Service Manual
Contents
1 INSTRUMENT INTRODUCTION .......................................................................................................... 3
6 TROUBLE SHOOTING....................................................................................................................... 39
1 Instrument Introduction
Principles and Intended use
The instrument Is using the principle of colorimetry, suitable for clinical routine biochemical tests. This device
has no external energy output, no risk of over-temperature, excess radiation output. The matters needing
attention mainly include mechanical motion risk and electric shock risk: when repairing or dissemble the cover
for other reasons, you must cut off the power and shutdown the mainframe.
HTI BioChem FC-200 is easy to operate, has friendly interface to user, It is a good choice for clinical
chemistry laboratories.
2. 38 reagent positions, position of single and double reagent can be set randomly.
3. Display and print results in different styles, including the storage of test results and info of patients.
4. End point, two point and kinetic method, test with double-wavelength is supported.
6. 8 standard filters: 340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm, Other wavelengths
7. The system will continuously detect and control during the operation, to avoid errors, Error message will
be displayed when detected errors, provide instruction to the user to choose correct buttons and
operations
The instrument mainframe is mainly composed of sample/reagent processing system, mixing system,
reaction system, and photoelectric detection system. The structure of each function module is as following:
The sample and reagent disc are located in a same big plate body, and there are 40 sample positions and 40
reagent positions.
The sample bracket supports the standard test tube, the original blood sampling tube, and the micro sample
Original blood sampling tube: Ф12×68.5, Ф12×99, Ф12.7×75, Ф12.7× 100, Ф13×75, Ф13×100;
The reagent disc can hold two sizes of reagent bottles (one big, another small), the volumes are 40ml and 18ml,
The system will automatically detect the positions of the working reagents. Before each operation, you can
check the position of each reagent should be placed, as well as the total volume of reagent required for this test.
The core functional components of the reagent/sample disc cooling system is composed of fan, radiator, Peltier
and other component parts. The cool side of peltier is close to the bottom of the refrigeration cabin but the hot
The driver component of the sampling system has two optical couplers, one located at the vertical direction,
another at the horizontal direction, to determine the original mechanical position of the vertical and horizontal
directions precisely. The system also has two stepping motors on the vertical and horizontal directions to control
the accurate movement of the sampling needle at the vertical and horizontal directions, which is driven by belt
and pulley.
The sampling needle with liquid sensor function will contact with the liquid, to aspirate the required volume.
When the reagent or sample is not enough, the instrument will give prompt message automatically. During the
operation of HTI BioChem FC-200, the aspiration cycle of the sampling needle is as follows:
1. The sampling needle enters the bottle of the reagent 1, aspirate reagent with required volume, and eliminate
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 5 of 43
Document Number SM-E-FC240
2. The sampling needle moves to the reaction cup and dispense reagent 1;
3. The sampling needle moves to the washing station, and perform washing;
4. The sampling needle moves to the sample cup to aspirate sample with preset volume, and eliminate the
return error;
6. Transfer the sample and reagent into the Reaction disc wells;
7. The mechanic arm moves back to the washing station, to wash the sampling needle and mixer;
8. A bit of air and 100μL distilled water is aspirated into the flowcell, rinse the flowcell to clean the remaining
9. The sampling needle aspirate 80μL reaction liquid and a bit of air into the flowcell, rinse the flowcell to prevent
10. Aspirate the remaining reaction liquid into the flowcell for measurement;
11. The mechanic arm moves back to its original position (the washing station); clean the sampling needle and
Generally, when testing with end-point method items and double-reagents items, you can set the stirring time
as 1 ~ 2 seconds.
When pre-heating the reagent, if the required volume of the reagent is more than 250 micro liters, the frozen
reagent should be treated with the pre-heating module, and the temperature is 30 ~ 35 ℃
after pre-heating.
Before adding into the reaction cup, the reagent mainly obtains heat from the tube of the middle heating section,
perform the pre-heating by the mixing of high temperature reagent and low temperature reagent.
