You are on page 1of 13

959

Functional and structural effects of percutaneous


edge-to-edge double-orifice repair under cardiac
cycle in comparison with suture repair
A Avanzini*, G Donzella, and L Libretti
Department of Mechanical and Industrial Engineering, University of Brescia, Italy

The manuscript was received on 14 February 2011 and was accepted after revision for publication on 3 June 2011.

DOI: 10.1177/0954411911414803

Abstract: Percutaneous procedures for double-orifice mitral valve repair using the MitraClipÒ
device (clip) have been recently introduced as new treatment options as alternatives to medi-
cal management and open-heart surgery, especially for patients with high estimated operative
risk. Similarly to the open-heart surgical technique, where suturing is used, the clip creates a
double-orifice configuration that not only improves the closing function of the valve, but also
significantly modifies its behaviour, particularly in the diastolic phase. While several clinical
trials have been conducted, and are ongoing, in order to assess the safety and effectiveness of
this technique, a deeper knowledge of the structural and functional effects on the valve, and
of the cyclic loads transmitted to the clip itself, would allow a comparison with other repair
techniques, and could serve as a foundation for possible further optimization of the clip
design. The effects of the MitraClipÒ device developed by Evalve Inc. were studied by means
of a finite element model of the mitral valve, specifically developed to study the structural
effects of the original, suture-based, edge-to-edge technique. A second model was developed
in order to simulate the effects of a suture with similar extension from the leaflet edge in a
direction to the annulus, in order to compare the two repair techniques. The mitral valve area
and transvalvular pressure gradient predicted by the models for the clip and the suture are
quite similar. Similar leaflet cyclic stresses, both in value and in location, were noted for the
two mechanisms of linking the leaflets, while minor differences were found in the load trans-
mitted to the suture and the clip, with slightly higher values for the clip. The model satisfacto-
rily allowed functional parameters (valve area and transvalvular pressure gradient) and
structural parameters (load, leaflet stress) to be determined. Overall, the structural effects of
the clip and the suture are quite similar under the cyclic loading conditions imposed by the
cardiac cycle.

Keywords: double orifice, MitraClip, percutaneous repair

1 INTRODUCTION enhancing the quality of life and long-term survival


for thousands of patients [1, 2]. The technique allows
Since 1991, the surgical double-orifice or edge-to- repairs in more complex valve pathology and avoids
edge repair technique has been used in patients with of valve replacement, and may thereby avoid compli-
leaflet prolapse to treat mitral valve regurgitation, cations such as thromboembolism, and the need for
long-term anticoagulation treatment, associated with
valve replacement. Mitral repair preserves the papil-
*Corresponding author: Department of Mechanical and lary muscles and chordae, and therefore left ventricu-
Industrial Engineering, University of Brescia, via Branze, lar function, and is associated with lower operative
38-25123 Brescia, Italy. and late mortality [3, 4]. Contraindications to surgery
email: andrea.avanzini@ing.unibs.it do exist in patients with high estimated operative risk

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
960 A Avanzini, G Donzella, and L Libretti

and multiple co-morbidities. In order to overcome obtained with a standard edge-to-edge suture tech-
these limitations, percutaneous procedures for heart nique during diastole and systole.
valve disease have recently been introduced [5].
Because of their minimally invasive nature, they are 2 METHOD
associated with a low risk of procedure-related com-
plications, but optimal patient selection and long-
The effects of the clip were studied by means of a
term effectiveness have still to be determined [6].
hyperelastic finite element model of the mitral valve
Regarding percutaneous mitral leaflet repair, one
specifically developed to study the structural effects
of the most promising approaches has been devel-
of the surgical technique. Two different models
oped by Evalve Inc. (Abbott Vascular, Santa Clara,
were implemented, in order to simulate the pres-
CA) using the MitraClipÒ system [7]. It consists of a
ence on the same mitral valve either of the clip,
MitraClipÒ device and a clip delivery system, which
modelled as an independent part with dimensions
is advanced through a steerable guide catheter. The
corresponding to its actual size, or of a suture with
procedure consists in positioning the clip following
similar extension along the leaflet edge. In the fol-
standard trans-septal access to the left atrium under
lowing, these models will be referred to as model 1
echocardiographic guidance, and joining together a
(MitraClipÒ device, with width 3.5 mm and length
portion of the anterior and posterior leaflets at the
7 mm) and model 2 (suture, with width 4 mm and
location of mal-coaptation.
length approximately 2.5 mm).
Several clinical trials [8] have been performed,
After valve geometry definition, structural analy-
and are ongoing, in order to assess the safety and
ses were carried out by means of the commercial
effectiveness of this technique in comparison with
software ABAQUS 6.8.1 (SimuliaÓ). The program
standard cardiac surgery. Up to now, the results
used for finite element modelling of the mitral valve
have demonstrated that percutaneous clip repair is
is a specialized program for structural analysis, par-
able to reduce mitral regurgitation, with a low mor-
ticularly suitable for the study of complex and
bidity and mortality rate and meaningful clinical
sophisticated models of biological apparatus.
benefits, in patients with both degenerative and
functional mitral regurgitation.
2.1 Mitral valve geometry
Similarly to the double-orifice repair with suture,
the presence of the clip creates a double-orifice con- The mitral valve apparatus mainly consists of two
figuration that can significantly modify the beha- valve leaflets, anterior and posterior, connected
viour of the valve, particularly in the diastolic phase. along the valve perimeter to a fibrous ring called the
It is therefore essential to assess the stress on the annulus. On the surface facing the ventricle, leaflets
leaflets induced by the clip, and the loads sustained are connected to papillary muscles by marginal and
by the clip during the cardiac cycle. A deeper knowl- strut chordae tendinae.
edge of the structural effects of this procedure on A scheme of the valve model adopted in the pres-
the valve, and of the loads transmitted to the clip ent work is represented in Fig. 1.
itself, would allow a comparison with other repair The complex anatomical structure of the mitral
techniques, and could serve as a basis for possible apparatus was represented by deriving the valve
optimization of clip design. Very limited data are
available in the literature concerning these topics,
and none are available referring to the application
of a clip. Only a few numerical studies that consid-
ered the presence of a surgical suture [9, 10], and
experimental measurements carried out in vivo and
in vitro with miniature load transducers joining the
leaflets [11–13], are available.
In order to fill this gap in the present knowledge,
a finite element model was developed to simulate
the clip and to calculate the corresponding trans-
mitted load and leaflet stresses during diastole and
systole. The functional parameters (mitral valve area
and transvalvular pressure gradient) and the struc-
tural effects (load transmitted to the clip and stress
state in the leaflets) due to the application of the
clip were then determined and compared with those Fig. 1 Mitral valve: atrial view

