Professional Documents
Culture Documents
CHE-524
➢ The tunnelling
current is therefore
a very sensitive
probe of this
separation.
Imaging of the surface topology may then be carried out in one of two
ways:
1) In constant height mode (in which the tunnelling current is monitored as
the tip is scanned parallel to the surface)
➢ At one point the tip will be directly above a surface atom and
the tunnelling current will be large
➢ whilst at other points the tip will be above hollow sites on the
surface and the tunnelling current will be much smaller.
“A plot of the tunnelling current v's tip position”
In practice, however, the normal way of imaging the surface is to maintain the
tunnelling current constant whilst the tip is scanned across the surface. This is
achieved by adjusting the tip's height above the surface so that the tunnelling
current does not vary with the lateral tip position. In this mode the tip will
move slightly upwards as it passes over a surface atom, and conversely, slightly
Constant current Mode
➢ In practice, however, the normal way of imaging the surface is to
maintain the tunnelling current constant whilst the tip is scanned
across the surface.
➢ This is achieved by adjusting the tip's height above the surface so
that the tunnelling current does not vary with the lateral tip
position.
In this mode the tip will move slightly upwards as it passes over a surface
atom, and conversely, slightly in towards the surface as it passes
over a hollow.
1) The position of the tip with respect to the surface must be very accurately
controlled (to within about 0.1 Å) by moving either the surface or the tip.
2) The tip must be very sharp - ideally terminating in just a single
atom at its closest point of approach to the surface.
Let us look at the region where the tip approaches the surface in greater
detail ....
Magnification
https://www.youtube.com/watch?v=P4ywnN0htQI
Q. Explain the basic principles of quantum tunneling through a
potential barrier.
They work differently to TEMs and SEMs too: they have an extremely sharp
metallic probe that scans back and forth across the surface of the specimen. As
it does so, electrons try to wriggle out of the specimen and jump across the gap,
into the probe, by an unusual quantum phenomenon called "tunneling." The
closer the probe is to the surface, the easier it is for electrons to tunnel into it,
the more electrons escape, and the greater the tunneling current.
➢ AFMs don't suffer from this problem because, although they use
still tuneling, they don't rely on a current flowing between
the specimen and a probe, so we can use them to make atomic-
scale images of materials such as plastics, which don't conduct
electricity.
Atomic Force Microscopy (AFMs)
➢ As the tip sweeps across the surface, the force between the atoms
from which it's made and the atoms on the surface
constantly changes, causing the cantilever to bend by minute
amounts.
➢ As the tip sweeps across the surface, the force between the atoms
from which it's made and the atoms on the surface
constantly changes, causing the cantilever to bend by minute
amounts.
➢ By measuring how far the laser beam travels, we can measure how
much the cantilever bends and the forces acting on it from moment to
moment, and that information can be used to figure out and plot the
contours of the surface.
➢ The atomic force microscope (AFM) probe is mounted onto a
flexible cantilever that is manipulated by a vertical piezoelectric into
interacting with the sample.
➢ The piezoelectric expands and exerts a force on the cantilever
proportional to the potential voltage.
➢ The force is balanced by those acting on the probe by its interaction
with the surface and the stain on the cantilever.
A Pizoelectric materials where if you apply a voltage it will deform
the crystal in it and if u apply a known voltage the crystals contract
and expands by some known amount and that how you can control
the matter of the cantilever.
AFM: How it works?
➢ The specimen to be
scanned (1) is
mounted on a drive
mechanism (2) that
can move it in three
dimensions.
As the sample (1) moves around the imaging point (5), the current that
tunnels between the spring cantilever (6) and the tunneling tip (8) is
constantly measured. These measurements are converted into data that can
be used to draw a detailed surface map of the specimen.
“ Wavelength”
It's just the same when it comes to tiny things. The smaller the object we want to
study, the shorter the wavelength and the higher the energy of the probe we need
to use.
Electrons in an electron microscope have shorter wavelengths than visible light,
which is why they can resolve smaller things.
