Contents .................................................................................................................... 2
Update information ................................................................................................... 3
General Information.................................................................................................. 4
Changing the pneumatic spring .............................................................................. 5
Changing and adjusting the inclination switch...................................................... 7
Changing the illumination potentiometer ............................................................... 8
Changing the suspension arm .............................................................................. 10
Removing the illumination unit.............................................................................. 11
Changing the stand casters................................................................................... 12
Checking and adjusting the maximum lamp voltage .......................................... 13
Setting the inclination switch ................................................................................ 13
Issue Comments
25.09.2003 Issue 1
- Cement 44
- GF 100
Removal:
• Remove OPMI from the suspension arm.
• Set the balancing screw to maximum 2 1
weight (turn cw).
• Remove the locking screw and the top
cover from the suspension arm.
• Remove protection flap (1 Fig. 1).
• Loosen setscrew (2 Fig. 1) and push out
pin (3 Fig. 1).
3
Note:
Hold the two halves of the suspension arm
together by hand while pushing out pin
(3 Fig. 1).
• Allow the halves of the arm to slowly move Fig. 1
apart. If necessary, remove the cable
to the illumination potentiometer from
the clip on the upper half of the arm.
• Remove the cover plate from below the
2
balancing screw.
• Remove support plate (1 Fig. 2) and
unscrew balancing screw (2 Fig. 2).
• Remove the pneumatic spring.
Fig. 2
Installation:
• Insert the pneumatic spring in the
suspension arm.
Fig. 3
Fig. 5
Fig. 6
Removal:
• Remove OPMI from the suspension arm. 3 2 1
• Set the balancing screw to minimum weight
(turn ccw).
• Remove the locking screw and the top
cover from the suspension arm.
• Remove protection flap (1 Fig. 7).
• Loosen setscrew (2 Fig. 7) and push out pin
(3 Fig. 7).
Note:
Hold the two halves of the suspension arm
together by hand while pushing out pin
(3 Fig. 7)
Fig. 7
• Allow the halves of the arm to slowly move
apart. If necessary, remove the cable to the illumination potentiometer from the clip
on the upper half of the arm.
Fig. 8
Installation:
• Insert the microswitch and make soldering
connections (cover connections with shrink
tubing) or route cable up to the illumination 1
unit. 2
• Slightly press the halves of the arm
together and secure the cable from the
illumination potentiometer in the clip on the
upper half of the arm, if necessary.
• Insert pin (1 Fig. 9) in lock (2 Fig. 9).
Bild 21
Fig. 9
Bild 22
Fig. 10
Removal:
• Remove OPMI from the suspension arm.
• Set the balancing screw to minimum weight (turn ccw).
• Remove the locking screw and the top cover from the suspension arm.
• Remove cable conduit (4 Fig. 10).
• Remove protection flap (3 Fig. 10).
• Loosen setscrew (2 Fig. 10) and push out pin (1 Fig. 10).
Note:
Hold the two halves of the suspension arm together by hand.
• Allow the halves of the arm to slowly move apart. If necessary, remove the cable leading to
the illumination potentiometer from the clip mounted on the upper half of the arm.
4 3 1
Bild 23
Fig. 11
Fig. 12
Fig. 13
Installation:
• Thread the cable through the suspension
arm and secure it position using cable ties.
• Mount the potentiometer. 1
• Re-attach the knob and the cap. 2
• Slightly press the halves of the arm together
and secure the cable from the illumination
potentiometer in the clip on the upper half of
the arm, if necessary.
• Insert pin (1 Fig. 14) in lock (2 Fig. 14).
Fig. 14
Removal:
• Remove OPMI from the suspension arm.
• Remove locking screw (1 Fig. 16), cover 4
(2 Fig. 16) and cable conduit (3 Fig. 16). 1
• Remove locking screws (4 Fig. 16) and
cover (5 Fig. 16).
• Remove cover (6 Fig. 16).
• Remove cover (7 Fig. 16).
7 5 6 2 3
Fig. 16
1 2
Fig. 17
Fig. 19
Mounting:
• Insert the release button for the lamp
module in the housing and re-attach the 2
halves of the housing, making sure that the
release button is correctly positioned. Fig. 20
• Align the halves of the housing with each
other and screw them together using screws
(1/2 Fig. 20).
• Insert the lamp module and check that it moves smoothly. If necessary, loosen the halves of
the housing and re-align them.
• Connect the light guide.
Mounting:
Proceed in the reverse order, making sure that
you correctly insert the locking washers.
Caution:
Do not exceed 12.5V during the adjusting procedure.