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 961

geometry directly from the model proposed in refer- The leaflets’ thickness was kept the same as in
ence [9]. In that work the annulus size was 5 cm2 the previous model (1.32 mm for the anterior leaflet
(corresponding to a largest diameter of about and 1.26 mm for the posterior leaflet [14]).
30 mm). This is a value in the range reported in the The model includes some simplifying assump-
literature for a healthy mitral valve; it also corre- tions concerning the valve geometry (i.e., absence of
sponds to the size of the prosthetic ring usually clefts in the posterior leaflet, physical separation of
implanted during repair intervention to restore phy- anterior and posterior leaflet in the commissural
siological valve dimension. However, the current clip region), but it still correctly reproduces the main
repair procedure does not include annuloplasty, and features of the mitral apparatus, and at the same
furthermore it is well known that many valves with time it guarantees computational efficiency and
degenerative disease may present a significantly accuracy, as shown by previous studies.
enlarged MVA compared with their original dimen- Both suture and clip models implicitly take into
sions. For this reason both model 1 and model 2 used account possible leaflets prolapse, as the repair
in the present study, which share the same valve geo- simulation introduces a local constraint (described
metry, were created by a scaling operation in which in the following) if and where prolapse is present. In
the annulus size was enlarged to 10 cm2 (correspond- addition, compared with the values reported in liter-
ing to a largest diameter of about 43 mm). In order to ature for a reference mitral valve [15, 16], the geo-
keep the same ratios as in the original model metry considered here refers to a valve having a
between the anterior-posterior diameter (B) and the higher annulus area (10 cm2 rather than 8 cm2) but
commissure-to-commissure diameter (A), and similar leaflet lengths (about 24 mm and 14 mm),
between the anterior and posterior leaflet length, a thus also addressing the condition of pure annulus
scaling factor of approximately 1.42 was assumed for dilatation. As a consequence, the model fits patients
these characteristic dimensions (see Table 1). with both pure annular dilatation and ruptured
It should be noted that the scaling operation chordae.
starting from a smaller valve does not really corre-
spond to the changeover from a healthy to a patho-
2.2 MitraClipÒ device and suture schemes
logical valve; it was used only for ease of creation of
the desired geometry starting from an existing The clip (shown in Fig. 2(a)) is a complex mechani-
model, taken as a reference. cal device consisting of many different components.

Table 1 Mitral valve dimensions


Model Annulus A (mm) B (mm) Anterior leaflet Posterior leaflet Annulus
area (cm2) length (mm) length (mm) perimeter (mm)

Present 10 42.74 28.14 23.9 14.3 116.95


Reference [9] 5 30 19.75 16.8 10 81.6

Fig. 2 (a) MitraClipÒ device; (b) model

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
962 A Avanzini, G Donzella, and L Libretti

