You are on page 1of 10

1213

Optimizing Imaging Parameters


Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

for MR Evaluation of the Spine


with Titanium Pedicle Screws

Cheryl A. Petersilge1 OBJECTIVE. This study examined the contribution and Interdependence of multiple
Jonathan S. Lewi& imaging parameters in clinical imaging sequences to aid practicing radiologists In
minimizing artifacts during MR imaging of the spine after implantation of titanium
Jeffrey L. Duerk1
pedicle screws.
Jung U. Yoo2
MATERIALS AND METHODS. A lumbar spine specimen with titanium pedicle
Alexander J. Ghaneyem3 screws implanted in the pedicle was imaged with a 1 .5-T scanner. Sequence type,
voxel volume, TE, and bandwidth varied. Different voxel volumes were achieved by
altering section thickness, field of view (FOV), and matrix size. Artifact size was mea-
sured on sagittal and axial images at the midpedlcle level. Artifact size was expressed
as a percentage of actual screw size, and mean artifact size was calculated for each
sequence. Analysis of variance without replication was done.
RESULTS. Mean artifact size ranged from 231% to 364% of actual screw size. Arti-
fact size was independent of voxel volume for voxels greater than 3 mm3 (p < .001).
Artifact size decreased significantly (p < .001) when voxel volume was less than I
mm3. When we increased slice thickness and maintained a constant voxel volume,
artifact size decreased. Decreases in artifact size correlated with a reduction in the
ratio of the FOV to the number of pixels in the frequency-encoding direction (Nx). Arti-
fact sizes were smallest when fast spin-echo sequences were used. Conventional
spin-echo sequences produced artifacts that were smaller than the artifacts produced
by the gradient-echo sequences. Decreasing the TE did not diminish artifact size for
conventional spin-echo images at larger voxel volumes.
CONCLUSION. Although voxel volume has been recognized as a factor that affects
artifact size, the role of other contributing factors-slice thickness, number of phase-
encoding steps, and FOVIN,-has not been evaluated before. Artifact reduction
proved to be dependent only on FOV/Nx. Artifact size was reduced by the use of fast
spin-echo sequences. With conventional spin-echo sequences, TE should be mini-
mized, although other technical factors may outweigh the gain in artifact reduction.

AJR 1996;166:1213-1218

MR imaging is currently the technique of choice for evaluation of the spine after
surgery. In the past, MR studies have been severely compromised in patients with
Received October 27, 1995; accepted after revi- spinal instrumentation made of stainless steel, especially low-nickel stainless
sion December 12, 1995. steel [1-5]. Today, several types of metallic implants are made of titanium. The
1 Department of Radiology, University Hospitals reason for the shift includes better biocompatibility and the physical advantages of
of Cleveland, Case Western Aeserve University,
titanium over stainless steel, including increased resistance to corrosion fatigue
11100 Euclid Ave., Cleveland, OH 44106. Address
correspondence to C. A. Petersilge. and lack of hypersensitivity response by the body [6]. Titanium also creates signif-
2Department of Orthopedics, University Hospi- icantly less distortion on MR images than does stainless steel [4, 5, 7-16].
fats of Cleveland, Case Western Reserve Universi- Improved imaging of the spine after surgery is one of the main reasons that tita-
ty, 11100 Euclid Ave., Cleveland, OH 44106.
nium rather than stainless steel pedicle screws are being used at our institution
3Department of Orthopedics, Loyola University and others in the United States.
Medical Center, 2160 South First Ave., Maywood,
IL 60153. A number of imaging factors decrease artifacts produced by metallic implants,
including shorter echo times, lower field strengths, higher readout bandwidths
0361-803X/96/1 665-1213
© American Roentgen Ray Society (Bws), and smaller voxel sizes [2, 17-22]. In previous studies, each of these
1214 PETERSILGE ET AL. AJR:166, May 1996

parameters has been isolated from all other parameters. To rim of high signal amplitude that followed the shape of the titanium
our knowledge, the contribution by, and the interdependence screw. We measured artifact along the short axis of the screw. This
of, each of these factors in clinical imaging after hardware axis corresponded to the frequency-encoding axis in each plane.
The artifact on gradient-echo images was so severe that we did not
implantation have not been previously studied.
measure it. One author made all measurements. For all sequences,
This study was undertaken to assist practicing radiologists
artifact size for each screw was measured at the midpedicle level on
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

in determining the optimal imaging parameters on a 1 .5-T


the image that contained the most welt-defined artifact. Artifact size
horizontal-bore MR imager for evaluation of the spine after was calculated and expressed as a percentage of actual screw size,
implantation of titanium hardware. using this equation:

Materials and Methods Artifact size on image (cm) 1 .


