Monitor Suspension Hardware Replacement
Instructions

Copyright © 2007 by General Electric Company, Inc.


All rights reserved.
GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100

THIS PAGE LEFT INTENTIONALLY BLANK

Published by the GE Company, Inc.


Engineering/Technical: Dan Beetcher

Page 2
GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100

Revision History

Revision Date Reason for change


1 09MAR2007 Initial release of document.

Revision History Page 3


GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100

THIS PAGE LEFT INTENTIONALLY BLANK

Page 4
GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100

Field Supplied Materials


• Standard Field Engineer’s toolkit.
• Torque wrench capable of applying 50 in-lbs (5.6 N-m).
• Loctite ® 271 or equivalent high strength threadlocker.

Furnished Materials (Kit 5198686)


This kit includes the following parts.
Item # Qty. Description Part #
1 1 Screw 46-170498P55
2 1 Setscrews 46-208563P4
3 1 Nylon Washer 1000904P525
4 1 Instructions (this document) 5212837-100

Personnel Requirements
Two service engineers for 0.5 hours of labor.

Procedure
1.) Remove monitor from monitor yoke.
2.) Refer to Figure 1. Remove button plug from the bottom of the monitor yoke.
3.) Remove and discard the setscrew and center screw.

Note: Removing the center screw detaches the yoke from the upper pivot. Make sure the yoke is
supported while removing the center screw.

4.) Install nylon washer onto new center screw (provided).


5.) Thoroughly coat the threads of the new center screw (provided) with Loctite ® 271 or
equivalent threadlocker (approximately 10 drops), make sure that none of the threadlocking
agent comes in contact with the nylon washer.
6.) Install the center screw with threadlocking agent though the monitor yoke and into the upper
pivot.
7.) Tighten center screw to obtain a clearance of 0.05±0.02 in. between the upper pivot and
monitor yoke.
8.) Apply Loctite ® 271 or equivalent threadlocker onto the threads of the hole where the setscrew
will be installed.
9.) Thoroughly coat the threads of the setscrew (provided) with Loctite ® 271 or equivalent
threadlocker (approximately 3 drops).
10.) Install setscrew and tighten to a torque of 50 in-lbs (5.6 N-m).
11.) Install button plug removed in step 2.
12.) Re-install monitor onto monitor yoke.
13.) Restore power to the system and perform a functional check of the system. Verify that an
image appears on the monitor during a fluoro exam.