In the present study its actual geometry has been Table 2 Material coefficient (units MPa)
simplified by considering the clip itself as consisting
C10 C20 C30 C40 C50
of the two arms that perform coaptation of the leaf-
lets. The clip was modelled as consisting of 8.733 3 1022 21.236 12.912 250.532 95.518
two symmetrical arms connected by a hinge in its
bottom part (Fig. 2(b)). When the device is closed,
the arms are almost parallel, and the distance
polynomial function, taken equal to 5; and Ci0
between them is equal to the sum of the leaflets
and Di0 are material parameters, determined by a
thickness. The upper parts of the arms are inclined
simple fitting procedure, automatically available in
at a small angle to allow a slight separation between
the Abaqus code, and described in more detail in
the leaflets, reproducing their V-shape. A radius is
reference [9].
present at the top part of the arms, to allow a
The values of the coefficients Ci0 are reported in
smooth transition from the area constrained by the
Table 2. Under the assumption of material incom-
clip to the free surfaces of the leaflets. The arm
pressibility, the coefficients Di were all set equal
thickness assumed for the clip is 1 mm.
to zero.
The suture has been simulated using link connec-
Once the strain-energy function is known, the
tors supported by coupling constraints for each of
stress state can be determined by taking the deriva-
the two leaflets, with the function of distributing the
tive of U with respect to a strain measure
transmitted load on an area corresponding to the
extension of the actual Teflon pledgets. ∂U
When comparing the constraint imposed by the s= (2)
∂e
two repairs in the present model, it should be noted
that the grasp length is not exactly the same for the where s and e are any work conjugate stress and
two techniques, and moreover that, in its upper strain measures. As previously mentioned, the leaf-
part, the clip exerts a unilateral constraint (from the let material consists mainly of water and therefore
outside) on the leaflets’ surfaces, which is not pres- has very little compressibility. Incompressibility is
ent in the case of the suture. in general a difficult phenomenon to handle com-
putationally; nevertheless, when using shell ele-
2.3 Finite element models ments in applications where the material is not
Based on the mitral valve and clip geometry, the finite highly confined, it is quite satisfactory to assume
element models described below were developed. that the material is fully incompressible. The clip is
made mainly of metallic materials whose stiffness is
2.3.1 Materials greater than that of the leaflets by orders of magni-
tude. Since the stress state in the device is not con-
Mitral valve leaflets consist of a soft biological tis- sidered in these simulations, its material is simply
sue, essentially made of a fibrous tissue network, assumed as infinitely rigid by assigning to the clip a
mainly collagen and elastin, saturated with a fluid value of Young’s modulus that is extremely high
that is mostly water. compared with the leaflet tissues stiffness.
As a result of the coupling between different con-
stituents, the typical stress–strain diagram may be 2.3.2 Contact interactions and boundary conditions
divided into a pre-transition region, in which the
stress strain curve is fairly flat, followed by a post- Valve leaflets are involved in a different kind of con-
transition region where the slope suddenly increases. tact interaction, whose correct definition is funda-
Many different theoretical approaches have been mental. Leaflet coaptation at valve closure during
proposed for constitutive modelling of this complex systole has been modelled considering frictionless
mechanical behaviour. In the present work a hypere- contact between their atrial surfaces. The clip is
lastic material constitutive law has been adopted, in considered as an independent geometric part whose
which the strain-energy potential U has been structure interacts with the leaflets. Contact
defined in the reduced polynomial form as between the leaflets’ ventricular surfaces and clip
arms has been considered as rough – that is, no
X
N  i XN
1  el 2i separation or relative sliding is allowed after contact
U= Ci0 I1 3 + J 1 (1)
i=1
D
i=1 i
is reached – for better simulation of the device’s
strong fixing action. In order to model precisely the
where I1 is the first deviatoric strain invariant; Jel is contact interaction in the clipping region and the
the elastic volume ratio; N is the order of the real kinematic constraints imposed by the clip, it

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 963

Fig. 3 Boundary conditions: (a) fixed annulus; (b) schematization of chordae tendinae constraint

was decided to adopt a surface-to-surface approach, modelled using rigid connector elements (link type)
taking as reference surfaces for contact pairs the to simulate suture stitches. In particular, the suture
inner surfaces of the clip, which actually interacts was simulated with six link connectors (for a total
with the valve leaflets, and the midplane surfaces of width of 4 mm) coupled to a small surface near the
the leaflets. The boundary conditions reported in leaflet edges. In this case, the total force transmitted
Fig. 3 were initially applied; in particular, a fixed to the suture is given by the algebraic sum of those
constraint (no displacement) was imposed on the of each link connector.
nodes of the fibrous annulus, and the effect of
the chordae tendinae was simulated by constraining 2.4 Analyses
the free-edge nodes of the leaflets to move on a plane
inclined at 26.5° with respect to the annulus plane. In order to simulate the surgical procedure and the
This boundary condition allows correct simulation of effect of clip positioning, FEM analyses were carried
the constraint imposed by the chordae tendinae on out considering two main steps: the first included
leaflet movement without physically modelling this leaflets approach and clip or suture application, and
apparatus, thus significantly reducing the model the second involved pressure application to simu-
complexity and the related convergence problem, late the cardiac cycle in diastole and systole.
especially for the contact interaction phases. Different phases of the FEM analyses, which are
slightly different depending on the presence of the
2.3.3 Mesh clip or of the 4 mm suture, are detailed below and in
Figs 4–6, in which, for convenience, only half the
The leaflet mesh consists of three- and four-node mitral valve is displayed.
reduced-integration shell elements with hourglass
control and finite membranal strain. In order to
2.4.1 Leaflets approach and MitraClipÒ device
obtain satisfactory convergence in all configurations
or suture application
examined, slightly different meshes were used for
the clip and suture models. In particular, for model Model 1 - MitraClipÒ device. The clip application is
1 (the clip), the anterior and posterior leaflets were described below (see Fig. 4).
meshed with 6736 and 3550 elements, respectively.
Each clip arm consists of 792 elements, for a total of (a) The hinge of the clip is positioned in corre-
1584. In order to calculate the load transmitted to spondence with the leaflet border in the middle
the clip, a link connector was introduced in the cen- of the valve. The clip is completely open, in a
tre of the device. For model 2 (the suture) the ante- position parallel with respect to the orifice
rior and posterior leaflets were meshed with 5506 plane; the hinge is constrained to a fixed posi-
and 3550 elements, respectively. The suture was tion. A pressure of about 30 per cent of systolic

