You are on page 1of 70

Manual No.

: M511-E020
Revision :C

X-RAY RADIOGRAPHY STAND


BR-120
OPERATION MANUAL

Read this manual carefully to ensure appropriate use,


and keep it with the unit.
About the symbols appearing in this operation manual
Throughout the text in this manual, warnings and other information essential when using this unit,
such as cautionary or prohibited items, appear classified as per the following:

Mark Description

States a direct danger that may cause death or serious injury if


it is not avoided.

States an indirect or potential danger that may cause death or


serious injury if it is not avoided.

States a danger that may cause slight or medium injury or


may cause damage in equipment or fire if it is not avoided.

States the information which helps to use the system


correctly.

Provides useful information about operation of this system.


Please read the description when required.

Indicates reference sections and pages.

Original version is approved in English.

BR-120 OPERATION MANUAL


0Preface

This operation manual must be carefully read prior to using the


"X-RAY RADIOGRAPHY STAND BR-120" (after referred to
as "device") and thoroughly followed for proper use.

This manual should be kept available for future reference.

Copyright

Shimadzu Corporation reserves the copyright of this manual


and the whole or a part of the contents may not be copied or
reproduced without the company's permission.

The contents of this manual may be changed for improvement


without notice.

© 2004-2006 Shimadzu Corporation. All rights reserved.


Preface

Operating Precautions

"Operating Precaution (for Both the Safety and the Prevention of


Danger) in the Use of Electric Medical Equipment"

1. Only an experienced technician should operate the equipment.

2. When installing the equipment, pay attention to the following items:

(1) Do not install it near water faucet or similar equipment.

(2) Install it away from potential sources of problems such as abnormal pressure, temperature or
humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.

(3) During transportation and operation of the equipment, avoid tilting, vibration and sharp impact
against it.

(4) Keep the equipment away from the areas where chemicals or gases are stored.

(5) Use only the correct electrical power source with matching frequency, voltage and current (or
wattage).

(6) Check the conditions of the battery power source (power and polarity) before operating the
equipment.

(7) Properly ground the equipment.

3. Before operating the equipment, pay attention to the following items:

(1) Check the conditions of switch contacts, polarity, dial settings, and meters, and make sure the
equipment performs correctly.

(2) Confirm that the ground is connected properly.

(3) Check all wiring for proper and correct connections.

(4) Pay attention when using more than one unit at a time, because it may lead to an incorrect
diagnosis and cause danger.

(5) Check the condition of the external electric circuit, which will be directly connected to a patient.

(6) Check the condition of the battery power source.

Introduction-2 BR-120 Operation Manual


Preface

4. While operating the equipment, pay attention to the following items:

(1) Do not over-exceed time or the amount of equipment use needed for diagnosis or therapy.

(2) Observe the equipment and patient continuously for early detection of problems.

(3) When a problem is detected with the equipment or patient, take proper action to stop the
equipment without harming the patient.

(4) Do not let the equipment touch the patient.

5. After operating the equipment, pay attention to the following items:

(1) Turn off the switches and return the dial to their original before use in the prescribed order. Then,
turn off the main power switch.

(2) Do not pull the power cable forcibly from the outlet.

(3) When storing the equipment, pay attention to the following items:

(i) Keep it away from the water.

(ii) Store it away from the potential causes of problems such as abnormal pressure, temperature
and humidity, draft, direct sunlight, dust, chlorine or sulphur gas.

(iii) During transportation and storage of the equipment, avoid tilting, vibration and sharp impact
against it.

(iv) Store the equipment away from areas where chemicals and gases are stored.

(4) Clean all attachments, cables and contacts, and store them in one place.

(5) Keep the equipment clean to avoid problems during the next use.

6. When the equipment is found to be out of order, do not try to repair it.
Display an appropriate sign to indicate that the equipment is out of order, and
contact Shimadzu Service Representative for repair.

7. Do not modify any part of the equipment.

8. Preventive maintenance

(1) The equipment and its parts should be periodically checked.

(2) If the equipment has not been in operation for an extended period of time, test it prior to actual
operation to make sure it works correctly and safely.

9. Concerning other items, operate properly according to the operation manual.

BR-120 Operation Manual Introduction-3


Preface

For Assurance of Safety during Operation

The responsibility for management of use and maintenance of


medical equipment lies in a user.
This device is restricted to be used by, or on the order of, a diagnostic
radiology technician or a person with a certificate indicating equal proficiency.

Never remodel the equipment.


In general, almost all of the modifications are strictly prohibited by the legal
stipulation of the country where the equipment is installed. Please contact the
Shimadzu Service Representative if any modification is necessary.

Perform periodic maintenace.


Regular maintenance and inspections must be arranged to retain the full
performance of the device. See Chapter 4 Maintenance for procedures.

Maintenance and inspections inside the equipment must be


performed by the expert.
Maintenance must be assigned to specially trained experts. The service is
available under a contract with Shimadzu Corporation. Contact Shimadzu
Service Representative for repair and maintenance.

Introduction-4 BR-120 Operation Manual


Preface

Guarantee

The guarantee period of this system is once year from the date of purchase.
Our copany is not responsible for the following failures and damages.

1. Failure/damage caused by installation, movement, maintenance and


repair performed by anyone other than our company and companies
specified by our company.

2. Failure/damage in a products of our company caused by a product of


other company which is not delivered by our company.

3. Failure/damage based on maintenance, repair, etc. using a spare part


other than the genuine parts specified by our company.

4. Failure/damage based on a result caused by disobedience to the cautions


and operating procedures described in this operation manual.

