and contact from polyethylene wear.3,49,52-54 The raised anterior lip, were developed to
stresses in order to maximize wear accelerated wear rates observed in improve anteroposterior stability
resistance. Bartel et al48 showed that these studies were thought to result and thus prevent uncontrolled slid-
the effect of contact load and stress from the higher contact stresses ing of the femoral component. In
distribution at the bearing surface placed through small contact areas some studies, these implants dem-
depends on the contact area over on the flat articular surface. A high- onstrated good clinical success and
which weight-bearing stresses are er incidence of eccentric and asym- a low incidence of polyethylene
transferred. Bearing-surface designs metric wear patterns also was noted, wear.58 Retrieval studies also con-
with small areas of contact between thought to result from soft-tissue firmed that these more conforming
the femoral and tibial components imbalances (Figure 5). The most designs exhibited lower wear rates
distribute weight-bearing forces over common wear patterns were typical- and fewer eccentric wear patterns
a smaller area, which in turn in- ly located anteromedially and an- than did less conforming de-
creases the total amount of stress per terolaterally, likely the result of pro- signs.3,49,53
unit area that is transmitted to the gressive subluxation of the femur Posterior-stabilized designs were
tibial polyethylene component. from posterior cruciate ligament developed because of the technical
Thus, bearing-surface designs with insufficiency.52-54 However, other re- difficulties encountered in trying to
small areas of contact may have less ports documented posterior wear correct significant clinical deformity
favorable wear characteristics. Con- patterns of the tibial insert caused by with partially conforming designs.
versely, designs with the largest area a tight posterior cruciate ligament They employ a highly conforming or
of contact allow weight bearing to be and/or external rotation of the tibial relatively congruent articulation.
distributed over a larger area, which tray.54,55 The fact that flat articulat- Posterior-stabilized designs use a tib-
decreases the contact stresses and ing geometries have not performed ial polyethylene post in the middle
maximizes wear resistance. inferiorly in clinical use suggests of the knee and a cam between the
The area of contact between the that other variables may be more femoral condyles to prevent poste-
femoral and tibial components is de- important in the survival of these rior subluxation of the tibia. These
termined by the articular geometry bearing-surface designs.50,51,55 designs decrease load carried by the
of the femoral component and the The concept that cruciate- surrounding soft tissues and likely
polyethylene bearing surface. Poorly retaining implants provide more nor- provide a more predictable kine-
conforming articulations have small mal knee kinematics also has been matic path of the knee; however, the
contact areas, increased contact challenged recently by in vivo fluo- trade-off is increased stresses at dif-
stresses, and theoretically less favor- roscopic measurements that have ferent knee interfaces, including
able wear characteristics. However, demonstrated more erratic kine- component-bone and modular tibial
poorly conforming articulations are matic patterns for cruciate-retaining interfaces.56,57
thought to reproduce more nearly implants.56,57 Dennis et al56 and Ko- Although these designs may in-
normally the physiologic move- mistek et al57 have shown, for exam- crease the likelihood of motion and
ments in the knee by allowing ple, that some cruciate-retaining wear at the modular backside inter-
motion to be dictated by the soft knees demonstrate paradoxical ante- face, many of them have demon-
tissues.52 Highly conforming, or rior femoral translation with flexion strated excellent long-term clinical
relatively congruent, articulations and therefore do not duplicate nor- success.12,13 However, the polyethyl-
maximize contact areas and wear re- mal femoral rollback. These au- ene tibial post has recently been
sistance, but they may decrease the thors56,57 have shown not only that identified as an additional source of
load carried by the surrounding soft posterior stabilized designs demon- polyethylene debris, and several
tissues and, subsequently, increase strate more reproducible in vivo ki- studies have reported that wear of
stresses at the component-bone in- nematics but also that these designs the polyethylene post has resulted in
terface.53 are unable to duplicate normal fem- fracture of the post or clinical
Cruciate-retaining implants were oral rollback. In addition, posterior instability58-62 (Figure 6). What re-
designed with the idea of reproduc- stabilized designs demonstrate mains unclear, however, is whether
ing physiologic movements in the greater condylar lift-off during deep damage to the tibial polyethylene
knee by allowing motion to be dic- knee flexion.57 post occurs as a result of normal
tated by the soft tissues and liga- Bearing surfaces with more con- function from wear on the posterior
mentous structures, particularly the forming polyethylene inserts were surface of the post or as a result of
posterior cruciate ligament.51 How- developed to prevent asymmetric pathologic motion (relative hyperex-
ever, several retrieval studies of rel- wear patterns and reduce wear rates. tension of the components) leading
atively flat cruciate-retaining tibial Partially conforming implants, using to anterior cam-post impingement.