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
964 A Avanzini, G Donzella, and L Libretti

Fig. 4 Clip application steps: (a) clip positioning; (b) arms closure; (c) clip locking; (d) catheter removal

Fig. 5 Complete model of mitral valve after clip application

Fig. 6 Suture application phases: (a) leaflets approach; (b) suture positioning

is applied on the ventricular side to bring the (b) Closure of the clip is simulated by rotating its
leaflets closer so to allow successive clip arms around the fixed hinge; at the end of the
application. closure its arms present an angle of about 15°

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 965

with respect to the vertical plane. The central were determined using the procedure described in
link, initially deactivated, is activated at the end reference [17], based on the fluid-dynamic FEM
of the closure, simulating the clip locking. model proposed in reference [18].
(c) The pressure applied on the leaflet surfaces In particular, under the hypothesis of a constant
during phase (a) is removed. At this point the blood flow rate across the valve, Q (l/min), Dp
regions of the leaflet surfaces that came into (mmHg) was determined as a function of MVA
contact with clip arms are constrained to the (cm2) by means of the equation
clip itself (as a result of rough type interaction).
 2
(d) The fixed position constraint imposed on the Q
hinge is removed. The device is then free to Dp = 0:16 (3)
MVA
move in the space, remaining only in contact
interaction with the leaflets. At the end of this This equation represents all possible working
phase, which corresponds to the catheter being points of the valve. Following the literature [19], a
removed, the valve–clip system reaches the typical value of Q = 13.7 l/min has been assumed in
equilibrium configuration. present analyses. By computing the MVA from the
valve FEM model at different pressure levels, it is
A complete view of the valve model with the clip possible to identify the valve’s effective working
after the application phases is shown in Fig. 5. point.
Model 2: Suture 4 mm width. At the beginning of Systolic phase. After the clip or suture application,
the analysis, the same boundary conditions as for pressure from the ventricular side is applied, main-
the clip are applied; then the following sequence is taining the chordae tendinae constraint. The
applied (see Fig. 6). applied systolic pressure has a value of 130 mmHg,
in the physiological range reported in the literature.
(a) A pressure of about 10 per cent of systolic is
applied on the ventricular side to bring the
leaflets closer. The applied pressure is smaller 3 RESULTS
than that necessary for clip application,
because a large area of leaflet coaptation is not 3.1 Model 1: MitraClipÒ device
required, the suture being created very close to
the leaflet edge. 3.1.1 Diastolic phase
(b) The links that simulate the suture are activated According to the procedure previously described,
in the model, and the pressure is deactivated. the predicted MVA for the valve repaired with the
The leaflets reach the equilibrium position clip is 1.83 cm2, and the transvalvular pressure gra-
without loads. dient is 8.94 mmHg.
The tensile force transmitted to the link in dia-
It should be noted that the initial application of a stole reaches a value of 0.39 N, increasing almost
pressure on the ventricular side is only a modelling linearly with applied pressure.
expedient for bringing the leaflets together, thus The Von Mises stress contour map of the leaflets
simulating the technique used by the surgeon in under the effect of diastolic pressure is reported in
order to proceed with the suture. The extent of this Fig 7. From this figure, it can be observed that stress
‘virtual’ pressure can change, depending on valve peaks are located near the top edges of the clip
geometry, but it does not enter in the simulation of arms. The points where the stress is maximum are
the cardiac cycle, because it is removed once the indicated as AC in the figure.
repair is applied.
3.1.2 Systolic phase
2.4.2 Simulation of the cardiac cycle
Because of the systolic pressure application, a com-
Diastolic phase. After either clip or suture application, pressive (closing) force is transmitted to the clip
the chordae tendinae constraint is removed for both arms with a peak of 0.43 N, whose existence pre-
models, on the assumption that the chordae have no sumes a perfect sealing between clip arms and leaf-
structural effects during diastole, and an increasing lets. The Von Mises stress contour map of the
pressure is applied on the atrial side of the valve. In leaflets under the effect of systolic pressure is
this phase, the leaflets open in a double-orifice con- reported in Fig 8. In this case, a large, highly
figuration. The mitral valve area (MVA) and transvalv- stressed zone can be observed in two symmetrical
ular pressure gradient (Dp) in the diastolic phase central parts of the anterior leaflet, in addition to

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
966 A Avanzini, G Donzella, and L Libretti

Fig. 7 Diastole with MitraClipÒ device: leaflets Von Mises stress distribution (MPa)

Fig. 8 Systole with MitraClipÒ device: leaflets Von Mises stress distribution (MPa)

the local effect generated by the clip. The maximum where sIa and sIIa are the alternating principal
stress occurs in these central parts, at the point indi- stress components, i.e. one half of the correspon-
cated as BC in the figure. dent stress ranges.
In Table 3, the in-plane principal stress values sI The equivalent alternating Von Mises stress is a
and sII are reported for points AC and BC defined parameter suitable for evaluating the structural
above, on both the atrial and ventricular sides of the effects due to fatigue on the leaflet tissues. The
leaflets, at the end of diastole, and at the end of sys- resulting values may be useful in comparing the
tole. The corresponding stress ranges during the predicted cyclic damage level in the highly stressed
cardiac cycle (i.e. the difference between the diasto- points highlighted above, and for the two repair
lic and systolic values) are reported along with the techniques examined.
equivalent alternating Von Mises stress, calculated From the above results, the presence of both
as membranal and flexural stresses on the leaflets dur-
ing the cardiac cycle can be observed. It should also
qffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
be noted that the highest stressed zone is point BC,
saMises = s2Ia + s2IIa sIa sIIa (4) on the atrial side of the leaflet body during systole,

