Number : 2014-12-006 ARKRAY Infinity, Inc. TEL +81-748-86-8732 FAX +81-748-86-8739 Technical Information Number: 2014-12-006 Date:15/December/2014 Page: 1 of 4 Products : HA-8180V Serial# / Lott: — (ALL) 2Blectrical 3. Reagent 4. Software 5, Others Subject: Problem cause and countermeasure for abnormal baseline 1. Outline Possible causes and countermeasures for the HA-8160V abnormal baseline in chromatograph is explained in | this document. Optical unit cell washing or optical unit assembly related parts replacement may be effective in | some cases, 2. Details In HA-B180V, when measurement value is supposed as abnormal, baseline drift because of dirt in flow channel is considered a problem cause. Baseline in chromatograph can be checked by the instrument diagnosis function. Enter the Service Mode and select the menu below, then execute Analysis section check. Baseline is printed out automatically. Next, check the printed result comparing with the normal waveform and the following five check points. Check baseline using Analysis section check measurement. (Service mode>>8.Instrument diagnosis menu>>4.Analysis section check) If the value exceeds the acceptable range below, Drift baseline, Noise detected, or T490:Background occurs. Attention is needed also when waveform is around the threshold. Note that the baseline vertical reference position is one division higher from the horizontal axis. Normal waveform Points to check on baseline 1) Baseline shift while Eluent-CV is fed to the column. (Only for Variant Mode) Check that the baseline shift while Eluent CV solution is fed does not exceed +/-0.5 division shown in the examples below. T 30 6 eee) ARKRAY Technical Support ‘Number : 2014-12-006 ARKRAY Infinity, Inc. TEL +81-748-86-8732 FAX +81-748-86-8739 Page: 2 of 4 2) Eluent switching timing At the time Eluent is switched from one to another, baseline shifts. Check that the shift amount is within the allowable range below. Eluent A > Eluent C: Allowable range is up to 0.5 division. Eluent C > Eluent 8: Allowable range i upto one-third division, 5 of. o 30 60 Goo) ° 30 60 (cee) Eluent B: Allowable range is up to one-third division. ° 30 (coe) 3) Waveform during Eluent-8 feeding Normally, small peak appears upon feeding Eluent-B. (Refer to the normal waveform.) This peak height must be within 1.5 division. Allowable range is up to 1.5 division. Allowable range is up to 1.5 division. 4 __ eee 7 6 | \ ——_ ° 30 (seo) ° 30 60 (coe) 4) Noise Check that noise peak height on the baseline is 5 dots or less. Noise around the white arrow in the example below can be neglected. Allowable noise peak is five dots on the Allowable noise peak is five dots on the chromatograph. chromatograph. Saots Sele J ots Beate {At —_—___ ° 20 60 (Gee) ° 30 (eee) Noise around this mark “O* can be neglected. ARKRAY Technical Support Number : 2014-12-006 ARKRAY Infinity, Inc. TEL +81-748-86-8732 FAX *81-148-86-8739 Page: 3 of 4 5) Baseline Check that baseline does not go up and down shown in the example below. Acceptable range is 0.5 division. ° 30 60 (eee) ° 30 60 (eee) ° 30 (eee) ° 30 Gee) [Possible problem cause] The followings are possible problem cause for baseline drift. ‘Cause 1) Optical unit cell is dirty. (In case the waveform looks like the examples 1, 2, 3, 5.) — Optical cell characteristics degradation induced by protein stain adhering to the optical unit after instrument long use or cell dry-up due to solution leakage. Cause 2) Light beam misalignment. (In case the waveform looks like the examples 1, 2, 3, 5.) + Optical cell deformation caused by flow channel higher pressure induced by optical cell output resistant pipe clogging by mechanical shock or long term use, or looseness of fixing screws for interference filter block or optical unit cell. Cause 3) Optical transmission wavelength shift (In case the waveform looks like the examples 1,2,3,5,) — Lens or interference filter characteristic degradation due to dust or soot adhering to lens or dew condensation on lens, Cause 4) Air bubble mixing (In case the waveform looks like the examples 2, 4, 5.) — Air mixing or fluid leakage from tube joint portion or optical cell Cause 5) Dirt of Eluent (In case the waveform looks like the examples 1, 2,3.) — Incase Eluent is dirty, ete. [Countermeasures] ‘When baseline drift occurs, following countermeasures can be taken. Countermeasure 1) Check for air bubble in the drain tube. (Cause 4 is suspicious.) — Check for air bubble in the instrument OUT tubing (one connected to ‘D1 in the rear) during measurement. if air bubbles enter the tube, joint looseness or fluid leakage from optical unit assembly or degasser maifunction can be considered. When air bubble mixing or fluid leakage is found, check the tube connection and fluid leakage from a cell. Countermeasure 2) Replace Eluent ( Cause 5 is suspicious.) — Eluent may get dirt. Replace Eluent with a new one. Also, replace fluid inside Eluent chamber. Then, check that the sensor in Eluent chamber is functional (Service mode>>2228 Sensor status) ARKRAY Technical Support Number : 2014-12-006 ARKRAY Infinity, Inc. TEL +81-748-86-8732 FAX *81-748 868739 Page: 4 of 4 Countermeasure 3) Wash the optical unit cell. (Cause 1 is suspicious.) —Wash the optical unit cell with Cell Washing Kit (Product code:80386) (Reference: Service manual Chapter? 1.1) Washing the optical unit cell) Countermeasure 4) Replace the optical unit, (Cause 1,2,3 is suspicious) —ln case the cause 3 hits, disassemble and clean the interference filter and the cell. If disassembling and cleaning is difficult, optical unit assembly or interference filter plus cell Note that countermeasures other than above may be required depending on the situation. <> When the measurement value shows abnormality, column replacement or low pressure tube line washing may be effective. ( Reference: Service manual Chaptet7 1.2) Washing the low pressure tube line) 3.Effect on Inventory Control Part Number Description Effective Lot ‘Compatibility Compatibility: Use,Rework,Scrap.N/A 4. Approved by 2 4.Marketing: Yoshiaki Shibuta rs Z a Date: eI P (2. Ce ri . 2.Development: Takeshi Takagi Tehoptes Get, Date: 20% 2.6 ‘3 Technical Support: Fumio Yamauchi Euimio Yemau che’ Date: 20/% /2./(- ARKRAY Technical Support

You might also like