Professional Documents
Culture Documents
MT Level III Exam Bank Answers
MT Level III Exam Bank Answers
.
A lower frequency than for small parts
1003 For best demagnetization results using a coil with A.C. and passing parts through it you should
.
C both a and b
1005 The magnetic flux in a part increases as applied field is increased. The flux moves along the virgin
. curve of the hysteresis curve
C only once when the part is completely demagnetized
1006 Small parts can be demagnetized in multiple lots by passing them through an A.C. coil provided they
. are
D both b and c
1007 For rings and hollow parts a handy means of A.C. demagnetization is
.
B a central conductor using decreasing current
1008 The most likely tool you would use to determine if a part has been adequately demagnetized is the
.
B field meter
1011. Prior to the use of magnetic particles to detect flaws by magnetic methods, flaws could be located in
iron pieces using
B ordinary magnetic compasses
1013 Total current in two parallel conductors can be determined by knowing the distance between the
. conductors and the
D magnetic force exerted per unit length of wire
1015 The units Webers per square metre (Wb/m^2) are used to quantify a
.
A magnetic induction field
1
ESTestMaker Answers
1017 In a ferromagnetic material, a region where all the "atomic moments" are aligned parallel to each other
. is called a
C domain
1018 A void in an otherwise homogenous magnetized material presents itself as a point magnetic dipole.
. This is the basis for
A magnetic particle flaw detection
1020 When discussing magnetization and its effects the magnetic induction field is usually represented by
. the letter
A B
1021 When discussing magnetization and its effects the magnetic field intensity, or as its also called, the
. magnetic field is represented by the letter
B H
1022 The phenomenon whereby the magnetization of a ferromagnetic sample lags the applied cyclic
. magnetic field is referred to as
C hysteresis
1023 When a hysteresis curve is generated for a ferromagnetic material, a magnetic field remains in the
. material when the applied field is returned to zero. This field is called the
C residual induction
1024 The value of H applied to bring the residual value of B to zero is called
.
C induction coercive force
1028 Magnetic particle testing has an advantage over liquid penetrant testing because it can
.
C locate subsurface defects
2
ESTestMaker Answers
1030 When a discontinuity causes the magnetic field lines to be forced out of the specimen into the air, the
. field in the air is called a
B leakage field
1035 In order to detect a flaw by magnetic particle testing the magnetic field must
.
D intercept the principal plane of the flaw
1038 If the ends of a long bar magnet are bent to form a circle and the ends perfectly fused together the
. result will be
D all of the above
1041 Inside a bar magnet magnetic flux lines are considered to flow from
.
B south to north
1042 The direction of magnetic lines of force around a current carrying conductor is
.
B at right angles to the conductor
1043 If a nail is brought to a bar magnet it is held by the magnet. If another nail is now brought to touch the
. first nail it is held to the first nail. Subsequent nails can be held in a long chain. This is a result of
A induced magnetism
1044 The end of the compass needle that points towards the northern region of the earth is called
.
D both a and b
3
ESTestMaker Answers
1046 A single line of force represents the unit of magnetic flux in a field. This is called a(n)
.
D maxwell
1049 A permanent magnet configured such that both north and south poles are adjacent is usually termed a
.
D horseshoe magnet
1050 The "image" of a magnetic field produced by placing a magnet under a paper and sprinkling iron
. fillings on the top of the paper is called
C a magnetograph
1051 In direct current flow electric circuits, the unit of current is the
.
A ampere
1052 In alternating current electric circuits the unit of current flow is the
.
A ampere
1054 Electric current which at regular intervals reverses its direction of flow is called
.
D both a and c
1060 The ability of a material to emit visible light while being irradiated by UV light is called
.
A fluorescence
1062 The clamping contacts on a large stationary magnetic particle testing unit are called
.
B heads
1063 When a part is magnetized such that the flux lines run essentially parallel to the long axis of the part it
. is called
C longitudinal magnetization
1067 The relationship relating electric current flow, voltage and resistance in electric circuits is
.
B Ohm's law
1069 To determine the direction of magnetic field lines in a conductor (assuming - to + current flow) you can
. use
A the left hand rule
1070 For a single straight conductor, assuming modern theory current flow, the fingers on the left hand point
.
A in the direction of the magnetic field lines
1071 In the left hand rule as applied to a solenoid, the fingers point in the direction of
.
C current flow
1072 In the left hand rule, as it applies to a solenoid, the thumb point in the direction of
.
A north
1073 The relative amount of flux linkage between two coils (as in a transformer) is called
.
C coupling coefficient
1075 The difference between a yoke and prods in magnetic particle testing is
.
B a yoke produces a longitudinal field and prods a circular
1076 As one increases the size of the MPI test system from small yokes and portable units to large bench
. units the single most significant difference is
C the amount of magnetizing current available
1077 When single phase AC is halfwave rectified it is sometimes called half wave direct current. The
. reference to direct current is because
C current flow direction does not change
1081 The number of reversals normally required to properly demagnetize a part using electromagnetic
. means is
C 10-30
1083 What shape is used for the particles used in magnetic particle testing?
.
D all of the above
1085 What is the result if magnetic particles are made of a material of high coercive force?
.
D all of the above
1086 Large sized particles are not used in wet magnetic particle testing because
.
D both a and b
1088 The reason agglomerations of fine magnetic particles tend to move through the liquid they are
. suspended in is
C both a and b
6
ESTestMaker Answers
1089 Maximum particle mobility is had for dry magnetic particles with _______ shape.
.
A spherical
1090 The reason for adding a portion of spherical shaped magnetic particles to a batch of elongated shaped
. dry powder magnetic particles is to
B increase mobility
1093 In MPI testing using dry powder, mechanical blowers can be used for
.
D both a and b
1094 Dry magnetic particle powders must be handled with care because
.
A the ferrous powders and pigments are considered nuisance dusts
1095 For sensitive wet method magnetic particle testing, coloured particles are typically 5 to 15 µm
. diameter. Unpigmented ferromagnetic oxides for the same sensitivity would be
A 0.5µm to 1.5µm
1096 In general, likelihood of flaw detection by MPI is statistical and increases with
.
A flaw size
1097 Which is not a factor that will affect the level of magnetic flux leakage from a flaw?
.
C part thickness
1098 What is the purpose of limiting the flux density generated in a piece being MPI tested?
.
A reduce the occurrence of false reporting of flaws
1099 For precise delineation of the contour and extent of a magnetic particle indication is required the best
. way to record it is
B tape transfer or lacquer
1101. When photography of fluorescent indications is performed ____________ are used to prevent
overexposure of film.
C special filters
1102. For photography of MPI indications the use of a tripod is recommended. Why?
B exposures are long and camera movement would ruin the shot
7
ESTestMaker Answers
1104. When a ferromagnetic material is placed in an inducting field H, and H is increased until the
incremental permeability is simply the free space permeability this point is called the
A magnetic saturation
1105. The small jumps in magnetism as a ferromagnetic material is exposed to an increasing induction field
are termed
D Barkhausen jumps
1106. In photography of NDT indications filters are often incorporated as part of the lens optics, why?
C both a and b
1107. The slope of the virgin magnetization curve at the origin of a hysteresis plot is called the
D initial permeability
1108. A closed magnetic path through one or more materials is considered a magnetic circuit. As such, it
can be considered analogous to an electric circuit, the magnetic equivalent to voltage is
C magnetomotance
1109. The difference in colour between an MPI indication and the background colour of the part tested is
critical in evaluation of a part. The colour difference is referred to as
A contrast
1110. Dry magnetic particles should be applied so they have a minimum velocity when they reach the test
part. The reason being
B to ensure they can be caught by leakage fields
1112. What is the name of the technique whereby a part is magnetized first and the particles applied after
the magnetizing force is removed?
A residual method
1113. What is the name of the MPI technique whereby the magnetic particles are applied to a part only
during the time the magnetizing force is applied?
B the continuous method
1114. The purpose of contact pads placed on the electrodes used in MPI is
D to improve electrical contact
1115. The localized overheating of a part being magnetic particle tested, arising from high resistance or
sparking at the electrical contact point is
A burning (or arcing)
1116. The general appearance of the test surface on which indications are to be viewed is termed
C background
1117. The buildup of magnetic particles due to excessive magnetization of the part under test is called
A furring
8
ESTestMaker Answers
1119. A five turn coil on a bench unit is 30cm diameter. If 2000amps are passed through it will result in
A 10,000 ampere turns
1120. The process whereby the magnitude and significance of an indication is determined is termed
C evaluation
1121. A "keeper" is used on a permanent magnet to complete the magnetic circuit. The purpose of this is to
A prevent loss of magnetism
1123. The non-relevant indication resulting when the surface of a magnetized part comes in contact with
another piece of ferromagnetic material is
C magnetic writing
1124. Which of the following materials has the highest relative permeability
A iron
1125. The two basic decisions made before an MPI test is carried out are
C type of current and type of particle to use
1126. To detect a flaw in a low carbon steel shaft, if you suspect the flaw is about 6mm (1/4") below the
surface which method would you use?
B half wave rectified A.C.
1127. Which is the best material for detecting defects wholly below the surface using MPI tests?
A dry powder
1128. Low carbon steels must have the continuous method of MPI testing used on them due to
C low retentivity
1129. The magnetizing current in the continuous wet method, when the bath is applied by flooding from a
nozzle, is usually applied
B immediately after flooding is stopped
1130. The normal duration required for a magnetizing "shot" in the continuous wet method is about
A 1/4 to 1/2 second
1132. When a central conductor is used inside a tube the type of magnetization employed for the testing of
that tube is
B circular
9
ESTestMaker Answers
1134. Field strengths of the magnetic field produced by a solenoid (or coil) can be changed by
C both a and b
1135. The guideline to use 40 amperes of magnetizing current per mm of part diameter applies to
A head shots
1136. If using the rule of thumb of 40 amps per mm of part diameter for a head shot, what current is required
to magnetize a 20mm diameter copper bar?
D none of the above
1137. For the same diameter, length and current used, the difference between using a non-magnetic and a
ferro-magnetic central conductor is
A magnitude of magnetic field on the outside surface of the conductor
1138. When performing MPI using prods, the current used depends on
D both a and b
1139. If 24000 ampere turns are required for coil shot using a 5 turn bench coil, what should your current
setting be?
B 4800 amps
1140. In what way does surface roughness hinder the MPI test on a part?
D all of the above
1141. What do the following cleaning methods all have in common? Shot blasting, sand blasting, wire
brushing and scraping
C all are mechanical methods
1142. Which cleaning method is not likely to be used to remove oil films prior to an MPI test?
B wire brushing
1143. The magnetic field produced by a yoke made of permanent magnets is _______ between the poles.
A longitudinal
1146. When permeability of a material is quoted without an additional qualifying term the permeability
referred to is
C maximum permeability
1147. Current for circular magnetization of a part whose maximum permeability is 500 will be ___________
compared to a part with maximum permeability of 4000.
A more
10
ESTestMaker Answers
1148. In MPI a handy rule of thumb relating current flow and suspected flaw orientation is; current flow
should be directed _______ major axis.
B parallel to suspected defects'
1149. The strength of a circular magnetic field is not diminished by ________ of a part.
C length
1150. A 2cm diameter pipe and a 2cm diameter bar of the same steel alloy are each circularly magnetized by
a 1000 amp D.C. head shot. The strongest magnetic field will occur on the surface of the __________
-.