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 967

Table 3 Model 1: MitraClipÒ device, leaflets principal extent than that generated by the clip. The maxi-
peak stresses (all values in MPa) mum stress occurs at the point indicated as AS in
the picture.
Diastole Systole Stress Equiv. alt.
range Von Mises
stress
3.2.2 Systolic phase
Point AC
sI: atrial side 0.19 20.05 0.24 0.17 Because, during systole, the leaflets are compressed
sII: atrial side 0.14 20.25 0.39 against each other [7], and the surgical suture is a
sI: ventricular side 20.04 0.12 0.16 0.07
sII: ventricular side 20.03 0.10 0.13 unilateral constraint that cannot oppose any inter-
nal action against compression, it is clear that the
Point BC
sI: atrial side 20.01 0.49 0.50 0.22 load transmitted by the suture can be considered to
sII: atrial side 20.03 0.23 0.26 be zero. The Von Mises stress contour map of the
sI: ventricular side 0.01 0.03 0.02 0.01
sII: ventricular side 0 0 0 leaflets under the effect of systolic pressure is
reported in Fig 10; the results are similar to those
obtained for the clip. In particular, the local effects
of the suture are not noticeable in this phase, and
substantially away from the clip. The local stress the maximum stress occurs in two symmetrical cen-
concentration due to the clip is therefore overcome tral parts of the anterior leaflet, in the point indi-
by the effect of the systolic pressure in the central cated as BS in the picture.
part of the leaflet body closer to the annulus. In Table 4, the same stress components as in
Table 3 are reported for points AS and BS.
Observing these results, it can be verified that the
3.2 Model 2: suture 4 mm width
low local effect exerted by the suture leads to small
stresses at point AS. As in the case of the clip, the
3.2.1 Diastolic phase
most highly stressed zone corresponds to point BS, in
The MVA and the transvalvular pressure gradient the central part of the anterior leaflet, with values very
(Dp) were determined using the same method as similar to that encountered in the model with the clip.
performed for the clip case. The predicted MVA
result is 2.03 cm2, and the transvalvular pressure
gradient is 7.30 mmHg. The total force transmitted 4 DISCUSSION
to the links at the end of the diastole is 0.25 N.
The Von Mises stress contour map of the leaflets The double-orifice technique is a relatively new sur-
under the effect of diastolic pressure is reported in gical repair technique, and the amount of published
Fig 9. The stress concentration effects near the data available is limited and often inconsistent
suture edges can again be noted, although at a lower between different authors, thus making it very

Fig. 9 Diastole with suture: leaflets Von Mises stress distribution (MPa)

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
968 A Avanzini, G Donzella, and L Libretti

Fig. 10 Systole with suture: leaflets Von Mises stress distribution (MPa)

Table 4 Model 2: suture, leaflets principal peak stres- (about 9 mmHg) are in agreement with the peak
ses (all values in MPa) pressure gradient of 7.09 6 2.88 mmHg measured in
post-operative conditions after placement of the clip
Diastole Systole Stress Equiv. alt.
range Von Mises (see Table 5). Considering the MVA, it should be
stress emphasized that different methods are used to mea-
Point AS
sure it, which can result in significantly different
sI: atrial side 0.12 0.06 0.06 0.03 values, so it is more difficult to find consistent data
sII: atrial side 0.02 20.03 0.05 about valve pre- and post-operative conditions and
sI: ventricular side 20.01 0.15 0.16 0.07
sII: ventricular side 20.02 0.01 0.03 MVA values. Again, referring to reported clinical
Point BS
data [20], the predicted value of 1.83 cm2 is within
sI - atrial side 0 0.49 0.49 0.21 the range measured with the pressure-half-time
sI: atrial side 20.02 0.19 0.23 method (3.01 6 1.42) and the Gorlin formula
sII: atrial side 0.03 0.03 0 0.01
sI: ventricular side 0.01 20.01 0.02 (3.05 6 0.90), although closer to the lower limit, but
outside the lower limit of MVA measured by plani-
metry. Similar reasoning applies to the suture
model: the values are in agreement with the mean
difficult to compare experimental data, owing to the
MVA, assessed after surgical suture double-orifice
absence of the precise suture length and position,
repair, reported to be 2.3–2.7 cm2 (again depending
and of detailed information regarding the pre-
on the measurement method) in reference [21].
operative and post-operative conditions of the valve
Concerning the load transmitted by the leaflets to
itself. This study simulates the implantation of the
the link mechanism, it is practical to consider the
clip as an independent part applied to the mitral
diastolic and systolic phases separately.
valve, which is a new technique, and there are no
Under the effects of diastolic pressure, the load
direct terms of comparison regarding the force
was 0.39 N for the clip and 0.25 N for the suture.
transmitted to the MitraClipÒ device or the result-
This difference is due to the fact that the region of
ing stress on the leaflets.
the leaflets interacting with the clip arms is larger
Considering the above limitations, it is still worth
than the area near the leaflet edges on which the
comparing the FEM results with published literature
suture is applied, resulting in a larger area of leaflet
regarding this kind of surgical procedure, referring
to both clinical data and experimental in vivo and
in vitro testing. Since the clip and the suture have Table 5 MVA and diastolic pressure gradient compari-
been applied on the same valve model, it is also son for the two link typologies
possible to compare the behaviour of the valve in
MitraClipÒ Suture
the presence of different repair techniques.
The clip results can be compared with reported Diastolic MVA (cm2) 1.83 2.03
clinical measurements [20]. For the transvalvular Peak diastolic transvalvular 8.94 7.30
pressure gradient (mmHg)
pressure gradient the values predicted by the model