5. Failure/damage caused by the ambient condition beyond the service


condition of this system such as the power supply and installation
requirements described in this operation manual.

6. Failure/damage caused by natural disaster such as fire, earthquake, flood


damage and thunderbolt.

The servicing after expiration of the guarantee period will be offered with charge.
Contact our Shimadzu Service Representative nearest to you.

BR-120 Operation Manual Introduction-5


Preface

Precaution in Use
When using this equipment, please observe the following precautions for the safety of
the operator and patient.

Considerations against exposure to X-rays


Improper use of the X-ray equipment might cause the operator or patient to be
accidentally exposed to X-ray radiation. During X-ray radiation, any person
other than the subject patient should not stay in the irradiation room. If
circumstances compel any non-subject person to stay in the room, enough
protection should be provided for that person. Only radio-therapeutic or other
qualified personnel should handle the equipment.

Precautions on mechanical safety


Take care not to catch hands of patients or operators between the Bucky
device and strut when moving the Bucky device.

Emergency Procedures
To stop the all movement of the equipment in an emergency, push the "STOP"
switch (Option). The power supply to the equipment is cut off. The "STOP"
switch (red switches) is located on the side of the main body.

Don’t splash this equipment with water because there is a risk of electric
shock.

This operation manual does not cover the X-ray high voltage generator, X-ray tube
support, and X-ray collimator.

Introduction-6 BR-120 Operation Manual


Preface

Notes for Users


Use the following cassette for the Bucky device. Other cassette can not be used.

cm or inch size cassette based on IEC standard 60406.

Attenuation equivalent is not greater than 1.6Vmm Al.


Less than 15mm thickness.
All parts of the cassette such as hinges, locks and marking device shall be
within the minimum radii for carners and edges.
Flatness shall be less than 1mm warp.

Action for Environment


Contact Shimadzu service representative when disposing the product or parts.
Materials in the system may be discarded and pollute the environment upon improper
disposal.

To all user of Shimadzu equipment in the European Union:

Equipment marked with this symbol indicates that it was sold on or after 13th August
2005, which means it should not be disposed of with general household waste. Note
that our equipment is for industrial/professional use only.
WEEE Mark

Contact Shimadzu service representative when the equipment has reached the
end of its life. They will advise you regarding the equipment take-back.

With your co-operation we are aiming to reduce contamination from waste electronic
and electrical equipment and preserve natural resource through re-use and recycling.
Do not hesitate to ask Shimadzu service representative, if you require further
information.

BR-120 Operation Manual Introduction-7


Preface

No Text

Introduction-8 BR-120 Operation Manual


Contents

Contents
Chapter 1 Outline
1.1 Applications .................................................................................. 2
1.2 Feature ......................................................................................... 3

Chapter 2 Configuration
2.1 Configuration ................................................................................ 6

Chapter 3 Operation
3.1 Prior to Operation ....................................................................... 10
3.1.1 X-ray High Voltage Generator................................................................ 10
3.2 Shut Down of Power................................................................... 11
3.3 Emergency Operation................................................................. 12
3.3.1 Emergency Stop..................................................................................... 12
3.3.2 Recovery from emergency stop ............................................................. 13
3.4 Radiography Stand - Vertical Movement of Bucky Device ......... 14
3.5 Operation of Cassette Tray ........................................................ 15
3.5.1 Names and functions of Cassette tray ................................................... 15
3.5.2 Installing a cassette................................................................................ 16
3.5.3 Removal of Cassette.............................................................................. 18
3.6 Hand Grip (Option) ..................................................................... 19
3.7 Cassette Holder (Option)............................................................ 22
3.8 Operation of Tracking Device (Option) ....................................... 23
3.8.1 Vertical Tracking with X-ray tube support CH-200 ................................. 23
3.9 Operation of Oppression belt (Option)........................................ 27
3.10 Operation of Handy Switch Unit (Option) ................................. 28

Chapter 4 Maintenance
4.1 Daily Inspection (Inspected by User).......................................... 30
4.1.1 Check Sheet for Daily Inspection ........................................................... 30
4.1.2 Fuse ....................................................................................................... 31
4.1.3 Cleaning and disinfection ....................................................................... 31
4.2 Periodical Inspection
(Inspection by Shimadzu Service Representative)..................... 32
4.3 Consumable Parts List ............................................................... 33
4.4 The Operating Life of the Device................................................ 34

BR-120 OPERATION MANUAL Contents-1


Contents

Chapter 5 Specifications
5.1 Specifications..............................................................................36
5.1.1 Classification of equipment .................................................................... 37
5.2 Environmental Condition.............................................................39
5.2.1 Conditions for Operation ........................................................................ 39
5.2.2 Transport and Storage ........................................................................... 39
5.2.3 EMC (Electro-Magnetic Compatibility) [For Europe] .............................. 40
5.3 Statement of Compliance [For Europe].......................................41
5.4 Product safety .............................................................................42
5.5 Symbols and Labels....................................................................43
5.5.1 Package for Transport............................................................................ 46
5.6 Reference axis of unit .................................................................47
5.7 List of Associated Equipment [For Europe].................................48

Appendix
Check List for Trouble ......................................................................52
Safety Check Registry .....................................................................53
Revision History ...............................................................................54

Dimension
BR-120 (503-06335) ........................................................................ 56

Contents-2 BR-120 OPERATION MANUAL


1
Outline
This chapter deals with the applications and features
of BR-120.
1 Outline

1.1 Applications
• The X-ray radiography stand BR-120 can be applied to X-ray radiography for chest
of patients usually in the standing attitude when combined with X-ray tube unit,
collimator unit, X-ray tube support and X-ray high voltage generator.
• BR-120 have a function as servo tracking radiography system when combined with
the X-ray tube support CH-200; X-ray tube support will follow the Bucky device
automatically.

