TRO内部备件的保养 No.

20230201

Table of Maintenance of TRO Analyzer


每1-2个月 Every 1-2 months
1 检查分析仪的管路和接头 Check the tubing and fittings of the analyzer
2 检查试剂消耗量 Check the consumption of reagents
每3个月 Every 3 months
1 更换指示剂 Replace the Indicator
2 检查T型过滤器 Check the T‐strainer
3 检查比色皿 Check the cuvette
每12个月 Every 12 months
1 更换缓冲剂 Replace the Buffer
2 检查盖帽组件 Check the Cap assembly
如何清洗T型过滤器 No.20230201

How to clean the T‐strainer

1. 关闭水源,按 SERVICE 停止流动


Turn off the source water, and press SERVICE to stop the flow.
2. 从夹子上取下 T 型过滤器。
Remove the T‐strainer from the clamp.
3.断开 T 型过滤器和调节器的顶部与管道的连接
Disconnect the top of the T‐strainer and regulator from the tubing.
4.清洁 T 型滤网并更换
Clean the T‐strainer screen and replace.
5. 拧紧 T 型过滤器的转鼓并将 T 型过滤器装回仪器。
Tighten the bowl of the T‐strainer and install the T‐strainer back to the
instrument.
6. Turn source water back on. Return to normal operation. Check for leaks.
重新打开源水。恢复正常操作。检查是否有泄漏
如何更换盖帽装置 No.20230201

How to replace the Cap assembly

1. 按 PRIME,然后按 清空试管。
Press PRIME and then to empty the tubes.
2. 2. 按下 SERVICE 后,从止回阀的连接处取下缓冲盖组件
After pressing SERVICE, removethe buffer cap assembly from the
connection at the check valve.
3. 更换缓冲盖组件
Replace the buffer cap assembly.
4. 对指示剂重复步骤 2 和 3
Repeat steps 2&3 for the indicator side.
5. 按 PRIME,然后按 用试剂填充试管
Press PRIME and then to fill the tubes with reagents.
6. 恢复正常运行
Return to normal operation.
如何清洗比色皿 No.20230201

How to clean the cuvette


1. 按SERVICE水流停止流动
Press SERVICE to stop the flow.
2. . 逆时针转动光学组件顶部(如所见对于顶部),直到比色皿“弹出”出来,但不要取下顶部
Turn the knurled top on the optics system counterclockwise (as viewedfor the top) until the cuvette just “pops” out, but
do not remove the top.
3.当比色皿“弹出”时,移动固定O形圈并取下比色皿。您可能需要一根坚硬的电线,例如弯曲的回形针来抓住
比色皿。如果可以清洁,请保留此比色皿以备将来使用。
When the cuvette “pops” out, move the retaining o‐ring& remove the
cuvette. You may need a stiff wire such as a bent paper clip to grasp the
cuvette. Retain this cuvette for future use if it can be cleaned.
4.用柔软的湿布清洁比色皿。如果比色皿是严重脏污或划伤请更换
Clean the cuvette by a soft wet cloth. Replace a new one if the cuvette is
badly soiled or scratched.
5. 恢复正常运行。检查是否有泄漏。
Return to normal operation.Check for leaks.