Figure 6 Figure 7
Retrieved polyethylene insert from a A, Retrieved polyethylene insert demonstrating evidence of wear on the
posterior-stabilized total knee undersurface of the insert. B, Retrieved tibial baseplate demonstrating stippling on
arthroplasty component demonstrating the top of the corresponding metal tibial tray.
fracture of the tibial polyethylene post.
(Adapted from Puloski SK, McCalden
RW, MacDonald SJ, Rorabeck CH, al debris and thus damage the mod- ysis following TKA; they noted a
Bourne RB: Tibial post wear in ular interface. Very good mid- and 16% incidence of osteolysis with one
posterior stabilized total knee
long-term clinical results have been of the first modular metal-backed
arthroplasty: An unrecognized source
of polyethylene debris. J Bone Joint
reported with several mobile-bearing implants approved for general use.
Surg Am 2001;83:390-397.) designs, but osteolysis still has been After examining the retrieved tibial
shown to develop in some of these components, the authors reported
implant designs.11,64,65 fatigue-type wear modes, such as pit-
Mobile-bearing implants were de- ting and delamination, on the artic-
signed to allow very high conformi- Component Design: The ular surface of the polyethylene com-
ty between articulating surfaces and Backside ponents but not on the underside.
to provide more nearly normally the Metal backing and modularity of On the undersurface of most of the
physiologic knee kinematics. These the tibial component were intro- polyethylene inserts, screw-hole im-
designs have a high degree of confor- duced in TKA to improve fixation pressions were often observed, and,
mity and, therefore, a large surface- and address aseptic loosening.9 Metal in retrieved implants in which the
contact area, which, as noted, is fa- backing was added primarily to modular insert was larger than the
vorable in terms of maximizing wear lessen stresses on cement and bone metal tray, a lip of polyethylene was
resistance.63 These designs also have in an attempt to reduce tibial compo- identified on the overhanging edge
a mobile tibial polyethylene bearing, nent loosening. Modularity provided and in the recess of the metal tray for
which permits rotation and/or trans- a means of augmenting cementless the posterior cruciate ligament. The
lation during weight bearing. The tibial component fixation; it also cre- authors reasoned that these screw-
theoretic advantage offered by ated an opportunity to use polyeth- hole impressions and nonloaded ar-
mobile-bearing implants over fixed- ylene inserts of different thicknesses eas could be accounted for only by
bearing surfaces is that of self- with varying degrees of constraint. loss of material in adjacent areas on
correcting, internally or externally Although metal backing and modu- the underside of the polyethylene in-
rotated tibial components. These de- larity provided improved tibial com- sert (Figure 7, A).
signs avoid the effect of posterome- ponent fixation, they limited the The findings of the study from
dial or posterolateral tibial polyeth- polyethylene thickness that could be Peters et al18 led to further investiga-
ylene wear. Although these implants used without making additional bone tions about the quality of the lock-
were designed for motion to occur resections. In addition, this compo- ing mechanisms between the tibial
between the polished metal and nent design change created a new in- tray and polyethylene insert for
plastic surfaces, these surfaces are terface for wear between the polyeth- modular knee arthroplasty designs.
not impervious to the effects of wear ylene insert and the metal tibial Parks et al6 initiated a study to me-
debris. Any debris that does get baseplate.5-10,27,28 chanically test the locking mecha-
trapped between these two relative- In 1992, Peters et al18 published nism used to connect the polyethy-
ly large surfaces can create addition- the first case report series of osteol- lene insert and the metal tibial tray
in nine types of commercially avail- abutting polyethylene surface and, of nine new femoral components
able total knee implants. The au- therefore, are more desirable in clin- commonly used in TKA. The study
thors reported appreciable motion ical use.60 was conducted using stereomicros-
with both the snap-fit and the These studies led several centers copy and light profilometry over
tongue-in-groove locking mecha- to reconsider the use of an all- multiple surface points, including
nisms. Engh et al7 subsequently con- polyethylene or nonmodular metal- the cam. The results showed that
ducted a retrieval study to determine backed tibial knee design, with the the roughness parameters measured
whether locking mechanisms be- intent of minimizing backside wear from the condylar and trochlear sur-
came unstable over time with in and interface motion. Two separate faces were similar from one manu-
vivo stresses. The same testing pro- centers performed matched-pair facturer to the other. However, the
tocol of Parks et al6 was repeated, us- analyses of all-polyethylene versus authors found that, in six of the nine
ing retrieved implants with the same metal-backed tibial components; components, the cam was uniform-
locking mechanisms. The authors they found no statistically signifi- ly rougher than the condylar or
demonstrated motion indexes an or- cant differences in clinical perfor- trochlear surfaces of the implant.
