You are on page 1of 8

CXDI Image Processing Software

LANMIX MLT (M)


Operation Manual Version 1.0

Before using the instrument, be sure to read this manual thoroughly.


Also, read the X-ray digital camera manual.
Store this operation manual in a safe place for future reference as necessary.
PLEASE NOTE

1. In no event will Canon be liable for direct or indirect consequential damage arising out of the
use of this product. Canon will not be liable for loss of image data due to any reason.
2. Please contact your Canon representative or distributor before installing this product.
3. Canon reserves the right to change the specifications, configuration and appearance of the
product without prior notice.

© CANON INC. 2003

All rights reserved.


Under copyright laws, this manual may not be copied, in whole or in part, without the written consent of
Canon.
Contents

1. Overview .........................................................................................................................................................1

2. Laplacian Edge Enhancement Processing.......................................................................................................2


1. Overview

The Canon CXDI image processing software LANMIX MLT (M) (called “LANMIX MLT (M)” below) is
used to add greater functionality to the image processing functions of the Canon CXDI series X-ray digital
cameras (called “CXDI” below).
Installing LANMIX MLT (M) on the control PC adds the following image processing function.
Laplacian Edge Enhancement Processing
The radiography image is made sharper based on the settings of two parameters.
These parameters are dynamic range adjustment and degree of enhancement.
In addition, at the time of installation, the serviceman can make settings so that the tabs for setting Laplacian
Edge Enhancement Processing are not displayed, and the parameters are fixed at the recommended values or
setting values.

This instruction manual describes the operating procedures for the above-mentioned image processing
function.
For other operating procedures, refer to the CXDI Series Operation Manual.

Note: LANMIX MLT (M) Version 1.0 supports CXDI System Software Version 5.1.
Other CXDI versions are not supported.

–1–
2. Laplacian Edge Enhancement Processing

Note: If the degree of enhancement of Laplacian Edge Enhancement Processing is set to


an excessive value, artifacts can occur, and the effective data of the image can be
damaged.
Refer to the image, and if the processing is applied too strongly, either set the
degree of enhancement to a smaller value or disable the setting.

Laplacian Edge Enhancement Processing makes the radiography image sharper by enhancing the edges
(contour) of radiography images.
You can check the effects of parameter changes on the QA mode screen.

Laplacian Edge Enhancement Processing Tab

(1) Enable the laplacian edge enhancement parameter.


Each time the “LAP. ENHANCEMENT” button is touched, the highlight (white text with a colored
background) display and normal display are switched. The parameter is enabled in the highlight display.

Laplacian Edge Enhancement Processing Setting Screen

–2–
2. Laplacian Edge Enhancement Processing

(2) Set the DYNMIC RANGE


Touch the or buttons to set the DYNMIC RANGE in the 1 to 20 range.
The displayed value on the left is the saved value, and the value on the right is the current value.
This value becomes the dynamic range of radiography images.
Larger setting values result in wider dynamic range, and smaller values result in narrower dynamic range.

(3) Set the value for EFFECT.


Touch the or buttons to set the value for EFFECT in the 1 to 20 range.
The displayed value on the left is the saved value, and the value on the right is the current value.
This value is the degree of enhancement for the edges of radiography image.
Larger setting values result in greater enhancement, and smaller values result in lower enhancement.

(4) To save the modified parameters, touch the “SAVE PARAM” button, and to return the modified
parameters to their original values, touch the “RESET” button. To exit the screen, touch the “OK”
button.
* “SAVE PARAM” is displayed when is pressed.

–3–
CANON INC.
20-2 Kiyohara-Kogyodanchi, Utsunomiyashi, Tochigiken, Japan
Telephone: (81)-28-667-5711

CANON MEDICAL SYSTEMS


15955 Alton Parkway, Irvine, CA 92618-3616, U.S.A.
Telephone: (1)-949-753-4160

CANON EUROPA N.V.


Bovenkerkerweg 59-61, 1185 XB Amstelveen, The Netherlands
Telephone: (31)-20-545-8926

PUB. L-IE-4087 (1.0) 0203P0.01 Printed in Japan