2 BR-120 Operation Manual


1.2 Feature

1.2 Feature
1
• Detail positioning is easy to perform as the Bucky device can be moved smoothly.
• The shorter size underneath the Bucky device than past eases the chest


radiographic of the wheelchair patient.
Plentiful option items allow the X-ray radiographic efficiency to improve.
1

BR-120 Operation Manual 3


1 Outline

No Text

4 BR-120 Operation Manual


2
Configuration
This chapter describes the configuration of the
device.
2 Configuration

2.1 Configuration
The X-ray radiography stand BR-120 consists of the following components.

Fig. 2.1

J Standard Components

① Stand (1 set)
Stand holds the Bucky device using the counter weight balance method, and allows to
position the Bucky device in accordance with the body size of a patient.

② Bucky device (1 set)


Bucky device accommodates a film cassette, and has a built-in moving X-ray grid
which removes scattered X-ray.

③ Brake release button


Bucky device can be moved manually by pressing button to release the brake lock.

J Optional Components (offered separately upon order)

④ Hand grip
Hand grip supports the attitude of a patient.

"3.6 Hand Grip (Option)" P19

6 BR-120 Operation Manual


2.1 Configuration

⑤ Cassette holder
Cassette holder is attached on Bucky device and holds a film cassette and grid plate.
Cassette holder is used to general radiography with film cassette not Bucky device. 2
"3.7 Cassette Holder (Option)" P22

⑥ Base plate 2
Base plate is installed when BR-120 cannot be mounted in the standard holes.

⑦ Front radiography handle


Front radiography handle supports the attitude of a patient.
2
⑧ Servo tracking device
By servo tracking device, the following functions are attached with CH-200 combined: 2
• X-ray tube unit focus of X-ray tube support tracks automatically the center on
height of Bucky device.
• Collimator for radiography radiation field operates automatically.
2
⑨ Oppression belt (not shown)
Oppression belt fixes the radiography area of a patient on the Bucky device.

"3.9 Operation of Oppression belt (Option)" P27


6
⑩ Handy switch unit
The collimator is controlled by handy switch unit. 2
"3.10 Operation of Handy Switch Unit (Option)" P28

BR-120 Operation Manual 7


2 Configuration

No Text

8 BR-120 Operation Manual


3
Operation
This chapter describes how to operate the device
and optional equipment.
3 Operation

3.1 Prior to Operation

3.1.1 X-ray High Voltage Generator

Before operating the radiography stand BR-120, perform the following procedure.

1 Set to ON ( ) the power switch "LINE" provided on the X-ray high


voltage generator combined with the BR-120, then press power ON
button.
When shutting down the power once and turning it on again, wait for 10 sec or
more after shutting down the power.

2 Select the technique key provided on the X-ray high voltage


generator to " ".

3 Set the radiography condition in accordance with the operation


manual of the X-ray high voltage generator.

Follow the operation manual for the X-ray high voltage generator provided.

10 BR-120 Operation Manual


3.2 Shut Down of Power

3.2 Shut Down of Power


To shut down the power, press the power OFF button provided on the X-ray high
voltage generator combined.

Follow the operation manual for the X-ray high voltage generator provided. 3

3
Wait at least 10 seconds or more between turning off the power and
turning it on again. Wait 10 seconds or more before turning the power on 3
after pressing the stop switch and releasing it. Otherwise, the unit may
malfunction.

BR-120 Operation Manual 11


3 Operation

3.3 Emergency Operation

3.3.1 Emergency Stop

In case that the servo tracking device (Optional) is provided with the unit, the stop
switch is installed on the switch panel. When an emergency has occured in which the
vertical movement of the X-ray tube support CH-200 cannot be stopped due to failure
in switches, etc., press the stop switch provided on the control panel of BR-120 or
press the power OFF button provided on the X-ray high voltage generator. The vertical
movement will be stopped.

J Message of Emergency Stop


When operated the emergency stop procedure, the screen on X-ray tube support CH-
200 as shown below is displayed.

Stop button is pressed. Please wait!

Do not turn the stop switch to operate until the following message is displayed.
In case that the stop switch is operated, the error in unit may occur.

After 10 seconds, the screen as shown below is displayed. When this message is
displayed, release procedure can be operated.

Stop button is pressed. Release OK!

12 BR-120 Operation Manual


3.3 Emergency Operation

3.3.2 Recovery from emergency stop

To recover the BR-120 from the emergency stop status, turn the stop switch
clockwise after displayed the screen above-mentioned.

Make sure that the unit does not move abnormally when the stop switch
is turned clockwise to recover from the emergency stop. lf the unit
3
moves abnormally again, stop the unit immediately using the emergency
stop procedure.

BR-120 Operation Manual 13


3 Operation

3.4 Radiography Stand - Vertical


Movement of Bucky Device

1 Press the ( )brake release button on the rear side of Bucky.


The brake is released while the button is pressed. the position of brake release
button is varied from the insatallation condition.

2 Move the the Bucky device up and down by hand after released
brake.
Pressing the brake release button, the brake is released, the Bucky can be
moved up and down manually.

Brake Release Button

Fig. 3.1

When moving the Bucky device up and down, confirm always safety of
the patient.