der of magnitude greater than those mance or functional and radiograph- These findings raise the concern that
of fresh, out-of-the-box implants. ic outcomes.66,67 Brassard et al68 the cam-post articulation might con-
These findings suggested that lock- performed a similar study compar- tribute to wear debris and thus
ing mechanisms lose stability under ing monoblock and modular Insall- should be optimized during implant
repetitive loads in vivo. The magni- Burstein posterior-stabilized im- manufacturing.
tude of relative motion that occurs, plants (Zimmer, Warsaw, IN); they
however, has since been shown to be reported no statistical difference be-
Surgical Factors
smaller under physiologic loads than tween the type of knee implant and
in the unloaded condition.8 functional scores. They also report- Good surgical technique is required
In a final study of the same group ed a lower incidence of radiolucent to minimize polyethylene wear and
of retrieved modular tibial compo- lines in the monoblock design. A re- subsequent osteolysis around total
nents, Rao et al28 stereoscopically port of a survivorship analysis of knee implants.72 Poor alignment of
examined wear of the tibial base- more than 11,000 knees from the knee components may occur at the
plate. Baseplate wear grades corre- Mayo Clinic database indicated a time of surgery in the coronal, sagit-
lated with time in situ but demon- significantly (P < 0.0001) higher 10- tal, or rotational planes. Failure to
strated an even stronger correlation year survival rate in knees that em- restore the mechanical axis of the
with an insert motion index. These ployed an all-polyethylene tibial limb at the time of surgery may re-
authors also identified an organized component.69 Most recently, Berend sult in increased loading forces
pattern of discrete markings on the et al55 reported that osteolysis has across the bearing surface, leading to
tibial baseplate, referred to as stip- not been a problem with the AGC early degradation of the polyethy-
pling (Figure 7, B). These marks were (Biomet, Warsaw, IN) nonmodular lene. Minor deviations from a neu-
organized in a fashion that strongly metal-backed design. tral mechanical axis may contribute
indicated rotation of the insert on to accelerated wear that is not evi-
the tibial baseplate. The existence of Femoral Counter-Surface dent until the implant has been
a dominant pattern of marks for Roughness functioning for many years. In a re-
each implant suggested that they The surface characteristics of the cent study that analyzed the influ-
were directly related to the design of femoral component, or femoral ence of alignment on the wear of
the locking mechanism. The speci- counter surface, also must be consid- retrieved tibial components, the
mens examined consisted of 10 ered when evaluating wear in TKA. location of wear strongly correlated
baseplates made of titanium alloy Experimental studies in total hip re- with the location of the mechanical
and 7 of cobalt-chromium alloy; the placement have demonstrated that axis.31
authors identified no significant dif- increased roughness of the femoral Imbalance in the sagittal plane
ference between the type of alloy head leads to accelerated polyethy- can result in a higher incidence of
and insert motion index, backside lene wear.70 Hypothetically, similar- eccentric and asymmetric wear
polyethylene wear score, or base- ly high wear rates will be observed patterns with cruciate-retaining
plate wear score. Some investigators from poorly polished femoral com- designs.52-54 Imbalance in the sagittal
have suggested that polished base- ponents in TKA; to date, however, plane also can allow hyperextension
plates of harder materials, such as no published studies are available to of the knee and lead to anterior cam-
cobalt-chromium alloy, are more support this hypothesis. post impingement with posterior-
scratch-resistant, which may reduce In a recent study, Puloski et al71 stabilized designs.52-54,59-62 Internal or
wear on the undersurface of the examined the surface characteristics external rotation of the femoral or
Figure 8 Figure 9
not been used routinely to evaluate from osteolysis can lead not only to Figure 10
osteolysis around knee implants, prosthetic loosening and clinical fail-
they hold the promise of improving ure of the implant but also to weak-
accuracy in the future. ness in the periprosthetic osseous
support about the implant, resulting
Management in periprosthetic or component frac-
The clinical significance of wear and ture79,80 (Figure 11).
