Number : 2014-10-001A
ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX +81-748-86-8739
Technical Information
Number: 2014-10-001A Di 23/October/2014 Page: 1 of 4
Products : Sampler erialt)/ Loti:41407001 (ALL)
2Blectrical 3. Reagent 4. Software 5. Others
Subject: Common sampler specification change
J. Outline
‘Common sampler specification has been changed. Details ofthis change are explained below.
2. Details
4. Applicable models
HA-8180, HA-8180V, HA-8180T, AX-4030, GA-1171, HA-8160(VP/TP), AX-4280
‘Samplers before and after specification change are compatible. No instrument setting change
is required.
Therefore, newly specification changed sampler fits an instrument which uses a sampler
before specification change.
2. Applicable sampler Serial No.
Serial No. 41407001 or later
Label position Label position
Unlike the sampler level applicable serial number, we do not designate applicable instrument serial
number as it is not fixed by the time of instrument shipment.
3. Change detail
A) Appearance
(4) Holes or the ight ecge:of the top bana right assaribhyiet assembly are removed
Before specification change ‘After specification change
ARKRAY Technical SupportNumber : 2014-10-0014
ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX 481-748-86-8739
Pag
22 of 4
(2) STAT port cover is changed to be same as a mixing specification sampler.
Roller unit fixing holes are covered with screws if the specification is other than mixing
specification.
Before specification change After specification change
B) Adjustment items in the Service Manual
(1) Measurement Side Lever P11. Adjustment
Hexagon head bolt to fix adjustment screw is eliminated. Adjustment procedure is unchanged.
After specification change
(2) Measurement Side Lever Interlock Adjustment
‘Adjustment procedure is unchanged.
(3) Belt tension adjustment
+Sample Rack Longitudinal Feed Belt Tension Adjustment
‘Adjustment is not needed as spring is newly adopted for belt tension adjustment.
*Sample Rack Lateral Feed Belt Tension Adjustment
Adjustment procedure is changed to use Belt tension adjustment jig (Product code :101814)
This jig can not be used for the former sampler. (For adjustment procedure, refer to the
appendix.)
(4) Lateral Feed Lever Angle Adjustment
‘Adjustment is no longer necessary as the lateral movement lever roller holding plate position
is fixed.
(8) STAT Port Holder Height Adjustment
‘Adjustment procedure is unchanged.
(6) Rack ID Sensor Sensitivity Adjustment
Adjustment procedure is unchanged.
(7) Sensor Panel Angle Adjustment
‘Adjustment procedure is unchanged.
(8) Longitudinal Feed Pulley Height Adjustment
Adjustment procedure is changed to use Pulley height adjustment jig (Product code
101815). This jig can not be used for the former top base rightleft assembly. (For adjustment
procedure, refer to the appendix.)
ARKRAY Technical Support‘Number : 2014-10-001A
ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX +81-748-86-8739
Page: 3 of 4
C) Service parts
(1)Chassis : Before spec. change type : 96447, After spec. change type : 101817
(2)Top base right assembly (Product code: 95345) and Top base left assembly (Product code :
95346) of before/after specification change are compatible.
‘As soon as the former parts run out, parts will be switched to the new specification one without
changing product order code.
(3)Regardless of specification change, Extension cable S6(Product order code :101816) is
additionally registered.
(4) No product order code for service parts other than chassis is added as they are compatible
before/after the change.
D) Others
(1)The hexagonal post which is taken out when an extension sampler is installed on the sampler
eject side is eliminated. At the same time, a projection is added on the left portion of the side
panel to prevent Lateral Feed Lever from overrunning beyond the initial sensor position.
‘After specification change (Projection)
4. Appendix
- System Configuration Sampler (Changed specification)
~ Adjustments instruction Sampler (Changed specification)
ARKRAY Technical SupportNumber : 2014-10-001A
ARKRAY Infinity, Inc.
TEL +81-748-86-8732 FAX +81-748-86-8739
Page: 4 of 4
3.Effect on Inventory Control
Part Number Description | Effective Lot ‘Compatibility
7 ‘Compatibility: Use,Rework,Scrap,N/A
4. Approved by _
4.Marketing: Yoshiaki Shibuta LDPE Date: 20K a
2.Development: _Kazwa shimaga Katya Shi mace Date: d0/4,(0.23
yy \
Technical Support: Fumio Yemaucni_ LA 'd letra, pate 2A, (2 2 P-
ARKRAY Technical Support