When moving the Bucky device up and down, pay attention so that the
Bucky device does not come into contact with adjacent equipment.

Do not hang on the Bucky device applying a weight of more than 20 kgf.
The Bucky device goes down suddenly, attitude of the patient deteriorates.

14 BR-120 Operation Manual


3.5 Operation of Cassette Tray

3.5 Operation of Cassette Tray


Cassettes of up to 35 cm × 43 cm (14" × 17") is available for cassette radiography.

3
3.5.1 Names and functions of Cassette tray

3
1

2
6
3

4 3
Fig. 3.2

3
Name Function

1. Cassette tray Stores a cassette in the system.

2. Clamp Fixes a cassette. 3


3. Clamp handle Used to fix or remove a cassette.

4. Shelf Supports a cassette

BR-120 Operation Manual 15


3 Operation

3.5.2 Installing a cassette

1 Lift up the clamphandle of casstte tray in the direction of arrow.

2 Draw out the clamp in the direction of arrow.

Clamp Handle Clamp

Fig. 3.3

3 Fix the shelf at the position in accordance with the casstte size.

4 Lay a casstte on the shelf correctly.

Cassette

Shelf

Fig. 3.4

16 BR-120 Operation Manual


3.5 Operation of Cassette Tray

5 Align the center mark with the center notch.

6 Push the clamp handle forward until the cassette touches the
backside of the bracket.

7 Press the clamp handle down after the cassette is secured.


3
Cassette

Notch 6

3
Center Mark
3

Fig. 3.5
3

Use the following cassette for the Bucky device.


Other cassette can not be used.

• cm or inch size cassette based on IEC standard 60406.


• Attenuation equivalent is not greater than 1.6Vmm Al.
• Less than 15mm thickness.
• All parts of the cassette such as hinges, locks and marking device shall be within
the minimum radii for carners and edges.
• Flatness shall be less than 1mm warp.

BR-120 Operation Manual 17


3 Operation

3.5.3 Removal of Cassette

1 Draw out the cassette tray.

2 Lift the clamp handle to release locking and slide back.


The clamp is released.

3 Draw the clamp and remove the cassette.

4 Remove the cassette tray when finishing cassette radiography.

Cassette

Clamp handle Cassette tray

Fig. 3.6

18 BR-120 Operation Manual


3.6 Hand Grip (Option)

3.6 Hand Grip (Option)


The hand grip is used to keep the patient in radiographic of patient body side.

Fig. 3.7 3

3
The hand grip is used to hold patient position.
Do not hang on the hand grip applying a weight of more than 20 kgf.
The Bucky device goes down suddenly, attitude of the patient
3
deteriorates.

Be careful for patient’s head etc. not to collide with hand grip.

BR-120 Operation Manual 19


3 Operation

J Positioning of the hand grip


1 Insert the prism of hand grip into either square holes at upper both
sides of Bucky device.
The angle of handle is changeable by 90 degree in accordance with inserting
angle.

Hand grip
Clamp handle

Bucky device

Fig. 3.8

2 Pull the hand grip revolver in the direction of arrow to release the
lock, and then change the tilt angle.It is possible to rotate the handle
by pulling.

Hand grip revolver

Fig. 3.9

Confirm that the handle is fixed, prior to instruct the position of patients.

20 BR-120 Operation Manual


3.6 Hand Grip (Option)

The vertical movement of Bucky device may be unbalanced when the


hand grip is not attached. (When operating the Bucky device, it is upward
light and, downward, becomes heavy.
3

BR-120 Operation Manual 21


3 Operation

3.7 Cassette Holder (Option)

1 Hook the cassette holder to the fixing bar, and then fix the lower
portion by magnet.

2 Clamp the cassette with the cassette clamp, and then fix the holder
with turning knob.

Fixing
Bar

Cassette Clamp

Magnet Bucky
Device
Bracket

Knob

Grid
Magnet
Cassette Holder
Bucky Cassette
Device
Cassette
Holder

Magnet
Magnet Bracket

Fig. 3.10

When the cassette holder is mounted on the Bucky device, operation is a little
heavily for vertical movement.

22 BR-120 Operation Manual


3.8 Operation of Tracking Device (Option)

3.8 Operation of Tracking Device


(Option)
3
3.8.1 Vertical Tracking with X-ray tube support CH-200

J Operation of X-ray high voltage generator 3


1 Press the radiography program
key on the touch panel to select
the vertical Bucky radiography. 3
Confirm that the technique key indicates
the vertical Bucky radiography.

3
J Operation of X-ray tube unit
1 Press the detent mode button on 6
operation panel to turn detent
mode ON.
3

3
2 Move the X-ray tube support and then stop it to the preset
longitudinal/transversal position.
Longitudinal directional position is predefined radiography distance, and
3
transversal is the center of Bucky device. Face the X-ray tube of X-ray tube
support toward the X-ray radiography stand.

BR-120 Operation Manual 23


3 Operation

J Operation of X-ray radiography stand

Confirm no patient is positioned where the X-ray tube support declines.


Touching X-ray tube support may cause injury.

Two kinds of tracking operation with X-ray tube support are as follows:
For setting of full-automatic tracking operation or semi-automatic one, make a choice
between the two on installation. Consult Shimadzu service representative for setting the
function.

Full-automatic operation
• Moving upward/downward the Bucky device of X-ray radiography stand
manually, the X-ray tube support tracks with X-ray radiography stand.

Semi-automatic operation
• Pressing both tracking switches of X-ray radiography stand and X-ray tube
support, the X-ray tube support tracks with X-ray radiography stand.