. . . . x x 100 = Anttfact size, (1)
Magnification correction factor 0.65 cm
We obtained a lumbar spine specimen that included the five lum-
bar vertebral bodies, the sacrum, and the coccyx. All paraspinal soft
tissues were stripped. The thecal sac, lower portion of the spinal
where 0.65 cm is the actual diameter of the titanium screw, and the
cord, and nerve roots remained. Using standard surgical technique, magnification correction factor is derived by measuring the width of
we placed pedicle screws that were 5.5 cm long with a 6.5-mm a 1-cm marker on the actual MR image and then dividing that mea-
outer diameter (Danek Group, Inc., Memphis, TN) bilaterally at L3, surement by 1 cm.
L4, and L5. The spine specimen was suspended in a water bath in For each sequence, the data from one screw consistently resulted
the prone position. The water bath was placed on a linearly polar- in a smatter artifact than the artifacts from the five other screws. On
ized lumbar spine coil with the spine parallel to the static magnetic review, we found this screw to be oblique to the main magnetic field,
field and imaged with a 1 .5-T Magnetom scanner (Siemens, Erlan- whereas the five other screws were all perpendicular to that field. We
gen, Germany). chose to omit data for this screw from our calculations.
We designed imaging parameters to vary 6w, TE, voxet volume, We combined the measurements from the sagittal and axial planes
and type of pulse sequence. Pulse sequences included conven- for each imaging sequence to yield 1 0 data points for each sequence.
tionat spin-echo, short-effective-echo-time fast spin-echo, and gradi-
Mean artifact size for each sequence was then calculated. We did an
ent-echo techniques. Imaging parameters for our sequences appear analysis of variance without replication using Microsoft Excel 4.0
in Table 1 . Additional parameters include an interslice gap equal to
20% of the slice thickness and two excitations. We used an echo-
train length of four in the fast spin-echo sequences. The axial and
sagittat images were obtained with the phase-encoding axis on- TABLE 1 : Imaging Parameters
ented along the length of the screw. We varied voxet volumes by
systematically altering section thickness, field of view (FOV), and TR TE Bandwidth
Sequence Type
matrix size. Parameters for our different voxel volumes are pre- (msec) (msec) (Hz/pixel)
sented in Table 2. We obtained images in the axial and sagittal
Conventional spin-echo 575 10 130
planes for all sequences. Coronal images were initially obtained, but
Conventional spin-echo 690 15 130
later omitted because the spear shape of the resulting artifact pre-
Fastspin-echo 1700 15 178
cluded its accurate measurement (Fig. 1).
Gradient-echo (3D FLASH) 20 5 260
was measured on the axial and sagittat images (Fig. 1).
Artifact
We defined the artifact as the area of signal void surrounded by a Note-3D FLASH = three-di mensionat fa st low-angle shot.

Fig. 1 -MR images after placement of titanium pedicle screws at L3, L4, and L5.
A-C, Axial (A), sagittal (B), and coronal (C) fast spin-echo MR images show significant anatomic detail despite screws. Marrow-replacing process is
clearly visible in middle vertebral body on coronal image. Arrows in A and B show artifact at midpedicle level.
AJR:166, May 1996 MR IMAGING OF SPINE WITH TITANIUM SCREWS 1215

(Microsoft, Redmond, WA) to determine statistical significance voxel volume resulted in lower artifact for the thicker sec-
among voxel volumes of 3, 4, and 6 mm3 for each imaging sequence tions (with smaller pixel sizes), although this decrease was
and between voxel volumes of 1 and 3 mm3 for each imaging not statistically significant (p > .05).
sequence. We used the same software to determine statistical signifi- Fast spin-echo sequences produced significantly smaller
cance among imaging sequences at each voxet volume.
artifacts than conventional spin-echo sequences at all voxel
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

volumes (p < .001). Artifact for the conventional spin-echo


Results sequences with the shorter TE sequence was greater than
Mean artifact size as a function of voxel volume for each with the longer TE sequence. We observed this phenomenon
sequence is plotted in Figure 2 and shown in Figure 3. Mean at all voxel volumes greater than 3 mm3. Statistical analysis
artifact size ranged from 231 % to 364% of actual screw size. revealed that mean artifact size at each voxel volume for the
For a given imaging sequence at constant section thickness, longer TE sequence was significantly different from that for
we found no statistically significant difference among voxel the shorten TE sequence, even though the corresponding
volumes of 3, 4, and 6 mm3 (p .05). These voxel volumes
> curves crossed between voxel volumes of 1 mm3 and 3 mm3.
correspond to FOVs, section thicknesses, and matrices
commonly used for MR imaging of the spine. We found a
Discussion
significant decrease in mean artifact size (p .001) between <

voxel volumes of 1 mm3 and 3 mm3 for each imaging When metallic objects are located within the imaging field,
sequence. When section thickness was 3 mm, the mean artifacts are created by differences in magnetic susceptibility
artifact size decreased from a range of 31 5% to 364% for 3- of the metal and adjacent tissues [2, 1 2, 1 7, 23]. These arti-
mm3 voxels to a range of 263% to 303% for 1 -mm3 voxels. facts vary according to the metal from which the implant is
With section thickness at 4 mm, artifact size increased from manufactured as well as its shape and orientation to the
231% for 1-mm3 voxels to 287% for 3-mm3 voxels. Thus, main magnetic field [1 , 7, 12, 1 3, 17, 18, 23, 24]. Artifacts
increasing section thickness while maintaining a constant due to magnetic susceptibility are greatest with fenromag-

Fig. 2.-Graph plotting artifact size


360
(expressed as a percentage of actual
screw size) against voxel volume for
340
each imaging sequence shows artifact
reduction only when voxel volume is 320
less than 3 mm3. --TR/TE 575110 CONVENTIONAl. SPIN ECHO

300
---6NOII5CONVEN11ONAL SPIN ECHO

ARTIFACT SIZE (%) 280 I-.-iioo’scnvn SECT)ON THICKNESS)FAST SPIN


ECHO
-6--llOQflS(4nvn SECTiON THICKJ4ESS)FAST SPIN
260 [ECHO

240

220

200
0 1 2 3 4 5 6

VOXELVOLUME(MM3)

Fig. 3.-Effect of larger voxel volume on MR Imaqing of titanjum screw in pedicle.


A-C, Axial MR images at voxel volumes of 1 mm (A), 3 mm” (B), and 6 mm3 (C) depict direct relationship between artifact size and voxel volume.
Especially note overall image degradation due to decreased resolution at larger voxel volumes.
1216 PETERSILGE ET AL. AJR:166, May 1996

netic materials such as stainless steel. Artifact caused by tens are held constant [25-28]. Gradient-echo sequences result
titanium is markedly smaller because of the metal’s much in the most severe magnetic susceptibility artifacts because the
lower magnetic susceptibility. technique does not use refocusing pulses [2, 19].
The difference in magnetic susceptibilities of adjacent tis- Our results show a consistent relationship between voxel
sues or of tissue and implant creates local inhomogeneities volume and artifact size for each imaging sequence. Voxel
in the magnetic field. These field inhomogeneities alter the volumes of 3-6 mm3, which are currently used for imaging
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

