NEONATAL ICU
PEDIATRIC ICU
ADULT ICU
E150pptcy0606 REV. A 2
NEWPORT BREEZE
DESIGN ELEMENTS
E150pptcy0606 REV. A 3
NEWPORT BREEZE
DESIGN ELEMENTS
Unique Standard Functions
Volume Controlled Ventilation
Pressure Controlled Ventilation
SIMV
Inverse I:E Ratio Ventilation
Enhanced Monitoring
PROXIMAL Pressure Monitoring
Internal Battery (1 Hour)
Built-in Nebulizer Drive
E150pptcy0606 REV. A 4
NEWPORT BREEZE
DESIGN ELEMENTS
Logical layout of patient controls:
E150pptcy0606 REV. A 5
NEWPORT BREEZE
DESIGN ELEMENTS
Enhanced Monitoring & Displays:
E150pptcy0606 REV. A 8
NEWPORT BREEZE
ADVANCED FUNCTION
Unique Flow System - DUOFLOW
SINGLE DUO-FLOW
DUO-FLOWSYSTEM
SINGLEFLOW
FLOWSYSTEM
SYSTEM HIGH
SYSTEM
HIGH
AA CONSTANT
CONSTANT
FLOW
FLOW
HIGH
AA HIGH
CONSTANT
CONSTANT
FLOW ONE WAY
ONE WAY
FLOW VALVE
EXHALATION VALVE
EXHALATION
VALVE
VALVE
CLOSED
CLOSED EXHALATION
EXHALATION
VALVE
PATIENT VALVE
PATIENT BB HIGH
HIGH
CONSTANT
CONSTANT
CLOSED
CLOSED
FLOW PATIENT
HIGH FLOW PATIENT
BB HIGH
CONSTANT
CONSTANT
FLOW
FLOW ONE WAY
EXHALATION ONE WAY
VALVE
EXHALATION VALVE
VALVE
VALVE
OPENED
OPENED
PATIENT EXHALATION
PATIENT A. CMV Inspiration EXHALATION
A. CMV Inspiration VALVE
B. CMV Expiration, Spontaneous Inspiration VALVE
A. CMV Inspiration B. CMV Expiration, Spontaneous Inspiration OPENED
A. CMV Inspiration
B. CMV Expiration, Spontaneous Inspiration and Expiration
and Expiration
B. CMV Expiration, Spontaneous Inspiration and Expiration E150pptcy0606 REV. A
and Expiration PATIENT 9OPENED
PATIENT
NEWPORT BREEZE
ADVANCED FUNCTION
Unique Flow System - DUOFLOW
Duoflow
Duoflow allows
allows you you to
to control
control Mandatory
Mandatory andand Spontaneous
Spontaneous Flow Flow
separately.
separately. Mandatory
Mandatory Flow Flow isis delivered
delivered during
during mandatory
mandatory inspiration
inspiration
only,
only, Spontaneous
Spontaneous Flow Flow isis directed
directed through
through the
the circuit
circuit inin between
between
mandatory
mandatory breaths.
breaths. Spont.
Spont. Flow
Flow serves
serves as
as the
the source
source gas gas for
for spontaneous
spontaneous
breathing
breathing inin the
the SPONT
SPONT or or SIMV
SIMV Modes.
Modes.
Flow ISF
ISF DuoFlow
DuoFlow
Flow
system
system
Time
Time
Mandatory
MandatoryFlow
Flow
Spontaneous
SpontaneousFlow
Flow
Mandatory
Mandatory Spontaneous
Spontaneous
Breath
Breath Breath
Breath
E150pptcy0606 REV. A 10
NEWPORT BREEZE
ADVANCED FUNCTION
Unique Flow System - DUOFLOW
Spontaneous
Spontaneous Flow Flow isis calibrated
calibrated from
from
Minimum
Minimum to to 28
28 L/min.
L/min. WhenWhen thethe knob
knob isis
turned
turned toto "flush",
"flush", approx.
approx. 58 58 L/min.
L/min. isis
available.
available. Set Set Spont.
Spont. Flow
Flow to
to meet
meet pt.
pt.
spontaneous
spontaneous inspiratory
inspiratory demands
demands during
during
spontaneous
spontaneous breathing
breathing inin SIMV
SIMV and
and
Spontaneous
Spontaneous Modes Modes ofof Ventilation.
Ventilation. IfIf
patient
patient effort
effort exceeds
exceeds Spont.
