FDR MS-3500 Service Manual

SPEC 026-201-05(1)E Spec-11

2. Dimensions and Weight of the


Machine
 External Dimensions and Weight
Width Depth Height Weight
Exposure stand Approx. Approx. 1270 mm Approx. 1974 mm Approx.
624 mm (50 in.) 2285 mm 370 kg
(24.57 in.) (Base plate approx. (77.72 in. to (815.7 lb.)
787 mm (30.98 in.)) 89.96)
Control cabinet Approx. Approx. 503 mm Approx. 530 mm Approx.
205 mm (19.8 in.) (20.87 in.) 20 kg
(8.07 in.) (44.09 lb)
Generator Approx. Approx. 315 mm Approx. 825 mm Approx.
(X-ray high voltage unit) 445 mm (12.4 in.) (32.48 in.) 70 kg
(17.52 in.) (154.3 lb)
Positioner control cabinet Approx. Approx. 350 mm Approx. 473 mm Approx.
160 mm (13.78 in.) (18.62 in.) 17 kg
(6.3 in.) (37.48 lb)
Hardware specification, dimensions and weight are subject to change without notice.
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-12

2.1 External View

Exposure stand
Install the controller unit and generator within 5 m (196.9 in.)
of the exposure stand.
Install the positioner control cabinet within 3 m (118.1 in.) of
the exposure stand.

Generator
(X-ray high voltage unit)
Controller unit
Positioner control cabinet
(Accessory sold separately)

DRMS1001_0_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-13

2.2 Mammographic Stand

Units: mm (in.)

DRMS1001_01_1_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-14

2.3 Control Unit

Units: mm (in.)
DRMS1001_221_01_E ai

2.4 Control Pad

Units: mm (in.)
DRMS1001_222_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-15

2.5 Foot Switch

Units: mm (in.)
DRMS1001_223_E.ai

2.6 Generator (X-ray High Voltage Device)


825(32.48)

Units: mm (in.)
DRMS1001_224_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-16

2.7 AWS: PC and Keyboard

Units: mm (in.)
DRMS1001_225_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-17

2.8 HUB Unit


The external view and dimensions are just examples, as this is
general-purpose electric equipment/component.

Units: mm (in.)
DRMS1001_226_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-18

2.9 24x30 Compression Plate (High)

183(7.20)

341(13.43)

Units: mm (in.)
Permissible range:
±10%
DRMS1001_227_H_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-19

2.10 Operation Table


Characteristics
- Can store the AWS PC.

Units: mm (in.)
DRMS1001_228_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-20

 Protective Shield (Optional)


Characteristics
࡮Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room.
࡮Can store the AWS PC.

600(23.6)

Units: mm (in.)
DRMS1001_229_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-21

 Protective Plate (Optional), Storage Rack (Accessory Sold


Separately)
Characteristics
࡮Comes with protective glass (equivalent to 0.5 mm Pb), and can be used in the study room.
࡮Can store the AWS PC.
࡮Can store each compression plate or accessories during exposure.

600(23.6)

Units: mm (in.)
DRMS1001_230_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-22

 Protective Glass Cover for Additional Height (0.5 mm Pb Equivalent)


(Optional) (Outside Japan)

310(12.2)
609(23.98) 23(0.91)

2140(84.25)
1900(74.8)

Units: mm (in.)
DRMS1001_231_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-23

2.11 FDR MS-3500 Options


(1) Compression plate
(1)-1 24x30 compression plate (Low)

341(13.4)

183(7.20)

(1)-2 24x30 compression plate (Shift)

341(13.4)

183(7.20)

Units: mm (in.)
DRMS1001_232_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-24

(1)-3 24x30 compression plate (SS)

314(13.4)

183(7.20)

(1)-4 24x30 compression plate (Flex)

341(13.4) 313(12.3)

Units: mm (in.)
DRMS1001_233_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-25

(1)-5 24x30H compression plate (FS)

341(13.4) 313(12.3)

215(8.46)

(1)-6 18x24 compression plate (Low)

Units: mm (in.)
DRMS1001_234_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-26

(1)-7 18x24 compression plate (High)

276(10.9)
312(12.3)

183(7.20)

(1)-8 18x24 compression plate (Small)

276(10.9)

311(12.2)

183(7.20)

Units: mm (in.)
DRMS1001_236_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-27

(1)-9 18x24 compression plate (Flex)

276(10.9)

(1)-10 18x24L compression plate (FS)

276(10.9)
310(12.2)

215(8.46)

Units: mm (in.)
DRMS1001_236_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-28

(1)-11 Rectangular spot compression plate

123(4.84)

183(7.20)

(1)-12 Compression plate for axilla radiography

264 (10.4) 313 (12.3)

183 (7.20)

Units: mm (in.)
DRMS1001_237_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-29

(1)-13 Compression plate for magnifications

251(9.88) 311(12.2)

126(4.96)

(1)-14 Rectangle spot magnification compression plate

311(12.2)

123(4.84)

Units: mm (in.)
DRMS1001_238_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-30

(1)-15 2D biopsy compression plate

183(7.20)

Units: mm (in.)
DRMS1001_239_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-31

(2) Magnification table

Magnification table (when stored)

Units: mm (in.)
DRMS1001_240_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-32

(3) Remote exposure hand switch

(4) Remote exposure foot switch

(5) Shielding plate

Units: mm (in.)
DRMS1001_241_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-33

(6) Under-armrest

Installed

Units: mm (in.)
DRMS1001_242_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-34

(7) Compression plate storage holder

(2.25)
57
* Can be attached to the wall surface in either horizontal or vertical orientation.