• During tracking operation, press either tracking switch of X-ray tube support
or X-ray radiography stand, so that the tracking operation will be stopped.

• If there is a possibility that the X-ray tube support contacts the X-ray
radiography stand, it once stops and continuous beep sounds
(Approx. 5 seconds).
If the tracking switch is pressed again, while the X-ray tube support will
sound beep, it descends at a low speed, and continues to descend until the
tracking switch is released. So, position the X-ray tube unit in contact with
X-ray radiography stand with care.

• It is not a malfunction that the X-ray tube support once goes beyond too far
and returns to the predetermined position for alignment.

24 BR-120 Operation Manual


3.8 Operation of Tracking Device (Option)

J For Oblique Projection Radiography


After tracking the X-ray tube support and X-ray radiography stand, operate the
following procedures for oblique projection radiography.

1 Adjusting the height of X-ray tube unit, tilt the unit with rotating a little
3
for upper or lower direction.

6
When the rotation angle is shifted over 4
Automatic
degree from axis, or the height of X-ray
tube unit is shifted the center of Bucky 3
device, the [Automatic] is displayed.

3
When not aligned the axis of X-ray tube unit with the center of Bucky device,
[Automatic]* is displayed on touch panel. When [Automatic]* is displayed, the
operator must have responsibility in carrying out exposure field adjustment and
performing X-ray radiography.
* For the specification of the products to the United States, [Manual], instead of
[Automatic], is displayed in the setting.

2 Press the collimator lamp button to light the collimator lamp up.

3 Lighting the collimator lamp, adjust the exposure field.


The collimator lamp lights for 30 seconds.

BR-120 Operation Manual 25


3 Operation

J Operation of X-ray high voltage generator (Radiography)


1 Set the X-ray radiography
conditions of X-ray high voltage
generator.

2 Hold down the radiography


preparation button.

3 When the radiography button or


radiography ready indicator lights,
Radiography Radiography
press the radiography button on Preparation Button
the operation panel while holding Button

down the radiography preparation


button.
• Radiography operation starts.
• The radiography button and the radiography indicator on the operation panel
light during radiography.
• The radiography is completed when the radiography indicator on the operation
panel lights off.

4 Release the radiography button.

In case that the tracking operation is not completed, [Automatic] is


displayed on touch panel. When [Automatic] is displayed, the operator
must have responsibility in carrying out exposure field adjustment and
performing X-ray radiography.

26 BR-120 Operation Manual


3.9 Operation of Oppression belt (Option)

3.9 Operation of Oppression belt


(Option)
1 Align the one end (Velcro tape) of oppression belt with Velcro tape 3
on back surface of Bucky device and then stick them firmly.

2 Tighten the oppression belt to contact the patient with Bucky device.

3 After tightening the belt, align the other end (Velcro tape) of 3
oppression belt with other Velcro tape on back surface of Bucky
device and then stick them firmly.

4 Remove the both ends of oppression belt from Bucky device after
3
use, and the keep it in the specified location.

3
Oppression belt
6

3
Velcro Tape Bucky device

Velcro Tape 3

BR-120 Operation Manual 27


3 Operation

3.10 Operation of Handy Switch Unit


(Option)
This handy switch unit is able to be operated leaf open/close of collimator and
illumination of lamp when combining with the ceiling tube support CH-200.

① Leaf open switch


Handy switch unit
Leaf open switch allows the throttle to be open.

② Leaf close switch


① Leaf close switch allows the leaf to throttle.

② ③ Leaf open switch


Leaf close switch allows the leaf to throttle.

③ ④ Leaf close switch


Leaf open switch allows the throttled leaf to be
④ open.
⑤ ⑤ Collimator lamp switch
Pressing the collimator lamp switch, the
collimator lamp will light up the collimator lamp.

The collimator leaf open/close operation is possible when the collimator key
switch of X-ray tube unit is in automatic mode ( ).
The leaves of collimator are designed to throttle collimation within the range
taking the cassette film size as the upper limit.

Automatic mode

Manual mode

Collimator

X-ray tube unit CH-200

28 BR-120 Operation Manual


4
Maintenance
This chapter describes points for daily and periodic
inspections as well as consumable parts.
4 Maintenance

4.1 Daily Inspection (Inspected by User)


Inspect the following items before operating the device, and if any abnormal points are
observed, please contact Shimadzu Service Representative:

4.1.1 Check Sheet for Daily Inspection

Date of inspection:

Inspected by:

1 Check the following points visually and perform the appropriate


procedure in case of any problems.
Verification Points Prizedure

□ Are all unnecessary objects removed from the vicinity of


Remove
the device?

□ Are the cables free of pinches, twists and scrapes? Fix the cables

Contact Shimadzu
□ Any breakage (dents, fractures, etc.) on the cover? Service
Representative

2 Check the following points by hand and perform the appropriate


procedure in case of any problems.
Verification Points Procedure

□ Any large shaking at overall structure?

□ Cassette tray is put in/out smoothly? Contact Shimadzu


Service
□ Cassette is fixed? Representative
□ Any jerk or creak while rotating of hand grip?

3 Turn ON the power supply and check the following points. Perform
the appropriate procedure in case of any problems.
Verification Points Procedure

□ Does the unit make any noise when moving ?