osteolysis around total knee im- The timing of surgical interven-
plants relates mainly to the extent of tion for an osteolytic lesion is con-
the disease process. Usually, patients troversial. The decision to operate
early in the evolution of osteolysis on a patient with osteolysis is based
are asymptomatic, and the amount on the presence of symptoms and
of wear and bone resorption is limit- the likelihood that, if surgery is not
ed. In such cases, osteolytic lesions performed, a well-fixed implant may
may be visible on radiographs but become loose during the patient’s
are of little clinical significance to lifetime. In general, we recommend
the patient. In these patients, serial surgery in most symptomatic pa-
radiographs are essential to evaluate tients and in asymptomatic patients
the progression of lesions. However, when the osteolytic lesion is rapidly
the rate of progression of osteolytic increasing in size and seems to be
lesions varies. In some cases, lesions eroding the cortical support for the
have been followed for several years implant. We would not routinely op-
with only minimal changes in the erate on asymptomatic patients who
radiographic appearance (Figure 10). have osteolytic lesions that do not
Because these lesions are often disturb cortical bone support. In the
asymptomatic and slowly progres- absence of instability, malalign- Anteroposterior radiograph of a large
sive, revision may not be necessary ment, and significant backside wear osteolytic lesion in the lateral aspect of
when they are first identified. Most in carefully selected patients, we the tibial plateau that appeared 9 years
after primary total knee arthroplasty.
cases of osteolysis, however, progress would consider component reten-
with time.15 Loss of bone support tion, modular polyethylene ex-
Figure 11
Posteroanterior (A) and lateral (B) radiographs of a patient with severe osteolysis around the tibial stem of his primary total knee
replacement, which resulted in fracture of the tibial baseplate of his implant. C, The removed tibial component. (Reproduced
from Naudie DDR, Rorabeck CH: Sources of osteolysis around total knee arthroplasty: Wear of the bearing surface. Instr
Course Lect 2004;53:251-259.)
change, and morcellized bone graft- nation of patient, implant, and sur- 273:232-242.
ing of the osteolytic defect. gical factors. Osteolysis is most 4. McKellop HA, Campbell P, Park SH,
et al: The origin of submicron poly-
When revision surgery is indi- commonly the result of the produc-
ethylene wear debris in total hip ar-
cated, effort should be made to iden- tion of biologically active polyethyl- throplasty. Clin Orthop Relat Res
tify the underlying cause of the wear ene debris. Polyethylene debris par- 1995;311:3-20.
and osteolysis. Malalignment and in- ticles may originate from both the 5. Wasielewski RC, Parks N, Williams I,
stability usually are apparent on articular bearing surface and modu- Surprenant H, Collier JP, Engh G: Tib-
physical examination and may be the lar component backside surfaces. ial insert undersurface as a contribut-
cause or the result of significant poly- ing source of polyethylene wear de-
Many factors related to the manu- bris. Clin Orthop Relat Res 1997;
ethylene wear. Most commonly, ac- facturing and design of the articular 345:53-59.
celerated polyethylene wear occurs
bearing surface influence the extent 6. Parks NL, Engh GA, Topoleski LD,
as a result of loss of material, which Emperado J: The Coventry Award:
of polyethylene wear generated over
in turn reduces ligament support to Modular tibial insert micromotion. A
time; surgeons should be cautious in
the knee and further aggravates both concern with contemporary knee im-
considering enhanced polyethyl-
malalignment and instability. This plants. Clin Orthop Relat Res 1998;
enes until further investigations are 356:10-15.
combination frequently results in
performed. Surgeons also should be 7. Engh GA, Lounici S, Rao AR, Collier
debris-generated bone defects, which
aware of backside wear and under- MB: In vivo deterioration of tibial
may require the use of bulk allografts
stand the problems associated with baseplate locking mechanisms in
and stemmed tibial components.81 contemporary modular total knee
modular baseplates and their lock-
With revision arthroplasty, every components. J Bone Joint Surg Am
ing mechanisms for polyethylene
effort should be made to restore 2001;83:1660-1665.
inserts. Alignment and stability in 8. Conditt MA, Ismaily SK, Alexander
alignment and stability if the pro-
the sagittal, coronal, and rotational JW, Noble PC: Backside wear of ultra-
cess of wear and osteolysis are to be
planes are important for durability high molecular weight polyethylene
avoided. When wear and osteolysis tibial inserts. J Bone Joint Surg Am
of the implant. Orthopaedic sur-
occur in the absence of significant 2004;86:1031-1037.