phase and frequency of the local spins. The spins are subse- the spine, create an artifact that is independent of the voxel
quently mapped to an erroneous location within the image, size for each imaging sequence. Diminution of artifact is not
creating distortions in the shape of the object along the fre- achieved until voxel volumes approach 1 mm3. This relation-
quency-encoded and slice-select axes. Loss of signal within ship between voxel volume and artifact size is explained by
the object also results [17, 1 9]. A rim of high signal intensity examining the relationship among FOV, the number of pixels
appears around the periphery of the metallic object as a in the frequency-encoding direction (Nx), and artifact size. To
result of mismapping a disproportionate number of spins to examine this relationship, we rewrote Equation 2, multiplying
that location [1 7]. Orientation of the phase- and frequency- both sides by Nx/FOV to yield:
encoding axes also affects artifact appearance [2, 7, 1 7, 19].
The spatial shift (Sx) seen on MR images is directly propor- &(BOFOV
x 2BWNx (3)
tional to the main magnetic field and inversely proportional to
the strength of the readout gradient:
where Bw = (xFOt’x and y = HzfTesIa.

x 2Gx (2) A review of our imaging parameters (Tables 1 and 2)


reveals that for voxel volumes from 3 to 6 mm3, we held con-
stant all parameters for a given sequence including FOV, Nx,
In Equation 2, x the shift of the interface (on the MR and Bw. When these parameters are constant, 5x (or image

image) between regions of different magnetic susceptibility distortion) will be constant. Voxel volumes from 3 to 6 mm3
in centimeters, LX is related to susceptibility differences, B0 were achieved by changing only the number of phase-
is the main magnetic field measured in Tesla, and Gx is the encoding steps. (The number of phase-encoding steps is not
readout gradient (Tesla/cm) [17]. a factor in Equation 3.)
Local eddy currents produced within the implant by the RF When voxel size is decreased to 1 mm3, the FOV/Nx also
pulse may also contribute to local field inhomogeneities and decreases. Thus, 5x must also decrease as described by

subsequent artifact [24]. In many imaging sequences, eddy Equation 3. We can conclude that the most important parame-
currents play a secondary role because they are very short ten in decreasing metallic artifact is not voxel volume but more
lived, decaying before the echo is acquired. specifically FOV/Nx, which describes the resolution along the
Local field distortions are reduced by shorter echo times, frequency-encoding axis. This relationship between artifact
lower field strengths, higher readout Bws, and smaller voxel vol- size and FOV/Nx is shown in Figure 4.
umes [2, 17-22]. Other imaging parameters such as TR do not A decreased FOV/Nx also explains why artifacts are
play a significant role in artifact reduction. The type of pulse smaller with thicker sections (Table 2). Because FOV, Nx, and
sequence chosen will also affect the extent of image degrada- Bw are parameters that are readily available for any MR imag-
tion by magnetic susceptibility artifact. In spin-echo imaging, ing sequence, and are routinely changed to customize
each 180 refocusing pulse compensates for dephasing. In fast sequences, one can easily use the relationship described in
spin-echo imaging, the multiple closely spaced 180* pulses pro- Equation 3 to compare expected artifact size for different
duce magnetic susceptibility artifacts that are less severe than imaging sequences. This relationship can also be used to
those in conventional spin-echo imaging, especially when short- determine the minimum resolution at which a sequence with a
effective-echo-time sequences are used and all other parame- different Bw must be run to maintain artifact size, assuming

360

340

320
-+-TR/TE 575/IOCONVENT1ONAI. SPIN ECHO

300 -a-NSPI500NVENT)ONALSPIN ECHO

ARTIFACTSIZE (%) 280 -.- I7oISi3nvn SECTIONTHICKNESS)FAST SPIN


ECHO
-#{232}-I7O5IS(4mm SECTiON THICIOESS)FA5T SPIN
260 ECHO

240
Fig. 4.-Graph plotting artIfact size
220 (expressed as a percentage of actual
screw size) against ratio of field of view
to number of pixels In frequency-en-
0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 coding direction (FOVINx for each Im-
aging sequence reveals effect of
FOV/N5
resolution on artifact sIze.
AJR:166, May 1996 MR IMAGING OF SPINE WITH TITANIUM SCREWS 1217

TABLE 2: Factors Varied to Create Different Voxel Volumes mm3. Why this unexpected relationship did not occur at 1
mm3 is also unclear. Explanation of this phenomenon awaits
No. of
Voxel Section Field of further investigation.
Phase- N a FOV/Nb
Volume Thickness View x Our results confirm a significant decrease in artifact from
(mm) (mm) Encoding
(mm3) titanium pedicle screws when fast spin-echo imaging was
Steps
compared with conventional spin-echo imaging, using clinically
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

6 3 256 128 256 1.00


relevant parameters. Our fast spin-echo sequence had the
4 3 256 192 256 1.00
advantages of multiple 180’ pulses and a higher Bw. The mdi-
3 3 256 256 256 1.00
1
3
150 256 256 0.59
vidual contribution of each factor was not studied because Bw
3d is not an independently variable parameter with available soft-
300 512 512 0.59
6 4 222 128 256 0.91 ware. Our results corroborate those of Tartaglino et al. [25],
4 4 222 192 256 0.91 although they compared T2-weighted conventional spin-echo
3 4 222 256 256 0.91 images with fast spin-echo images whereas we compared Ti -

