INTERNAL CALIBRATION PROCEDURE

Approved by: H. Woodyat Date: 2004


BMTS-CP-011, Issue 1, January 2004

INTERNAL CALIBRATION PROCEDURE AND RESULTS


Linear Shrinkage Mould measured against BS 1377-2:1990. Figure 5.7

ITEM DETAILS
Bodycote Reference Serial Number Description Range
Linear Shrinkage Mould- Soil N/A

Procedure

1. Equipment/Consumables required

REFERENCE EQUIPMENT REQUIRED/USED


Description Bodycote Reference Date of Last Calibration Stated
Uncertainty
Digital Calipers readable to
0.01mm
Thermometer

2 Method
2.1 Wash the mould with distilled water to remove the adhering oil particles and dry in oven.
2.2 Measure the length of the mould at two ends and centre to the nearest 0.01 mm. (a)
2.3 Measure the wall thickness at four corners to the nearest 0.01 mm. (b)
2.4 Measure the width at the centre of both the ends to the nearest 0.01mm. (c)
2.5 Measure the wall thickness at the centre of both sides at both the ends to the nearest 0.01mm. (d)
2.6 Measure the height at both the ends. Take two measurements at sides of each end to the nearest 0.01mm.
(e)
2.7 Measure the diameter at both the ends and centre to the nearest 0.01mm. (f)

Page 1 of 2 BMTS–CP-011
INTERNAL CALIBRATION PROCEDURE
Approved by: H. Woodyat Date: 2004
Results of Calibration

Mass or Dimension Measured Reading Number Specification


1 2 3 4 Mean
Essential Dimensions to section 6.5.2.3 Figure 7
Length, mm (0.01) a 139 - 141
Wall thickness, mm(0.01) b 3
Width, mm (0.01) c 40
Wall thickness, mm (0.01) d 6
Height, mm (0.01) e 20
Diameter, mm (0.01) f
Radius, Mean f/2, mm (0.01) 12.0 – 13.0

Date of Calibration Ambient Temperature (°C)

Suitability for Use/Restrictions on Use

Name of Operator Signature of Operator Checked by, Name Checked by, Signature

Page 2 of 2 BMTS–CP-011