Monitor Suspension Hardware Replacement
Instructions
Page 2
GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100
Revision History
Page 4
GE HEALTHCARE MONITOR SUSPENSION HARDWARE REPLACEMENT INSTRUCTIONS
REVISION 1 DIRECTION 5212837-100
Personnel Requirements
Two service engineers for 0.5 hours of labor.
Procedure
1.) Remove monitor from monitor yoke.
2.) Refer to Figure 1. Remove button plug from the bottom of the monitor yoke.
3.) Remove and discard the setscrew and center screw.
Note: Removing the center screw detaches the yoke from the upper pivot. Make sure the yoke is
supported while removing the center screw.