Contact
□ Does the interlock switch operate correctly? Shimadzu
Service
□ The brake can be released at up/down movement of Representative
Bucky device

30 BR-120 Operation Manual


4.1 Daily Inspection (Inspected by User)

4.1.2 Fuse
4
When the brake does not work well even when the power is ON, check the fuse on the
downside of column.

For replacement, the following fuse or the equivalent should be used. 4


• Fuse 313001 (072-01664-20) Rated 250V 1A (slow blow)

Prior replacing the fuse, "LINE" switch on the X-ray high voltage
4
generator should be turned OFF to shut off the power.

6
4.1.3 Cleaning and disinfection
4
Clean the hand grips using a clean cloth and alcohol or disinfectants.

4
Do not splash or spray any disinfectant onto the system.
Any disinfectants that leak into unit may cause failure or trouble with the
system.
4

BR-120 Operation Manual 31


4 Maintenance

4.2 Periodical Inspection


(Inspection by Shimadzu Service
Representative)
Even without any problem in daily inspection, the following items should be inspected
periodically in every 6 months by Shimadzu Service Representative or other
certificated persons.

Table 4.1 Periodical inspection (Inspection performed by Shimadzu Service Representative)

Item Inspection points

Loosening of bolts Bolts fixing and roller of each part

・Is the wire rope severed?


・Is the wire rope scratched or snagged?
・Is the wire pulley scratched?
Wire rope ・Are there any scratches or obstructions around the rail in the
wheels?
Remove
・Apply an aerosol oil (RO-300 or equivalent) to the wire ropes on
the
periodical inspection.
cover
and Power supply Is input/Output voltage appropriate?
check
Pulley, shaft,
Wear, deformation, breakage, looseness, dropping
pins, screws, etc.

Bearing Wear, rust oil shortage

Magnet Are there any damage or dirt on the magnet surface?

32 BR-120 Operation Manual


4.3 Consumable Parts List

4.3 Consumable Parts List


4
The consumable parts used for this device are as follows:
When exchanging the parts, use the regular parts with following part number:
4
Table 4.2 Consumable parts list

4
Required
Name Part number
quantity

Consumable Fuse, 313001


072-01664-20 1
parts (250VAC, 1A, SLOWBLOW)

4
If any damages are found on wire ropes at inspection, replace it immediately.
Contact Shimadzu Service Representative for replacement.

BR-120 Operation Manual 33


4 Maintenance

4.4 The Operating Life of the Device


This device's operating life of the device is approximately 10 years (on average).
An overhaul is necessary to use the device beyond its operating life in order to ensure
the safety.

Please contact Shimadzu Service Representative before the device exceeds its
operating life.

Do not use this device past its operating life without an overhaul.

34 BR-120 Operation Manual


5
Specifications
This chapter describes specifications.
5 Specifications

5.1 Specifications

Item Description

Construction

① Bucky Device
② Stand h

Bucky device BF-11VP

Distance between
Bucky device guathal h = 627 - 2147 mm
hub and floor surface

By manual.
Operation Electromagnetic lock provided is released while power is
supplied.
Distance between
Bucky device guathal 247 mm
hub and film upper end

Distance between
Bucky device front 40 mm
face and film

cm 18 × 24, 24 × 24, 24 × 30, 35 × 35, 35 × 43


Application film
inch 8 × 10, 9.5 × 9.5, 10 × 12, 11 × 14, 14 × 14, 14 × 17
size
Film can be located in either vertical or horizontal direction.

Cassette fixing size Center reference

Except
North America
North America

Grid ratio R = 12 : 1 R = 10 : 1
X-ray grid Grid density N = 40 Lines/cm N = 40 Lines/cm
(It functions as a
moving grid within Focusing
Fo = 170 cm Fo = 150 cm
irradiation field.) *1 distance

Intermediate
AL AL
material

Dimensions 479 mm × 438 mm 479 mm × 438 mm

Tray which can be QJC or QJS (POERSCH Metal Manufacturing Co.)


combined Size sensing tray

36 BR-120 Operation Manual


5.1 Specifications

Item Description

Installation conditions
5
Outer dimension Refer to 503-06335

Required space Approx. 650 mm (L) × 400 mm (D)

Required ceiling height 2350 mm or more 5


Mass Approx. 120 kg

Single phase, 100V, 0.1kVA, 50/60Hz


Power Supply
(Supplied from X-ray high voltage generator)
5
*1 When radiography is performed without an object for time (40msec or less), lead stripes shadow
may appear.

Note: Contents of the specifications are subjected to change for further improvement without notice. 5
J Dose Information
Attenuation by Items in the X-ray Beam
5
Measurement conditions:

• X-ray tube voltage: 100 kV




X-ray tube current: 1.8 mA
X-ray beam: Half value layer is 3.7 mmAl
6
• Front panel: 0.63 mmAleq.

5
5.1.1 Classification of equipment

J The class of protection against electric shock 5


Class I equipment

5
It means that CLASS I EQUIPMENT, that is, equipment in which protection
against electric shock does not rely on BASIC INSULATION only, but which
includes an additional safety precaution in that means are provided for the
connection of the equipment to the protective earth conductor in the fixed wiring
of the installation in such a way that accessible metal parts can not become live
in the event of a failure of the basic insulation. (International standard
IEC60601-1 Second edition 1988).

BR-120 Operation Manual 37


5 Specifications

J The degree of protection against electric shock


Type B equipment

It means TYPE B EQUIPMENT, that is, equipment providing a particular degree


of protection against electric shock, particularly regarding:
- allowable leakage current;
- reliability of the protective earth connection.
(International standard IEC60601-1 Second edition 1988).

J The degree of protection against harmful ingress of water


Ordinary equipment

The main body of this unit is not protected against immersion of liquid.
Never use this unit in a place in which immersion of liquid may occur.
Never spill liquid on the surface or in the inside of this unit. Otherwise,
electrical shock may occur. When liquid is split, contact our office or
representative described on the back cover of the operation manual.

