1. 按 PRIME,然后按 清空试管。
Press PRIME and then to empty the tubes.
2. 2. 按下 SERVICE 后,从止回阀的连接处取下缓冲盖组件
After pressing SERVICE, removethe buffer cap assembly from the
connection at the check valve.
3. 更换缓冲盖组件
Replace the buffer cap assembly.
4. 对指示剂重复步骤 2 和 3
Repeat steps 2&3 for the indicator side.
5. 按 PRIME,然后按 用试剂填充试管
Press PRIME and then to fill the tubes with reagents.
6. 恢复正常运行
Return to normal operation.