MEDICON ALT/GPT Reagent AU480/AU680 Reagent ID: 707

Pr. Code: 1418-0087 Serum / Plasma

Specific Test Parameters


General LIH ISE HbA1c Calculated Tests Range

Test Name: ALT ▼ ◄ ► Type: Serum ▼ Operation Yes ▼

Sample Volume 4 uL Dilution 10 uL OD Limit


Pre-Dilution Rate 1 ▼ Min OD 0.5 Max OD 2.5
Rgt. Volume R1 (R1-1) 80 uL Dilution 20 uL Reagent OD limit
First Low 0.75 High 2.5
Last Low 0.75 High 2.5
R2 (R2-1) 20 uL Dilution 0 uL
Dynamic Range Low 8 High 800
Common Rgt. Type NoneΦ Name Correlation Factor A 1 B 0
Wavelength Pri 340 ▼nm Sec 660 ▼nm Factor for Maker A 1 B 0
Method RATE ▼
Reaction Slope - ▼ Onboard Stability Period 30 Day Hour
Measuring Point 1 First 14 Last 27 LIH Influence Check # ▼
Measuring Point 2 First Last Lipemia ▼
Linearity Limit 15 % Icterus ▼
Lag Time Check ▼ Hemolysis ▼

Parameters Specific Test Parameters


General LIH ISE HbA1c Calculated Tests Range

Test Name: ALT ▼ ◄ ► Type: Serum ▼

Value/flag # ▼ Low High


Level Panic Value
Specific Ranges: From To Low High
Sex Year Month Year Month L H # #
❑ 1. # ▼ # # # # # #
❑ 2. # ▼ # # # # # #
❑ 3. # ▼ # # # # # #
❑ 4. # ▼ # # # # # #
❑ 5. # ▼ # # # # # #
❑ 6. # ▼ # # # # # #
7. No Demographics # #
8. Not Within Expected Values
Unit U/L Decimal Places #

Parameters Calibration Parameters


Calibrators Calibration Specific STAT Table Calibration
General ISE

Test Name: ALT ▼ ◄ ► Type: Serum ▼ ❑Use Serum Cal.

Calibration Type: AB ▼ FORMULA: Y= AX+B ▼ Counts: # ▼


<Calibration Parameters> Factor Range
Calibrator OD Conc. Low High Slope Check ▼
Point 1 # ▼ + # #
Point 2 ▼ Allowance Range Check
Point 3 ▼
Point 4 ▼ ❑ Reagent Blank
Point 5 ▼ ❑ Calibration
Point 6 ▼
Point 7 ▼ Advanced Calibration
Point 8 ▼ Operation ▼
Point 9 ▼
Point 10 ▼ Interval (RB/ACAL) ▼

<Point Cal. For No. of Correction Points ▼ Use Master Curve ▼ ❑ Lot Calibration
Master Curve> OD Range
Calibrator OD Conc. Low High Stability
Point-1 ▼ Reagent Blank 999 Day Hour
Point-2 ▼ Calibration 999 Day Hour

MB Type Factor 1-Point Calibration Point ▼ ❑ with Conc-0

+ MEDICON MEDI-CAL (p.c.: 1578-0891) or OLYMPUS System Calibrator Cat.No: 66300 (p.c.: 4478-0891).
# User defined.
Values set at U/L. To work in SI units (μKat/L) divide by 60.