Professional Documents
Culture Documents
EMMI Manual
EMMI Manual
RD
SEMESTER ELECTRICAL
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR
Department Of Electrical Engineering
III
RD
Sem Electrical Engg.
Electrical Measurement & Measuring Instruments Lab
________________________________________________________________________
INDEX
Sr.
No.
Name of Experiment Page No.
1 Measurement of medium resistance by using
Voltmeter Ammeter method.
02
2 Measurement of medium resistance by using
Wheatstones bridge.
04
3 Measurement of high resistance by using loss of
charge method.
07
4 Measurement of low resistance by using Kelvins
double bridge.
09
5 Measurement of unknown inductance by using
Hays bridge.
12
6 Measurement of unknown inductance by using
Owens bridge.
15
7 Measurement of unknown inductance by using
Maxwell bridge.
17
8 Measurement of unknown capacitance by Desauty
bridge.
20
9 Measurement of unknown capacitance by Schering
bridge.
22
10 Measurement of 3-phase power by the one-
watt meter method.
24
11 Measurement of 3-phase power by the two-watt
meter method.
26
12 Measurement of Reactive power in 3- phase circuit
by Wattmeter method.
28
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 2
Experiment No-01
Aim:- Measurement of medium resistance by the Voltmeter , Ammeter method.
Apparatus:- DC ammeter(0-50mA)
DC Voltmeter (0-50V)
Dc power supply (0-30V)
Variable Resistance -100 ohm.
Connecting wires.
Circuit Diagram:-
V
A
V
A
(0-30v)
(0-50mA)
R
+
_
(0-30v)
(0-50mA)
(0-50v)
R
+ - -
+
-
(0-50V)
100 Ohm
100
Ohm
Fig.(1) Fig. (2)
Theory: -
Two types of the connections are done in this method. Ammeter voltmeter
method is shown in the figure (2) voltmeter and ammeter are connected in series, where
ammeter measures the total current flowing through the circuit and voltmeter measures
the voltage across the unknown resistance .The voltmeter should have ideally infinite
resistance and ammeter should have ideally zero resistance so that it will measure total
current flowing through the unknown resistance. But practically it is not possible and
measured value Rm of the resistance is the sum of resistance of ammeter and actual
resistance.
Rm =R
1
+Ra
Where
R
1
=Actual resistance.
Ra=Resistance of the ammeter.
It is clear from the expression that the value of measured resistance is equal to actual
resistance when ammeter has zero resistance.
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 3
Observation Table:-
For CASE 1 voltmeter Ammeter Method
Voltage (volt) Current (Amp) Resistance
(calculated)
Resistance
(measured)
For CASE 2 Ammeter voltmeter Method
Voltage (volt) Current (Amp) Resistance
(calculated)
Resistance
(measured)
Procedure:-
1) Make the connections as per circuit diagram.
2) Switch on the supply and note down the readings of ammeter and voltmeter.
3) Calculate the value of the unknown resistance by ohms low.
4) Perform the procedure for the other case similarly.
Result: - Hence the measured value of the unknown resistance is
found to be _________.
Viva Questions:-
1) What are the other methods of measurement of medium resistance?
2) What are the disadvantages of this method?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 4
Experiment No.-02
Aim: - Measurement of the medium resistance by using wheatstone bridge.
Apparatus: -
Power supply (0-32V D.C)
Resistor: - R
1
=1K, R
X
=1K, R
3
=1K, R
4
=10K.
Unknown resistor=100,
Wheatstone bridge kit.
Digital multimeter-1no,
Patch chords.
Circuit Diagram:-
G
R3
R1
R4
Rx
0-32v
a
b
c
d
I1
I2
I3
I4
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 5
Theory-
Wheat stones bridge is very important device used in the measurement of medium
resistances. It is an accurate and reliable instrument .The wheat stone bridge is an
instrument based on the principle of null indication and comparison measurements .
The basic circuit of a wheat stone bridge is shown in fig . it has four resistive
arms, consisting of resistances R
1
,R
X
,R
3
and R
4
together with a source of emf and a null
detector , usually a galvanometer G or other sensitive current meter is used. The bridge is
said to be balanced when there is no current through the galvanometer or when the
potential difference across the galvanometer is zero. This occurs when the voltage from
point a to point d equals the voltage from point d to point b or by referring to other
battery terminal, when the voltage from point a to point c equals the voltage from
point c to point b.