geons must understand the factors
malalignment or instability, an at- 9. Engh GA, Ammeen DJ: Epidemiology
that contribute to wear and osteoly- of osteolysis: Backside implant wear.
tempt should be made to identify the
sis to maximize the longevity of Instr Course Lect 2004;53:243-249.
implant type. The surgeon may
TKA. 10. Conditt MA, Thompson MT, Usrey
choose to obtain the product identi- MM, Ismaily SK, Noble PC: Backside
fication numbers from the hospital wear of polyethylene tibial inserts:
References Mechanism and magnitude of materi-
records. Although this is not practi-
al loss. J Bone Joint Surg Am 2005;87:
cal for most surgeons, this informa- Evidence-based Medicine: Prospec-
326-331.
tion can help determine whether tive randomized studies of level I/II 11. Huang CH, Ma HM, Liau JJ, Ho FY,
failure occurred in association with are limited (refs 56, 57 and 68) to Cheng CK: Osteolysis in failed total
polyethylene that was not manufac- multicenter analysis and in vivo knee arthroplasty: A comparison of
tured or sterilized according to the analysis. The majority of references mobile-bearing and fixed-bearing
knees. J Bone Joint Surg Am 2002;84:
most currently desirable methods. are case-controlled or cohort studies
2224-2229.
Whenever a modular implant is (level III/IV) and expert opinion (lev- 12. O’Rourke MR, Callaghan JJ, Goetz DD,
present, backside wear also should el V). Sullivan PM, Johnston RC: Osteolysis
be considered as a potential contrib- associated with a cemented modular
Citation numbers printed in bold posterior-cruciate-substituting total
utor to the osteolytic process. Recog-
type indicate references published knee design: Five to eight-year follow-
nizing that backside wear may be a
within the past 5 years. up. J Bone Joint Surg Am 2002;84:
contributing source of debris is the 1362-1371.
first step in eliminating backside 1. Lonner JH, Siliski JM, Scott RD: Pro- 13. Lachiewicz PF, Soileau ES: The rates
wear during the revision surgery. dromes of failure in total knee arthro- of osteolysis and loosening associated
When backside wear is present on plasty. J Arthroplasty 1999;14:488- with a modular posterior stabilized
492. knee replacement: Results at five to
the insert, modular insert exchange
2. Sharkey PF, Hozack WJ, Rothman fourteen years. J Bone Joint Surg Am
may not be the best management op- RH, Shastri S, Jacoby SM: Insall 2004;86:525-530.
tion during revision surgery.82 Award paper: Why are total knee ar- 14. Fehring TK, Murphy JA, Hayes TD,
throplasties failing today? Clin Roberts DW, Pomeroy DL, Griffin
Orthop Relat Res 2002;404:7-13. WL: Factors influencing wear and
Summary 3. Collier JP, Mayor MB, McNamara JL, osteolysis in press-fit condylar modu-
Surprenant VA, Jensen RE: Analysis of lar total knee replacements. Clin
The generation of polyethylene the failure of 122 polyethylene inserts Orthop Relat Res 2004;428:40-50.
wear and development of osteolysis from uncemented tibial knee compo- 15. Naudie DDR, Rorabeck CH: Sources
around TKA are caused by a combi- nents. Clin Orthop Relat Res 1991; of osteolysis around total knee arthro-
plasty: Wear of the bearing surface. UHMWPE wear debris in failed total al: Overview of polyethylene as a
Instr Course Lect 2004;53:251-259. knee arthroplasties: A comparison be- bearing material: Comparison of ster-
16. Howie DW, Vernon-Roberts B, Oake- tween mobile bearing and fixed bear- ilization methods. Clin Orthop Relat
shott R, Manthey B: A rat model of re- ing knees. J Orthop Res 2002;20: Res 1996;333:76-86.
sorption of bone at the cement-bone 1038-1041. 43. Collier JP, Sperling DK, Currier JH,
interface in the presence of polyethy- 30. Kuster MS, Stachowiak GW: Factors Sutula LC, Saum KA, Mayor MB: Im-
lene wear particles. J Bone Joint Surg affecting polyethylene wear in total pact of gamma sterilization on clini-
Am 1988;70:257-263. knee arthroplasty. Orthopedics 2002; cal performance of polyethylene in
17. Jacobs JJ, Roebuck KA, Archibeck M, 25(2 suppl):S235-S242. the knee. J Arthroplasty 1996;11:
Hallab NJ, Glant TT: Osteolysis: Ba- 31. Collier MB, Engh CA Jr, Engh G: The 377-389.