Spont. Flow,
Flow, the
the
patient
patient may
may draw
draw mixed
mixed gas gas from
from the
the
Reservoir
Reservoir Bag.Bag. E150pptcy0606 REV. A 11
NEWPORT BREEZE
ADVANCED FUNCTION
Trigger Mechanism - 3 methods
Pressure
Pressure Trigger
Trigger Level
Level -- AA
trigger
trigger pressure
pressure isis set set on
on the
the
electronic
electronic manometer.
manometer. When When thethe
manometer
manometer needle needle crosses
crosses this
this
set
set level,
level, mandatory
mandatory inspiration
inspiration
will
will begin
begin ifif timing
timing isis
appropriate.
appropriate. This This control
control must
must bebe
manually
manually adjusted
adjusted to to
accommodate
accommodate PEEP PEEP levels.
levels.
E150pptcy0606 REV. A 12
NEWPORT BREEZE
ADVANCED FUNCTION
Trigger Mechanism - 3 methods
TIME
TIME--When
Whentiming
timingisis
right,
right,according
accordingtotothe
the
“SET
“SETRATE”
RATE”setting,
setting,
the
theBreeze
Breezeisistriggered
triggered
“ON”.
“ON”.
MANUAL
MANUAL--isisavailable
availablewith
withthe
the“MANUAL
“MANUAL
INFLATION”
E150pptcy0606 REV.INFLATION”
A
button.
button. 13
NEWPORT BREEZE
CLASSIFICATION
Electronically Controlled
Electrically and Pneumatically Driven
Time Cycled
Flow or Pressure Limited (Flow or Pressure may be held constant)
Pressure, Time or Manually Triggered
Output Control of Flow = Solenoid/pneumatic interface valve combo
Modes (Mechanical modes may be Pressure or Volume Controlled):
Control (CMV)
Assist/Control
SIMV / IMV (Continuous Flow with ISF)
Spontaneous/CPAP (Continuous
E150pptcy0606 REV. A Flow) 14
NEWPORT BREEZE
CONTROLS & SYSTEMS
Breath Control Panel Pressure Control Panel
Mode Selector
Preset
FIO2 Control
PEEP/CPAP Control
Mandatory Flow.
PIP Control
Inspiratory Time
PAW Monitor
Respiratory Rate
Patient Effort Indicator
Spontaneous Flow Trigger Level Control
Electronic Manometer
Trigger Level Indicator
E150pptcy0606 REV. A 15
NEWPORT BREEZE
CONTROLS & SYSTEMS
Additional Controls Rear Panel Controls
Manual Inflation
Pressure Relief Valve
Alarm Loudness
Power Entry Module
Nebulizer ON/OFF
Analog Output
Apnea Alarm Control
RS232
Low CPAP Alarm
E150pptcy0606 REV. A 16
NEWPORT BREEZE
PATIENT SET-UP
The
TheBreeze
Breezecancanbe
beset-up
set-upwith
withsmall
small (12
(12--15
15mm mmID)
ID)
or
orlarge
large(22
(22mmmmID)
ID)bore
bore circuit
circuittubing
tubingtoto
accommodate
accommodate ventilation
ventilationofofdifferent
differentsizes
sizes ofof patients.
patients.
Make
Makesure
suretotoappropriately
appropriatelyconnect
connectthetheproximal
proximal
sensing
sensingline
lineand
andthe
theexpiratory
expiratorydrive
driveline.
line.
E150pptcy0606 REV. A 17
NEWPORT BREEZE
PATIENT SET-UP
Volume Control
Set both the inspiratory time and the mandatory
flow to adjust tidal volume.
If the patient has a low inspiratory If the patient has an active
drive and can tolerate a longer respiratory drive with a naturally
inspiratory time, "I" time can be short natural inspiratory phase,
lengthened and flow lowered to or if you wish to decrease the
achieve the desired tidal volume. potential for auto-PEEP by
Flow 350 lengthening expiratory time, flow
mL's can be increased and "I" time
Inspiratory Time shortened to match the patient's
needs.
E150pptcy0606 REV. A 18
NEWPORT BREEZE
PATIENT SET-UP
Pressure Control
In all mechanical Pressure Controlled Modes, the
PIP knob is activated and can be set to limit the peak
inspiratory pressure for each breath.
E150pptcy0606 REV. A 22
NEWPORT BREEZE
PATIENT SET-UP
Alarms
Non-Patient Related Alarms
Cannot be silenced with the Alarm Silence Button.
Continuous ventilator alarm indicates electrical or
mechanical system failure.
Continuous mixer alarm indicates low inlet gas
pressure for either air or oxygen.
E150pptcy0606 REV. A 23