1 4 130 256 256 0.53 weighted conventional spin-echo with short-effective-echo-


4d
260 512 512 0.53 time fast spin-echo images.
aN number of pixels in frequency-encoding direction. In conclusion, magnetic susceptibility artifact from titanium
bFOv/N = ratio of field of view to number of pixels in frequency-encoding
pedicle screws was minimized by using the smallest possi-
direction. ble FOV/Nx We found that the number of phase-encoding
cSagittal plane.
plane. steps did not alter artifact size. Artifact was reduced by the
use of fast spin-echo sequences. When conventional spin-
echo sequences must be used, the smallest feasible FOV/
Nx should be chosen, and TE selected with care, because
that other imaging parameters-such as sequence type- the benefits achieved with the shortest possible TE may be
remain constant. outweighed by other technical factors.
Distortion also occurs along the slice-select axis [17] and
will be manifested in the MR image as artifact. This study
REFERENCES
found that with section thicknesses between 3 and 4 mm,
the decrease in artifact size from the increased in-plane nes- 1 . Ballock AT, Hajek PC, Byrne TP, Garfin SR. The quality of magnetic reso-
olution outweighs any artifact increase from greaten section nance imaging, as affected by the composition of the halo orthosis. J Bone
thickness. The effect of section thickness could be best Joint SurgAm 1989:71-A:431-434
2. Czervionke LF, Daniels DL, Werhti FW, et al. Magnetic susceptibility an-
defined by examining a wider range of section thicknesses;
facts in gradient-recalled echo MR imaging. AJNR 1988:9:1149-1155
however, section thicknesses greater than 4 mm on less than 3. Lyons CJ, Betz AR, Mesgarzadeh M, Revesz G, Bonakdarpour A, Clancy
3 mm are not practical for imaging the spine. Therefore, we M. The effect of magnetic resonance imaging on metal spine implants.
did not evaluate them. Spine 1989;14:670-672
4. Rupp R, Ebraheim NA, Savolaine ER, Jackson WT. Magnetic resonance
Additionally we did not evaluate the effect of orientation of
imaging evaluation ofthe spine with metal implants. Spine 1993;18:379-385
the pedicle screw to the main magnetic field as a means of 5. Sullivan PK, Smith JF, Rozzelle AA. Cranio-orbital reconstruction: safety
reducing artifact. The relationship between the size of arti- and image quality of metallic implants on CT and MRI scanning. Plast
fact produced by a cylindrical metallic device and the onien- Reconstr Surg 1994:94:589-596
tation of the device to the main magnetic field is described 6. Williams DF. Titanium as a metal for implantation: Part 2. Biological prop-
erties and clinical applications. J Med Eng Technol 1977:1:266-270
elsewhere [12]. Mueller et al. [12] found that artifact size is
7. Augustiny N, von Schulthess GK, Meier D, Bosiger P. MR imaging of large
greatest when the device is perpendicular to the static mag- nonferromagnetic metallic implants at 1 ST. J Comput Assist Tomogr
netic field and decreases in size as the long axis of the 1987:11:678-683
device becomes parallel to the static field. Similarly, we 8. Ebraheim NA, Coombs RE, Jackson WT, Zeiss J. The effect of metallic
implants on magnetic resonance imaging. J Bone Joint Surg Am 1991;
observed a decrease in artifact size from the one screw that
73-A: 1397-1398
was obliquely oriented rather than perpendicularly oriented 9. Ebraheim NA, Savolaine ER, Stitgen SH, Jackson WI Magnetic reso-
to the main magnetic field. However, we did not believe that nance imaging after pedicular screw fixation of the spine. C/in Orthop
changing the orientation of the screw relative to the main 1992;279:133-1 37
magnetic field is a practical imaging strategy once the screw 10. Ebraheim NA, Savolaine ER, Zeiss J, Jackson WT. Titanium hip implants
for improved magnetic resonance and computed tomography examina-
has been implanted within a patient.
tions. C/in Orthop 1992:275:194-1 98
We were surprised that decreasing TE did not reduce the 11 . Mirvis SE, Geisler F, Joslyn JN, Zrebeet H. Use of titanium wire in cervical
mean artifact size for conventional spin-echo sequences at spine fixation as a means to reduce MR artifacts. AJNR 1988:9:1229-1231
voxel volumes of 3-6 mm3. The reasons are not clean. We 12. Mueller PR, Stark DD, Simeone JF, et al. MR-guided aspiration biopsy:
needle design and clinical trials. Radiology 1986:161:605-609
suspect that the RF pulse truncation that produces the short
13. New PFJ, Rosen BR, Brady TJ, et at. Potential hazards and artifacts of ferro-
echo-time sequence may be responsible. RF pulse trunca- magnetic and nonferromagnetic surgical and dental materials and devices in
tion results in broadening the slice profile, essentially nuclear magnetic resonance imaging. Radiology 1983;147: 139-1 48
increasing section thickness, and thus voxel volumes are 14. Teitelbaum GP, Bradley WG Jr, Klein BD. MR imaging artifacts, ferromag-
netism, and magnetic torque of intravascular filters, stents, and coils.
larger than would be anticipated from the imaging parame-
Radiology 1988; 166:657-664
tens. This enlargement of voxel volume may account for the
1 5. Tominaga T, Shimizu H, Koshu K, Kayama T, Yoshimoto T. Magnetic reso-
unexpected increase in magnetic susceptibility artifact for nance imaging of titanium anterior cervical spine plating systems. Neuro-
the short-TE sequence at voxel volumes greaten than 3 surgery 1995:36:951-955
1218 PETERSILGE ET AL. AJA:166, May 1996

16. Torpey B, Dormans JP, Drummond DS. The use of MAt-compatible tita- rate on magnetic resonance imaging. Magn Reson Med 1987:5:278-285
nium segmental spinal instrumentation in pediatric patients with intraspi- 23. Fache JS, Price C, Hawbolt EB, Li 0KB. MA imaging artifacts produced
naltumor. JSpinalDisordl995;1:76-81 by dental materials. AJNR 1987:8:837-840
17. Ludeke KM, Aoschmann P. Tischler A. Susceptibility artifacts in NMR 24. Camacho CA, Plewes DB, Henkelman AM. Nonsusceptibitity artifacts due
imaging. Magn Reson Imaging 1985;3:329-343 to metallic objects in MA imaging. J Magn Reson Imaging 1995;5:75-88
18. Laakman AW, Kaufman B, Han JS, et at. MR imaging in patients with 25. Tartaglino LM, Flanders AE, vintiski 5, Friedman DP. Metallic artifacts on
metallic implants. Radiology 1985;157:711-714 MA images of the postoperative spine: reduction with fast spin-echo tech-
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