D none of the above, fields will be equal
1151. When a hollow non-magnetic central conductor is used to induce a circular magnetic field in a ferro-
magnetic cylinder maximum field strength is at the
C magnetic material's inside surface
1152. When A.C. is used instead of D.C. for a head shot on a steel tube, the field on the inside surface of the
tube is
D both are exactly zero
1153. The best way to determine magnetizing current required for irregularly shaped pieces is
C empirically (trial and error)
1154. When using the prod method of MPI, defects are normally looked for
A between the prods
1155. In order to maintain the same field strength at the midpoint of prods spaced 8 inches apart as was had
with a 6 inch spacing you must
A increase the magnetizing current
1156. Which is not a consideration when choosing current type to use for prod inspection methods of MPI?
B safety
1157. Capacitive discharge methods of producing a magnetizing current do so for only a short duration.
C typically 10 to 100ms
1159. Typical pocket sized magnetic field indicators are used to measure
B external residual fields
1160. A low cost rugged tool used to measure residual magnetic field strength of ferromagnetic parts is a
C magnetic field indicator
1161. A good magnetic field indicator (MFI) should have which characteristic?
D all of the above
11
ESTestMaker Answers
1163. The standard steel ring 7/8" thick having a 1 1/4" diameter central hole and twelve 0.007" diameter
holes arranged with increasing depth from the outside edge of the ring is called a(n)
B Ketos ring
1165. Although settling time prior to measuring wet fluorescent magnetic particle concentration is now
increased to 1 hour for most standards, complete settling usually takes
D 24 hours
1166. When viewing the oil carrier of wet fluorescent particles (ie. the liquid without the particles) under
blacklight it usually has a _________ tint.
B bluish
1168. Indications from medium to deep surface breaking seams found using A.C. MPI
B usually look alike
1171. Because in general, there exists a direct relationship between indication width and seam (defect)
depth, a minimum magnetizing current level can be set to
B detect significant flaws
1172. Very fine MPI particles which fall into a crack opening can cause the flux lines to remain in the part,
this could result in
D missing the crack all together if nonfluorescent particles are used
1173. When lighting requirements are specified for a blacklight viewing of fluorescent MPI, 20 lux white light
intensity is
A a recommended maximum
1174. What method is used to ensure proper magnetization levels are used?
D all of the above
1175. An advantage of MPI over LPI (Liquid Penetrant Inspection) is that ___________ using MPI.
C coated surfaces can be tested
1176. The difference between magnetic particle testing and magnetic flux leakage testing is
A the type of sensor used to detect fields
12
ESTestMaker Answers
1177. The "standard depth of penetration" resulting from the skin effect is caused by
B alternating current
1179. The main purpose of a ferrite core in the coil type flux sensitive devices for magnetic field testing is
B to amplify the flux density
1180. In addition to flux density amplification, ferrite cores in coil type flux sensitive devices provide
C both a and b
1181. The crystals used in Hall detector type flux sensitive devices are
D semiconductors
1182. The main advantage of the Hall detector as a flux sensing device is
A the small size of its active area
1183. Which materials are not used to make wet method particles for MPI testing?
B red iron oxide
1185. A ferro-probe (or Foerster microprobe) has a high frequency alternating current in a coil mounted on a
_______ core.
A ferrite
1186. Most non-fluorescent magnetic particles used in the wet method are
B unpigmented ferromagnetic iron oxides
1187. For a given mass of fluorescent magnetic particles, 60µm sized particles would be more visible by
containing more fluorescent pigment than 125µm particles. The reason for this is the smaller particles
C have more surface area
1188. Which is not an advantage of oil based vehicles versus water based vehicles of wet magnetic
particles?
B low fire hazard
1189. Which is not an advantage of water based vehicles versus oil based vehicles of wet magnetic
particles?
C lower electrical hazard
1190. When fine fluorescent magnetic particles are seen floating on the surface of their water carrier like
dust and no amount of mixing disperses them it indicated
B insufficient or lack of wetting agent
13
ESTestMaker Answers
1192. Mineral seed oil is sometimes added to a part after MPI testing with water or oil based baths to provide
D longer term rust prevention
1193. When preparing a bath from dry concentrated MPI particles, in either oil or water vehicle, if the tank is
not equipped with a re-circulation pump uneven distribution resulting from hand stirring can be avoided
by
A premix the quantity into a paste in a blender or small container before adding to the bath
1196. Settling volume determination used to establish MPI bath particle concentrations usually uses which
implement?
B a centrifuge tube
1197. Bath concentrations for magnetic particle testing are usually specified as having
C a range of values
1198. The practical upper limit temperature of the part surface performing MPI on a part using a water
vehicle is
C 100°C.
1199. The result of vibrating a sample of magnetic particle bath being tested for concentration is to
B decrease apparent concentration by compaction
1200 The source of contaminants that can cause erroneous concentration readings for MPI baths is
.
D all of the above
1201 Blacklight intensity in post 1990 specifications nearly always quote minimum irradiance in
.
D µW/cm²
1202 The unit of illuminance used to quantify white light for visual inspections is
.
A lux
1203 The SI that is now used as the standard measure of the wavelength of light is the
.
B nm (nanometer)
1204 As light intensity decreases in the visible range the retina's cone shaped cells shut down and the
. ______ shaped cells assume the role of light collection.
C rod
1205 The cells commonly referred to as the "cone shaped cells" and the "rod shaped cells" are to be found
. in the
A eye
14
ESTestMaker Answers
1206 Eye glasses with blue-absorbing filters are used to view fluorescent indications because
.
C both a and b
1207 Maximum contrast when viewing magnetic particle indications would occur for
.
A bright daylight conditions
1209 To reduce the discomfort of eye fatigue when performing fluorescent MPI you can
.
C both a and b
1210 In MPI testing, the difference between the amount of light reflected by the surface of a test piece and
. the amount reflected by an MPI indication is called
B brightness contrast
1211. If two objects, such as an MPI indication and its background test surface have the same apparent
brightness but reflect light at different wavelengths they would exhibit
A colour contrast
1212 Dark adaption from bright daylight to full darkened conditions for a blacklight viewing booth will require
. about
B 2 minutes
1213 In automated scanning systems used for defect detection in MPI testing unwanted background lighting
. is illuminated by
C filters
1214 The purpose of the deep purple filter used on the mercury vapour lamp is to
.
D both a and b
1215 Which of the following can result in an increased output from a blacklight?
.
B line variations in voltage supply
1218 The purpose of running the circulation pump prior to performing the MPI settling test is to
.
C ensure even particle distribution
1219 The first thing you would do if your mercury arc blacklight did not have adequate intensity when tested
. is
B clean the filter
15
ESTestMaker Answers
1220 Which is a means of determining if the applied magnetic field is in the correct orientation?
.
D all of the above
1224 In the wet MPI method, timing of the magnetizing current with respect to the flowing of bath over a test
. piece is especially important if the test piece is of low retentivity because
C indications could be washed off if flow continues after the shot
1225 The primary advantage of the television imaging systems over flying spot systems used for MPI
. scanning systems is
A cost and availability
1226 The glass commonly used as a filter for blacklights is Kopp 41 (Kopp Glass Inc. Pittsburg, PA.,USA).
. It effectively
D both a and c
1227 Automated scanning systems rely on the _______ to establish if an MPI indication is real or not.
.
B linear nature or patterns of defects
1228 The problem with using drawings or written descriptions as a means of recording an MPI indication is
.
A they are highly dependant on the writing or drawing skills of the inspector
1229 The tape transfer method of recording an MPI indication is easiest to use with
.
C dry magnetic particle method
1230 A major problem with trying to take a tape transfer of an MPI indication where the vehicle was oil
. instead of water is
A longer drying time
1231 Which method of recording MPI indications is different from the others?
.
A a photograph
1232 When using fixing coatings to adhere MPI indications to the test piece surface the part is usually dried.
. What is the reason for this?
C to allow the coating to adhere to the part
1233 Which of the following methods of recording MPI indications is different from the other 3?
.
A magnetic rubber
16
ESTestMaker Answers
1234 The magnetic rubber method of producing permanent records of MPI indications
.
B requires very long duration magnetization shots
1235 The recommended way to make high quality recordings of MPI indications is by
.
C photography
1236 The magnetic particle method does not work if the magnetic flux density is below
.
D none of the above, there is no sudden cut off point below which MPI does not work
1237 When the central conductor MPI method is used to inspect a ring shaped object, or short cylinder
.
C both inside and outside surface and subsurface flaws can be found if the wall thickness is not too great
1238 Rating of yokes is either by the force needed to pull one leg off the specimen or by the lifting power of
. the yoke. The lifting force of a yoke is
B twice the pull off force
1239 When lifting power of an A.C. yoke is specified in kg. you must also specify a minimum
.
C prod spacing
1240 When rotating field MPI techniques are used it is required that you use
.
B the wet method
1241 On which of the following surfaces would you not likely use the dry MPI method?
.
C wet or oily surfaces
1242 The region outside a magnet, where the magnet exerts its influence is the
.
A magnetic field
1243 The following are properties of what? a) they begin and end on poles, b) they appear to repel each
. other, c) their direction is that in which a free north pole would travel.
B magnetic lines of force
1244 If a ferromagnetic material is not taken to magnetic saturation on its hysteresis loop it will follow
. ________ when the applied field is reversed.
B a minor loop
1246 During magnetization of an initially unmagnetized specimen the total number of domains in the
. specimen
C decreases
1247 Steels used in transformer cores and electromagnets would have which characteristics?
.
D all of the above
17
ESTestMaker Answers
1248 The super magnets, such as samarium-cobalt and neodynium-iron are virtually impossible to de-
. magnetize by any means other than raising their temperature over the curie point because of
C high coercivity
1249 As compared to inspecting at the middle of a long tube, when inspecting the end of a tube using the
. longitudinal MPI method, ______ are required of the coil.
A higher fields
1251 The purpose of a silicon controlled rectifier (SCR) in the circuitry for pulsed or capacitor discharge
. magnetization is
A to ensure the pulse in unipolar
1252 For field applications, in order to induce a longitudinal field in a section of piping, if a fixed coil or split
. coil is not available, you would probably use
B several turns of welding cable around the pipe
1253 As paint coating thickness increases on a ferromagnetic plate, the force required to pull the yoke off
. the plate
B decreases
1254 What is the disadvantage of doing MPI by the residual method when looking for sub-surface
. indications?
B excessive surface noise can mask small indications
1256 "Banding" of magnetic particles when using the dry powder method with prods indicates
.
B excessive current
1257 Most portable field indicators, such as the "pie" gauge are used to
.
C indicate general direction of surface fields
1258 Irregularly shaped parts that have been magnetized are demagnetized by
.
D there is no effective way to completely demagnetize an irregular part other than heating over the curie
temperature
1259 Soft iron bars placed across the pole ends of horseshoe magnets are called
.
B keepers
1260 To detect radial cracking on a bolt flange 2cm thick, 30cm diameter with an inside diameter of 25cm
. you could probably use
B a central conductor for circular magnetization
18
ESTestMaker Answers
1262 Welding machines, magnetic chucks, contact with electromagnetic lifting devices and permanent
. magnets are the cause of what problem?
B unwanted residual magnetism in ferromagnetic parts
1263 If a UV filter used for photography of fluorescent indications was made of fluorescent material what
. would the effect be?
A cause fogging of the image
1264 When taking photographs of NDT indications it is often advised to take several exposures of the same
. view but using different f-stops. The reason is
B to compensate for unforeseen conditions
1265 The problem with de-magnetizing a piece by raising its temperature to over the curie point is it
.
A is often impractical
1266 Ferromagnetic materials that contain very high residual fields are
.
D none of the above, retentivity has nothing to do with ease of demagnetization
1267 What is the advantage of photography of MPI indications over other archival methods?
.
D all of the above
1270 When using reversing D.C. to demagnetize a part about 30 reversals of decreasing amperage are
. used in a time of
C 30 seconds
1271 In most cases A.C. demagnetization methods are adequate for parts having a cross-section of not
. more than
B 50mm
1272 The coil 30 point step down D.C. demagnetization is not an acceptable method for
.
C small sized, soft steel objects at high production rates
1275 Operator safety as it concerns MPI testing equipment usually concerns the
.
D both a and b
19
ESTestMaker Answers
1276 The safety hazard associated with petroleum distillates used as vehicles for MPI is
.