For bridge balance;
I
1
=I
3=
E/(R
1
+R
3
) (1)
I
2
=I
4
=E/(R
X
+R
4
) (2)
E=emf of battery.
Combining equ (1) and (2) we get
R
X
*R
3
=R
1
*R
4
OR
R
X
=(R
1
*R
4
)/R
3
Where R
X
is the unknown resistance, R
1
, R
3
and R
4
are called the ratio arms.
Observation Table: -
Ratio Arm
Resistor
Std. Arm
Resistor
Measured Rx Calculated
Rx
R
1
R
3
R
4
Procedure: -
1) Connect the patch chords as per the circuit diagram.
2) Note the resistance of R
1
,R
4
and R
3
using multimeter.
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 6
3) Switch on the power supply and adjust the resistance R4 such that galvanometer
shows the zero deflection.
4) Disconnect the supply & measure the value of R
X
.
5) Now calculate the value of unknown Resistor R, Using formula
Rx=(R
1
*R
4
)/R
3
.
Result:- Unknown Resistance found to be __________.
Viva Questions: -
1) What are the other methods used for measurement of medium resistance?
2) Why we use this method for measurement of medium resistance?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 7
Experiment No-03
Aim:- Measurement of the high resistance by using loss of charge method.
Apparatus:-
Multimeter 1no
Voltmeter (0-30v)-1no
Capacitor-10uf-1no
Resister-100K-1no
Power supply-(0-30v)-1no
.
Circuit Diagram:-
V
(0-30V)
S1
S2
C=10uf
R
(0-30V)
Theory:-
In this method the resistance which is measured is connected in parallel with
the capacitor C and the electronic voltmeter V. The capacitor is the charged up to some
suitable voltage by means of the battery having the voltage V and is then allowed to
discharge through the resistance.
The terminal voltage is observed over the considerable period of the time
during discharge.
Let,
V=initial voltage on the charged capacitor.
v=instantaneous discharging voltage.
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 8
t=the charging time of the capacitor.
V=v exp (-t/C*R)
or
V/v=exp (-t/C*R)
or
Insulation resistance
R=t/(Clog V/v)
=0.4343*t/(C log
10
V/v)
R=0.4343*t/(C log
10
V/v)
Where R is a resistance of R
1
& R in parallel
R
1
represents the leakage Resistance &
R is the unknown resistor.
The test is then repeated with the unknown resistance R disconnected & the
capacitor discharging through R
1
R=(R*R1)/(R+R1)
Observation Table:-
S.
NO.
Time (sec) V
(withoutR)
V(with R) Log
10
(V/v)
without R
Log
10
(V/v)
with R
Procedure:-
1) Connections is make as per the circuit diagram.
2) Close the switch S
1
and keep S
2
open till the capacitor charge upto V volts.
3) Now open the switch S
1
& allow the capacitor to discharge by its Owen
leakage resistance
4) Note down the reading of the voltmeter verses equal interval of the time.
5) Calculate the unknown resistance using the formula.
6) Now close the switch S
1
& S
2
till the capacitor charge upto V volts.
7)Now open the switch S
1
& allow the capacitor to discharge through the unknown
Resistance.
Result:- Unknown value of high resistance is calculated by using loss of charge
method.
Viva Questions:-
1) Why this method is called as loss of charge method?
2)What kind of errors occurs while performing this practical?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 9
Experiment No:-04
Aim: - Measurement of the low resistance by using Kelvins Double bridge .
Apparatus: -
Regulated dc supply-1no
Standard resistance coil-1no
Kelvins double bridge kit.
Digital multimeter-1no,
Patch chords.
Components Used: -
Q=q=10K, R
X1
=5, R
X2
=10, P=p=100,
R
B
=11K,, S=Pot of 1k.