19. Farahani K, Sinha U, Sinha 5, Chiu LC-L, Lufkin RB. Effect of field niques. Radiology 1994;190:565-569
strength on susceptibility artifacts in magnetic resonance imaging. Com- 26. Metki PS, Mulkern Av, Panych LP, Jotesz FA. Comparing the FAlSE
put Med Imaging Graph 1990:14:409-413 method with conventional dual-echo sequences. J Magn Reson Imaging
20. Young IA, Cox IJ, Bryant DJ, Bydder GM. The benefits of increasing spa- 1991 1:319-326
tial resolution as a means of reducing artifacts due to field inhomogene- 27. Jones KM, Mulkern AV, Mantelto MT, et at. Brain hemorrhage: evaluation
ities. Magn Reson Imaging 1988:6:585-590 with fast spin-echo and conventional dual spin-echo images. Radiology
21 . Vinitski 5, Mitchell 0G. Einstein SG, et at. Conventional and fast spin-echo 1992;182:53-58
MA imaging: minimizing echo time. J Magn Reson Imaging 1993:3:801-507 28. Norbash AM, Glover GH, Enzmann DR. Intracerebrat lesion contrastwith spin-
22. vinitski 5, Griffey A, Fuka M, Matwiyoff N, Prost A. Effect of the sampling echo and fast spin-echo pulse sequences. Radiology 1992;185:661-665
This article has been cited by:

1. Laura Eisenmenger, Aaron J. Clark, Vinil N. Shah. 2019. Postoperative Spine. Radiologic Clinics of North America 57:2, 415-438.
[Crossref]
2. Uwe H.W. Schütz. 2019. Moderne radiologische Diagnostik in der Wirbelsäulenchirurgie. Der Orthopäde 48:1, 5-43. [Crossref]
3. Yimin Ma, Panli Zuo, Mathias Nittka, Xiaoguang Cheng, Hongyi Shao, Chen Wang. 2018. Comparisons of slice-encoding metal
artifact correction and view-angle tilting magnetic resonance imaging and traditional digital radiography in evaluating chronic hip
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

pain after total hip arthroplasty. Journal of Orthopaedic Translation 12, 45-54. [Crossref]
4. Panayiotis Mavroidis, Nada Boci, Spiros Kostopoulos, Constantin Ninos, Dimitrios Glotsos, Georgia Oikonomou, Athanasios
Bakas, Violeta Roka, Georgios K. Sakkas, Antonios Tsagkalis, Vasiliki Chatzigeorgiou, Georgios Batsikas, Georgios Zaimis,
Dionisios Cavouras, Eleftherios Lavdas. 2017. Investigation of the Possibility to Reduce Susceptibility Artifacts in MRI Knee
Examination. Journal of Medical and Biological Engineering 37:5, 644-652. [Crossref]
5. Aaron F. Cipriano, Jiajia Lin, Christopher Miller, Alan Lin, Mayra C. Cortez Alcaraz, Pedro Soria, Huinan Liu. 2017. Anodization
of magnesium for biomedical applications – Processing, characterization, degradation and cytocompatibility. Acta Biomaterialia
62, 397-417. [Crossref]
6. Brian G. Jones, Geoffrey T. Fosgate, Eric M. Green, Amy M. Habing, Bianca F. Hettlich. 2017. Magnetic resonance imaging
susceptibility artifacts in the cervical vertebrae and spinal cord related to monocortical screw–polymethylmethacrylate implants in
canine cadavers. American Journal of Veterinary Research 78:4, 458-464. [Crossref]
7. F. Caranci, G. Leone, L. Ugga, E. Cesarano, R. Capasso, S. Schipani, A. Bianco, P. Fonio, F. Briganti, L. Brunese. 2017. Imaging
of post-surgical treatment and of related complications in spinal trauma. MUSCULOSKELETAL SURGERY 101:S1, 63-73.
[Crossref]
8. S. Sivaram Kaushik, Robin Karr, Matthew Runquist, Cathy Marszalkowski, Abhishiek Sharma, Scott D. Rand, Dennis Maiman,
Kevin M. Koch. 2017. Quantifying metal-induced susceptibility artifacts of the instrumented spine at 1.5T using fast-spin echo
and 3D-multispectral MRI. Journal of Magnetic Resonance Imaging 45:1, 51-58. [Crossref]
9. J. P. Dillenseger, S. Molière, P. Choquet, C. Goetz, M. Ehlinger, G. Bierry. 2016. An illustrative review to understand and manage
metal-induced artifacts in musculoskeletal MRI: a primer and updates. Skeletal Radiology 45:5, 677-688. [Crossref]
10. Cameron A. Elliott, Richard Fox, Robert Ashforth, Sita Gourishankar, Andrew Nataraj. 2016. Magnetic resonance imaging
artifact following anterior cervical discectomy and fusion with a trabecular metal cage. Journal of Neurosurgery: Spine 24:3, 496-501.
[Crossref]
11. Jin Ho Hong, Ha Young Lee, Young Hye Kang, Myung Kwan Lim, Yeo Ju Kim, Soon Gu Cho, Mi Young Kim. 2016.
Improvement of Fat Suppression and Artifact Reduction Using IDEAL Technique in Head and Neck MRI at 3T. Investigative
Magnetic Resonance Imaging 20:1, 44. [Crossref]
12. C Zachriat, P Asbach, K I Blankenstein, I Peroz, F H Blankenstein. 2015. MRI with intraoral orthodontic appliance—a
comparative in vitro and in vivo study of image artefacts at 1.5 T. Dentomaxillofacial Radiology 44:6, 20140416. [Crossref]
13. Amit Gupta, Naveen Subhas, Andrew N. Primak, Mathias Nittka, Kecheng Liu. 2015. Metal Artifact Reduction. Radiologic
Clinics of North America 53:3, 531-547. [Crossref]
14. Saverio Pollice, Mario Muto, Tommaso Scarabino. 2015. Post-therapeutic imaging findings. European Journal of Radiology 84:5,
799-806. [Crossref]
15. S. Piesnack, G. Oechtering, E. Ludewig. 2015. Options for the reduction of magnetic susceptibility artifacts caused by implanted
microchips in 0.5 Tesla magnetic resonance imaging. Tierärztliche Praxis Ausgabe K: Kleintiere / Heimtiere 43:2, 83-92. [Crossref]
16. Jan L. M. A. Gielen, J. Veryser, P. Van Dyck. Shoulder Imaging 299-373. [Crossref]
17. Jeanne M Gunzinger, Gaspar Delso, Andreas Boss, Miguel Porto, Helen Davison, Gustav K von Schulthess, Martin Huellner,
Paul Stolzmann, Patrick Veit-Haibach, Irene A Burger. 2014. Metal artifact reduction in patients with dental implants using
multispectral three-dimensional data acquisition for hybrid PET/MRI. EJNMMI Physics 1:1. . [Crossref]
18. Jing Li, Hong-jun Ai. 2014. The responses of endothelial cells to Zr61Ti2Cu25Al12 metallic glass in vitro and in vivo. Materials
Science and Engineering: C 40, 189-196. [Crossref]
19. Elisabetta Panfili, Laura Pierdicca, Luca Salvolini, Luigi Imperiale, Jeffrey Dubbini, Andrea Giovagnoni. 2014. Magnetic resonance
imaging (MRI) artefacts in hip prostheses: a comparison of different prosthetic compositions. La radiologia medica 119:2, 113-120.
[Crossref]
20. O Tymofiyeva, S Vaegler, K Rottner, J Boldt, AJ Hopfgartner, PC Proff, E-J Richter, PM Jakob. 2013. Influence of dental
materials on dental MRI. Dentomaxillofacial Radiology 42:6, 20120271. [Crossref]
21. Aaron F. Cipriano, Tong Zhao, Ian Johnson, Ren-Guo Guan, Salvador Garcia, Huinan Liu. 2013. In vitro degradation of four
magnesium–zinc–strontium alloys and their cytocompatibility with human embryonic stem cells. Journal of Materials Science:
Materials in Medicine 24:4, 989-1003. [Crossref]
22. Erika J. Ulbrich, Reto Sutter, Ranieri F. Aguiar, Mathias Nittka, Christian W. Pfirrmann. 2012. STIR Sequence With Increased
Receiver Bandwidth of the Inversion Pulse for Reduction of Metallic Artifacts. American Journal of Roentgenology 199:6, W735-
W742. [Abstract] [Full Text] [PDF] [PDF Plus]
23. Rashmi S. Thakkar, John P. Malloy, Savyasachi C. Thakkar, John A. Carrino, A. Jay Khanna. 2012. Imaging the Postoperative
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