D all of the above
1277 When working over tanks of petroleum distillates used for MPI particle vehicles the safety precaution
. that would be taken is
C ensure proper ventilation
1278 Some blacklights come with thin outer glass filter envelope around the discharge lamp. Should this
. outer shell break the inner lamp continues to work. If the outer shell cracks you should
B turn it off and discard it immediately
1279 Even a cool well ventilated environment, there is still a safety risk of explosion when performing
. fluorescent MPI. What is the source of this risk?
C high intensity discharge lamps
1280 The ultraviolet light used in NDT for viewing fluorescent indications is called
.
D all of the above
1281 Prior to the main electrodes arcing what must occur in the mercury vapour blacklight lamps?
.
C the mercury must vaporize
1282 The purpose of the aluminized surface on the inside surface of the outer glass bulb of the mercury
. vapour lamp is to
B produce a spot profile of light
1283 When voltage drops more than 15% or power has been switched off to the blacklight the arc ceases
. and attempts to immediately restart it will fail because
C mercury vapour pressure must reduce by cooling before the arc will re-start
1284 Bulb radiance output from a mercury vapour blacklight can rise suddenly as a result of
.
A line voltage surge
1285 The phosphor coating on the glass filter envelope of blacklight tubes emits
.
B UV-A and visible light
1286 The reason blacklight tubes are not used for critical fluorescent viewing is
.
A their irradiance is mush lower than the medium pressure mercury vapour lamps
1287 In MPI testing (and any work with electric apparatus) risk of dangerous electric shock is increased in
. the presence of water or moisture on or around the body because
C the body's electrical resistance is lowered allowing higher currents to flow
1288 From an MPI safety point of view, if you cannot eliminate the possibility of a spark occurring you
. should
B control the environment so an explosion cannot occur
1289 Adherence to the empirical rules set down for MPI will
.
C often lead to over magnetization and poor coverage
20
ESTestMaker Answers
1291 Which of the blocks listed below is not a magnetic particle inspection reference standard?
.
C I.I.W. calibration block
1292 The number of holes 0.07mm diameter found in the ketos tool steel ring standard is
.
C 12
1295 When viewing parts with blacklight both direct and reflected blacklight striking the eye should be
. avoided because
C vision becomes clouded due to ocular fluorescence
1296 The mercury found in the mercury vapour lamp in its cold or off state is
.
B condensed droplets of liquid mercury
1297 Blacklights used for fluorescent indication viewing are usually mounted in fixtures for what reason?
.
D all of the above
1299 UV lamps used for viewing fluorescent indications tend to operate under higher than optimum design
. conditions. As a result
C life expectancy is lower than manufactures' estimates
1300 The unit of light intensity is the lux or footcandle which is based on the human eye's response to light.
. Why is it an inappropriate unit for UV radiation?
A UV radiation is not sensed by the eye
1301 The reason high quality UV light filters, having well defined upper and lower limits, are not used for
. measuring UV intensities in NDT applications is primarily because
B of cost
1302 UV meters should be calibrated about every six months, this is because
.
D both a and b
21
ESTestMaker Answers
1305 When welds are required to have only partial penetration use of HWDC yokes often give rise to
. nonrelevant indications. This can be overcome by
A using A.C.
1306 When using dry fluorescent powders outdoors, ambient lighting must be
.
B minimized
1307 To detect fatigue cracks in a field weld you would most likely use
.
B an A.C. yoke
1308 For crack detection on a painted steel surface that has had the paint removed, MPI as compared to
. Liquid Penetrant Inspection is
B far better
1309 Particle concentrations for magnetic particles applied from aerosol cans are
.
B known and accurate only when the can is full
1310 When performing MPI with A.C, the resulting eddy currents provide
.
A a demagnetizing effect
1311. In order that they be held by weak leakage fluxes, dry magnetic particles must have
B high permeability
1313 Which of the following would risk erroneous test results for an MPI test?
.
D all of the above
1314 If the magnetizing current ammeter on your bench unit is no longer reading current when you hit the
. "on" switch for the current, the problem is
C both a and b could cause the problem
1316 Rejecting all parts in an MPI inspection which contain relevant indications is
.
D both a and b
1317 Which of the following flaws in a highly stressed steel component is most likely to be cause for
. concern?
A a shallow vee shaped scratch
22
ESTestMaker Answers
1319 MPI indication formation for dry powder methods will best be formed on _______ surfaces.
.
C smooth, flat (horizontal)
1320 For dry powder MPI methods, inspection on flat horizontal surfaces is preferred because
.
A particles are not pulled off by gravity
1321 Cold working steel can result in non-relevant MPI indications, these result from
.
B permeability changes from increased harness
1322 Although a welding machine can be used to supply many amps of current it is not used for practical
. MPI testing for which reason?
B voltages are too high
1323 Which NDT method would you use to locate shrinkage in a steel casting?
.
C radiography
1324 Which NDT method would you use to detect tight cracks in tungsten forgings
.
B liquid penetrant
1326 When the contact heads of an MPI bench unit are used to direct current along cable to test a large or
. remote piece, the spacer placed between the heads is
C an insulator
1327 For MPI applications, horseshoe permanent magnets have their advantage in
.
B portability
1328 Which NDT method would you use to locate defects in 15mm diameter monel tubing?
.
C ultrasonics
1329 Which of the following NDT methods would be best suited to inspect thin wall (1mm) 15mm diameter
. monel tubing?
C eddy current
1330 Magnetic flux lines intercepted by a discontinuity lying just below the surface flow
.
D all of the above
23
ESTestMaker Answers
1332 The most likely cause of no current reading when you apply a prod shot to a rough casting (assuming
. you have turned the unit on) is
C poor prod/casting contact
1333 Dry magnetic particle techniques would not be well suited to finding
.
B grinding cracks
1334 In MPI testing the dry powder equivalent to the wet particle pressurized sprayer is the
.
B powder blower
1335 The primary difference between magnetic particle testing and other methods of testing using
. magnetics, such as electromagnetic resonance, electron paramagnetic resonance and eddy current
testing is
B frequency of field oscillation
1336 "Dam" structures are used in MPI testing using _______ techniques.
.
C magnetic rubber
1338 Magnetic particles used in wet methods have _________ to prevent clumping and high background
. buildup.
D both a and b
1339 Low retentivity and low coercivity are requirements for magnetic particles to
.
C decrease background
1340 Two long conductors (wires) are placed parallel and near to each other. When current flows in the
. same direction in both wires there will exist a
B net attractive force between the wires
1341 Two long conductors (wires) are placed parallel and near to each other. When current flows in
. opposite directions in both wires, there will exist a
B net repulsive force between the wires
1344 The magnitude of a magnetic induction field at a perpendicular distance 'r' from the centre of a long
. straight conductor with a current 'I' amperes is
C both a and b
1345 When a current is passed through a long circular conductor of radius R such as a wire the maximum
. magnetic induction field occurs at
C R (i.e. the conductor's surface)
24
ESTestMaker Answers
1346 The most significant factor contributing to a material possessing magnetic properties is
.
B unpaired electrons in the structure
1348 In ferromagnetic materials the ratio of the magnetic induction field to the magnetic field intensity is
. called the
B permeability
1349 The internal magnetic intensity of a specimen brought into an external magnetizing field will always
. __________ the applied field.
A be less than
1350 Magnetic circuits are equivalent to electric circuits hence have an equivalent equation to Ohm's law,
. the term equivalent to electrical resistance in magnetics is
B reluctance
1351 In a magnetic circuit, which is equivalent to an electric circuit, the magnetic equivalent to electrical
. current is
D magnetic flux
1352 The ability of the magnetic particle method to detect subsurface defects is not determined by
.
A specimen size
1353 For colour photography of both visible and fluorescent indications it is recommended to use
.
B normal daylight film
1356 Polarizing filters are used in MPI indication photography for what purpose?
.
C to reduce highlights and reflections
1358 When photographing MPI indications it is often a good practice to expose extra frames using higher
. and lower f-stops to compensate for unforeseen conditions. This practice is called
A bracketing
1359 Which of the following is an advantage of video tape recording of MPI indications?
.
C availability to do subsequent image enhancement
25
ESTestMaker Answers
1360 The fact that only some magnetic domains resume a random orientation when the magnetizing field is
. removed accounts for what?
A retentivity
1362 The requirement for multiple switching of polarity and simultaneously reducing current amplitude is
. had for
A D.C. demagnetization methods
1363 In the reversing D.C. method of demagnetization the current is gradually reduced to
.
A zero
1365 The zone of transition between one magnetic domain and the adjacent magnetic domain is called
.
C the Bloch wall
1366 Compared to particles used in wet MPI, those used in dry MPI methods are usually
.
C of higher magnetic permeability
1367 An excessive number of fine small particles in an MPI test will result in
.
A excessive background
1368 What is an advantage to having some residual magnetism in dry magnetic powder particles?
.
B detection of weak leakage fields from subsurface defects
1369 The demagnetization factor results wherever magnetic field lines leave a test piece. In general the
. demagnetization factor
D both a and b
1371 The yoke break test was designed to evaluate flux strength in a steel plate through various coating
. thicknesses. The force to pull the yoke off the plate
D both a and b
1372 An off centred central conductor with a pulsing current supply has the advantage in MPI testing of
. tubular products of
C inducing eddy currents which reduce non-uniformity of field
26
ESTestMaker Answers
1374 NDT specifications may be revised under which of the following conditions?
.
C both a and b
1376 The reliability of finding deep defects wholly below the surface has decreased with the use of the
. harder alloyed steels used in pressure vessels. This is due to
C permeability decreases with increasing hardness
1377 The collection of magnetic particles caught by a leakage field in a magnetic particle test indicates a
. discontinuity's
D all of the above
1378 Very wide surface cracks may not produce a powder pattern because
.
A the surface opening is too wide for the particles to bridge
1380 Magnetic particle testing would give best results when used on
.
B soft steel
1381 Which of the following might be considered a disadvantage of magnetic particle testing?
.
C need to demagnetize parts
1382 Eddy current testing can also be used to detect surface and sub-surface defects on ferromagnetic
. materials but is usually inferior to magnetic particle testing because of
D both a and b
1384 The void that occurs at the top centre portion of an ingot is usually called a
.
C shrink cavity
1386 Blowholes not welded shut in subsequent rolling processes can result in
.
A seams or laminations
1387 Seams in bar stock formed from under-filled rolls are usually
.
C nearly normal to the surface of the bar being formed
1388 Seams in barstock formed as a result of over filled rolls are usually
.
D at an acute angle to the surface of the bar
27
ESTestMaker Answers
1390 If a compass needle is brought near a bar magnet the needle will orient itself ______ to the magnetic
. lines of force at any point.
A parallel
1391 The Weber is the MKS form of magnetic flux. The cgs unit is called the maxwell. 1 weber = ________
. maxwells
D 10^8
1396 The negative magnetizing force required to reduce the flux density in a part to zero after saturation is
. called
B coercive force
1398 Short duration currents resulting from sudden changes in electric or magnetic condition within a circuit
. are called
D transient currents
1399 The degree to which a material holds a residual magnetic field is termed
.
D retentivity
1400 The opposition to the formation of a magnetic flux in a magnetic circuit is referred to as
.
B reluctance
1401 When an electric current, a magnetic field or an electrical potential drops off to zero, the reduction in
. amplitude of these quantities is given the same term. That term is
A decay
28
ESTestMaker Answers
1406 The wavelength of blacklight used in fluorescent magnetic particle testing is usually around
.
B 3650Å
1408 When a current is passed directly through a part the magnetization that results is considered to be
.
B circular
1409 A shot of magnetizing current passed through a part while clamped between the clamping contacts of
. a stationary unit is referred to as a(n)
D head shot
1410 A magnetic particle cluster resulting from a true leakage field but not originating from a metal flaw is a
. (n)
D non-relevant indication
1412 In a D.C. series circuit what resistance is required to cause a voltage drop of 170 volts when 5 amps
. are flowing?