In order to ensure the accuracy and repeatability of test results, the suggested sample volume should be more
than 4μL.
The sampling washing system and the stirring washing system use the same materials, called "Washing station";
use water-spring-type washing method to wash the surface of needle and the stirring rod.
When the operator has performed an illegal operation or errors of procedure occurred, make the instrument
perform wrong actions, the waste liquid in the waste bottle may overflow, it will be drained from the mandatory
this structure is simple and practical, and the risk of movement has also been reduced in a certain degree. Oar-
shaped mixing head can achieve rapid mixing effect through the micro DC motor.
Using air bath, three fans inside, to make internal hot air rotate and circulate up and down;
Using high quality insulation materials externally will effectively resist the interference to the internal
using over-temperature protection, the system will cut off the power if the heating air exceeds a certain
The Reaction disc pot body is the same with the reagent/specimen disc body with external insulation, but the
external insulation is mainly to ensure that the Reaction disc internal temperature control easy to be
implemented, because if the effect of insulation is not good, the reaction disc outside will be featured with fast
External insulation
optical assembly
1.2.7 Reaction Cup
There are 81 reaction positions. Use high-quality plastic semi-permanent reaction cup, with jointly linked
Standard configuration: 340, 405, 450, 510, 546, 578, 630, 670 nm filters, total of 8 wavelengths. Interference
filter, front spectrum, has a separate dark current detection channel. Using 2,000 hours long-life tungsten
Photodiode component
Optical block
The light source uses a direct thermal fan, without any separate air path, use the negative pressure to
The whole machine heat emission adopts downdraught system; the whole machine can aspirate air from
HTI BioChem FC-200 uses color LCD screen to display WINDOWS software, with friendly user interface.
Operator can operate the software step by step according to the related instructions.
Technical parameters
and double-wavelengths
Reagent positions 40
Specimen positions 40
Sensitivity 0.0001Abs
Instrument unsealing
Unpack the package of the instrument and remove the materials for transport. Save the packing box and the
packaging materials well, for your convenience in the future of re-packaging the instrument.
2) Take away the packaging materials, and take out the instrument from the plastic packaging bag.
3) Check the items in the box and confirm that the following items are within the box:
Operator’s Manual
Packing List
Warranty certificate
Accessories: power cord, printer, print cable, LCD monitor, mouse, keyboard, RS-232 serial cable, spare
Note: If the items are inconsistent with the packing list or have any defects, please contact with the
supplier.
Installation Environment
HTI BioChem FC-200 auto chemistry analyzer must be installed by authorized professionals. The instrument
must be installed in a clean dust-free operation room, to avoid shock, humidity, strong magnetic fields and direct
light. The working environment is of 15 ℃~ 30℃, the relative hum idity is 35~
-condensing).
The power supply must be 220V/50 Hz, and should be well grounded. If the laboratory voltage change is greater
than +10%, it is recommended to install an external regulator above 1500-watt. As this machine is a full
computer-controlled precision instrument, it is proposed to install an UPS above 2000W (uninterruptible back-
up power supply).
The back-face of the instrument should be 0.5 m to the wall at least, in order to ensure the heat dissipation
requirements.
If it is a desktop machine, the water bucket should be placed at the right lower adjacent to the analyzer, pay
attention to ensure that the waste liquid pipe should always be downward, to avoid the situation such as upward
In order to ensure the normal operation of the instrument, it is prohibit to placing the instrument in the following
places:
Note:
●The AC power supply must be well grounded (zero ground voltage <5V).
●The AC power must be stable and it is prohibited to share power supply with the appliances of high
●When unplugging the power cord, please seize the plug itself, rather than the power cord.
