‘Number : 2014-04-003 ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX +81-748-86-8739
Technical Information
Number: 2014-04-003 Date:14/April/2014 Page: 1 of 2
Products : HA-8180series — Serial# / Lot#: (ALL)
1.Mechanical 2.Electrical 3. Reagent 4. Software
Subject: HA-8180 series Tube pump wiring standardization
1. Reason for Change
97099:Tube pump wiring used in HA-8180 family has changed its tubes for the purpose of common use by
multiple different instruments
2. Details
1. Details
For the purpose of common use by multiple different instruments, 97099 : Tube pump wiring was
| standardized, As a result, tubes connected to the pump were changed.
Previously, the tube is as shown in Figure 4. After the change, longer tube shown in Figure 2 comes with the
pump. Cut the tubes into appropriate length before use.
Note that the tube is commonly used in AX-4030, AU-4050, and HA-8180 family instruments.
2. Tube cut position instruction
Cut the tubes of Figure 2 pump at 40mm on the left tube and 30mm on the right tube as shown in Figure 3.
Then leave a 30mm length tube piece (shown in Figure 4). This results in the pump and the tube shown in
Figure 1
Figure 1 Pump with cut tube Figure 2 Delivered Tube
30mm
Figure 4
ARKRAY Technical SupportNumber : 2014-04-03 ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX *81-748-86-8739
Page: 2of 2
3. Approved by
‘4.Marketing: Yoshiaki Shibuta S pie 3 Date:.20/-4- fag
2. Development: Takeshi Takagi Lohohe haad Date: 20% 22
Technical Support: Fumio Yamauchi Furnto Yamauchi Date: 20/4. .
ARKRAY Technical Support