B 34 ohms
1413 What is the current through a 50 ohm resistor if the potential across the resistor is 12 volts?
.
A 240 mA
1415 The now outdated "Conventional current flow" has electrons moving
.
B from positive to negative
1416 The left hand rule for determining magnetic field direction applies to
.
C modern current theory
1418 The lagging of flux density (B) behind the magnetizing force (H) is called
.
D hysteresis
1419 In a transformer, if the voltage in the secondary winding is stepped down, the current in the secondary
. will be
A stepped up
1420 Penetration of the magnetic field, all other factors being equal, is greatest for
.
A half wave rectified single phase AC
1421 If Is and Ip are currents, Es and Ep voltages and Ns and Np number to turns in the secondary and
. primary windings of a transformer, which is not true.
C Es/Ep = Np/Ns
1422 Alternating current used as a magnetizing source tends to concentrate the magnetic flux near the test
. piece surface because of
C the skin effect
1425 The temperature at which a ferromagnetic material changes to a paramagnetic material is the
.
B Curie point
1426 Upon cooling a ferrous alloy below its Curie point, in the absence of an external field it
.
C becomes ferromagnetic but unmagnetized
1430 To eliminate the circular magnetism that exists in a part by applying a longitudinal field may require
. several "shots" if the part
C has a large L/D ratio
1431 Substances whose molecular structure is such that an unbalanced antiparallel electron spin exists are
. said to have ferrimagnetism. These substances are called
A ferrites
30
ESTestMaker Answers
1432 Aircraft, automotive and railroad components that have been inservice for some time but not near an
. artificially produced magnetic field are often found to be strongly magnetized. This is a result of
B the earth's magnetic field
1434 Yoke demagnetization can be more effective than coil methods in some cases because
.
D the field can be more concentrated
1435 To detect residual circular magnetism in a circular part using a field meter you must
.
B make a saw cut into the part
1436 A ball bearing is to be tested by MPI. Which is the best option for magnetization?
.
B 3 shots at mutual right angles using induced current
1437 Which is not a problem that might result if the wrong choice of particles is made in MPI testing?
.
B oversensitivity
1438 The single most important property to consider in selection of magnetic particles is
.
D none of the above
1439 Magnetic particles used for wet MPI purposes are usually colour coated ___________.
.
C ferromagnetic oxides
1440 The particle shape of dry magnetic particles that is best suited to detection of weak leakage fields from
. sub-surface defects is
B elongated
1441 Clumping of wet magnetic particles on the test specimen is usually a result of
.
C too high a retentivity of the particles
1442 The size of most dry powder magnetic particles used in general flaw detection is
.
C a mixture of shapes and sizes
1445 The shape of particles used in the wet magnetic particle method is
.
D not important
1447 Which is not a colour of magnetic particles that should be used on grit blasted surfaces?
.
B white
1448 The advantage of the oil based magnetic particles as compared to the water based is
.
D none of the above
1449 A combination of the residual and continuous MPI methods are sometimes used if
.
C the part has relatively high retentivity
1451 The condition reached in a ferromagnetic material subjected to an A.C. magnetic field such that the
. incremental permeability is one during part of each cycle is termed
B A.C. saturation
1452 What is the condition reached in a ferromagnetic material subjected to a D.C. magnetizing current
. when the incremental permeability is unity?
A D.C. saturation
1453 The portion of an electromagnetic circuit that is inside the coil (winding) is called the
.
B core
1454 When a part is magnetized by a current flow of very brief duration it is said to be
.
C flash magnetized
1456 When an electric current is forced to flow across a gaseous gap the result is
.
B an arc
1458 The closed path followed by a magnetic flux line or group of flux lines is called a
.
B magnetic circuit
1464 To detect the #3 hole (0.7" dia. 0.21" deep) in a Ketos ring using a central conductor, which would
. detect the hole using the least current?
A half wave A.C.
1465 Sensitivity of the residual method of MPI testing when the test part is immersed
.
C increases with time immersed
1466 The standard 1/2 second "shot" duration in the continuous wet method assumes
.
A correct particle concentration
1468 When a tubular part is MPI tested using a head shot where the current is passed through the tube
. itself the field strength on the inside surface of the tube compared to the outside is
B zero
1469 Shafts, drums and girders can be _______ magnetized by coiling a flexible cable around the part and
. passing a current through the cable.
A longitudinally
1470 As result of a head shot, the magnetic lines of force take the general form of
.
C concentric rings about the part axis
1471 Equations for ampere turns required for a coil shot are usually qualified by conditions. Which is not a
. condition usually stipulated?
B maximum part length
1473 Recommended D.C. current for a head shot is based on a part's diameter. Rule of thumb for a head
. shot is _______ amps / mm part diameter.
C 40
1474 Given a solid steel shaft 1m long and 20 cm diameter, what is the current recommended for a 5 turn
. coil shot assuming the bar is not centred in the coil?
B 1000 amperes
1475 Given a hollow steel shaft 1m long and 10cm diameter, what magnetizing current would you use with a
. 5 turn coil with the part at the bottom of the coil?
D 600 amperes
33
ESTestMaker Answers
1476 Given a steel bar 1 foot long and 3 inches diameter, what current should you use for the coil shot with
. a 3 turn coil (part placed at bottom of coil)?
B 2000 amps
1477 When it may be possible to damage parts exposed to magnetic particles or the liquid vehicle the
. inspector may have to
C plug holes and mask surfaces
1478 Vapour degreasing would not be an effective cleaning method for which material?
.
B mineral deposits
1479 A problem with performing an MPI test on a part that does not have its thin layer of paint or plating
. removed is
D all indications have the "fuzzy" subsurface appearance
1480 A coil shot is not effective if the ratio of the long to short dimension is less than
.
B 2
1482 When using 2 A.C. fields to provide a swinging field MPI test ________ to be effective.
.
C the 2 fields must be 50ø to 130ø out of phase
1483 In the swinging field method of MPI testing the application of the test medium (particles) is
.
B during the field application
1484 Calculations for coil currents for longitudinal magnetization of cylindrical parts assume the cross-
. sectional area of the coil is at least 10 times the part cross-section. What minimum radius would a coil
need to be to test a rod 20cm diameter?
A 31.6cm
1485 Although it is possible to magnetize a material whose maximum permeability is 100, a limiting factor
. might be
B current output of your equipment
1486 A circularly magnetized bar 2 inches in diameter carrying 1000 amps will have a field strength
. _________ a bar of the same material having a diameter of 4 inches carrying 1000 amp current.
D twice that of
1487 It may be possible to detect cracks on the inside surface of cylinders magnetized using a head shot.
. This is because
A cracks have depth
1488 When a central conductor is used to induce a circular field in a cylinder, maximum field on the cylinder
. is had on the cylinder inside surface when the central conductor is
D none of the above, central conductor material is not the controlling factor
1489 For very large parts where even stationary MPI units could not supply sufficient current for head or coil
. shots, it may be possible to secure higher strength fields by means of
A prods or yokes
34
ESTestMaker Answers
1490 In MPI testing of irregular shaped objects using D.C. head shots, field strength is
.
B greater at small diameters or sections
1492 When you move half the prod spacing from the midpoint of the prod centre-line the field is
.
C much less than on the centre-line
1494 Another term for the capacitive discharge method of producing magnetizing current is
.
A rod shot
1495 Prolonged agitation of wet fluorescent magnetic particles in large stationary bench units will result in
.
B reduced sensitivity due to abrading of fluorescent coating
1497 The sensing magnet in a magnetic field indicator brought next to a part with an associated external
. magnetic field will try to rotate ______ to the resultant vector of the two fields.
C parallel
1498 The pointer, attached to the sensing magnet in a magnetic field indicator points _______ to the field
. resulting from the interaction of the sensing magnet and the magnetic field of the part tested.
D perpendicular
1500 Using a Ketos ring and holding all test parameters constant except for fluorescent particle
. concentration; hole detectability _________ with increasing particle concentration
D increases to a maximum then decreases
1502 Magnetic particle inspection quantitative quality indicators can be used for
.
B detecting circular magnetism
1503 A Gauss-meter is a device used in conjunction with a ___________ to determine field strength.
.
A Hall-effect probe
35
ESTestMaker Answers
1504 A part that has been demagnetized using an A.C. demagnetization coil may be found to contain
. significant magnetization after being dropped or hammered. This is due to
C subsurface domains remagnetize the outer domains
1505 As magnetizing current is increased in the low to medium current range, magnetic particle indications
. width
A increases linearly
1507 To avoid an MPI bath washing an indication away when the current is removed you can
.
C both a and b
1508 Many daylight magnetic particles have fluorescence. The advantage of this is
.
D both a and b
1509 If very fine particles are used in an MPI test where a large crack occurs they could fill the opening.
. The bridging that results will
A short circuit the flux leakage lines causing them to remain within the part
1511. As it relates to the amplitude of the induced magnetic field, the _______ current value is most
significant.
A peak
1512 What is the only acceptable means of determining if magnetization is adequate when the residual
. method is used?
B a test piece (test standard)
1513 When sufficient magnetization is applied to a piece such that magnetic particles stick to minor surface
. field leakages not associated with discontinuities the part is said to be
A over magnetized
1515 The magnetic field properties that most affect flux leakage are
.
A field strength, permeability and angle to defect
1516 The magnitude of the voltage induced between the ends of a wire moving through a magnetic field is
.
B directly proportional to velocity of the wire crossing the flux lines
1517 A simple coil used as a flux sensitive device, when oriented parallel to the surface and positioned
. statically over the flux leakage field of a crack, will have
D zero voltage
36
ESTestMaker Answers
1518 Conductive metals do not make good Hall sensors because their
.
A Hall coefficients are too low
1519 The frequency of the voltage applied to the crystal in most Hall detectors used in flux sensitive devices
. is in the ________ range.
C kiloHertz
1520 In a magnetodiode flux sensor, the property that changes with magnetic field intensity is
.
A resistance
1521 The purpose of arranging a Hall detector at the mouth or apex of a ferrite ring (ferrite C-core) is to
.
C concentrate flux leakage for better sensitivity
1523 Dry powders cannot be used in water suspensions for the wet method of MPI because
.
D both a and b
1524 Light intensity requirements for daylight fluorescent dry magnetic particles are
.
C not specified in present standards or specifications
1525 Application of a light lacquer or penetrant developer can be used in dry powder MPI tests to enhance
. contrast but the disadvantage is
D both b and c
1527 The problem you would likely encounter if you tried to view daylight fluorescent dry powder in similar
. conditions used for wet fluorescent MPI (UV light in a darkened area) is
B excessive background fluorescence
1528 Fe3O4 (subscript numbers) is the chemical formula for which iron oxide?
.
A black
1529 True material magnetic permeability is not important in magnetic particles until
.
C the L/D ratio of the joined particles exceeds 1
1530 If a settling test is performed on fluorescent wet particles, and after allowing the sample to sit
. unagitated over night (10-15 hours) a thin, brightly fluorescent layer forms on top of the sediment,
B particle breakdown is indicated
1531 Higher flash point oil based liquids are preferred for the carriers in wet MPI testing because of
.
D both a and b
37
ESTestMaker Answers
1532 What is it about fluorescent magnetic particles that requires conditioning agents be added to the water
. used as their carrier?
B their pigments and binders are water repellent
1534 In order that indication to background contrast be maintained so as to ensure consistency of defect
. detection in an MPI test what is required?
C maintain particle concentration in the bath
1535 Small amounts of sodium nitrate or sodium chromate in the water based MPI bath are used to
.
C protect pieces against rusting
1536 Addition of a small quantity of mineral oil to an oil based MPI bath is occasionally used to provide rust
. protection. The risk, however, is that
B viscosity will rise above required limits
1537 What can oil vehicles used for MPI baths be contaminated with that water vehicles cannot?
.
