Professional Documents
Culture Documents
Argon Flows PMP UV-Pro From SN 48P0163 & 56Q0001
Argon Flows PMP UV-Pro From SN 48P0163 & 56Q0001
Required equipment:
This instruction may only be carried out by persons trained and authorised in maintaining
and repairing the device by Oxford Instruments Analytical GmbH.
Prolog:
We recommend to check and to adjust the argon flows on the UV-Pro at each maintenance.
Moreover, this procedure should be performed if any changes have been done on the argon block
of the UV-Pro e.g. exchange of a solenoid valve.
Note! This argon flows are only valid for UV-Pros with serial numbers beginning from
48P0163 & 56Q0001.
Adjusting the Argon Flows
Fig. 1-1
Use a slit screw driver to adjust the flows on the corresponding valve (see Fig. 1-3)
Scale Division
Valve TestControls selection Flow Flowmeter
(only OI flowmeters)
20 l/h V100-140.12
1V6 Argon on 10 mm
50 sl/h (steel ball)
140 l/h V100-140.12
1V7 Argon on + Argon high 115 mm
380 sl/h (steel ball)
-2-
Adjusting the Argon Flows
Adjusting UV Flows
Connect the corresponding flowmeter (see Tab. 1-2) to the tube which is
connected to the mini optic chamber (see Fig. 1-2)
Use a slit screw driver to adjust the flows on the corresponding valve
(see Fig. 1-3)
Fig. 1-2
Scale Division
Valve TestControls selection Flow Flowmeter
(only OI flowmeters)
80 l/h V100-140.12
1V5 UV flush 70 mm
220 sl/h (steel ball)
5 l/h V100-140.09
1V4 nothing 30 mm
15 sl/h (plastic ball)
Fig. 1-3
-3-