Circuit Diagram: -
G
P Q
p
q
r
E
a
m n
S
R
R
b
c
d
Rb
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 10
Theory: -
Kelvins bridge is a modification of wheatstone bridge and always used in
measurement of low resistance. It uses two sets of ratio arms and the four terminal
resistances for the low resistance. As shown in above fig. The first set of ratio P and Q &
second set of ratio arms are p and q are used to connect to the galvanometer to a pt d at
an Approx. potential between points m and n to eliminate the effects of connecting
lead of resistance r between the known std. resistance s and unknown resistance R.
The ratio P/Q is made equal to p/q. under balanced condition there is no current flowing
through galvanometer which means voltage drop between a and b, Eab equal to the
voltage drop between a and d, Eamd.
Now
Eab = Eamd
Eab=(P*Eac/P+Q); Eac=I[R+S+[(p+q)r/p+q+r]] ---------(1)
Eamd= I[R+ p/p+q[ (p+q)r/p+q+r]]
=I[R+p*r/(p+q+r)] ----------------------- (2)
For zero deflection->
Eab=Eamd
[ P/P+Q]I[R+S+{(p+q)r/p+q+r}]=I[R+pr/p+q+r] --3)
Now, if
P/Q=p/q
Then equa (3) becomes
R=P/Q=S -(4)
Equation (4) is the usual working equation. For the Kelvins Double Bridge .It indicates
the resistance of connecting lead r. It has no effect on measurement provided that the two
sets of ratio arms have equal ratios. Equation (3) is useful however as it shows the error
that is introduced in case the ratios are not exactly equal. It indicates that it is desirable to
keep r as small as possible in order to minimize the error in case there is a diff. between
the ratio P/Q and p/q.
R=(P*S)/Q
Observation Table: -
P (ratio arm
resistor)
Q (ratio arm
resistor)
Standard resistor
S
R measured
value
Calculated
R
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 11
Procedure: -
1) The circuit configuration on the panel is studied.
2) The unknown resistance is connected as shown.
3) The value of P,Q was selected such that
P/Q=p/q
4) S was adjusted for proper balance i.e., galvanometer to shows zero deflection.
5) The value of Unknown Resistance R=(P*S/Q) was calculated.
Precautions-
1) Check all the connections before turning ON the power supply.
2) Note the readings accurately.
Result- The observed value of unknown resistance is __________.
Viva Questions:
1) Why this method is called as double bridge method?
2) Can this method be beneficial for measurement of low value of
Resistance or not?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 12
Experiment No :-5
Aim: - Measurement of the unknown Inductance by using Hays bridge.
Apparatus:- Multimeter
LCR meter
Hays bridge kit,
Patch cords.
Components Used: -
R
2
=100, R
3
= 4.7K, R
4
= 100, C
4
= 1uf
Circuit Diagram:-
G
R3
A
D
B
C
L1
R1
R4
R2
C4
I1
I2
I1
I2
E
E1
E3
E2
E4
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 13
Observation:-
R
2
= __________.
C
4
= __________.
For Unknown Inductance Lx1:-
Calculated - Lx1 = (R
2
R
3
C
4
)
/(1+
2
R
4
2
C
4
2
).
Lx1 =________mH.
Measured - Lx1 = ________mH.
Quality factor (Q)=1/R
4
C
4
.
For Unknown Inductance Lx2:
Calculated - Lx2 = (R
2
R
3
C
4
)
/(1+
2
R
4
2
C
4
2
).
Lx2 =________mH.
Measured - Lx2 = ________mH.
Quality factor (Q)=R
4
C
4
.
Theory:-
The hays bridge is the modification of the Maxwell Bridge. This bridge uses a
resistance in series with the standard capacitor. The bridge has four resistive arms in
which the arms one is consists of the resister R
1
, Lx .The arm 2 is consists of the variable
resistance R
3
.The low value of the resistance is obtain by the low resistive arms of the
bridge. The value of R
4
and C
4
is the standard value of the capacitor and resistance.
By using the unknown inductance having a resistanceR
1
. R
2
, R
3
,R
4
-is the known
non-inductive resistance and C
4
is standard value of the capacitor. The unknown value of
inductance and Quality factor of the Bridge is obtained by formula.
Lx = (R
2
R
3
C
4
)
/(1+
2
R
4
2
C
4
2
)
Quality factor (Q)=(1/R
4
C
4
)
For value of Q greater 10, the term (1/Q)
2
will be smaller & hence neglected
.