D water
1538 A marked increase in background blue fluorescence of the oil vehicle of an MPI bath as compared to
. the reference sample indicates
A probable oil contamination of bath
1539 When using magnetic particle concentrations at the lower limit of the recommended range in a
. recirculating system, you will
D all of the above
1542 The upper limit for part temperature when using MPI in an oil vehicle is
.
D primarily a health consideration
1543 The quantity measured in blacklight intensity has units of µW/cm². The quantity is
.
C irradiance
1544 The quantity measured in white light viewing has units of lux. The quantity is
.
D illuminance
1545 If the eye has no sensitivity to the power irradiated (eg. UV light) photometric units quantifying that
. light are
A meaningless
38
ESTestMaker Answers
1546 In the retina the cells responsible for collecting light to visualize images are the
.
D cone and rod shaped cells
1547 An eye that has become accustomed to low light levels is said to exhibit __________ response.
.
A scotopic
1548 An eye that has adapted to bright daylight is said to be exhibiting ________ response.
.
B photopic
1549 The reason you can sometimes detect faint objects in the dark by not looking directly at them is
. because
D both a and b
1552 The mercury vapour arc was selected as a blacklight source because its spectrum
.
C has several peaks, one of which is at 365nm
1553 At very low light levels (<0.03 cd/m²) colour differentiation is best for
.
D none of the above
1554 The component in wet fluorescent MPI testing that is most likely to cause dermatitis is
.
B solvent based carrier
1555 Which vision requirement is a critical prerequisite for fluorescent MPI viewing?
.
A near vision
1556 Non-imaging scanners used for indication detection in visible MPI testing systems detect
.
B a decrease in reflected light
1557 Blacklight output for a given bulb bares what relation to line supply?
.
A varies directly
1560 When performing the settling test for the wet method concentration determination you find the
. concentration too high. To rectify the situation you would probably
B add more carrier
39
ESTestMaker Answers
1563 If cleaning the filter does not correct low intensity of your blacklight during an intensity test your next
. step would logically be to
C replace the bulb
1564 In MPI, the thin metal devices with artificial subsurface discontinuities etched into them, which are
. used placed on the part to determine field direction and adequacy are
B shim gages
1565 Automated inspection systems for MPI have their advantage primarily in
.
B consistency and reproducibility
1566 When multidirectional magnetization is used to do an MPI test on a part ________ after the current is
. stopped.
D 1 field remains
1568 The advantage of the He-Cd laser used by the flying spot scanning system for automated fluorescent
. MPI, as compared to television imaging systems is
D all of the above
1569 Prior to taking a tape transfer recording of a wet fluorescent MPI indication you must
.
C ensure the surface is completely dry
1570 Which of the following is a process used for pattern recognition in automated MPI scanning methods?
.
D all of the above
1571 The pressure sensitive tape transfer technique recording MPI indications is used on
.
D all of the above
1572 Rubber replica materials such as alginates, when used to record MPI indications have their advantage
. in that
C indications can be lifted from locations not directly viewed by the inspector
1573 To accelerate drying of oil based vehicles off an MPI test piece so a tape transfer of indications can be
. made you could
A rinse the part with an oil solvent
1574 In the MPI indication recording technique that utilizes fixing coatings the purpose of the spray coating
. is to
D both a and b
40
ESTestMaker Answers
1578 When performing the residual technique of MPI you should use
.
D either b or c
1579 The reason the central conductor for MPI tests are often made of copper is
.
C to decrease heating effects
1580 When a flexible cable is used in the "laid-on-cable" technique for MPI of welds, the cable is arranged
. such that the return loop of the cable is
A parallel to the outgoing loop and current flow is in the same direction
1581 Lifting force or pull-off force required by an A.C. yoke is less than required by a D.C. yoke because
.
B skin effect concentrates the flux at the surface
1582 The purpose of lead contact covers on prods for MPI testing is to
.
C reduce arc burns
1585 Laboratory studies of magnetic hysteresis properties of ferromagnetic materials are done using
.
A a Rowland ring
1587 Which of the following has the highest flux density at magnetic saturation?
.
C iron
1589 Both ferro and ferrimagnetic materials support relatively large _______ for relatively small _______.
.
C B, H
41
ESTestMaker Answers
1590 Cobalt-samarium is considered one of the "super magnets" having a coercivity of several million
. Oersteds. How would you easily demagnetize it?
A heat it above its curie temperature
1592 The double coil system, whereby a tube to be MPI tested is passed through 2 coils arranged in
. tandem was designed to
B minimize the L/D ratio of the material
1593 Using an AC yoke, detectability of weld cracks decreases with increasing number of weld bead passes
. on a weld crown due to
B concurrent leakage fields
1594 A reason to magnetize a part circularly after it has been magnetized in the longitudinal direction, other
. than to find defects is to
A eliminate external fields on the end of the part
1598 When inspecting a part with several diameters by passing current directly through the part you would
.
C use several shots starting with low current for small diameters and increasing current for larger
diameters
1599 The advantage of the Burmah-Castrol strip over other portable surface field indicators is
.
B it places the simulated flaw as close as possible to the test surface
1601 The purpose of the soft iron keeper used in conjunction with a permanent horseshoe magnet is
.
D both a and b
1603 If a part is MPI tested using D.C. and an indication is located how would you estimate depth if it was
. not surface breaking?
A use HWDC or A.C and see if it still shows
42
ESTestMaker Answers
1604 Placing an A.C. yoke onto a large part, then energizing the yoke and removing the yoke about half a
. meter from the part before switching off the yoke is a(n) _____________ technique.
B demagnetization
1606 The result of the inability of a flawed region to sustain the same level of magnetization as the
. surrounding material is
B magnetic flux leakage field
1610 When a D.C. bench unit, that does not have the step down demagnetisation option built in, is used to
. demagnetize a part with a cable wrap, how is current reversal achieved?
B manually switching cable connections
1612 A problem that if not corrected for in photography of fluorescent test indication, will overexpose a film,
. is
A film sensitivity to UV light
1613 Almost anything MPI tested can be demagnetized to an acceptable level using reversing D.C.
. methods; but its disadvantage is
A low production rates
1614 When demagnetizing a ring shaped object by passing it through a coil you would
.
D roll it through the coil
1616 MPI field meters are now provided with two units on the meter face, they would be
.
C Gauss and milli Tesla (B & mT)
1617 The field strength of an A.C. magnetizing coil is about 36% of the surface value at a depth of 1mm into
. a steel piece. About what percentage of the surface field strength would exist 4mm from the surface?
D 2%
43
ESTestMaker Answers
1618 Safety hazard associated with blacklights used in fluorescent viewing of indications arises from
.
D both a and b
1619 Because of its reddening effect on the skin, UV-B ultraviolet light is called _________ ultraviolet.
.
C erythemal
1620 At concentrations below their lower explosive limit, petroleum distillates used in MPI testing will
.
A be safe to use
1621 Why are the high intensity discharge lamps (mercury vapour) used in fluorescent viewing a safety
. hazard?
B the risk of explosion upon impact
1623 What is not housed in the quartz tube of the standard high pressure mercury vapour lamps used for
. lamps used for fluorescent indication viewing?
C a large value resistor
1624 Why do the main electrodes of the mercury vapour blacklight not arc immediately upon switching on
. the current to the lamp?
A there is insufficient mercury vapour to conduct an arc
1625 If a blacklight is used for frequent but only intermittent inspections it should be _______ between uses.
.
C left on continuously or in the standby mode if so equipped
1626 Which of the following affect the life expectancy of a mercury vapour lamp?
.
D all of the above
1627 When long blacklight tubes are used instead of mercury vapour lamps the source of UV radiation is
.
A mercury vapour
1628 Although total blacklight energy output of a bank of blacklight tubes may be about equal to a 100W
. mercury vapour lamp their irradiance is lower because
B they cannot be focused
1630 Electric discharge from MPI equipment is usually only a problem when
.
C an explosive air-gas mixture exists in the area
1631 An artificial discontinuity in a sample used to verify the effectiveness of a test system is a
.
B reference standard
44
ESTestMaker Answers
1633 The split prism test block utilizes which method of magnetization?
.
B a central conductor
1634 The result of using "shared flux" discontinuity standards, such as the "pie gage", on convex surfaces is
.
C excessive magnetization will be used
1635 Why is it actually incorrect to call the Hall detector meter a "Gaussmeter"?
.
D both b and c
1636 On some small portable magnetic particle units there are 3 cable connection terminals. They are for
.
B A.C., halfwave rectified and common
1637 The ballast in the power supply for the blacklight is a(n)
.
A autotransformer
1638 The requirement for an autotransformer to regulate current flow in a mercury vapour UV lamp is
. because
B the arc tube shows negative resistance and would destroy itself
1639 The advantage of the 125W mercury vapour lamp over the 100W variety is
.
D all of the above
1640 The large 400W flood lamp blacklights are used primarily for
.
B locating large and medium sized indications
1641 Normal life expectancy for a 100W spot mercury arc lamp is about
.
C 1000 hours
1642 A 40W tubular UV lamp and a 400W mercury arc bulb have ___________ at 50cm from the bulb.
.
A similar beam intensities (µW/cm²)
1643 The reason a 40W UV tube lamp and a 400W mercury arc lamp have similar intensities at 500mm is
.
C 40 and 400W are merely electrical not light ratings
1644 The purpose of some manufactures providing a perforated metal strip to be placed over the sensor of
. the UV meter is to
B extend the range of the meter
1646 Which of the following normal specification requirements can the "single leg" A.C. yoke technique
. meet?
D all of the above
45
ESTestMaker Answers
1648 The settling test for magnetic particle suspensions is appropriate for
.
A agitated suspensions in large MPI "wet" units
1649 In order to use a permanently magnetized block with a network of cracks as an MPI standard for
. evaluating a particle suspension you would need
D either a and c or b and c
1650 The great advantage of the prism block standard for evaluating the sensitivity and or effectiveness of
. an MPI suspension is
B speed compared to the settling test
1652 Even if a part is MPI tested by indirect means (ie. a central conductor) arcing can occur where the test
. part contacts other metal pieces. The source of this arc is
C eddy currents
1653 The requirement for dry magnetic particles to have a high permeability is especially important for
.
C residual magnetism techniques
1655 The written document outlining the test parameters and acceptance criteria for a specific product is a
.
B specification
1656 Which of the following provides the NDT inspector with a brief outline of the steps to perform a test on
. a specific item?
C procedure
1657 Safety factors built into component design while still vitally important, are reduced to minimum values
. when
B weight is a major consideration
1658 In general, there is never a requirement for parts to be better than they were designed to be. As a
. result, caution must be taken to guard against
C over-inspection
1659 A defect oriented with its major plane at right angles to the principal tensile stress is _________
. compared to the same defect not a right angles to this stress.
A more likely to be harmful
1660 For a given linear defect lying completely below the surface, detectability _____________.
.
A increases with increasing permeability
46
ESTestMaker Answers
1661 Mill scale on steel plate can form non-relevant MPI indications because
.
B the scale has slight permeability
1662 Under controlled conditions, the width of an indication over a discontinuity that is wholly below the
. surface indicates
C defect depth from test surface
1663 Localized cold working of steel can result in non-relevant indications. These result from local
. increased hardness and are removed by
B annealing or normalizing
1664 The problem with inspecting bearing race rings with a head-shot across the diameter of the diameter is
.
D all of the above
1665 A mixture with metallic properties composed of 2 or more elements, of which one of at least is a metal
. is a(n)
A alloy
1669 If a discontinuity is shallow and broad like a scratch, the flux lines may stream-line below the
. discontinuity. This will result in
D a lack of indication
1670 Work hardening tends to decrease magnetic permeability. Work hardened aluminium can be made
. more inspectable by MPI by
D none of the above
1673 A metal that can be easily bent or permanently distorted without rupturing is considered
.
A malleable
1674 Defects found between prods are aligned _________ the line between the prods.
.