●If the instrument is found to have smoke, smell or abnormal sounds, immediately turn off the power,
Instrument connections
1. Interface of the whole machine
Ground Interface
AC Power Input
CPU
If the color mark is still remained, you can connect according to the color mark, if the color identification is
missed, then connect in line with the following figure to ensure proper connection.
3 Functional Description
HTI BioChem FC-200 function box
The main board is machine's central control unit, including the mid-position machine, and five hypostasis
machine modules;
The mid-position machine is responsible for the channel scheduling of the five control modules;
Module 1 is for the movement control of the reagent disc motor and the filter wheel motor;
Module 2 is for the movement control of the reaction disc motor and the auto-rinsing motor;
Module 3 is for the control of the sampling needle up and down movement motor, the sampling needle rotating
Module 4 is for the control of the stirring rod up and down movement motor and the stirring rod rotation DC
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 18 of 43
Document Number SM-E-FC240
motor;
Module 5 is an independent board, it is a temperature control module, which can independently complete the
temperature control for the reaction disc constant temperature heating, sampling needle pre-heating, auto-
JM4:Auto-rinsing motor
J1: The interface to serial port transfer board, communicate with computer;
U15: The interface to the temperature control board, and communicate with it;
JP4: The interface to the liquid feeding and dispensing alarming board;
Temperature Control Board performs the whole machine temperature control, it can complete the temperature
control for the reaction disc constant temperature heating, sampling needle pre-heating, auto-rinsing pre-
The control parameters will be loaded through the serial port by the main board, and the temperature data will
Photoelectric conversion board is to convert the photodiode current signal into voltage signal, and perform the
J1
The photoelectric amplification board is for the second stage amplification of the photoelectric signal, the
magnification is adjusted by the digital potentiometer, the specific regulation should follow the corresponding
Input the former two stage amplified signal into the AD conversion chip for analog-digital conversion, the control
signal is from the main board, and the output signal will return back to the main board;
J2
J1
The machine has 10 correlation-type infrared photoelectric sensors, commonly known as optic-coupler;
The 10 correlation-type infrared photoelectric sensors will be fixed on board, connected to the main board by
3pin cable, supporting the main board for collecting the position signals of various fixed optic-couplers.
J1
This board will connect to the alarm sensors of the wash bottles and the waste bottle, the connector is designed
as audio interface, connect to the main board through the 10pin flat cable, for the alarm circuit signal processing;
It is designed with two-way cleaning liquid bottle alarm inputs, two-way waste liquid bottle alarm inputs, which
JP1
The liquid level detection board is to perform the liquid level detection when the sampling needle is aspirating
the reagent or specimen, and the test principle is capacitance. When moving the sampling needle up and down,
from up and down movement to contact with the liquid surface, the capacitance of the sampling needle will
change, the adjusted voltage signal will change, refers to the liquid surface information;
The crash-proof signal detection of the sampling needle up and down movement will be completed by the
The liquid level detection signal and the up and down crash-proof signal will be connected to the main board
TP1
It is the Serial Port transfer board of the whole instrument, one end will be connected to the main board serial
converter with 3 pin cable, and the other end will be connected to the computer serial port through an external
serial cable.
RS232
J1
J1
4 Parts Replacement
Instrument assembly
HTI BioChem FC-200 is a high- precision instrument, if the instrument has trouble and can not be resolved with
the method in this manual, it should be repaired by professionals; If it is confirmed that the internal parts of the
instrument have some troubles which should be replaced for repairing, please follow the below steps to
1. Turn off the instrument power switch (unplug the power cord to prevent from turning on the power switch
accidently).
2. Remove the eight screws which fix the rear-board of the machine with a screwdriver and remove the rear
3. After removing the rear-board, you can see the instrument's internal hardware boards, pump valves, the
The lamp light source will gradually age to make energy attenuation when used for a long time, if the system has
given out alarm message, or the total working time is estimated to be over 1000 hours, need to replace the light
source lamp.