sic science. Clin Orthop Relat Res correlation between limb alignment 44. Bohl JR, Bohl WR, Postak PD, Green-
2001;393:71-77. and polyethylene wear in retrieved to- wald AS: The Coventry Award: The
18. Peters PC Jr, Engh GA, Dwyer KA, tal knee arthroplasties. J Bone Joint effects of shelf life on clinical out-
Vinh TN: Osteolysis after total knee Surg Am, in press. come for gamma sterilized polyethy-
arthroplasty without cement. J Bone 32. Wrona M, Mayor MB, Collier JP, Jen- lene tibial components. Clin Orthop
Joint Surg Am 1992;74:864-876. sen RE: The correlation between fu- Relat Res 1999;367:28-38.
19. Berry DJ, Wold LE, Rand JA: Extensive sion defects and damage in tibial poly- 45. McKellop H, Shen FW, Lu B, Camp-
osteolysis around an aseptic, stable, ethylene bearings. Clin Orthop Relat bell P, Salovey R: Development of an
uncemented total knee replacement. Res 1994;299:92-103. extremely wear-resistant ultra high
Clin Orthop Relat Res 1993;293:204- 33. Tanner MG, Whiteside LA, White SE: molecular weight polyethylene for to-
207. Effect of polyethylene quality on wear tal hip replacements. J Orthop Res
20. Engh GA, Parks NL, Ammeen DJ: Tib- in total knee arthroplasty. Clin 1999;17:157-167.
ial osteolysis in cementless total knee Orthop Relat Res 1995;317:83-88. 46. Muratoglu OK, Mark A, Vittetoe DA,
arthroplasty: A review of 25 cases 34. Landy MM, Walker PS: Wear of ultra- Harris WH, Rubash HE: Polyethylene
treated with and without tibial com- high-molecular-weight polyethylene damage in total knees and use of high-
ponent revision. Clin Orthop Relat components of 90 retrieved knee pros- ly crosslinked polyethylene. J Bone
Res 1994;309:33-43. theses. J Arthroplasty 1988;3 suppl:
Joint Surg Am 2003;85(suppl 1):S7-
21. Lewis PL, Rorabeck CH, Bourne RB: S73-S85.
S13.
Screw osteolysis after cementless to- 35. Li S, Burstein AH: Ultra-high molecu-
47. Muratoglu OK, Bragdon CR, Jasty M,
tal knee replacement. Clin Orthop lar weight polyethylene: The material
O’Connor DO, Von Knoch RS, Harris
Relat Res 1995;321:173-177. and its use in total joint implants.
22. Whiteside LA: Effect of porous- J Bone Joint Surg Am 1994;76:1080- WH: Knee-simulator testing of con-
coating configuration on tibial osteol- 1090. ventional and cross-linked polyethyl-
ysis after total knee arthroplasty. 36. Ahn NU, Nallamshetty L, Ahn UM, ene tibial inserts. J Arthroplasty
Clin Orthop Relat Res 1995;321:92- et al: Early failure associated with the 2004;19:887-897.
97. use of Hylamer-M spacers in three pri- 48. Bartel DL, Bicknell VL, Wright TM:
23. Ries MD, Guiney W Jr, Lynch F: Os- mary AMK total knee arthroplasties. The effect of conformity, thickness,
teolysis associated with cemented to- J Arthroplasty 2001;16:136-139. and material on stresses in ultra-high
tal knee arthroplasty: A case report. 37. Ries MD, Bellare A, Livingston BJ, Co- molecular weight components for to-
J Arthroplasty 1994;9:555-558. hen RE, Spector M: Early delamina- tal joint replacement. J Bone Joint
24. Robinson EJ, Mulliken BD, Bourne tion of a Hylamer-M tibial insert. Surg Am 1986;68:1041-1051.
RB, Rorabeck CH, Alvarez C: Cata- J Arthroplasty 1996;11:974-976. 49. Hirakawa K, Bauer TW, Stulberg BN,
strophic osteolysis in total knee re- 38. Won CH, Rohatgi S, Kraay MJ, Gold- Wilde AH, Borden LS: Characteriza-
placement: A report of 17 cases. Clin berg VM, Rimnac CM: Effect of resin tion of debris adjacent to failed knee
Orthop Relat Res 1995;321:98-105. type and manufacturing method on implants of 3 different designs. Clin
25. Schmalzried TP, Jasty M, Rosenberg wear of polyethylene tibial compo- Orthop Relat Res 1996;331:151-158.
A, Harris WH: Polyethylene wear de- nents. Clin Orthop Relat Res 2000; 50. Meding JB, Ritter MA, Faris PM: Total
bris and tissue reactions in knee as 376:161-171. knee arthroplasty with 4.4 mm of tib-
compared to hip replacement prosthe- 39. Currier BH, Currier JH, Collier JP, ial polyethylene: 10-year followup.