B parallel to
47
ESTestMaker Answers
1675 The small regions between rivet holes are best MPI tested using
.
C horseshoe magnets
1676 When using magnetic rubber techniques to record MPI indications the magnetization is applied by
.
D any of the above can be used
1677 In the settling test for MPI, 3 components can occur, lint and dust, metal grindings and magnetic
. particle oxides. Which will be the first to settle out.
B grindings
1678 The magnetization experienced by a ferromagnetic material brought into the influence of an outside
. magnetic field is called
B inductance
1679 When a HWDC yoke is equipped with articulating legs it is usually advantageous to have it also
. equipped with
C variable current control
1680 When checking the ammeter output of a conventional MPI unit the calibrated ammeter reading is
. checked against the unit's ammeter reading at a minimum of
C 3 settings
1681 When small amounts of carbon are added to pure iron the result is
.
C steel
1682 Although hysteresis measurements have sometimes been used to detect flaws, it is more often used
. to
D determine material properties
1684 When two electric charges are brought to rest relative one another, the magnetic force between them
. is
A zero
1685 If two long parallel conductors, 1 metre apart, each exerts a force of 2 x 10^-7 N/m of length of wire,
. the total current flowing in the wires is
A 1 ampere
1686 The magnetic induction that results in a torque of 1 N.m on a coil of the magnetic moment 1 A.m²,
. when placed with its axis perpendicular to the magnetic induction field is defined as
C 1 Wb/m²
1687 The magnitude of a magnetic induction field at a perpendicular distance 'r' from the centre of a long
. straight conductor with a current 'I' amperes is
D none of the above
1688 The use of closed loops to represent magnetic fields is based on an empirical evidence that
.
B magnetic mono-poles do not exist
48
ESTestMaker Answers
1689 A material having all electrons paired can still show slight magnetic effects by electron perturbation
. when introducing an external magnetic field. Such a material is considered
B diamagnetic
1690 The summation of vectors representing the dipole moment per unit volume is represented by
.
C M
1691 The ratio of the magnetic field intensity H, to the dipole moment vector M is referred to as the _____ of
. a material.
C magnetic susceptibility
1693 Initial permeability is defined as the slope of the hysteresis curve at the
.
A origin of the virgin curve
1694 The largest value of B/H ratio along the virgin hysteresis curve is called the
.
A maximum permeability
1695 The magnetic field intensity required to reduce the magnetization "M" to zero is called the
.
B magnetization coercive force
1697 The ratio of magnetomotance to reluctance in a magnetic circuit gives the quantity
.
A magnetic flux
1699 To overcome detectability problems associated with magnetic field directionality in magnetic particle
. testing, magnetization of a test piece is required in at least __ directions.
B 2
1701 Which of the following is usually accepted as being responsible for discovering magnetic particle
. testing as a means for locating surface cracks in magnetic materials?
A W. E. Hoke
1702 In very repetitive mass production tests, all variables of the test are controlled so that
.
C only defects considered important are located
49
ESTestMaker Answers
1703 Which service defect would magnetic particle testing be used for?
.
D none of the above
1705 Austenitic stainless steels contain large quantities of iron but magnetic particle testing is not
. recommended for them because
B they are non-magnetic
1707 Magnetic particle testing is preferred over ultrasonic and radiographic testing for locating
.
B shallow surface cracks
1709 Failure to remove the shrink cavity that forms at the top of an ingot will result in voids called
. __________ in the finished product.
C pipe
1710 When certain elements of an alloy concentrate in the last-to-solidify liquid the discontinuity resulting is
.
C segregation
1711. Liquid steel that splashes on the cool wall of an ingot mold and does not subsequently get reabsorbed
by the main body of the pour, forms
A scabs
1712 During the primary processing procedure of rolling an ingot, if the rolls become over filled the resulting
. discontinuity is a
B lap
1714 A magnetic pole exerting a force of one dyne on a pole of equal magnitude one cm away defines one
.
A electromagnetic unit (emu)
1715 What is the repulsive force between 2 like poles, one 30 emu. the other 40 emu, placed 6 cm apart in
. air?
C 33 dynes
1716 What is the net force between two bar magnets each 30cm long place in line with their north poles 10
. cm apart (making the south poles 70 cm apart), assume a pole strength of 140 emu is concentrated at
each pole.
B 175.5 dynes repulsion
50
ESTestMaker Answers
1717 If the field intensity (H) in an induction field of an area in air is found to be 50,000 oersteds, what is the
. flux density in that field?
A 50,000 gauss
1718 Alignment of "domains" in a ferromagnetic material by the application of an external magnetic field is
. called
D magnetization
1719 The potential (voltage) measured across a Hall detector is inversely proportional to
.
C the thickness of the Hall detector plate
1721 When a part is subjected to an increasing magnetizing force (H) until the induced field no longer
. increases other than that which would occur in air the point on the hysteresis curve is called
A the saturation point
1722 The typical hysteresis curve is "S" shaped the point at the top of the "S" is the saturation point. The
. opposite end of the "S" is termed their
B negative saturation point
1725 The elimination of the reverse half of an AC cycle by electronic means results in
.
A half wave rectification
1726 The opposition to alternating current flow that results from a solenoid with an iron core is called
.
C inductive reactance
1727 3 phase A.C. from commercial supplies has the phases separated by
.
C 120 degrees
1728 In electromagnetism, the product of the number of coil wraps made by a conductor and the current
. passing through that conductor is given as
C ampere turns
1730 The inductance that results when 20 volts are produced by a change in current of 200mA in 0.1
. seconds is
C 10.0 henrys
51
ESTestMaker Answers
1731 A central conductor used in magnetic particle testing is used to _______ in a part.
.
B generate a circular magnetic field
1734 In magnetic particle testing, a short duration magnetizing current passed through a solenoid
. surrounding a part is called a(n)
A coil shot
1735 Any patch or cluster of magnetic particles occurring on a test piece that cannot be accounted for by a
. leakage field is considered
C a false indication
1737 If several magnetizing forces are applied to a piece simultaneously and one or more varies in direction
. or amount the result is
D both a and b
1739 A transformer in which the ratio of the primary to the secondary voltage equals the ratio of the number
. of turns in the primary and secondary coils is a(n) ___________ transformer.
A ideal
1740 When the primary and secondary windings of a transformer are combined into a single tapped
. winding, at the expense of circuit isolation the result is
B an autotransformer
1741 The sum of the currents flowing into a junction of an electric circuit is equal to the sum of the currents
. flowing out of the junction. This statement describes
A Kirchoff's first law
1742 The algebraic sum of the potential differences around a closed loop circuit is zero. This statement
. describes
B Kirchoff's second law
1743 Using modern theory current flow (negative to positive), if a conductor carried a current out of the
. plane of this paper (screen) i.e. current flow towards you; the magnetic lines of force would move
__________.
A clockwise
1744 Field intensity at the centre of a typical 5 turn coil used for MPI would best be approximated by which
. equation?
C H = 2(π)NI/10r
52
ESTestMaker Answers
1745 A transformer, if connected to a D.C. supply on the primary side will result in
.
D both a and b
1747 Half-wave rectified AC is commonly used for magnetic particle inspection but cannot also be used for
. demagnetizing a part because
C the current does not reverse
1748 Three phase MPI units have a significant advantage over single phase units in
.
B reduced current input requirements
1750 In addition to bringing a ferromagnetic material to its curie point, in order to demagnetize the part by
. heating what else must be done?
C cool in the absence of an external field
1752 The reason for using several small decreases in current when demagnetizing a part by
. electromagnetic means is to
C ensure the reverse field overcomes the coercive force
1753 In order that a yoke with pulsating half wave D.C. be used to demagnetize a piece you must
.
D ensure the unit has a current reversing circuit
1754 It is recommended that longitudinal magnetization be performed as the last step in a 2 step operation.
. The purpose of this is
C both a and b
1755 Fe2 O3 is composed of 2 types of atoms having anti-parallel electron spins that are unbalanced. The
. result is
C ferrimagnetism
1756 Unreasonable requirements for demagnetization levels called for in specifications should be
.
B modified in light of what needs to be done
1757 The minimum number of reversals ideally required to demagnetize a part using D.C. demagnetization
. is
A 1
53
ESTestMaker Answers
1759 Demagnetization using oscillating currents in a D.C. energized solenoid system operate at a frequency
. of
D resonance determined by the circuit
1760 Ring shaped parts passed through an A.C. coil demagnetizer should be
.
A "rolled" through the coil
1761 Extremely fine dry powder magnetic particles are well suited to detect defects resulting in weak
. leakage fields; however, they are not suitable for production type inspections because
C excessive background masks indications
1762 To reduce background and detect only large flaws using dry powder MPI you could
.
C both a and b
1763 Agglomerations of fine particles in the wet MPI method have the advantage of
.
D all of the above
1764 The particle shape best suited for dry magnetic particle detection of broad shallow defects is
.
B elongated
1766 The purpose of applying liquid penetrant developer or chalk in alcohol to the surface of a magnetic
. particle test piece is to
C provide a high contrast background for viewing
1767 In addition to the magnetic particles what else is mixed with the paste preparations purchased for the
. wet visible magnetic particle method
D all of the above
1769 The advantage of water suspended magnetic particles as compared to oil suspended is in
.
D both a and c
1770 The fluid containing magnetic particles for the wet method is sometimes referred to as
.
B magnetic ink
1771 Failure to stop flooding a part with the magnetic particle bath at the same time the magnetizing current
. is cut off results in a poor test because
B weakly held particles can be removed
1772 The development of the most effective MPI technique for a specific part is
.
B empirically derived
54
ESTestMaker Answers
1773 Indications not well defined, such as sub-surface indications might be are usually termed
.
A diffuse
1776 The ferromagnetic portion of an electro-magnet, not surrounded by windings, forming a fixed magnetic
. circuit and serving to complete that circuit is also called a(n)
B yoke
1778 The equations which when combined with the appropriate boundary condition can be solved to
. describe all electro-magnetic phenomena are ____________ equations.
A Maxwell's
1779 What is the term given to a test piece similar to the product being tested but having artificial or natural
. discontinuities similar in size or extent to the maximum acceptable in the product?
D all of the above
1780 A magnetic flux indicator having a cross-shaped discontinuity below an adjustable cover is a(n)
.
B Berthold penetrameter
1782 The choice of type of current to use for an MPI test is determined by which factor?
.
B location of defect (on or below surface)
1783 For a given output of magnetizing force, which is best adapted to portability and least cost in
. equipment?
B A.C.
1784 The peak current on a full wave rectified A.C. MPI unit reading 1000 amps would be
.
C 1414 amps
1785 The effect of increasing hardness of steel on the detectability of subsurface defects is to
.
C both a and b
1786 Maximum sensitivity from the continuous wet method is obtained when
.
D all of the above
55
ESTestMaker Answers
1787 If the principal direction of potential defects in a part to be MPI tested is not known you should
.
D do both longitudinal and circular magnetizations
1788 Which is not an advantage of using the central conductor technique to test tubular or hollow products?
.
B both axial and circumferential flaws are detected
1790 When the magnetizing current in an MPI test is passed through an auxiliary conductor to produce a
. useful magnetic field in the test part it is called
A indirect magnetization
1792 The equations It = 45000 / L/D and It = 35000 / L/D+2 are both guidelines for coil shots, the condition
. that differs for each is
B coil to part diameter ratio
1793 Equations are often quoted recommending a specific number of ampere-turns for longitudinal
. magnetization of parts. It is usually a guide to ensure what minimum flux density in the part?
C 100 lines per mm²
1794 Given a solid steel shaft 1.5m long and 50cm diameter, what current setting should you use if you had
. a 5 turn coil and the part is not centred in the coil?
C 3000 amperes
1795 Given a hollow steel pipe section 75cm long, 25cm diameter with a 2cm thick wall, what magnetizing
. current would you use with a 3 turn coil if the part is placed at the bottom of the coil?