Spine. Radiologic Clinics of North America 50:4, 731-747. [Crossref]


24. Lali H. Sekhon. 2012. Postoperative Imaging of Spinal Disk Arthroplasty Devices. Seminars in Spine Surgery 24:1, 48-56.
[Crossref]
25. Alister J Hart, Keshthra Satchithananda, Alexander D Liddle, Shiraz A Sabah, Donald McRobbie, Johann Henckel, Justin P
Cobb, John A Skinner, Adam W Mitchell. 2012. Pseudotumors in Association with Well-Functioning Metal-on-Metal Hip
Prostheses. The Journal of Bone and Joint Surgery-American Volume 94:4, 317-325. [Crossref]
26. F. H. David, J. Grierson, C. R. Lamb. 2012. Reducing susceptibility artefacts in magnetic resonance images of the canine
stifle following surgery for cranial cruciate ligament deficiency. Veterinary and Comparative Orthopaedics and Traumatology 25:6,
488-497. [Crossref]
27. Thorsten Ernstberger, Gabert Heidrich, Hans-Michael Klinger, Mike Herbert Baums. 2012. Score based assessment of implant-
related post fusion MRI artifacts focused on different interbody disc spacers: An in vitro study. Open Journal of Clinical Diagnostics
02:02, 23-29. [Crossref]
28. Thorsten Ernstberger, Gottfried Buchhorn, Gabert Heidrich. 2012. Score Evaluation of MRI Sequence Type Related Artifacts
after Interbody Fusion with Metallic Implants—A Spine Specimen Study. Open Journal of Medical Imaging 02:02, 57-63. [Crossref]
29. Jang Gyu Cha, Wook Jin, Min Hee Lee, Dong Hoon Kim, Jai Soung Park, Won Han Shin, Boem Ha Yi. 2011. Reducing
Metallic Artifacts in Postoperative Spinal Imaging: Usefulness of IDEAL Contrast-enhanced T1- and T2-weighted MR Imaging
—Phantom and Clinical Studies. Radiology 259:3, 885-893. [Crossref]
30. Huinan Liu. Biodegradable Metals and Responsive Biosensors for Musculoskeletal Applications 115-137. [Crossref]
31. T. Ernstberger, G. Buchhorn, G. Heidrich. 2010. RETRACTED ARTICLE: Magnetic resonance imaging evaluation of
intervertebral test spacers: an experimental comparison of magnesium versus titanium and carbon fiber reinforced polymers as
biomaterials. Irish Journal of Medical Science 179:1, 107-111. [Crossref]
32. Cormac Farrelly, Amir Davarpanah, Stephen Brennan, Mathew Sampson, Stephen J. Eustace. 2010. Imaging of Soft Tissues
Adjacent to Orthopedic Hardware: Comparison of 3-T and 1.5-T MRI. American Journal of Roentgenology 194:1, W60-W64.
[Abstract] [Full Text] [PDF] [PDF Plus]
33. T. Ernstberger, G. Heidrich, G. Buchhorn. 2010. Imaging characteristics of metallic interbody spacers: in vitro score evaluation
of susceptibility artifacts considering different MRI sequences. Journal of Biomedical Science and Engineering 03:02, 181-186.
[Crossref]
34. Thorsten Ernstberger, Gottfried Buchhorn, Gabert Heidrich. 2009. Intervertebral test spacers and postfusion MRI artifacting: A
comparative in vitro study of magnesium versus titanium and carbon fiber reinforced polymers as biomaterials. Central European
Journal of Medicine 4:4, 496-500. [Crossref]
35. Thorsten Ernstberger, Gottfried Buchhorn, Gabert Heidrich. 2009. Artifacts in spine magnetic resonance imaging due to
different intervertebral test spacers: an in vitro evaluation of magnesium versus titanium and carbon-fiber-reinforced polymers as
biomaterials. Neuroradiology 51:8, 525-529. [Crossref]
36. P. Stradiotti, A. Curti, G. Castellazzi, A. Zerbi. 2009. Metal-related artifacts in instrumented spine. Techniques for reducing
artifacts in CT and MRI: state of the art. European Spine Journal 18:S1, 102-108. [Crossref]
37. Thorsten Ernstberger. 2009. Implant-related MRI artifacts of determined interbody test spacers: artifact calculations due to
implant parameters in a porcine spine model. Health 01:03, 207-210. [Crossref]
38. Sinchun Hwang, David M. Panicek. Imaging Techniques: Magnetic Resonance Imaging 31-52. [Crossref]
39. M.E. Bowers, G.A. Tung, N. Trinh, E. Leventhal, J.J. Crisco, B. Kimia, B.C. Fleming. 2008. Effects of ACL interference screws
on articular cartilage volume and thickness measurements with 1.5 T and 3 T MRI. Osteoarthritis and Cartilage 16:5, 572-578.
[Crossref]
40. Kenneth B. Baker, Michael D. Phillips. Deep Brain Stimulation Safety: MRI and Other Electromagnetic Interactions 453-472.
[Crossref]
41. R. Guillevin, J.-N. Vallee, F. Lafitte, C. Menuel, N.-M. Duverneuil, J. Chiras. 2007. Spine metastasis imaging: review
of the literature. Journal of Neuroradiology 34:5, 311-321. [Crossref]
42. Thorsten Ernstberger, Gabert Heidrich, Christian Dullin, Gottfried Buchhorn, Eckhardt Grabbe. 2007. Preclinical evaluation by
flat-panel detector-based volumetric CT versus MRI of intervertebral spacers implanted in a porcine model. The Spine Journal
7:3, 360-367. [Crossref]
43. Thorsten Ernstberger, Gabert Heidrich, Gottfried Buchhorn, Bioengineer. 2007. Postimplantation MRI with cylindric and cubic
intervertebral test implants: evaluation of implant shape, material, and volume in MRI artifacting—an in vitro study. The Spine
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