Visible light has wavelengths ranging from about 7 x 10-7 metres for red to
about 4 x 10-7 metres for violet. Electrons in a typical electron microscope have
wavelengths measured in picometres. One picometre is 10-12 of a metre,
meaning that electron microscopes can resolve things hundreds of thousands
of times smaller than optical microscopes
Electron microscopes can resolve atoms, which are about 10-10 metres across.
To get a glimpse of what is inside a proton, which is about 10-15 metres across,
a wavelength of 10-16 metres would be needed.
.Electron microscopes can resolve atoms, which are about 10-10 metres across. To
get a glimpse of what is inside a proton, which is about 10-15 metres across, a
wavelength of 10-16 metres would be needed.
➢ It is termed a scanning
electron microscope because
the image is formed by
scanning a focused electron
beam onto the surface of the
specimen in a raster pattern.
Scanning Electron Microscope: SEM
How does SEM Works
Generation of electrons: Thermoionic emission
Acceleration of electrons: Anode
Electromagnetic lenses
Sample Chamber
Signals from the Samples
http://virtual.itg.uiuc.edu/training/EM_tutorial/
Electron microscopes use signals arising from the interaction of an electron
beam with the sample to obtain information about structure,
morphology, and composition.
1.The electron gun generates electrons.
2.Two sets of condenser lenses focus the electron beam on the specimen and
then into a thin tight beam.
3.To move electrons down the column, an accelerating voltage (mostly
between 100 kV-1000 kV) is applied between tungsten filament and anode.
4.The specimen to be examined is made extremely thin, at least 200 times
thinner than those used in the optical microscope. Ultra-thin sections of 20-
100 nm are cut which is already placed on the specimen holder.
5.The electronic beam passes through the specimen and electrons are
scattered depending upon the thickness or refractive index of different parts
of the specimen.
6.The denser regions in the specimen scatter more electrons and therefore
appear darker in the image since fewer electrons strike that area of the screen.
In contrast, transparent regions are brighter.
7.The electron beam coming out of the specimen passes to the objective lens,
which has high power and forms the intermediate magnified image.
8.The ocular lenses then produce the final further magnified image.
“ Wavelength”
Parts of Electron microscope
EM is in the form of a tall vacuum column which is vertically mounted. It has
the following components:
1.Electron gun
•The electron gun is a heated tungsten filament, which generates electrons.
2.Electromagnetic lenses
•Condenser lens focuses the electron beam on the specimen. A second condenser
lens forms the electrons into a thin tight beam.
•The electron beam coming out of the specimen passes down the second of magnetic
coils called the objective lens, which has high power and forms the intermediate
magnified image.
•The third set of magnetic lenses called projector (ocular) lenses produce the final
further magnified image.
•Each of these lenses acts as an image magnifier all the while maintaining an
incredible level of detail and resolution.
3.Specimen Holder
•The specimen holder is an extremely thin film of carbon or collodion held by a metal
grid.
4.Image viewing and Recording System.
•The final image is projected on a fluorescent screen.
https://www.youtube.com/watch?v=d7ch1X
SmOgI
Principle of Light Microscope
http://virtual.itg.uiuc.edu/training/LM_tutoria
l/
http://virtual.itg.uiuc.edu/training/AFM_tutor
ial/
https://www.youtube.com/watch?v=ljTEG-B-kGc
https://www.slideshare.net/drbhargava5745
/scanning-electron-microscopy-sem-lecture
Basic Principle
Auger electron spectroscopy (AES) uses the energy of emitted electrons to
identify the elements present in a sample, similar to XPS.
Like XPS, AES measures the kinetic energy (Ek) of an electron to determine
its binding energy (Eb). The binding energy can be found from eq. 1, ΔΦ is
the difference in work function between the sample and the detector material.
Eb = hν − Ek + ΔΦ (1)
First, an electron beam comes in with sufficient energy to eject a core electron
creating a vacancy ( Figure a).