Units: mm (in.)
DRMS1001_243_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-35

2.12 Accessories Sold Separately


(1) Biopsy Positioner
(1)-1 Positioner

(1)-2 Positioner Control Cabinet

Units: mm (in.)
DRMS1001_246_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-36

(1) -3 Control panel

(1)-4 Biopsy compression plate

Units: mm (in.)
DRMS1001_247_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-37

(1)-5 Needle holder C

(1)-6 Needle holder M

Units: mm (in.)
DRMS1001_248_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-38

(1)-7 Needle holder Y

(1)-8 Needle holder M8

Units: mm (in.)
DRMS1001_249_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-39

(1)-9 Needle holder E

(1)-10 Biopsy spacer

Units: mm (in.)
DRMS1001_250_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-40

(1)-11 Needle guide cover

Units: mm (in.)
DRMS1001_251_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-41

(2) Lateral approach kit


(2)-1 Lateral adapter

(2)-2 Lateral biopsy compression plate

Units: mm (in.)
DRMS1001_252_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-42

(3) Decoration label

594㸦23.39㸧

1670
(65.75)

Decoration label
attachment surface

Units: mm (in.)
DRMS1001_254_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-43

3. Installation Space and Servicing


Space
■ Mammographic Stand/Control Unit
● Front/rear direction
R 7.6
70 )
(2

0
R .87
20 )
(7

Units: mm (in.)
DRMS1002_1_01_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-44

● Height direction

C arm variable
range
(10.43)
265

Units: mm (in.)
DRMS1003_1_01_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-45

4. Range of Movement
■ Mammographic Stand
REFERENCE
Although the machine is set in the factory so that the C arm stops at 2,350 mm when it is
moved up to its uppermost position, the uppermost position of the C arm can be adjusted
within a range of 2,050 to 2,350 mm by repositioning the C arm elevation upper-limit
switch.
The corresponding adjustment range of the uppermost position of the FPD upper surface
is 1,200 to 1,500 mm.

Arm uppermost adjustable range


1503


350

C arm adjustable range

+90° -90°
1935~2285(factory default)

460~810

+180° -180°
1150~1500

+120° -120°
974

690

224

Lowermost position at which c arm can swivel 180° Lowermost position Uppermost position
of FPD upper surface of FPD upper surface

Unit:mm
DRMS1060_1 ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-46
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-47

5. Anchoring Locations

5.1 Mammographic Stand

Th
rou
gh
-ho
le

Units: mm
DRMS1003_21_01_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-48

5.2 Generator (X-ray High Voltage Unit)


300(11.81) recommended
(Min. 200(7.87))

Units: mm (in.)
DRMS1003_22_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-49

5.3 Control Unit

Units: mm (in.)
DRMS1003_21__01_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-50

5.4 Operation Desk and Storage Rack

Units: mm (in.)
DRMS1003_23_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-51

6. System Configuration Example


CAUTION
Do not use a multi-socket/extension cord for the quality and safety of the product.

 NOTES 
- The FDR-3500DRLH can be installed in an exposure chamber (patient
environment).
- Appropriately furnish the exposure chamber and the operation chamber to
visually and acoustically check the patient conditions.

6.1 Installing only the Mammographic Stand in


the Exposure Chamber
Control Pad Cable: approx. 9m

Do not place peripheral devices in Power Supply, I/F cable: approx. 4m


the safety guaranteed area.

Do not place obstacles on the flow


line of a patient.

2503 or more

DRMS1005_1_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-52

6.2 Installing Everything Except the Operation


Desk in the Exposure Chamber
Power and I/F cable length
: Approx. 11m Length of the cable for the control pad : Approx. 10m

Power and I/F cable length Power and I/F cable length
: Approx. 5m : Approx. 5m

No obstacle is allowed to be installed No peripheral devices are allowed to be installed


on the flow line of a patient. in the safety guaranteed area.

3090 or more

DRMS1005_2_en.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-53

6.3 Installing the Operation Desk in the Exposure


Chamber
WARNING
When the operation desk is to be installed in the exposure chamber, put the
optional protective board for radiation protection. Place the operation desk at a
distance of 1,500 mm or more from the X-ray focus.

Do not place peripheral devices in


the safety guaranteed area

Do not place obstacles on the flow


line of a patient.

Focus has to be on the same vertical


line of center of the protective board Unit:mm
DRMS1008_1_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-54
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-55

7. Machine Specifications
Specifications of the FDR MS-3500 including processing capabilities are described
below.
They are subject to change without prior notice.

