Monitoring Gaseous Pollutants: Wet Chemical Methods Ozone
Monitoring Gaseous Pollutants: Wet Chemical Methods Ozone
S ample Microproces s or
Inlet
O3 Output
P articulate
Filter
Detec tor
exhaus t
EC9810 Operation
The EC9810 ozone monitor is a US EPA equivalent method (U.V.)
photometer. This measurement principal is based on the ozone’s
characteristics of absorbing U.V.light at a wavelength of 253.7 nm.
Two measurements of a U.V. lamp’s light intensities are taken in a
measurement cycle. In the first, an air sample is passed directly to
the sample cell, the second is directed through the ozone scrubber.
The scrubber function is to selectively remove the ozone from the
sample for the reference portion of the measurement cycle,leaving
all other potential interference’s intact. These interference’s are
then compensated by the differential process of subtracting the
measurement cycle from the reference cycle to produce an ozone
measurement.
Theory of operation cont.
– To avoid interference from a number of compounds absorbing light at
254 nm, the predominant wavelength emitted by the UV lamp, such
compounds include aromatics HC,NO, SO2,H2O and others including
instability of the light source. The scrubber selectively removes O3 but
passes the interfering compounds. Thus, the intensity ratio described in
the previous paragraph is a function of only ozone absorption.
Ozone Analyzers in use in
India
• Envirotech commissioned the first Ozone
Analyzer at Hyderabad in 2006
• Early users of Ozone Analyzers supplied by
Envirotech:
– Tata Power, Mumbai
– UP Pollution Control Board – Kanpur, Varanasi and
Agra
– Sri Ramchandra Institute, Chennai
– Annamalai University, Chennai
– NIT, Trichy
Siting of Monitoring Station
Note that ozone is a reactive gas that is absorbed by vegetation,
measurements should therefore be carried out well away from plant
life. O3 Networks should sample 3-5m above ground level and more
than 20 m away from trees. Sampling manifolds should be made of
material that is inert to ozone, such as glass, Teflon,FEP, residence
time of the sample gas should be as short as possible. The sample
lines to the analyzer should be kept as short as possible to avoid
losses, and should be thermally isolated to prevent condensation.
Any ambient NO present in the sample air will react with some of
the ozone during the residence time in the sample line.
EC9810 Specifications
• Ozone Range: 0 - 20 ppm
• Resolution: 0.001 ppm
• Precision : 0.001 ppm
• Zero drift 30 day: 0.001 ppm
• Span drift 30 day: 0.002 ppm/C0 or 0.5%
• Temperature range: 5 - 40 C0
• EPA designated range: 0 - 50 to 0 - 1000 ppb
• Sample flow: 0 - 0.5 to 0 - 0.75 slpm