ses. J Appl Biomater 1994;5:185-190. Mayor MB: Effect of fabrication meth- Clin Orthop Relat Res 2001;388:112-
26. Schmalzried TP, Callaghan JJ: Wear in od and resin type on performance of 117.
total hip and knee replacements. tibial bearings. J Biomed Mater Res 51. Ritter MA, Worland R, Saliski J, et al:
J Bone Joint Surg Am 1999;81:115- 2000;53:143-151. Flat-on-flat, nonconstrained, com-
136. 40. Benson LC, DesJardins JD, LaBerge M: pression molded polyethylene total
27. Engh GA, Ammeen DJ: Polyethylene Effects of in vitro wear of machined knee replacement. Clin Orthop
wear. Clin Orthop Relat Res 2002; and molded UHMWPE tibial inserts Relat Res 1995;321:79-85.
404:71-74. on TKR kinematics. J Biomed Mater 52. Feng EL, Stulberg SD, Wixson RL: Pro-
28. Rao AR, Engh GA, Collier MB, Louni- Res 2001;58:496-504. gressive subluxation and polyethy-
ci S: Tibial interface wear in retrieved 41. Berzins A, Jacobs JJ, Berger R, et al: lene wear in total knee replacements
total knee components and correla- Surface damage in machined ram- with flat articular surfaces. Clin
tions with modular insert motion. extruded and net-shape molded re- Orthop Relat Res 1994;299:60-71.
J Bone Joint Surg Am 2002;84:1849- trieved polyethylene tibial inserts of 53. Benjamin J, Szivek J, Dersam G, Pers-
1855. total knee replacements. J Bone Joint selin S, Johnson R: Linear and volumet-
29. Huang CH, Ho FY, Ma HM, et al: Re- Surg Am 2002;84:1534-1540. ric wear of tibial inserts in posterior
lated particle size and morphology of 42. Collier JP, Sutula LC, Currier BH, et cruciate-retaining knee arthroplasties.
Clin Orthop Relat Res 2001;392:131- syndrome: A report of 5 cases. ship to technical considerations dur-
138. J Arthroplasty 2003;18:942-945. ing total knee arthroplasty. Clin
54. Lewis P, Rorabeck CH, Bourne RB, 63. Kilgus DJ: Polyethylene wear in mobile- Orthop Relat Res 1994;299:31-43.
Devane P: Posteromedial tibial poly- bearing prostheses.Orthopedics 2002; 73. Collier MB, Jewett BA, Engh CA Jr:
ethylene failure in total knee replace- 25(2 suppl):S227-S233. Clinical assessment of tibial polyeth-
ments. Clin Orthop Relat Res 1994; 64. Kaper BP, Smith PN, Bourne RB, Rora- ylene thickness: Comparison of radio-
299:11-17. beck CH, Robertson D: Medium-term graphic measurements with as-im-
55. Berend ME, Ritter MA, Meding JB, et results of a mobile bearing total knee planted and as-retrieved thicknesses.
al: Tibial component failure mecha- replacement. Clin Orthop Relat Res J Arthroplasty 2003;18:860-866.
nisms in total knee arthroplasty. 1999;367:201-209. 74. Fukuoka Y, Hoshino A, Ishida A: A
Clin Orthop Relat Res 2004;428:26- 65. Buechel FF Sr: Long-term followup af- simple radiographic measurement
34. ter mobile-bearing total knee replace- method for polyethylene wear in total
56. Dennis DA, Komistek RD, Colwell ment. Clin Orthop Relat Res 2002; knee arthroplasty. IEEE Trans
CE Jr, et al: In vivo anteroposterior 404:40-50. Rehabil Eng 1999;7:228-233.
femorotibial translation of total knee 66. Udomkiat P, Dorr LD, Long W: 75. Short A, Gill HS, Marks B, et al: A nov-
arthroplasty: A multicenter analysis. Matched-pair analysis of all-polyeth- el method for in vivo knee prosthesis
Clin Orthop Relat Res 1998;356:47- ylene versus metal-backed tibial com- wear measurement. J Biomech 2005;
57. ponents. J Arthroplasty 2001;16:689- 38:315-322.