J The degree of safety of application in the presence of a


flammable anesthetic mixture with air or with oxygen or nitrous
oxide
Equipment not suitable for use in the presence of a flammable anesthetic mixture with
air or with oxygen or nitrous oxide

Risk of explosion if used in the presence of flammable anesthetics.

J The mode of operation


Continuous operation with intermittent loading

38 BR-120 Operation Manual


5.2 Environmental Condition

5.2 Environmental Condition


5
5.2.1 Conditions for Operation
5
Use the BR-120 in the following condition so that it can offer its full performance.

J Radiography condition 5
• Temperature : +10 to +40°C
• Humidity : 30 to 85%
• Atmospheric pressure : 700 to 1060hPa
5
J Power supply
• Single-phase, 100V, 0.1kVA, 50/60Hz
5
J Grounding
• Grounding to the ground resistance of less than 100Ωwith annealed copper wire of
1.6mm diameter for grounding wire. 6

5
Do not expose the device to drastic temperature or humidity fluctuations
even under the above mentioned conditions.
Such a rapid ambient fluctuation causes the occurrence of condensation
resulting in damages rust and corrosion in the device.
5
Also, condensation within the device may cause rust and corrosion.
Under a low temperatures, frosts may damage the internal circuit.
5

5.2.2 Transport and Storage

The environmental conditions for transport and storage are described as below.

Item Condition
Ambient temperature -10°C to 60°C
Ambient humidity 10 to 95 %
Atmospheric pressure 700 to 1060hPa

BR-120 Operation Manual 39


5 Specifications

Rust and corrosion may be generated caused by dew inside this unit.
The internal circuits may be damaged by frozen dew caused by low
temperature. Pay rigid attention when this unit is stored in a place with
drastic difference in temperature and humidity such as warehouse.

5.2.3 EMC (Electro-Magnetic Compatibility) [For Europe]

• This system belongs to Group 1 and Class A equipment in accordance to


EN60601-1-2:1993.
• The system uses radio-frequency energy only for its internal function and is not
intended to deliver enrgy to patient. But little leakage radio-frequency energy
does harm to high-sensitive equipment.
• The system main power line in the clinical site should be connected to the
domestic power sources which are separated from the public main network.

40 BR-120 Operation Manual


5.3 Statement of Compliance [For Europe]

5.3 Statement of Compliance


5
[For Europe]
This unit conforms the following international standards.
5
J 93/42/EEC
• EN60601-1: 1990 + A1: 1993 + A2: 1995 +A13: 1996
• IEC60601-1: 1988 + A1: 1991 + A2: 1995 5
• EN60601-1-1: 2001
• IEC60601-1-1: 2000


EN60601-1-2: 1993
IEC60601-1-2: 1993
5
• EN55011: 1998 (group1, class A)
• IEC801-2: 1991/EN61000-4-2: 1995 (contact: level 3, air: level 3)
• IEC801-3: 1984 with reference to EN61800-3: 1996/A11: 2000
Clause 5.3.3.2
5
• IEC801-4: 1988/EN61000-4-4: 1995 (power lines: level 3, signal lines:
level 3)
• IEC1000-4-5: 1995/EN61000-4-5: 1995 (power lines: class 3) 6
• EN61000-4-6: 1996 (power lines: level 2, signal lines: level 2)
• EN61000-4-8: 1993 (level 2)
• EN60601-1-3: 1994
• IEC60601-1-3: 1994 5
• EN60601-2-32: 1994
• IEC601-2-32: 1994

BR-120 Operation Manual 41


5 Specifications

5.4 Product safety

Do not operate this unit if there is any uncertainty as to the proper


functioning of the system. refer all servicing to qualified service
personnel.

This instrument must grounded! To minimize the shock hazard, make


sure of performing the ground work according to Installation Manual
(M511-E319).

The operator must set the focal spot to skin distance as large as possible
in order to keep the absorbed does to the patient as low as reasonably
achievable.

42 BR-120 Operation Manual


5.5 Symbols and Labels

5.5 Symbols and Labels


5
J Symbols
Symbols used on this system are shown and described as follows.
5
Symbol Location Meaning

In name plate Alternating current


5

Inside the device, where


protective earth conductor Protective earth ground 5
in power cord is connected

In name plate Safety classification: Type B 5

On warning and caution


labels
Attention:
Refer to Operation Manual
6

5
J Labels
Locations where labels are attached are shown as follows.

Fig. 5.1

BR-120 Operation Manual 43


5 Specifications

① Name Plate A (Name Plate of BR-120)

② Certification Label A (for UL)

③ Warning Label, Fuse

④ Caution Label

⑤ CE Label (for CE)

44 BR-120 Operation Manual


5.5 Symbols and Labels

⑥ EU Representative Label (for CE)

5
⑦ Label EMC
5

⑧ Label TÜV USA (for UL) 5

5
⑨ TÜV GM Label (for CE)

⑩ WEEE Mark Label (for CE)

BR-120 Operation Manual 45


5 Specifications

5.5.1 Package for Transport

Label that is described about storage environment and transport environment on


package for transport is shown.

