You are on page 1of 9

Philips Healthcare

NEWSLETTER SERVICE

Issued by : CT Service Innovations Reference No. : CT3855


Code number : D-File : 621372
CSIP Level 1 Date : December 2014

CT SERVICE NEWS
Introducing the New Series of Standard, Extended, and Bariatric Patient Tables
A new series of patient tables is in forward production for Brilliance, Ingenuity, and iCT systems. The new patient tables
include several new hardware components.
The purpose of this newsletter is to introduce the new series of patient tables and the main differences compared to the
previous series of patient tables.
• Applies To
• Main Differences of the New Series of Patient Tables
• How to Identify the New/Previous Series of Patient Tables
• Release to Production
• Training and Related Documentation

Logistics: Refer to Release to Production


This document and the information contained in it is proprietary and confidential information of Philips Healthcare ("Philips") and may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated
without the prior written permission of the Philips Legal Department. Use of this document and the information contained in it is strictly reserved for current Philips personnel and Philips customers who have a current and valid license
from Philips for use by the customer’s designated in-house service employee on equipment located at the customer’s designated site. Use of this document by unauthorized persons is strictly prohibited. Report violation of these
requirements to the Philips Legal Department. This document must be returned to Philips when the user is no longer licensed and in any event upon Philips’ first written request.

This document and the information contained in it is strictly reserved for current Philips Healthcare ("Philips") personnel, Philips licensed representatives and Philips customers who have purchased a valid service agreement for use by
the customer’s designated in-house service employee on equipment located at the customer’s site. Use of this document by unauthorized persons is strictly prohibited. This document must be returned to Philips when the user is no
longer licensed and in any event upon Philips’ first written request.
CSIP Level 1 © 2014 Koninklijke Philips N.V All Rights Reserved 1
CT Service News CT3855

Applies To

Table 1: Affected Systems

System Name System Identifier

Brilliance CT 16-slice Configuration (Air) 728246

Brilliance CT Big-Bore Configuration 728243

Brilliance CT Big-Bore Configuration Radiology 728244

Brilliance iCT 728306

Brilliance iCT SP 728311

Ingenuity Flex 728317

Ingenuity CT 728326

Ingenuity Core 128 728323

Ingenuity Core 728321

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 2
CT Service News CT3855

1.1 Main Differences of the New Series of Patient


Tables
Previous
Part New Patient
Patient Table Comments / Details
Description Table P/N
P/N

Motion force changed from 40 lbs to 80 Ibs to resolve the


intermittent horizontal stops
Horizontal 459800623441 453567139465 System software V3.6 and above is prerequisite for the AMC motor
Controller (AMC) (ACS) controller

Note: This controller cannot be programmed in the field.


Three vertical PCBs (vertical monitor board, vertical relay board,
and the Inverter Regen assembly) are consolidated into one PCB,
Vertical
459800623431 453567484781 which will improve the reliability of the couch
Controller
Note: This controller cannot be programmed in the field.
The 20:1 ratio significantly reduces the risk of uncontrolled vertical
movement.
Vertical 459800623661 459800321371
Motor/Brake Note: The new vertical motor assembly is NOT
Assembly (Ratio 20:1) (Ratio 10:1)
interchangeable with the previous Brilliance motor
assembly

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 3
CT Service News CT3855

1.1.1 Horizontal Control


• AMC motor controller replaces the ACS motor controller.
• The motion force is changed from 40 lbf to 80 lbf to resolve sudden horizontal stop motion.

Note: The AMC motor controller can only support systems with software version 3.6 and above.

Figure 1: New AMC Motor Controller Figure 2: ACS Motor Controller

For more diagnostic and troubleshooting information about the AMC controller, refer to the "Ingenuity Series /
Ingenuity Core Patient Support (New Type of Patient Table) Theory and Troubleshooting Guide".

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 4
CT Service News CT3855

1.1.2 Vertical Control


• New type of vertical control assembly
• Three vertical PCBs (vertical monitor board, vertical relay board, and the Inverter Regen assembly)
are consolidated into one to improve quality

Figure 3: New Vertical Control Assembly Figure 4: Previous Vertical Control Assembly

Inside of the new vertical control assembly Inside of the previous vertical control assembly

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 5
CT Service News CT3855

1.1.3 Vertical Motor/Brake Assembly (20:1)


ATTENTION: The new vertical motor assembly is NOT interchangeable with the previous Brilliance
motor assembly.

How to Identify the Two Types of Vertical Motor/Brake Assembly


There is a thermal switch on the new type of Motor/Brake assembly. It is connected to the invertor
assembly, and its function is to protect the Motor/Brake assembly from overheating.
In addition, note the ratio listed on the motor label.

Figure 5: 20:1 Vertical Motor/Brake Assembly

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 6
CT Service News CT3855

Figure 6: 10:1 Vertical Motor/Brake Assembly

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 7
CT Service News CT3855

1.2 How to Identify the New/Previous Series of


Patient Tables
There are several ways to identify the new patient table:
• From the presence of the new AMC Horizontal controller (see Horizontal Control on page 4)
or
• The patient table’s serial number (see table on on page 9)
or
• From the system configuration
Check the system configuration as follows:
1. Log on as Service.
2. From the desktop, open Service Tools.
3. Go to System Configuration, and click Verify.
4. Under the Configuration Parameter column, find uCT.Couch.Controller.Firmware.
5. Under the Current Value column:
• Value=2650 —AMC Controller (New Patient Table)
• Value=3211 — ACS Controller (Previous Patient Table)
Figure 7:

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 8
CT Service News CT3855

1.3 Release to Production


The new series of patient tables is cut into production as follows:

First Patient Table Serial Standard Patient Extended Patient


Bariatric Patient Table
Number Table Table

900502 (Non RoHS)


Brilliance and Ingenuity 941001 921001
901001 (RoHS)

iCT 911001 951001 931001

1.4 Training and Related Documentation


All related training courses (for Brilliance, Ingenuity and iCT systems) have been updated. No extra or
specific training is required for the new series of patient tables.
The following new service documents are introduced:
• 4598 000 73161 Patient Support Repair and Replacement Manual Rev E
• 4598 006 19221 Patient Support Troubleshooting Guide Rev A

Rev A CSIP Level 1 WARNING: THIS PAGE CONTAINS COPYRIGHTED MATERIALS THAT ARE CONFIDENTIAL AND OR PROPRIETARY. ANY RELEASE OR DISTRIBUTION OF THIS MATERIAL, WITHOUT PERMISSION, IS A VIOLATION OF LAW. 9