BUREAU VERITAS (K) LIMITED

CALIBRATION CERTIFICATE
REQUESTED BY:

ADDRESS:

EQUIPMENT: CHEST FREEZER SINGLE DOOR (ICE CREAM)


MODEL: WDD 5005
LOCATION: STORES
SERIAL NO: 016062500001
MANUFACTURER UGUR
LAB NO:
STICKER NO. BVL.IND.4564
CALIBRATED ON: 06.11.2019
CERTIFICATE NO:

CALIBRATION LOCATION: BV LABORATORY CLIENT SITE 

STANDARD EQUIPMENT USED


Standard reference by Lutron Model: TM 947USB + set of thermocouple K
CALIBRATION PROCEDURES
The oven was calibrated in accordance with BV-TP-TEMP-003 standard calibration
procedures for the calibration ovens, fridges, incubators and freezers
Ref material: JCGM 100:2008 Evaluation of measurement data — Guide to the expression of
uncertainty in measurement
TRACEABILITY
The freezer has been calibrated against reference standard mentioned above whose
Calibration is traceable to national standards.

Environmental conditions: Calibration was carried out under controlled environmental


conditions as indicated below:
Temperature: 22.0°C ±1°C
Relative Humidity: 55 % ±10%

RESULTS VALIDATION: The validity of this certificate expires on NOVEMBER 2020.

Calibrated by: J. JUMA 08.11.2019


Checked by: M. GAKUBU 11.11.2019
BUREAU VERITAS (K) LIMITED

CALIBRATION DATA

Time (min) Controller Temperature Corner (˚C) Center (˚C)


0 -18.7 -19.5 -19.5
5 -18.7 -19.5 -19.5
10 -18.7 -19.6 -19.6
15 -18.6 -19.6 -19.6
20 -18.6 -19.7 -19.7
25 -18.6 -19.7 -19.7
30 -18.6 -19.6 -19.6
Uncertainty computation
Parameter Center(˚C) Corner(˚C)
Average temperature -19.60 -20.29
Temperature correction -0.96 -1.64
Temperature fluctuation 0.20 0.50
Temperature variation -0.69
Expanded uncertainty ±0.86

UNCERTAINTY STATEMENT:
The reported expanded uncertainty is stated as expanded uncertainty of the measurement
multiplied by a coverage factor K=2, providing a confidence level of approximately 95%.

REMARKS:
The results were correct during the time of calibration, and will remain valid depending
on care and handling exercised during equipment use.
Issuance of this certificate indicates that the equipment calibrated complies with specified
requirements. When the equipment does not comply with requirement, no certificate shall
be issued instead a calibration report will be issued.