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 969

constraint. It is interesting, however, to note that in the fact that the maximum stresses are located
the peak opening loads are of the same order of reasonably far away from the link, and therefore
magnitude as reported by Nielsen et al. [11] of are not influenced in a significant way by the link
0.46 N in their in vivo ovine model. The Nielsen typology itself.
model used direct measurement of the load on the In conclusion, the comparison of the clip and
leaflet via a miniature force transducer, and the standard suture techniques suggests that their func-
measured peak load of 0.46 N occurred after indu- tional and structural effects on the mitral valve are
cing ischemia. Similar values of peak suture tension similar, so that in this sense they are comparable
(0.30 6 0.18 N) are reported in reference [12]. During from a clinical point of view.
systole, a differential compression force of 0.43 N is
transmitted to the clip, owing to the effective area of 4.1 Study limitations
its arms exposed to the ventricular side pressure,
which tends to close it. This result is valid under the Any model necessarily embodies simplifications and
hypothesis that blood cannot penetrate between the limitations. For the present study a first set of lim-
clip arms and the side of the ventricular leaflets side itations concerns assumptions about the geometry
– that is, that a perfect sealing is obtained between of the mitral leaflets. In particular, in the commis-
these two elements. Nevertheless, there is no open- sure regions, the irregular shape of the leaflets,
ing load transmitted to the clip due to systolic pres- tissue folding and redundancy, and the transition
from leaflet to chordae tissue were not considered.
sure application. No load at all is transmitted to the
As a consequence, the stress distribution in this area
suture, as it represents a unilateral constraint that
and the shape of the orifice of the valve opening
cannot bear compression loads. These results agree
may in reality be slightly different.
with those reported by Umana et al. [22], who com-
It has also been shown that annulus dilatation
ment: ‘During the high pressure of systole, the leaf-
reduces leaflet mobility in a sutured valve, thus
lets are naturally pushed together with the suture
increasing the leaflet stresses slightly, whereas the
serving mainly to limit the excursion of either leaflet
MVA and transvalvular pressure are not significantly
into the atrium.’
affected by this change [9]. The same behaviour is
In Table 6, the load peak values transmitted to
to be expected with the clip, but more analyses are
the link, representing the maximum and minimum
necessary to ascertain it, and also to study more
forces during the cardiac cycle, are shown.
severe pure annulus dilation compared with that
A symmetrical push–pull alternate load cycle with
assumed in the present work.
a maximum value of 0.39 N acts on the clip, while a
Another fundamental aspect is represented by the
pulsating tensile load cycle with a maximum value
clip representation. In the present study the clip
of 0.25 N is applied to the suture. Examining the
was modelled as an independent part whose geome-
problem from the viewpoint of the linking mechan-
try reproduces the actual shape of the device.
ism, these loads constitute external forces due to the
However, in order to investigate the clip–leaflets
leaflets, which add to the internal forces due to the
interaction more thoroughly, a solid (3D) element
link application, including suture pretension and
model instead of present shell element model would
intrinsic forces within the clip due to the clip design.
be more appropriate; this could also help in investi-
Considering the leaflet stresses, very similar maxi-
gating details of the local stress distribution around
mum equivalent alternating Von Mises stress values
the clip arms. As the main interest of the study con-
during the cardiac cycle resulted for both the clip
cerned the cyclic load transmitted to the link due to
and the suture cases, equal to 0.22 MPa and
the diastolic and systolic pressure application, some
0.21 MPa, respectively (see Tables 3 and 4). These
simplifications have been introduced in the clip
stresses are located in the central part of the ante-
model. In particular, at this stage of the research,
rior leaflet on its atrial side, in two symmetrical cen-
the mechanics of the clip itself have not been mod-
tral zones of it. The reason for this similarity lies
elled, so that internal forces, intrinsic to the clip
design itself, have been not taken into account.
Concerning the leaflet material, the choice of a clas-
Table 6 Link loads peaks (N) due to diastolic and sys- sical hyperelastic constitutive law under the hypoth-
tolic pressure application esis of isotropy has been discussed elsewhere [9,
MitraClipÒ Suture 17]; however, it has to be emphasized that the mate-
rial parameters were assumed to be uniform across
Diastole 0.39 0.25 the valve leaflets, and the different mechanical
Systole 20.43 0
behaviours of the anterior and posterior leaflet