Therefore Lx= R2*R3*C4
Basic AC bridges consist of four arms, source excitation and a balanced detector.
Commonly used detectors for AC bridges are:
(1) Head phones
(2) Vibration galvanometers
(3) Tunable amplifier detectors
Vibration galvanometer is extremely useful at power and low audio
frequency ranges. Vibration galvanometers are manufactured to work at various
frequency ranging from 5 KHZ to 1 KHZ. But one most commonly used between
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 14
200HZ.
Advantage-1) This Bridge gives very simple expression for unknown for High Q coil.
2) This bridge also gives a simple expression for Q factor.
Disadvantage- 1) The hays bridge is suited for the measurement of the High Q inductor.
2) It is used to find the inductor having the q value of the smaller then 10.
Procedure:-
1) Study the circuit provided on the front panel of the kit.
2) Connect unknown inductance L
X1
in the circuit. Make all connections to
complete the bridge.
3) Put the supply ON
4) Set the null point of galvanometer by adjusting variable resistance R
3
.
5) Note value of R
2
, R
3
, and C
4
by removing connection by patch cords.
6) Calculate theoretical value of L
X1
using L=R
2
R
3
C
4
7) Measure value of LX
2
by LCR meter and compare it.
8) Repeat process for LX
2
.
Result:- The unknown inductance is measured using Hays bridge and is found to
be___
Viva Questions:-
1) What is the Q factor of the coil?
2) Which bridges are used for measurement of inductances?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 15
Experiment No :-6
Aim:- Measurement of the unknown inductance by using OWENS bridge method
Apparatus:-
Digital multimeter, Patch chords.
Components used:
R
3
=1K(pot), C
4
=1uF, R
2
=1K(pot),
L
1
= L
2
= ------
Galvanometer, 12VAC source.
Circuit Diagram:-
G
A
D
B
C
R1L1
R3
C4
R2
C2
Theory:-
Bridge are used for the accurate measurement of electrical quantities viz;
esistance, Capacitance, Inductance, Storage Factor, Loss factor etc. Depending upon the
excitations used , the bridge are classified as AC bridges & DC bridges.
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 16
Owens Inductance Capacitance Bridge comes under category of AC bridges and it is
used for measurement of an Inductance in terms of capacitance. General form of AC
bridges consists of four arms of impedances & AC excitation.
Let
L
1
= unknown self inductance of resistance
R
2
=variable non- inductive resistance
R
3
= fixed non- inductive resistance
C
2
=variable standard capacitor
C
4
=fixed standard capacitor
And
At balance condition,
(R
1
+jwL
1
) (1/jwC
4
)=(R
2
+1/jwC
2
)*R
3
Separating the real & imaginary terms, we obtain:
L
1
=R
2
R
3
C
4
&
R
1
=R
3
*C
4
/C
2
Procedure:-
1. Study the circuit provided on the front panel of the kit.
2. Connect unknown inductance L
X1
OR L
X2
by patch chords
3. Switch ON power supply
4. By varying the pot R
2
& R
9
make bridge balance
5. Switch off the power supply, disconnect the patch chords & measure the value of
R
3
& R
2
6. Find the value of unknown inductance by using formula
7. L
1
=R
2
R
3
C
4
Observations:
Measured value of L Calculated value of L
Result:-
Using Owens Bridge we can calculate the value of unknown inductance &
we found there is short difference between theoretical and practical value of L.
VIVA QUESTIONS:-
1) Draw the Phaser diagram of OWEN Bridge?
2) What are the balancing conditions of any a.c. bridge?
DEPARTMENT OF ELECTRICAL ENGINEERING III
RD
SEMESTER ELECTRICAL
1
G.H.RAISONI COLLEGE OF ENGINEERING, NAGPUR 17
Experiment no 7
Aim:- Measurement of the unknown inductance by using Maxwells bridge.
Apparatus:-
Digital multimeter,
Patch chords.
R
2
=100, R
4
=1K, R
3
=4.7K,
C
4
=1f, L
X2=
318mH, L
X1
=73 mH
Circuit Diagram:-
G
R3
A
D
B
C
L1
R1
R2
C4
R4
E
I1
I3
I2
I4
Theory:-
The Maxwell