D 5000 amperes
1796 Given a steel bar 1 foot long and 2 inches diameter, what current should your coil shot use with a 5
. turn coil and the part centered in a coil 300mm diameter?
B 1600 amps
1797 If MPI is to be used on a part by means of prods and geometry prevents prod spacing of more than
. 45mm you should
D use another form of inspection
1799 A 1/2 metre long pipe 5cm diameter is to be longitudinally magnetized using a 5 loop coil shot. If the
. loop is 400mm radius what current is used when the pipe is centred in the loop?
B 1250
1800 Should an irregularly shaped part have a long to short dimension (L/D) ratio of less than 2, in order to
. longitudinally magnetize the piece with a coil shot you would
A end pieces should be added to extend the length
56
ESTestMaker Answers
1801 When a part is subjected to the magnetic field of a D.C. yoke and the circular magnetization from an
. A.C. head shot, the result is
B a swinging field
1802 What is the reason A.C. yokes are constructed of laminated transformer sheet?
.
A to reduce eddy current loses
1803 A problem with indication formation when using the swinging field technique is
.
C both a and b
1804 Effective permeability μeff can be approximated for coil magnetization of cylindrical objects by which
. relation?
B 6 (L/D) -5
1805 In order that the standard equation for current be used, what is the maximum diameter bar that can be
. tested by coil shot when centred in a coil 40cm diameter? i.e. cross-sectional area of bar not to
exceed 1/10 the coil.
B 63cm
1806 When steels of high permeability are MPI tested the formulae for calculating magnetizing current to
. use
B over estimate requirements
1807 If a long rectangular bar, 2cm X 6cm is circularly magnetized using the 40 A (D.C) per mm about 2400
. amps would be used, where would maximum field strength occur?
B in the middle of the 6cm face
1808 Four bars of the same alloy steel are cut and to be circularly magnetized using a 1000 amp head shot.
. Which bar will have the greatest surface field strength?
A 10mm diameter 1000mm long
1809 You are given 3 conductors each 2cm diameter and 30cm long. Two are non-magnetic, one solid the
. other hollow; and one is ferromagnetic and solid. Which will have the strongest field at a distance of 3
radii from the centre if each is carrying 1000 amps D.C.?
D none of the above, all will be equal
1810 Calculations for coil shot current requirement stipulate length to diameter ratios from 2 to 15. If the
. part has an L/D ratio of less than 2 you should
C use an extension on the piece of the same material as that tested
1811. Calculation for coil shot current requirements stipulate they be used for length to width ratios of 2 to
15. If the L/D is greater than 15 you should
C use the value 15 in the equation
1812 When D.C. is passed through an irregularly shaped part the current is
.
A distributed evenly over the cross-sections
1814 Circular magnetism is preferred over longitudinal magnetization for what reason?
.
B fewer poles result which could cause confusion
57
ESTestMaker Answers
1815 A critical limitation to the settling volume test for determining wet MPI bath concentrations is
.
B the application to aerosol spray preparations
1816 When performing A.C. yoke MPI on a part having a restrictive geometry such that both legs of the
. yoke cannot be used to straddle the area, an optional technique would be to use
B the single leg technique
1817 Why is it impractical to design a magnetic field indicator using a single strong magnet and a spring to
. return the indicator needle to zero?
C both a and b
1818 The NDT method having the worst record for reliability and repeatability of crack detection in
. ferromagnetic parts is
B magnetic particle
1819 Because the permeability of air is _______ the magnetic field indicator can be used to measure both
. the magnetic induction field in air and the magnetic field strength in air.
B 1
1820 Because the magnetic permeability of air is 1 a magnetic field indicator can be used to measure both
. the magnetic induction field in air and the
A magnetic field strength in air
1821 The Ketos ring has been a Standard in the MPI test method since 1941; however, it is not a reliable
. standard because of variations in
A permeability
1823 Specifications (such as Mil-STD 1949A) have been modified to allow 1 hour settling time instead of the
. previous 1/2 hour because
D both b and c
1824 Misinterpretations result from observations to meet the specification for fluorescent baths which states
. "for fluorescent baths, examine the liquid directly above the precipitate with UV light. The liquid shall
be essentially nonfluorescent." The misinterpretation is caused by
B fluorescence of submicron particles that have not settled
1825 Due to load demands on power supplies ___________ is the preferred current for MPI on larger parts.
.
A three phase full wave rectified
1826 When plotting width (or height) of MPI indication versus magnetizing current for surface breaking
. seams in billets, a more linear curve results when using
C larger sized particles
1827 When magnetic rubber replicating techniques are used, the orientation of a crack into the part can be
. determined by
A symmetry or asymmetry of particle accumulation in the setting medium along the indication length
1828 The number of amperes of magnetizing current or ampere-turns to use on a test piece will be specified
. by a
D a specific specification
58
ESTestMaker Answers
1829 Occasionally a thick MPI indication will form as a result of wet particle attraction to deep seams. Once
. the current is turned off the indication collapses and smears the background. This may be prevented
by
C both a and b
1830 Magnetic particle baths used for billet inspection tend to have a greater proportion of larger particles
. than small. The reasons for this is
B the fine shallow defects located by small particles are not important in billet inspections
1832 When performing an MPI test on a coated part, maximum coating thickness that can be tolerated is
.
A determined by demonstrating detectability of minimum allowable defects
1833 A sharp magnetic particle indication has formed on testing a large (3m diameter) submerged arc
. circumferential seam. The indication runs for the complete circumference and an acid etch shows the
indication to occur at the weld-parent metal fusion line, but no visible non fusion is seen. What is the
probable source of the indication?
A magnetic permeability difference between metals
1834 At the skin depth, also considered the standard depth of penetration, the magnetic field intensity
. resulting from a 60 Hz A.C. in steel is
C 37% of its surface value
1835 Maximum curvature of external magnetic flux leakage fields occurs for
.
A narrow surface breaking defects
1837 Which of the following flux sensitive devices is not time dependant?
.
B a Hall detector
1839 The reason for using a high frequency A.C. voltage on the semiconductor crystal in a Hall sensor is to
.
C eliminate the thermoelectric voltage component
1840 Which of the following compounds is a commonly used semi-conductor material used as a Hall
. detector sensor crystal?
A InSb
1841 In the semiconductor magnetodiode flux sensing device the p-zone and n-zone are separated by the
.
C recombination zone
1842 Which form of artificial lighting could not be used to view daylight fluorescent dry magnetic particles?
.
B sodium vapour lights
59
ESTestMaker Answers
1843 The dry powder method of MPI would not be used for
.
C grinding cracks
1845 At temperatures of around 150°C. fluorescent powders lose their visible contrast. This is because
.
D both a and b
1846 A typical minimum light intensity of 1000 lux is recommended for viewing non-fluorescent dry powder.
. This may be barely adequate if
A the test surface is dark or covered with scale or rust
1847 AMS 3040 is a specification for properties of magnetic particles properties. One test requires a glass
. cylinder of specified dimensions be stuffed with the powder to be tested. The powder in the cylinder is
used as the core of a transformer coil. For 11 kA/m input on the primary, a minimum of 2.5V must
occur in the secondary with the correct circuitry. Since the volume of the cylinder limits the amount of
powder the test limits the amount of
A inert pigment and resins in the powder
1848 The difference between red iron oxide (alpha) and brown iron oxide (gamma) apart from colour is
.
D all of the above
1849 For fine wet particles, those with high permeability are no more effective than those of only moderate
. permeability due to
C the shape demagnetization factor
1850 Most specifications for establishing properties of oil carriers for wet MPI will state
.
C maximum viscosity at a specified temperature
1851 In wet method of MPI testing when water is used as the carrier, conditioners are used to
.
D all of the above
1852 When too much antifoaming agent is added to the water vehicle for wet fluorescent MPI testing what is
. the result?
B loss of the bath's wetting ability
1853 Sodium nitrate and sodium chromate are the traditional rust inhibitors added to water based MPI
. baths. They are no longer recommended because
C of their high level of toxicity
1854 Use of anti-freeze in water based MPI baths is not feasible as a method to prevent freezing because
.
C the viscosity would exceed recommended limits
1856 To overcome the sticky mass that occurs as a result of water contamination in an oil bath for MPI you
. can easily
D heat the bath with an immersion heater to boil off the water
60
ESTestMaker Answers
1857 Maintaining a strict control on particle concentration in the MPI bath eliminates which variable in an
. MPI test?
A indication to background contrast
1859 Some specifications require the sample of MPI bath be de-magnetized before performing the settling
. test. What is the reason for this demagnetization?
A magnetic condition of bath can affect settling rate
1860 In general, centrifuge tubes having narrower settling stems (1.5mL, 1.0mL, and 0.2mL stems
. available), tend to __________ for the same bath.
B show higher concentrations
1861 Online continuous monitoring methods are not often used to establish MPI bath concentrations
. because
C both a and b
1862 The lower limit for part temperature when using MPI in an oil vehicle is
.
D variable from oil to oil and depends on viscosity at a given temperature
1866 __________ is concerned with the measurement of E-M radiation in the visible region only.
.
B photometry
1867 The variable response exhibited by an average eye as it crosses over from scotopic to photopic vision
. is termed
A mesopic vision
1868 At the area around the point of the retina where visual axis strikes there is a slight depression called
. the fovea. The cells in this area are almost exclusively ____________-.
C cone shaped
1869 Under good viewing conditions for fluorescent MPI, indication contrast ratio can be a high as
.
C 200:1
1870 Reasonable probability of MPI indication detection would require a brightness contrast ratio of at least
.
A 2:1
61
ESTestMaker Answers
1871 A pure red MPI indication with wavelengths 600nm on a background of pure blue at 470nm having
. equal brightness
D all of the above
1872 As ambient lighting changes from bright daylight conditions to dim blacklight viewing conditions
. maximum colour sensitivity of the eye changes from
B yellow to green
1873 The preferred gas for use in vapour arc lamps used for blacklight production is
.
B mercury
1874 For extended viewing of visible magnetic particle indications of critical tests requiring 2000-5000 lux
. lighting the effect may be
A eyestrain
1875 Laser light sources are often used in non-imaging detector systems for MPI because
.
B they produce narrow and intense beams
1876 Photoelectric sensors used in scanners for non-imaging MPI detector systems are
.
C rarely the same brightness or colour sensitivity as the human eye
1878 In MPI of aerospace components, copper penetration resulting from prod or electrode contact will
. cause
A rejection of the part
1880 If the results of an MPI test make you suspect the indication seen is nonrelevant, originating from
. hardness or alloy differences, you would
B verify using penetrant or ultrasonic methods
1881 A letter or number hardstamped into a part and ground off prior to MPI testing can result in
.
D both a and b
1882 To determine if magnetizing fields are present and balanced in a multidirectional operation you would
. use a(n)
B MPI shim gage
1883 Mercury arc UV bulbs should not be handled with bare hands because
.
B acidity of fingers can cause glass failure
1884 Calibration of an MPI machine used to apply magnetizing current to a part requires
.
A comparing current readings on the machine to a calibrated ammeter
62
ESTestMaker Answers
1885 Fully automated magnetic particle inspection systems can only be cost justified if
.
D all of the above
1886 The magnetic particle testing portion of an automated system can be a time consuming step. To
. decrease the inspection time in such a setup you would
A use multidirectional magnetization
1888 The flying spot scanning system requires _________ to permit the system to discriminate rejectable
. discontinuities from background noise.
B pattern recognition algorithms in the CPU
1889 Automated MPI systems that include the ability to scan and sort indications require some form of
. ___________ processing or algorithms.
A pattern recognition
1891 For tape transfer recording of fluorescent particle indications, best results are had if
.
C both a and b
1892 If taking a tape transfer of an MPI indication where a water carrier has been used, the drying of the
. surface can be accelerated by
B gentle rinsing with acetone
1893 Rubber replica records of MPI indications are bulky. Archival is often accomplished by
.