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
970 A Avanzini, G Donzella, and L Libretti

tissue were treated as negligible. In reality, in addi- highly stressed zones are located far away from the
tion to the anisotropy related to variation in the link, in a position that is not affected by the link
principal fibre angles, the regional variation of fea- mechanism.
tures such as collagen density or fibre tortuosity Some differences were found regarding the load
may have significant effects on the leaflet material transmitted to the link. In particular, in diastole, the
behaviour, which have not yet been investigated for results showed that the loads are of the same order
the present model. of magnitude, but about 50 per cent higher for the
Leaflets soft tissues also exhibit a viscoelastic clip. A closing load is applied to the clip in systole,
behaviour, but this was not considered in this work, with an absolute peak value comparable to diastole,
owing to the lack of reliable stress relaxation data whereas no load is applied to the suture, owing to
for human heart valve tissues. its physical configuration.
Finally, in this study the focus was on determin- As a consequence, under the effects of cyclic
ing the effects of the clip on both the functional and pressure variation during the cardiac cycle, the clip
structural behaviour of the mitral valve. A key point is subjected to an alternating push–pull external
for the proposed procedure is the use of a Bernoulli- load, whereas the suture is subjected to a pulsating
type equation to couple the applied pressure, flow tensile external load. This consideration neglects the
rate and resulting valve opening. Although the equa- effects of the load induced by the link application
tion coefficients for the specific case of the mitral phase, including internal forces intrinsic to the clip
valve were derived by numerical fluid-dynamic design itself.
simulations, these are in turn based on simplified
assumptions on the three-dimensional behaviour of
the blood flow between the left atrium and ventricle, FUNDING
which in reality is strongly dependent on the atrio- This work was supported by Evalve, Inc. (Abbott
ventricular geometry and kinematics. Vascular, Santa Clara, CA).

Ó Authors 2011
5 CONCLUSIONS

In the present work, a finite element model is pre- REFERENCES


sented for studying the effects of the percutaneous
double-orifice repair technique of the mitral valve 1 Maisano, F., Torracca, L., Oppizzi, M., Stefano, P.
using the MitraClipÒ device. The model, which is L., D’Addario, G., La Canna, G., Zogno, M., and
based on previous work studying the structural Alfieri, O. The edge-to-edge technique: a simplified
method to correct mitral insufficiency. Eur. J. Car-
effects of the open-heart suture-based technique,
diothorac. Surg., 1998, 13(3), 240–246.
allows determination of the MVA and transvalvular 2 Alfieri, O., Maisano, F., De Bonis, M., Stefano, P.
pressure gradient, as well as the structural effects of L., Torracca, L., Oppizzi, M., and La Canna, G. The
clip application. In particular, the load transmitted double-orifice technique in mitral valve repair: a
to the link created with the repair and the stress simple solution for complex problems. Eur. J. Car-
state in the leaflets were evaluated, considering both diothorac. Surg., 2001, 122(4), 674–681.
the diastolic and systolic phases of the cardiac cycle. 3 Kuduvalli, M., Ghotkar, S. V., Grayson, A. D., and
Details of the complex interaction between the Fabri, B. M. Edge-to-edge technique for mitral
valve repair: medium term results with echocardio-
clip and the leaflet areas were also taken into account
graphic follow-up. Ann. Thorac. Surg., 2006, 82(4),
by modelling the clip as an independent part with 1356–1361.
dimensions corresponding to its actual size. 4 Fucci, C., Cicco, G. D., Chiari, E., Nardi, M.,
For comparison purposes, the same valve model Faggiano, P., Procopio, R., Coletti, G.,
was also used to simulate the effects of a suture Rambaldini, M., and Lorusso, R. Edge-to-edge
having a width on the leaflet edge similar to that of mitral valve repair for isolated prolapse of the ante-
the clip. rior leaflet caused by degenerative disease. J. Cardi-
Considering the MVA and the pressure gradient ovasc. Med., 2007, 8(5), 354–358.
5 Tops, L. F., Kapadia, S. R., Tuzcu, E. M.,
across the valve, the results for the MitraClipÒ and
Vahanian, A., Alfieri, O., Webb, J. G., and Bax, J. J.
the standard suture are quite similar. Percutaneous valve procedures: an update. Curr.
The leaflet stresses resulting from pressure varia- Probl. Cardiol., 2008, 33(8), 417–457.
tions during the cardiac cycle, which continuously 6 Davidson, M. J. and Cohn, L. H. Surgeons’ per-
affect the mitral valve during working, are also very spective on percutaneous valve repair. Coron.
similar for the two repair techniques, as the most Artery Dis., 2009, 20(3), 192–198.