Journal 7:3, 353-359. [Crossref]


44. Mi-Jung Lee, Sungjun Kim, Sung-Ah Lee, Ho-Taek Song, Yong-Min Huh, Dae-Hong Kim, Seung Hwan Han, Jin-Suck Suh.
2007. Overcoming Artifacts from Metallic Orthopedic Implants at High-Field-Strength MR Imaging and Multi-detector CT.
RadioGraphics 27:3, 791-803. [Crossref]
45. Thorsten Ernstberger, Gabert Heidrich. 2007. Postfusion Magnetic Resonance Imaging Artifacts Caused by a Titanium, Cobalt-
Chromium-Molybdenum, and Carbon Intervertebral Disc Spacer. Journal of Spinal Disorders & Techniques 20:2, 154-159.
[Crossref]
46. Lali H. S. Sekhon, Neil Duggal, James J. Lynch, Regis W. Haid, John G. Heller, K Daniel Riew, Kevin Seex, Paul A. Anderson.
2007. Magnetic Resonance Imaging Clarity of the Bryan®, Prodisc-C®, Prestige LP®, and PCM® Cervical Arthroplasty Devices.
Spine 32:6, 673-680. [Crossref]
47. T. Ernstberger, G. Heidrich, T. Bruening, S. Krefft, G. Buchhorn, H.M. Klinger. 2007. The interobserver-validated relevance of
intervertebral spacer materials in MRI artifacting. European Spine Journal 16:2, 179-185. [Crossref]
48. Thorsten Ernstberger, Gabert Heidrich, Wolfgang Schultz, Eckhardt Grabbe. 2007. Implant detectibility of intervertebral disc
spacers in post fusion MRI: evaluation of the MRI scan quality by using a scoring system—an in vitro study. Neuroradiology
49:2, 103-109. [Crossref]
49. Tim A.J. Hopper, Branimir Vasilić, James M. Pope, Catherine E. Jones, Charles L. Epstein, Hee Kwon Song, Felix W. Wehrli.
2006. Experimental and computational analyses of the effects of slice distortion from a metallic sphere in an MRI phantom.
Magnetic Resonance Imaging 24:8, 1077-1085. [Crossref]
50. Elmar M. Merkle. 2006. Pneumobilia: where to look for on hepatic MR imaging?. European Radiology 16:10, 2366-2368.
[Crossref]
51. Elmar M. Merkle, Brian M. Dale, John Thomas, Erik K. Paulson. 2006. MR liver imaging and cholangiography in the presence
of surgical metallic clips at 1.5 and 3 Tesla. European Radiology 16:10, 2309-2316. [Crossref]
52. Elmar M. Merkle, Rendon C. Nelson. 2006. Dual Gradient-Echo In-Phase and Opposed-Phase Hepatic MR Imaging: A Useful
Tool for Evaluating More Than Fatty Infiltration or Fatty Sparing. RadioGraphics 26:5, 1409-1418. [Crossref]
53. Craig A. Harris, Lawrence M. White. 2006. Metal Artifact Reduction in Musculoskeletal Magnetic Resonance Imaging. Orthopedic
Clinics of North America 37:3, 349-359. [Crossref]
54. Lali H.S. Sekhon, Paul A. Anderson. 2006. Disc Replacement: Postoperative Imaging. Seminars in Spine Surgery 18:2, 99-108.
[Crossref]
55. Douglas P. Beall, Hal D. Martin, Justin Q. Ly, Scot E. Campbell, Suzanne Anderson, Moritz Tannast. 2006. Postoperative Imaging
of the Hip. Radiologic Clinics of North America 44:3, 343-365. [Crossref]
56. Thomas R. McCauley. 2005. MR Imaging Evaluation of the Postoperative Knee. Radiology 234:1, 53-61. [Crossref]
57. Shannon H. Kolind, Alex L. MacKay, Peter L. Munk, Qing-San Xiang. 2004. Quantitative evaluation of metal artifact reduction
techniques. Journal of Magnetic Resonance Imaging 20:3, 487-495. [Crossref]
58. A. Guermazi, Y. Miaux, S. Zaim, C.G. Peterfy, D. White, H.K. Genant. 2003. Metallic Artefacts in MR Imaging: Effects of
Main Field Orientation and Strength. Clinical Radiology 58:4, 322-328. [Crossref]
59. Carolyn M. Sofka, Hollis G. Potter, Mark Figgie, Richard Laskin. 2003. Magnetic Resonance Imaging of Total Knee Arthroplasty.
Clinical Orthopaedics and Related Research 406, 129-135. [Crossref]
60. Hideki Matsuura, Takashi Inoue, Hiromu Konno, Makoto Sasaki, Kuniaki Ogasawara, Akira Ogawa. 2002. Quantification of
susceptibility artifacts produced on high-field magnetic resonance images by various biomaterials used for neurosurgical implants.
Journal of Neurosurgery 97:6, 1472-1475. [Crossref]
61. Wayne William Gibbon, Gillian Long, Dominic Anthony Barron, Philip James O'Connor. 2002. Complications of orthopedic
implants: Sonographic evaluation. Journal of Clinical Ultrasound 30:5, 288-299. [Crossref]
62. A. S. Malik, O. Boyko, N. Aktar, W. F. Young. 2001. A COMPARATIVE STUDY OF MR IMAGING PROFILE OF
TITANIUM PEDICLE SCREWS. Acta Radiologica 42:3, 291-293. [Crossref]
63. John D. Port, Martin G. Pomper. 2000. Quantification and Minimization of Magnetic Susceptibility Artifacts on GRE Images.
Journal of Computer Assisted Tomography 24:6, 958-964. [Crossref]
64. ROBERT W. GAINES. 2000. The Use of Pedicle-Screw Internal Fixation for the Operative Treatment of Spinal Disorders*.
The Journal of Bone and Joint Surgery-American Volume 82:10, 1458-1476. [Crossref]
65. David Rubin. MR Imaging Technique and Principles 1-36. [Crossref]
66. Claude Pierre-Jerome, Arzu Arslan, Svein Ivar Bekkelund. 2000. MRI of the spine and spinal cord: Imaging techniques, normal
anatomy, artifacts, and pitfalls. Journal of Manipulative and Physiological Therapeutics 23:7, 470-475. [Crossref]
Downloaded from www.ajronline.org by 101.255.45.218 on 04/28/19 from IP address 101.255.45.218. Copyright ARRS. For personal use only; all rights reserved