A secondary electron (imaging electron) of higher energy drops down to fill the
vacancy (see Figure b) and emits sufficient energy to eject a tertiary electron
(Auger electron) from a higher shell (see Figure c).
The shells from which the
electrons move from lowest
to highest energy are
described as the K shell, L
shell, and M shell.
The peak seen in the spectrum is labeled according to the shells involved in the
movement of the electrons. For example, an electron ejected from a gold atom
could be labeled as Au KLL or Au KLM.
The intensity of the peak depends on the amount of material present, while
the peak position is element dependent.
Auger transitions characteristic of each elements can be found in the
literature.
https://www.yourarticlelibrary.com/microeco
nomics/working-principle-of-a-electron-
microscopes-with-diagram/26479
Primary Electron Emission
L1 or 2s
X-ray
Photoelectron
K or 1s
Secondary Electron Emission (Auger Electron)
– Low Energy “hole” produced
– High Energy electron fills hole
– Secondary Electron can absorb excess energy
and enter vacuum
Auger
• Called Auger Electron
Electron
L2,3 or 2p
L1 or 2s
K or 1s
Auger Electron Spectroscopy (Auger spectroscopy or AES) was developed in the
late 1960's , deriving its name from the effect first observed by Pierre Auger, a
French Physicist, in the mid-1920's. It is a surface specific technique utilising
the emission of low energy electrons in the Auger process and is one of the
most commonly employed surface analytical techniques for determining the
composition of the surface layers of a sample.
The Auger Process & Auger Spectroscopy
The Auger process is illustrated using the K, L1 & L2,3 levels. These could be
the inner core levels of an atom in either a molecular or solid-state
environment.
I. Ionization
The Auger process is initiated by creation of a core hole - this is typically
carried out by exposing the sample to a beam of high energy electrons
(typically having a primary energy in the range 2 - 10 keV). Such electrons
have sufficient energy to ionise all levels of the lighter elements, and higher
core levels of the heavier elements.
In the diagram above, ionisation is shown to occur by removal of a K-shell
electron, but in practice such a crude method of ionisation will lead to ions with
holes in a variety of inner shell levels.
In some studies, the initial ionisation process is instead carried out
using soft x-rays ( 1000 - 2000 eV ). In this case, the acronym XAES is
sometimes used. As we shall see, however, this change in the method of
ionisation has no significant effect on the final Auger spectrum.
II. Relaxation & Auger Emission
The ionized atom that remains after the removal of the core hole electron is,
of course, in a highly excited state and will rapidly relax back to a lower
energy state by one of two routes :
In the Auger process illustrated, the final state is a doubly-ionized atom with
core holes in the L1 and L2,3 shells.
Experimental Setup:
KE = ( EK - EL1 ) - EL23
The expression for the energy can also be re-written in the form :
KE = EK - ( EL1 + EL23 )
It should be clear from this expression that the latter two energy terms could be
interchanged without any effect - i.e. it is actually impossible to say which electron
fills the initial core hole and which is ejected as an Auger electron ; they are
indistinguishable.
K L2,3
K L1 L1 K L1 L2,3
L2,3
In general, since the initial ionisation is non-selective and the initial hole may therefore
be in various shells, there will be many possible Auger transitions for a given element -
some weak, some strong in intensity. AUGER SPECTROSCOPY is based upon the
measurement of the kinetic energies of the emitted electrons. Each element in a sample
being studied will give rise to a characteristic spectrum of peaks at various kinetic
energies.
This is an Auger spectrum of Pd metal - generated using a 2.5 keV electron beam to
produce the initial core vacancies and hence to stimulate the Auger emission process.
The main peaks for palladium occur between 220 & 340 eV. The peaks are situated on
a high background which arises from the vast number of so-called secondary electrons
generated by a multitude of inelastic scattering processes.
Auger spectra are also often shown in a differentiated form : the reasons for this are
partly historical, partly because it is possible to actually measure spectra directly in this
form and by doing so get a better sensitivity for detection. The plot below shows the
same spectrum in such a differentiated form.