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015
Effects of percutaneous edge-to-edge double-orifice repair under cardiac cycle in comparison with suture repair 971

7 Kar, S., Feldman, T., and St Goar, F. Mitral valve 18 Redaelli, A., Guadagni, G., Fumero, R.,
repair using the MitraClip: from concept to reality. Maisano, F., and Alfieri, O. A computational study
Cardiac Interventions Today, 2008, May/June, 39–45. of the hemodynamics after ‘edge-to-edge’ mitral
8 Feldman, T., Kar, S., Rinaldi, M., Fail, P., valve repair. J. Biomech. Eng., 2001, 123(6), 565–570.
Hermiller, J., Smalling, R., Whitlow, P. L., 19 Muntinga, H. J., van den Berg, F., Knol, H. R.,
Gray, W., Low, R., Herrmann, H. C., Lim, S., Niemeyer, M. G., Blanksma, P. K., Louwes, H.,
Foster, E., and Glower, D. Percutaneous mitral and van der Wall, E. E. Normal values and repro-
valve repair with the MitraClip system: safety and ducibility of left ventricular filling parameters by
midterm durability in the initial EVEREST (Endo- radionuclide angiography. International Journal of
vascular Valve Edge-to-Edge REpair Study) cohort. Cardiac Imaging, 1997, 13(2), 165–171.
J. Am. Coll. Cardiol., 2009, 54(8), 686–694. 20 Herrmann, H. C., Rohatgi, S., Wasserman, H. S.,
9 Dal Pan, F., Donzella, G., Fucci, C., and Block, P., Gray, W., Hamilton, A., Zunamon, A.,
Schreiber, M. Structural effects of an innovative Homma, S., Di Tullio, M. R., Kraybill, K.,
surgical technique to repair heart valve defects. J. Merlino, J., Martin, R., Rodriguez, L., Stewart, W.
Biomech., 2005, 38(12), 2460–2471. J., Whitlow, P., Wiegers, S. E., Silvestry, F. E.,
10 Votta, E., Maisano, F., Soncini, M., Redaelli, A., Foster, E., and Feldman, T. Mitral valve hemody-
Montevecchi, F. M., and Alfieri, O. 3-D computa- namic effects of percutaneous edge-to-edge repair
tional analysis of the stress distribution on the leaf- with the MitraClip TM device for mitral regurgita-
lets after edge-to-edge repair of mitral regurgitation. tion. Catheter. Cardiovasc. Interv., 2006, 68(6), 821–
J. Heart Valve Dis., 2002, 11(6), 810–822. 828.
11 Nielsen, S. L., Timek, T. A., Lai, D. T., 21 Frapier, J. M., Sportouch, C., Rauzy, V.,
Daughters, G. T., Liang, D., Hasenkam, J. M., Rouviere, P., Cade, S., Demaria, R. G., Davy, J. M.,
Ingels, N. B., and Miller, D. C. Edge-to-edge mitral and Albat, B. Mitral valve repair by Alfieri’s tech-
nique does not limit exercise tolerance more than
repair: tension on the approximating suture and
Carpentier’s correction. Eur. J. Cardiothorac. Surg.,
leaflet deformation during acute ischemic mitral
2006, 29(6), 1020–1025.
regurgitation in the ovine heart. Circulation, 2001,
22 Umana, J. P., Salehizadeh, B., DeRose, J. J. Jr.,
104(12 Suppl. 1), I-29–I-35.
Nahar, T., Lotvin, A., Homma, S., and Oz, M. C.
12 Timek, T. A., Nielsen, S. L., Lai, D. T., Tibayan, F.,
‘Bow tie’ mitral valve repair: an adjuvant technique
Liang, D., Daughters, G. T., Beineke, P.,
for ischemic mitral regurgitation. Ann. Thorac.
Hastie, T., Ingels, N. B. Jr, and Miller, D. C. Mitral Surg., 1998, 66(5), 1640–1646.
annular size predicts Alfieri stitch tension in mitral
edge-to-edge repair. J. Heart Valve Dis., 2004, 13
(2), 165–173. APPENDIX 1
13 Jimenez, J. H., Forbess, J., Croft, L. R., Small, L., Notation
He, Z., and Yoganathan, A. P. Effects of annular
size, transmitral pressure, and mitral flow rate on A valve commissure-to-commissure
the edge-to-edge repair: an in vitro study. Ann. diameter
Thorac. Surg., 2006, 82(4), 1362–1368. B valve anterior-posterior diameter
14 Kunzelman, K. S., Reimunk, M. S., and material parameters for hyperelastic
Ci0
Cochran, R. P. Flexible versus rigid ring annulo-
constitutive law
plasty for mitral valve annular dilation: a finite ele-
ment model. J. Heart Dis., 1998, 7, 108–116. Di0 material parameters for hyperelastic
15 Kunzelman, K. S., Cochran, R. P., Verrier, E. D., constitutive law
and Eberhardt, R. C. Anatomic basis for mitral valve I1 first deviatoric strain invariant
modeling. J. Heart Valve Dis., 1994, 3, 491–496. Jel elastic volume ratio
16 Maisano, F., Redaelli, A., Soncini, M., Votta, E., MVA mitral valve area
Arcobasso, L, and Alfieri, O. An annular prosthesis
N order of polynomial strain-energy
for the treatment of functional mitral regurgitation:
finite element model analysis of a dog bone-
function
shaped ring prosthesis. Ann. Thorac. Surg., 2005, Dp transvalvular pressure gradient
79(4), 1268–1275. Q blood flow rate across valve
17 Avanzini, A. A computational procedure for predic- U strain-energy potential
tion of structural effects of edge-to-edge repair on sia alternate principal stress components
mitral valve. J. Biomech. Eng., 2008, 130(3), 031015- saMises equivalent alternating Von Mises stress
1:10.

Proc. IMechE Vol. 225 Part H: J. Engineering in Medicine


Downloaded from pih.sagepub.com at The University of Iowa Libraries on June 19, 2015

You might also like