67. Randall V. Olsen, Peter L. Munk, Mark J. Lee, Dennis L. Janzen, Alex L. MacKay, Qing-San Xiang, Bassam Masri. 2000. Metal
Artifact Reduction Sequence: Early Clinical Applications. RadioGraphics 20:3, 699-712. [Crossref]
68. Lawrence M. White, Jae K. Kim, Mitesh Mehta, Naeem Merchant, Mark E. Schweitzer, William B. Morrison, Carol R.
Hutchison, Allan E. Gross. 2000. Complications of Total Hip Arthroplasty: MR Imaging—Initial Experience. Radiology 215:1,
254-262. [Crossref]
69. Ch. Josten, P. Verheyden, S. Katscher. Wirbelsäulenoperationen im offenen MRT 202-210. [Crossref]
70. Alan S. Hilibrand, Thomas S. Dina. 1998. THE USE OF DIAGNOSTIC IMAGING TO ASSESS SPINAL ARTHRODESIS.
Orthopedic Clinics of North America 29:4, 591-601. [Crossref]
71. Jeffrey C. Wang, Warren D. Yu, Harvinder S. Sandhu, Vivian Tam, Rick B. Delamarter. 1998. A Comparison of Magnetic
Resonance and Computed Tomographic Image Quality After the Implantation of Tantalum and Titanium Spinal Instrumentation.
Spine 23:15, 1684-1688. [Crossref]
72. Ansgar Rudisch, Christian Kremser, Siegfried Peer, Anton Kathrein, Werner Judmaier, Herwing Daniaux. 1998. Metallic Artifacts
in Magnetic Resonance Imaging of Patients With Spinal Fusion. Spine 23:6, 692-699. [Crossref]
73. Stephen Eustace, Binod Shah, Michael Mason. 1998. IMAGING ORTHOPEDIC HARDWARE WITH AN EMPHASIS ON
HIP PROSTHESES. Orthopedic Clinics of North America 29:1, 67-84. [Crossref]
74. Jeffrey G. Ojemann, Erbil Akbudak, Abraham Z. Snyder, Robert C. McKinstry, Marcus E. Raichle, Thomas E. Conturo. 1997.
Anatomic Localization and Quantitative Analysis of Gradient Refocused Echo-Planar fMRI Susceptibility Artifacts. NeuroImage
6:3, 156-167. [Crossref]
75. S. Eustace, R. Goldberg, D. Williamson, E.R. Melhem, O. Oladipo, E.K. Yucel, H. Jara. 1997. MR imaging of soft tissues
adjacent to orthopaedic hardware: Techniques to minimize susceptibility artefact. Clinical Radiology 52:8, 589-594. [Crossref]
76. Colin S. Poon, John Doumanian, Gordon Sze, Michele H. Johnson, Carl E. Johnson, Felix Lin. Spine and Spinal Canal 535-662.
[Crossref]
77. Jean Palussière, Vincent Dousset. Imagerie des métastases vertébrales 73-99. [Crossref]

You might also like