You are on page 1of 1

HBCH/MUL/2021-22/CAP/LT-01

TATA MEMORIAL CENTRE


HOMI BHABHA CANCER HOSPITAL, SANGRUR
PURCHASE DEPARTMENT

Technical Specifications of "Portable Suction Machine " Qty: 20 nos.


Please quote all the models including the state of art equipment which meets the basic minimum specifications.

DEPARTMENT: Anaesthesia, Mullanpur Compliance Remarks


Technical Specifications of "Portable Suction machines"Qty; 20 Nos. Yes No
Please quote all the models including the state of art enquipment which meets the basic
minimum specifications.
Model Name:
Make:
Year of Introduction:
Probable year of end of support:

Should be designed for draining blood and other fragmetic secrtetions in operation
1 room and emergency room in hospital.
2 Should be fitted with oil immersed noiseless motorized vacuum pump.
3 Cabinet made from stainless steel.
Two glass jars on the top of having minimum capacity of 2 Ltrs fitted with rubber air
4 tight lids and overflow safety device.

5 Should be suppplied with adequately long pressure tubing providing required pressure

6 Should have vacuum control by knob.


7 Should be mounted on four caster wheels.
8 Should have motor of 1/4 HP capacity and power consumption not more than 250 Wat
9 Should have vacuum at least between 100 mm Hg to at least 575 mm Hg (-75 kPa).
International standards: the unit should comply with international standards and
10 should have CE or FDA marking.
Should perform yearly calibration and preventive maintenance (2 nos) during
warranty & if entered into AMC (includes unlimited breakdown calls)/CMC (which
will include spare replacement, consumable spares, breakdown calls)/Testing &
11 measuring equipment used should be traceable to SI units through
National/international standards (As per NABH norms).

Please specify list & cost of consumables/consumable Spares(i.e.spares need to be


12 replaced at regular intervals, maybe quartely/half yearly/yearly such as annual
maintenance kit etc.) if any.

13 Please specify pre installation requirements (electrical, HVAC etc.)


Demo of the quoted model, will be required if so desired by the user, after the
14 opening of the technical bid and prior to opening of financial bid. This is for technical
evaluation.
Local Service Support: Should have local office and service support/service engineer
15 for attending the breakdown calls.
Reponse Time: Should not be more than 12 hrs from lodging a breakdown complaint
16 on toll free or by email.

17 Warranty: 2 Years
18 AMC: Quote Rate for 8 years after warranty.
Back to back assurance to be taken by the supplier from the OEM to supply spares
19 for minimum 10 years & to be submitted.

20 Repairs work of the machine should be done at customer site.

1) If 2 years warranty, 8 years AMC and buyback conditions wherever applicable are not adhered to, TMC reserves right to
reject offer.

2) Please mention whether UPS / Stabilizer is required for the equipment. If so, should be included in the offer.

3) Only yes or no should be mentioned in the compliance report and additional info in remarks column.