57. Komistek RD, Scott RD, Dennis DA, 696. 76. Cadambi A, Engh GA, Dwyer KA,
Yasgur D, Anderson DT, Hajner ME: 67. Najibi S, Iorio R, Surdam JW, Whang Vinh TA: Osteolysis of the distal fe-
In vivo comparison of femorotibial W, Appleby D, Healy WL: All- mur after total knee arthroplasty.
contact positions for press-fit posteri- polyethylene and metal-backed tibial J Arthroplasty 1994;9:579-594.
or stabilized and posterior cruciate- components in total knee arthroplas- 77. Nadaud MC, Fehring TK, Fehring K:
retaining total knee arthroplasties. ty: A matched pair analysis of func- Underestimation of osteolysis in pos-
J Arthroplasty 2002;17:209-216. tional outcome. J Arthroplasty 2003; terior stabilized total knee arthroplas-
58. Plante-Bordeneuve P, Freeman MA: 18:9-15. ty. J Arthroplasty 2004;19:110-115.
Tibial high-density polyethylene 68. Brassard MF, Insall JN, Scuderi GR, 78. Berry DJ: Recognizing and identifying
wear in conforming tibiofemoral Colizza W: Does modularity affect osteolysis around total knee arthro-
prostheses. J Bone Joint Surg Br 1993; clinical success? A comparison with a plasty. Instr Course Lect 2004;53:
75:630-636. minimum 10-year followup. Clin 261-264.
59. Puloski SK, McCalden RW, Mac- Orthop Relat Res 2001;388:26-32. 79. Benevenia J, Lee FY, Buechel F, Par-
Donald SJ, Rorabeck CH, Bourne RB: 69. Rand JA, Trousdale RT, Ilstrup DM, sons JR: Pathologic supracondylar
Tibial post wear in posterior stabi- Harmsen WS: Factors affecting the fracture due to osteolytic pseudotu-
lized total knee arthroplasty: An un- durability of primary total knee pros- mor of knee following cementless
recognized source of polyethylene de- theses. J Bone Joint Surg Am 2003; total knee replacement. J Biomed
bris. J Bone Joint Surg Am 2001;83: 85:259-265. Mater Res 1998;43:473-477.
390-397. 70. Lancaster JG, Dowson D, Isaac GH, 80. Huang CH, Yang CY, Cheng CK: Frac-
60. Mikulak SA, Mahoney OM, dela Rosa Fisher J: The wear of ultra-high molec- ture of the femoral component associ-
MA, Schmalzried TP: Loosening and ular weight polyethylene sliding on ated with polyethylene wear and os-
osteolysis with the press-fit condylar metallic and ceramic counterfaces teolysis after total knee arthroplasty.
posterior-cruciate-substituting total representative of current femoral sur- J Arthroplasty 1999;14:375-379.
knee replacement. J Bone Joint Surg faces in joint replacement. Proc Inst 81. Engh GA, Herzwurm PJ, Parks NL:
Am 2001;83:398-403. Mech Eng [H] 1997;211:17-24. Treatment of major defects of bone
61. Callaghan JJ, O’Rourke MR, Goetz 71. Puloski SK, McCalden RW, Mac- with bulk allografts and stemmed
DD, Schmalzried TP, Campbell PA, Donald SJ, Bourne RB, Rorabeck CH: components during total knee arthro-
Johnston RC: Tibial post impinge- Surface analysis of posterior stabi- plasty. J Bone Joint Surg Am 1997;79:
ment in posterior-stabilized total lized femoral components used in to- 1030-1039.
knee arthroplasty. Clin Orthop Relat tal knee arthroplasty. J Arthroplasty 82. Engh GA, Koralewicz LM, Pereles TR:
Res 2002;404:83-88. 2003;18:822-826. Clinical results of modular polyethy-
62. Mauerhan DR: Fracture of the poly- 72. Wasielewski RC, Galante JO, Leighty lene insert exchange with retention of
ethylene tibial post in a posterior RM, Natarajan RN, Rosenberg AG: total knee arthroplasty components.
cruciate-substituting total knee ar- Wear patterns on retrieved polyethy- J Bone Joint Surg Am 2000;82:516-
throplasty mimicking patellar clunk lene tibial inserts and their relation- 523.