A photographing the indication in the replica
1895 When tape transfers are used as a means of MPI test records the adhesive often dries out and tape
. becomes brittle thereby limiting the time it can be archived. Under optimum conditions the tape
transfer record might last
C 5 - 10 years
1897 The reason for long duration magnetization shots when using the magnetic rubber method is
.
B high viscosity of uncured rubber
1898 Photography as a means of high quality recording of MPI indications is very costly because
.
D all of the above
63
ESTestMaker Answers
1899 The "laid-on-cable" technique used in MPI of welds requires the cable be spaced away from the test
. surface and the return loop be about ten times this space away from the first loop. The applied current
is determined by
A the spacing of cable off the test surface
1901 On irregular shaped specimens, the magnetic field is more likely to follow the surface contours if
.
D A.C. is used
1902 Lead contact covers used on prods for MPI testing should only be used in well ventilated areas
. because
B over heating causes noxious fumes
1903 The process whereby a magnetic tape (similar to that used in a tape recorder) is placed on a test
. piece during magnetization then removed and processed for analysis and results viewed on stripchart
or oscilloscope screen is called
A magnetography
1906 The region on the hysteresis loop's virgin curve where the flux density B increases in a non-linear
. fashion with increasing field intensity H, is termed the
A Rayleigh region
1907 In magnetic studies using a Rowland ring, by removing the effect of H from B you can plot
.
B a magnetization curve
1908 Ferrites are used as cores for "pick-up coils" in magnetic flux leakage amplifiers. Their advantage for
. this purpose originates because
A amplification occurs without inducing eddy currents
1909 In low carbon steels, chemical composition, inclusion density, residual stress and heat treatment can
. determine which magnetic property?
D all of the above
1910 The measure of how far from unity (1) permeabilities of dia and paramagnetic materials is / are termed
.
A magnetic susceptibility
1911. From a magnetizing point of view, the worst case scenario an MPI inspector would encounter when
performing a circular magnetization on a part is
A the material is initially magnetized to maximum remnance in the opposite direction to his inspection
1912 The miniature Hall detectors which have Hall sensor and regulated power supply on a chip have the
. advantage over sensing coils that
D both a and b
64
ESTestMaker Answers
1913 The reason some specifications do not permit MPI by passing current directly through the part is
.
B risk of arc burns
1914 In pulsed or capacitor discharge magnetization techniques the effectiveness of a lower amplitude
. current will be greater for deep flaws than a high amplitude current if the pulse length of the low
amplitude current is longer. This is due to
B the limiting depth of penetration by eddy currents
1917 When indications are not formed on a test piece being tested by MPI what is the reason?
.
D any of the above
1918 In order for a component to achieve a minimum specified service life after an NDT inspection has
. accepted it, it is important for the NDT engineer to allow for
B sub-critical crack growth
1921 A threshold exists in the detectability of flaws using MPI for which parameter?
.
D all of the above
1922 The magnitude of the magnetic leakage field is determined primarily by the
.
B flux density in the region
1923 For a code or specification to stipulate a minimum flux density be achieved in a test piece for MPI
. testing is
A not practical since it cannot be easily measured
1924 To compensate for variations in relative magnetic permeabilities, codes and specifications
.
B use conservative estimates of permeability
1925 The shortcoming of stipulating particle concentration as an indicator of magnetic particle suspensions
. is
B no information is had about the ability of the suspension to form indications
1926 Which artificial test piece is the most reliable for evaluating performance of an MPI particle
. suspension?
A pre-magnetized pieces such as the German MTU
65
ESTestMaker Answers
1927 Completely automatic cameras used in auto mode are not recommended for photography of MPI
. indications because
C they cannot normally compensate for the extreme exposure conditions
1930 The best all around UV filter for fluorescent indication photography has its cut off at ____________.
. This allows an image most resembling that seen by the human eye in a darkened booth.
B 390nm
1931 Polarizing filters can be used for photography of fluorescent indications if UV filters are not available
. since they effectively block UV light, however they have the drawback that
D both a and b
1932 Reciprocity failure is a problem that exists when performing photography of fluorescent indications. It
. is a result of
B long exposure times
1933 Alternate north and south poles, which can result from magnetic chucks or lifting magnets, are
. recognized by
A alternate north and south poles in close proximity
1934 Reversing D.C. demagnetization is often incorporated into the design of large bench units for what
. reason?
B economic
1936 In reversing D.C. magnetization of a part using a coil, the test piece is
.
A held stationary in the coil
1937 Pulsating reversing D.C. coil demagnetization has the advantage over A.C. coil demagnetization in
. that
A penetration is far greater
1939 Which of the following demagnetizing methods would be used in high production rate demagnetization
. of small, medium hardness steels parts?
D none of the above are recommended
1940 Coil demagnetization of parts with L/D ratios less than 3 is a problem. This is solved by
.
B adding ferromagnetic pole pieces to the object's ends
66
ESTestMaker Answers
1941 Specifications requiring residual fields less than 0.3 mT are often unreasonable because
.
C in most cases, subsequent machining or welding, or instrumentation is not affected by this level of field
1942 The presence of a small axial field strength at the end of a tube indicates
.
C the vector sum of fields has a slight axial component
1943 The lowest temperature at which a liquid gives off gases sufficient to form an ignitable vapour at the
. liquid's surface is the
A flash point
1945 Above the upper explosive limit volatile petroleum distillates used in MPI testing will
.
D will present no explosive hazard as concentrations are too rich to burn
1948 What allows current to flow in the high pressure quartz tube of a mercury vapour lamp upon initial start
. -up?
B argon gas
1949 The pressure inside the quartz tube of a mercury vapour regulates, to a large extent, quality of the
. emission spectrum. What is the pressure, in atmospheres, of the gas in the quartz tube of a blacklight
used for fluorescent indication viewing?
B 10
1950 In the standard mercury vapour lamp, the quartz cartridge is housed in an outer bulb filled with
.
C an inert gas at lower than atmospheric pressure
1951 Although UV-A produced by blacklights used in fluorescent indication viewing for NDT is not believed
. to be a health threat, you should
C both a and b
1952 When the electric field strength exceeds the breakdown value for air the result is
.
A a spark
1953 After about 4 minutes warmup time the blacklight has reached its peak irradiance. After that, when left
. running with a constant voltage
A output slowly decreases to an equilibrium value
1954 The "blacklight blue" glass filter envelope around the ultraviolet light tubes is coated on its inside
. surface with
C bluish white phosphor
67
ESTestMaker Answers
1955 When MPI is performed in grain elevators or flour mills there is a danger of
.
B an explosive air-dust mixture being ignited
1956 In order to establish a correlation between the response of an NDT test system and the severity of a
. discontinuity _______ are used.
D both a and b
1958 To avoid excessive flux densities resulting from traditional empirical formulae used in MPI testing
.
B reference standards are used
1959 Experiments on ketos rings indicated a lack of uniformity (1986 experiments). For the same
. conditions, different rings would show different numbers of hole indications. This was accounted for by
_____ differences between rings.
B annealing
1960 When an object is tested using a D.C. coil, the so called "quick break" magnetization utilizes ________
. to detect transverse defects at the ends of the test object.
B eddy currents
1961 Photosensitive eyeglasses darken in the presence of UV radiation. This makes them
.
C useless for blacklight viewing of indications
1962 The current to the blacklight is regulated by an autotransformer. The autotransformer has the unique
. feature of having
A both primary and secondary coils combined into a single tapped winding
1963 A single rod conductor 10cm long moving at 10m/s through a 10,000 gauss magnetic field at right
. angles to the field will induce what voltage between the ends of the rod?
B 1V
1964 A wire 10cm long cuts a magnetic field of 10,000 gauss at 45° moving 10m/s. What voltage is induced
. between the wire ends?
B 0.7 V
1965 The intensity of blacklight irradiance (µW/cm²) from a 400W flood lamp mercury arc lamp as compared
. to a 100W spot mercury arc lamp (both with plain filter) as measured at the beam centre at the same
axial distance is
D 1/10 the intensity
1966 Apart from its health hazard, a potential risk of breaking the tube containing the mercury in the
. mercury vapour lamps is had from
C mercury contacting aluminum or other metals can cause cracking
1967 Gross errors or differences may exist between UV output measured from a calibration lamp and the
. lamp in question even though the lamps are nominally the same e.g. 100W spot. Why?
B the lamp spectra are different
68
ESTestMaker Answers
1969 Even if all UV sensors were calibrated and found to be identical, measurements of a given UV lamp at
. the same distance would have different values if different manufacturer's meters were used. This is
because
B the area of the UV sensor varies for each manufacturer
1970 The most important variable when testing a piece coated with a thin non-conductive layer (such as
. paint) with an A.C. yoke is
B getting sufficient flux leakage out of a defect site
1971 For underwater MPI it has been found that the A.C. yoke can be used in the "single leg" technique.
. The magnetic field produced for this is
C radial
1972 Small cracks in steel plate that has been sandblasted are best tested for using
.
B MPI
1973 Which property of magnetic particle suspensions is not an important factor considered by the
. inspector?
C cost per unit tested
1974 When using the prism block standard for evaluating a magnetic particle suspension, increased quality
. is indicated by
A length of indication
1975 The addition of extension pole pieces to A.C. yokes has the effect of
.
B decreasing field strength midway between the poles
1976 Although it has been proven that some defects can be located through some types and thicknesses of
. paint, a reason to remove paint for MPI testing is
A paint type and thickness is not known
1977 When inspecting pipes using A.C. and rectified A.C. circumferential magnetization, pipes should be
. insulated from each other and insulated from metal racks they might be supported by. This is because
B arcing can result from the induced eddy currents
1979 A nonlegal document that is written by a recognized association of some authority, that recommends
. actions to achieve certain goals is considered a
B standard
1980 A standard when made enforceable by making it a legal document (law) is then considered a
.
B code
69
ESTestMaker Answers
1983 In MPI testing, a design constriction in the metal path through which the flux must flow results in
.
B non-relevant indications
1984 If a defect is suspected in a region of MPI non-relevant indications, such as from splines, you would
.
A demagnetize the piece and remagnetize at a current level just below that required to form the non-
relevant indications
1985 Metal's crystalline structures that are reversible through transformations are termed
.
A allotropic
1986 Austentic or gamma iron is normally stable only at high temperatures unless _____________.
.
B alloyed with another metal such as chromium or nickel
1987 The non-relevant MPI indication found on curved cold rolled plate and having an incomplete criss-
. cross pattern 45° to the axis of curvature are caused by
C Luder lines
1988 When a metal is heated to a predetermined temperature (below melting point) and holding it there for
. a period of time then cooling it at a slow rate, it is said to be
A annealed
1989 A binary alloy having a lower melting point than its neighbouring compositions is considered
.
C a eutectic alloy
1992 After an MPI test post cleaning of a part that is irregular shaped and made of soft steel can be
. especially difficult because
A of demagnetization being incomplete
1994 When two permanent magnets are used for an MPI test you must ensure
.
B unlike poles contact the test surface
1995 When a specimen cannot be placed in a fixed coil on a bench unit, a flexible cable can be wound
. around the piece and the two ends connected to the respective poles of the bench unit. The cable is
sometimes called
B wander cable
70
ESTestMaker Answers
1998 When performing an MPI settling test on a bath, lint or dust, and metal grindings can make up part of
. the settled volume. If this occurs
C subtract the dust and grindings volume from the total settled volume
1999 The MPI technique utilizing magnetic particles in a highly viscous vehicle is termed the ________
. technique.
A wet slurry
2001 When using a calibrated ammeter to verify the MPI unit ammeter accuracy the calibrated ammeter is
.
A connected in series with the output circuit
2002 When measuring halfwave DC with the calibrated ammeter used to verify the accuracy of the MPI
. units' ammeter, the current values read on the calibrated ammeter are multiplied by
D 2
71