Professional Documents
Culture Documents
Physical Variable, X
Output, O
Measuring
Instrument / Sensor
Variable
Interfering Input, X
Input
Example of Desired, Interfering & Modifying Inputs
Strain Gauge
Measures strain of a specimen
Desired Input: Strain
Interfering Input: Temperature
Change the gauges resistance even if there is no
strain on the gauge
Changes the gauge resistance due to the strain resulting from the differential
expansion of the gauge and the specimen (since they are made of materials with
different thermal expansion coefficients)
Modifying Input: Temperature
Change factor is dependent on temperature
Change in gauge resistance= Gauge Factor×Gauge resistance when unstrained×Unit Strain
Corrective Measurement for Undesired Inputs
Method of Signal Filtering:
Use filters to block the undesired inputs
The filters may be applied to input signal as well as output signal
Final
Unavoidable Interfering Output
TF_I1
Input, I_I1
Zero Signal
Intentionally Introduced
TF_I2
Interfering Input, I_I2
Performance Characteristics of Instruments
How do we describe the quality of measurement?
Performance characteristics are quantitative bases for comparing the
performance of an instrument with possible alternatives
Two Classes of characteristics are used to evaluate/compare instruments
Static Characteristics:
Attributes associated with static measurement
Set of criteria that are used when we measure a quantity that is constant or
changes very slowly
Dynamic Characteristics:
Attributes associated with dynamic measurement
Set of criteria that are used when we measure a quantity that is rapidly varying
with time
Note: Static characteristics also influence the quality of measurement under
dynamic environment. But we treat them separately.
Static Characteristics of Instruments
Static Characteristics
Linear
qout
Nonlinear
qIn
Static Characteristics of Instruments
Range: The Range of an instrument is the lowest and highest values it can
measure within its specification.
Span : the Span of an instrument is the high end of the range minus the low end
of the range
Resolution: Resolution is the smallest amount of input signal change that the
instrument can detect reliably
Example:
The lowest point of calibration = a units
The highest point of calibration = b units
Instrument’s range : a to b
Instrument’s span : b – a
Note: We should select instruments such that its range includes all values a
process will normally encounter, including expected (Disturbances and
possible failure)
Static Characteristics of Instruments
Accuracy: The accuracy of an instrument describes how close the measurement
approaches the true value of the process variable.
Accuracy is expressed in many ways
“Accuracy within ±× %” means “accurate to within ±× % of instrument span at all point
of the scale”
% True value
𝑴𝒆𝒂𝒔𝒖𝒓𝒆𝒅 𝑽𝒂𝒍𝒖𝒆 − (𝑻𝒓𝒖𝒆 𝑽𝒂𝒍𝒖𝒆)
× 𝟏𝟎𝟎
𝑻𝒓𝒖𝒆 𝑽𝒂𝒍𝒖𝒆
% Full Scale Deflection
𝑴𝒆𝒂𝒔𝒖𝒓𝒆𝒅 𝑽𝒂𝒍𝒖𝒆 − (𝑻𝒓𝒖𝒆 𝑽𝒂𝒍𝒖𝒆)
× 𝟏𝟎𝟎
𝑴𝒂𝒙𝒊𝒎𝒖𝒎 𝑺𝒄𝒂𝒍𝒆 𝑽𝒂𝒍𝒖𝒆
Measured variable
% span
Static Characteristics of Instruments
Precision: Ability of an instrument to reproduce a certain set of readings within
given accuracy.
This can be best example by an example
Shooting Target
High Accuracy Signifies that the means is close to the true value
High precision mean that the standard deviation 𝝈 is small
Static Characteristics of Instruments
Static Sensitivity: define as Ratio of change in the output or response of
instrument to a change of input or measured variable.
∆𝒐𝒖𝒕𝒑𝒖𝒕
∆𝒊𝒏𝒑𝒖𝒕
= 𝒔𝒕𝒂𝒕𝒊𝒄 𝒔𝒆𝒏𝒔𝒊𝒕𝒊𝒗𝒊𝒕𝒚 Slope of calibration curve
qout
∆ 𝒐𝒖𝒕𝒑𝒖𝒕
∆ 𝒊𝒏𝒑𝒖𝒕
qIn
The resistance value of a Resistance Thermometer (RTD) change when the
temperature increases. Therefore, the unit of sensitivity for resistance
thermometer is Ohm/ºC.
For a mercury-in-glass thermometer. If mercury level moves by 1 cm when the
temperature changes by 10 ºC. its sensitivity will be 1 cm/10ºC in other words it is
1mm/ºC.
Example: The output of a platinum resistance thermometer (RTD) is recorded as
follows. Calculate the sensitivity of the RTD.
Input ( ºC ) Output (Ohm)
Answer:
Draw an input versus output graph and the sensitivity is 0 0
the slope of the graph. 100 200
Slope of graph = (400-200)Ohm/(200-100)ºC = 2 Ohm/ºC
200 400
The above data obviously produces a linear relationship. 300 600
For a change in temperature of 100 ºC, the change in
resistance is 200 Ohm.
400 800
Hence the measurement sensitivity = 200/100= 2 Ohm/ºC
Static Characteristics of Instruments
Dead Zone: is the largest range of value of a measured
variable to which the instrument does not respond.