1. Turn off the power to make the lamp and lamp box cool for 30 minutes.
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 28 of 43
Document Number SM-E-FC240
Note:
Must wait more than 30 minutes after shutdown, even after 30 minutes, you must be careful to touch the shell
body of the light source lamp, because the hot lamp and lamp box will cause skin burns.
3. Unscrew the three cross screws on the top of the light source body shell.
4. Wait until the light source shell will not burn your skin because of high temperature, pinch the light source
5. Remove the light source components, replace new light source lamp, connect the light source line well, and
then put back the supporting base and lock the three cross screws unscrewed before.
Note:
When assembling or disassembling the light source, even if the lamp has no heat, please do not touch
the translucent lamp surface with your hands, because this may change the characteristics of the light
source due to grease, dirt and other reasons, if inadvertently make the translucent surface stained with
grease, dirt, you can wipe it with a clean cotton swab dipped with absolute ethanol.
screw
cable
6. The light source lamp has been positioned by screws, no need for fine-tuning, but because of the differences
of various lamps, you need to adjust the received light energy, it is recommended to replace all the reaction
cups here, in order to avoid the large differences between the old and new cups. Method 1: turn on the
instrument, enter into the "HTI BioChem FC-200 auto chemistry analyzer management software" → "Status" →
"Reaction disc", click on “Change Cuvette”" button, and then you can replace it based on prompt, (method 2: or
from the " HTI BioChem FC-200 auto chemistry analyzer management software," → "Maintenance" → "Blank"
and then click on "Change Cuvette" button, but note that this method requires a user authorization; Method 3:
7. Users usually operate only with method 1 and 2, after correct replacement of the reaction cup, from the " HTI
BioChem FC-200 auto chemistry analyzer management software" → "Maintenance" → "Blank" and then click
on “Check Blank” button, record the proximate light intensity AD value of each wavelength, and then enter into
"Maintenance"→ “Configure”, adjust the magnitude ratio of each wavelength, make the light intensity AD value
of each wavelength to be about 55000, but for the wavelengths over 510nm, its AD value could be about 57000
(if it can not successfully auto-configure the magnitude ratio, you must configure it manually, the AD value is
proportional to the configured value), after completing the configuration, please perform the blank check once
more, please observe whether it is appropriate or not, you may need to repeat several times before it meets the
requirements.
8. After confirming that there is no problem, you can tighten the light source cover of the worktable.
Note: When adjusting the location of the light source lamp, do not let the light directly shine your eyes, because
2) The method to replace the reaction cup is the same with the steps mentioned in section 4.2. Method 1: turn
on the instrument, enter into the "HTI BioChem FC-200 auto chemistry analyzer management software" →
"Status" → "Reaction disc", click on “Change Cuvette”" button, and then you can replace it based on prompt,
(method 2: or from the " HTI BioChem FC-200 auto chemistry analyzer management software," →
"Maintenance" → "Blank" and then click on "Change Cuvette" button, but note that this method requires a user
3) If the instrument is with auto-rinsing system, when replacing the reaction cup, carefully unscrew the reaction
cup fixing screw; do not let it slide inside. If it has been slid inside, the proposed operation is to shutdown firstly,
HIGH TECHNOLOGY, INC. PROPRIETARY AND CONFIDENTIAL Page 31 of 43
Document Number SM-E-FC240
and rotates the reaction disc in place by hand or takes the falling screw out with tweezers or magnetic
screwdriver.
4) If the instrument is not with auto-rinsing system, when replacing the reaction cup, pay attention to take it out
vertically to prevent the reaction cup orientation column breaking accidently; when placing the new reaction
cup, please put it vertically, after it has been positioned well, press the cup joint slightly, to check whether it
is flat or not;
5) Because that the reaction cups may have differences in different batches, it is proposed that you should
observe the Cuvette blank after replacing the reaction cup; from " HTI BioChem FC-200 auto chemistry
analyzer management software," → "Maintenance" → "Blank" and then click on “Check Blank” button,
observe the proximate light intensity AD value of each wavelength, to determine whether need to adjust the
magnitude ratio;
entrance cover-plate
Pump\valve Replacement
Firstly, disassemble all the pump\valve components requiring for replacement, and then replace the specific
pump\valves.