7.1 Exposure Unit Specifications


Power Requirements
Input voltage Input current (standby) Input current (normal) Input current (peak) Frequency
AC200V (Single phase) 2.4A
AC208V (Single phase) 2.3A
AC220V (Single phase) 2.18A 16.5A 90A 50 Hz/60 Hz
AC230V (Single phase) 2.09A
AC240V (Single phase) 2.0A

Requirements for Maximum Allowable Impedance between Substation Equipment and Circuit Breaker
Maximum allowable Maximum allowable
Input voltage Input voltage
impedance impedance
AC200V (Single phase) 0.143Ω AC230V (Single phase) 0.165Ω
AC208V (Single phase) 0.150Ω AC240V (Single phase) 0.172Ω
AC220V (Single phase) 0.158Ω

Requirements for Circuit Breaker Connected to Insulation Distance between L and N


Equipment Creepage 3mm (0.11 in.) or longer
Number of electrodes 2 Clearance 1.6mm (0.06 in.) or longer
Rated insulation voltage Ui (V) 600 (V) or more
Rated current (A) 50 (A) (AC240V)
* Install a breaker upstream of the device that does not trip when

or 130A flows for 10 seconds, and that trips when 1500A flows
continuously for 0.1 seconds.

Operating Characteristic Curves Temperature Characteristic Curve


temperature
Reference ambient
time variation (%)

Percentage of

Operating characteristic curves


operation

Maximum (60A~63A)
Reference ambient temperature ( 0°C)
Maximum (40A~50A) Reference ambient temperature ( 5°C)
Ambient temperature (°C)
Operation time

Minimum

Total shutdown
time
Maximum

Delayed trip Instantaneous trip

Current (Percentage of rated current)


* Recommended product : NF63-CV (50A) (Mitsubishi)

DRMS1008_2_E ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-56

Environmental Requirements
(1) Operating conditions
Temperature : 20°C to 30°C
Humidity : 30% RH to 75% RH (No dew condensation)
Air pressure : 750 hPa to 1060 hPa
(2) Non-operating conditions
Temperature : 15°C to 35°C
Humidity : 10% RH to 80% RH (No dew condensation)
Air pressure : 750 hPa to 1060 hPa
* This equipment cannot be used at an altitude of over 3000 m (9842 ft).

Attenuation Equivalent
• Collimator : At least 0.4 mmAl (to the conditions of 49kV radiation)
• Exposure table : 0.1 mmAl (to the conditions of 30kV radiation)
: 0.21 mmAl (to the conditions of 30kV radiation)
• Biopsy protection cover : 0.18 mmAl (to the conditions of 30kV radiation)

Operating Mode
Continuous operation with intermittent loading
Continuous Operation
60 shots/hour (28kV, W/Rh, 71mA, 4 consecutive hours)
40 shots/hour (28kV, W/Rh, 71mA, 8 consecutive hours)
However, maintain the interval shown in the “Minimum exposure interval for energy input to X-ray
tube” graph at least.

SID (Focal Spot to Image Receptor Distance)


65 cm (25.59 in.)

Characteristics of X-ray Exposure


Tube voltage range 22kV to 49kV (at 1kV increments)
Tube voltage accuracy ±1kV
mAs setting range Large focal spot: 2mAs to 600mAs
Small focal spot: 2mAs to 300mAs
Tomosynthesis: 16mAs to 90mAs
(Depending on load conditions such as tube voltage)
mAs setting accuracy ±10%+0.2mAs
Tube current Large focal spot: 20mA to 200mA
Small focal spot: 20mA to 61mA
Nominal maximum output power 7kW (35kV x 200mA) * Load time: 40ms
Nominal maximum output power (at 30kV, 40ms) 5kW (168mA)
Maximum output power for making exposures at 2.8kW (28kV x 100mA)
1-minute intervals Load time 1s
Total allowable mAs per hour (at 35kV) total 7560mAs
Anode material 10%Re/Tungsten face on molly alloy
Large focal spot: 0.3
X-ray tube focal spot size Small focal spot: 0.1
* Per IEC 60336, determined with the slit camera method.
0.63mm Be
W/50μmRh, W/700μmAl

DRMS1008_3_E.ai
FDR MS-3500 Service Manual
SPEC 026-201-05(1)E Spec-57

Cooling curve of the X-ray system (tube) Minimum exposure interval for energy input to
----- Tube [°C]
----- Ambient temperature
X-ray system (tube)

Min. exposure interval [sec]


Temperature [°C)]

■ 2D imaging
● Tomosynthesis imaging

Input for exposure [kV x mA x sec]

Time [min]
Condition: 25°C, large focal spot

X-ray Conversion Method


Direct conversion

X-ray Conversion Material


a-Se

236.4 mm (9.31 in.) (vertical) x 296.4 mm (11.7 in.) (horizontal)

Pixel Resolution
50 x 50 μm/pix

Number of Output Pixels


4728 (vertical) x 5928 (horizontal) = 28,027,584 pixels

Output gradation
(1) FDR-3000DRL
16bit
(2) FDR-3000AWS

DRMS1008_5_E.ai