Professional Documents
Culture Documents
RO-EDI
1 OBJECTIVE:
To Establish Standard Operating Procedure for operating RO-EDI System.
2 SCOPE:
This shall be applicable to operate Single pass RO-EDI in Purified Water Generation system.
3 RESPONSIBILITY:
3.1 Doing : Hydrocons Systems Pvt. Ltd.
3.2 Checking : HSPL. / Client
4 PURPOSE:
5 PROCEDURE:
RO System : I
Turn on the two way switch to normal AUTO mode
Press log in to operator and also go to manual mode
First check the level of the Dosing tanks And Soft water storage tank, if level is sufficient then
start the system.
Open Actuated valve AV-06 and also open ADV-01. By pressing in HMI screen.
RO System : II
Set the Reject flow 0.8 m3/hr on RM-04 maintain PH 7-8 at RO-II so we will get flow 2.2 m3/hr at
RO-II outlet(On RM-05).
Check the RO-II outlet conductivity is in limit less than 10.
After open valve DV-08, DV-09 and DV-11 ( DV-11 Always open)
Through the valve DV-10 set the flow on Rotameter (RM-06)such manner should be 200 L/Hr
Open the valve BTADV-01 and closed the valve ADV-03
NOTE-All Diaphragm valve operation will be manually.
EDI : I
Note:
If proper flow rate is not maintained in feed line of EDI then EDI will be burn. Hence proper
feed flow rate to be maintained as per above at feed, concentrate & electrolyte chambers.
The high conductivity of the feed water having more than <20 µs/cm then the scaling will
take place in EDI which will damage EDI unit permanently.
All instrument of RO-EDI system should be calibrated stringently every year.