Note: Pay attention to the bonding wire on the pump valve, so as not to break; be careful of liquid spilling when
Cables replacement
When replacing the cables, please pay attention to the labels on them, do not replace them with the cable
of incorrect labels, if you really have no correct label on the cables, you can write the correct label on a piece of
paper and cover it onto the original label, then twist it with transparent tape.
When plug and unplug the cables, please do not shake around.
HTI BioChem FC-200 is a precise clinical analyzer, in order to maintaining the instrument in a good condition;
you must do the routine maintenance work well. The maintenance of HTI BioChem FC-200 is very simple, but
Basic Maintenance
The cleaning of the instrument surface must be performed with some neutral detergents and damp cloth.
Check the sampling needle to make sure that it is in good condition, no impairment, and no dirt; wipe it with
Check the stirring rod to make sure that it is in good condition, no impairment, and no dirt; wipe it with some
connection of the pipeline is normal, and whether need to add water or drain.
Rinse the washing station of the cleaning needle and the stirring rod
Clean the ball spline shaft and the ball bearing of the sampling needle component (with grease lubrication)
Clean the guide rail of the stirring rod component (with grease lubrication)
Rinse the washing station: Add several milliliter of diluted bleach (5:1000), clean the inner surface of the
1. Turn off the instrument power, and open the rear end plate of the instrument.
2. Pull out the connector piece between the valve and the transferring jointer, as shown in the following picture.
3. Prepare a syringe of 10ml and aspirate the distilled water, install a cuvette head on the syringe and insert it
into the pipeline just pulled out, push the syringe to inject the water into the washing station (the needle must
be on the top of the washing station at this time), repeat the above step for more times until the water flow
washing tank
In order to avoid the breeding of bacteria, we recommend users to clean the washing station (the washing
stations of the sampling needle and the stirring rod) regularly. Infuse the diluted bleach liquid of 5:1000 into the
washing station and keep it for 15 to 20 minutes, then wipe it with a medical swab, after wiping, rinse with plenty
of distilled water to ensure that there is no cotton residue in the washing station.
station
Sampling needle
washing station
6 Trouble Shooting
Common Trouble Shootings
Trouble Solutions
Analyzer can not start up - Check whether the power plug is loose
The lamp has no light - Check whether the power is normal before replacing the lamp
Operating software can not be - Shutdown and wait for about 10 seconds, then restart the
initialized machine
properly connected
Bubble and spattering liquid gush - If the liquid-path is not used for a long-term, this happens during
Absorbency/blank AD value is
- Check whether the light source is emitting light normally
displayed incorrectly
the estimation
error
Poor reproducibility of the results -Some reaction cup may be contaminated, replace it
good
pipeline
Quality control is not within the - Check the period of validity of the quality control liquid, and
target value verify that the quality control liquid is not contaminated
There are a lot of air bubbles in - Check the connection air-tightness of the syringe T-way
the speed)
There are bulbs in syringe - Confirm that the piston is tight and no-leak.
water)
needle
contaminated
cup
Liquid leaks from the analyzer - Check whether the draining pipeline of the waste liquid is
● If there is any instrument failure, please contact with the dealer in time to obtain technical support!
As for the calculation item, recommend to use the following codes to fill in its item name and item code.
7. LDL (Low Density Lipoprotein) = CHO (Total Cholesterol)-HDL (High Density Lipoprotein)-TRIG
(Triglyceride)/2.2
Note:
This manual may contain other proprietary notices and copyright information, and the related
provisions must be observed. The information in this manual may contain some technical
This manual may be changed or updated without notice. HTI can also improve and/or change the