保管環境 温 度:-10 - 60°C


相対湿度:10 - 95%
気 圧:700 - 1060hPa
輸送環境 温 度:-10 - 60°C
相対湿度:10 - 95%
気 圧:700 - 1060hPa

Storage Environment Temperature : -10 - 60°C


Relative Humidity : 10 - 95%
Atmospheric Pressure : 700 - 1060hPa
Transport Environment Temperature : -10 - 60°C
Relative Humidity : 10 - 95%
Atmospheric Pressure : 700 - 1060hPa

46 BR-120 Operation Manual


5.6 Reference axis of unit

5.6 Reference axis of unit


5
The reference axis which employed in normal operation of X-ray radiography stand
means the incident X-ray axis from X-ray tube unit to Bucky device (Film) vertically as
shown below.
At this point, 90° as X-ray axis angle is displayed on the touch panel of X-ray tube 5
support CH-200, and the angle indication varies in proportion as the angle which the X-
ray tube unit is tilted.
When the vertical Bucky radiography technique is selected on the X-ray high voltage
generator, the same indication is displayed on the touch panel of X-ray tube support 5
CH-200 correspondingly.

5
Vertical Bucky radiography technique
5

6
X-ray axis

BR-120 Operation Manual 47


5 Specifications

5.7 List of Associated Equipment


[For Europe]

X-ray High Voltage


Equipment
BR-120 CH-200 Generator
Model
UD150B-40 /L-40
Manufacturer Shimadzu Shimadzu Shimadzu
Single phase :
Single phase : • AC200V, 1kVA, 50/ Three-phases :
Supply mains • AC100V, 0.1kVA, 50/ 60Hz • AC400V, 80kVA, 50/
60Hz • AC100V, 0.2kVA, 50/ 60Hz
60Hz
Classification Class I, Type B equipment
Mode of Continuous operation with
operation Intermittent loading
• Ambient temperature : • Ambient temperature :
+10 to 40 ℃ +5 to 35 ℃
Condition of • Relative humidity : • Relative humidity :
use 30 to 85% 45 to 85%
• Atmospheric pressure : • Atmospheric pressure :
700 to 1060hPa 700 to 1060hPa
Storage environment
• Temperature :
-10 to 60 ℃
• Relative humidity :
10 to 95%
• Atmospheric pressure :
Protective 700 to 1060hPa
packing Transport environment
• Temperature :
-10 to 60 ℃
• Relative humidity :
10 to 95%
• Atmospheric pressure :
700 to 1060hPa
• Hand Grip • Servo tomo option • Photo timer option
• Compression belt • Servo tracking option
• Cassette holder
Option
• Front radiography
handle
• Interlocking device

48 BR-120 Operation Manual


5.7 List of Associated Equipment [For Europe]

Equipment
Model
X-ray Collimator R-30H 0.6/1.2P324DK-120PF BK-200 5
Manufacturer Shimadzu Shimadzu Shimadzu

Supply mains
Single phase :
• AC12V, 0.1kVA, 50/
• Power is supplied from
X-ray high voltage
Single phase :
• AC200V, 1kVA, 50/
5
60Hz generator
60Hz
• DC12V, 6VA
Classification
Mode of
Class I, Type B equipment
Continuous operation with
5
operation Intermittent loading
• Ambient temperature :
+10 to 40 ℃ 5
Condition of • Relative humidity :
use 30 to 85%
• Atmospheric pressure :
700 to 1060hPa 5
Storage environment
• Temperature :
-10 to 60 ℃
• Relative humidity : 6
10 to 95%
• Atmospheric pressure :
Protective 700 to 1060hPa
packing Transport environment 5
• Temperature :
-10 to 60 ℃
• Relative humidity :


10 to 95%
Atmospheric pressure :
5
700 to 1060hPa
• Hand grip
• Grip switch 5
Option None None • Servo tomo option
• Compression band
• Cassette holder

BR-120 Operation Manual 49


5 Specifications

No Text

50 BR-120 Operation Manual


Appendix
Check List for Trouble

Fill following blanks and contact the SHIMADZU Service Representative.

Hospital :

Phone :

FAX :
Serial Number:

Date of Installation:

Item Check

□ Someone injured

□ When is the trouble occurred?

□ Is there a sign of the trouble?

□ Is there a power failure or thunderbolt when the


trouble is occurred?

□ The device has been watered?

□ The device has been given an impact upon?


□ The number of patients a day with the device?

□ When is the last periodic inspection?

52 BR-120 Operation Manual


Safety Check Registry
6
□ Name

□ Phone
6
□ Extension number
Hospital
□ FAX

□ Address
6
□ Room number

Device
□ Name
6
□ Serial number
□ Date of installation

□ Affiliation 6
Instructor □ Phone

□ FAX

6
This device's
maintenance and Instructor
Date Description
operation staff 6
Signature Signature

• At the time of installation or when changing the designated operator, the


SHIMADZU Service Representative will explain this manual's safety description
of this device's maintenance to operation staff and record the new information.

• This manual is important. We request all maintenance personnel to keep it in a


safe place close to the device.

BR-120 Operation Manual 53


Revision History

This manual's revision history is shown in the table below.

Revision Date Changes

First revision Sep. 2004 First edition released

Application film size revised in


A Nov. 2004
"5.1 Specifications"

"3.8 Operation of Tracking Device (Option)"


B Apr. 2006 revised
"5.5 Symbols and Labels" revised
Warning is revised that [Manual], instead of
[Automatic] is displaying for the specification
C Sep. 2006 of the products to the United States in item
"For Oblique Projection Radiography" on
3.8.1 heading

54 BR-120 Operation Manual


Dimension
BR-120 (503-06335)

56 BR-120 Operation Manual


7

BR-120 Operation Manual 57


No Text

58 BR-120 Operation Manual

You might also like