Del Mar Photonics - MCP products

Microchannel Plates
MCP is a specially fabricated plate that amplifies electron signal similar to secondary electron multiplier (SEM). Unlike SEM, MCP has several million
independent channels and each channel works as independent electron multiplier. In other words, one can imagine MCP as an assembly of millions
miniature SEMs. MCP consists of a two-dimensional periodic array of very-small diameter glass capillaries (channels) fused together and sliced in a thin
plate. A single incident particle (ion, electron, photon etc.) enters a channel and emits an electron from the channel wall. Secondary electrons are
accelerated by an electric field developed by a voltage applied across the both ends of the MCP. They travel along their parabolic trajectories until they in
turn strike the channel surface, thus producing more secondary electrons. This process is repeated many times along the channel; as a result, this cascade
process yields a cloud of several thousand electrons, which emerge from the rear of the plate. If two or more MCPs are operated in series, a single input
event will generate a pulse of 108 or more electrons at the output.

Микроканална плате
МЦП је специјално произведен плоча која појачава електронски сигнал слично на секундарну електрона мултипликатор (СЕМ). За разлику од
СЕМ, МЦП има неколико милиона независних канала и сваки канал ради као независни електронски мултипликатор. Другим речима, може се
замислити МЦП као скупштини милиона минијатурних СЕМС. МЦП се састоји од дводимензионалног периодичног низ веома-малих пречника
стаклених капилара (канала) спајају и нарезане на танке плоче. Један инцидент честица (јонска, електрон, фотон, итд) улази у канал и емитује
електрон из зида канала. Секундарни електрони се убрзавају електричним пољем развијен од стране напоном преко обе стране МЦП. Они
путују дуж својих параболичне путање све док се са друге стране погађа површину канала, чиме се производи више секундарних електрона.
Овај процес се понавља више пута дуж канала; као резултат, овај процес каскадни даје облак неколико хиљада електрона, која произилазе из
задњег дела плоче. Ако два или више МЦПс се ради у серији, један улаз догађај ће генерисати пулс 108 или више електрона на излазу.
Since the individual channels confine the pulse, the spatial pattern of electron pulses at the rear of the plate preserve the pattern (image) particles incident
on the front surface. The output signals are typically collected in any of several ways, including metal or multimetal anodes, resistive anode (one- or two-
dimensional), wedge and strip anode, Delay-Line Readout or on a phosphor screen deposited on a fiberoptic or other substrate.

Будући да су поједини канали ограничити пулс, просторни образац електрона импулса на задњем делу плоче сачувати образац (имаге) честица
инцидент на предњој површини. Излазни сигнали су обично прикупљају у једном од неколико начина, укључујући метала или са више метала
аноде, отпора аноде (једногодишње или дводимензионална), клин и стрип аноде, Делаи-Лине сатима или на екрану фосфором депонована на
оптичких влакана или други супстрат.

Microchannel Plates have a combination of unique properties like high gain, high spatial resolution and high temporal resolution. They can be used in a
large variety of applications including, imaging spectroscopy, electron spectroscopy and microscopy, mass spectrometry, astronomy, molecular and
atomic collision studies, cluster physics etc. Most of these applications require only some of MCP properties, for example Time-of-Flight Mass
Spectrometry require high temporal resolution of MCPs, imaging of single atoms in field ion microscopes or X-ray imaging of the Sun require mainly
spatial resolution. Particle analyzers may be produced by using a MCP detector at the output of a electrostatic and/or magnetic dispersion system. Very
high sensitivity optical, UV and EUV and X-ray spectrometers can also be produced with appropriate filtering and dispersive elements. The same
microchannel plate technology is used to make visible light image intensifiers for night vision goggles and binoculars.

Микроканална Плоче имају комбинацију јединствених карактеристика као што су висока добит, високе просторне резолуције и високе
временске резолуције. Они се могу користити у великом броју апликација укључујући, обраду спектроскопија, електронску спектроскопија и
микроскопија, масеном спектрометријом, астрономију, молекуларне и атомском судара студије, касетне физика итд Већина ових апликација
захтева само неке од МЦП особинама, на пример Тиме- од-Лет Маса спектрометрија захтевају високу временску резолуцију од МЦПС,
снимање појединачних атома у пољу јонских микроскопе или Кс-зрацима сунца потребна углавном просторну резолуцију. анализатори
Партицле могу се произвести коришћењем МЦП детектор на излазу једног електростатичким и / или магнетног дисперзија систему. Врло
висока осетљивост оптички, УВ и ЕУВ и Кс-раи спектрометре може се произвести са одговарајућим филтрирање и дисперзивни елементи.
Иста технологија Микроканална плоча се користи за прављење видљиве светлости слике интенсифиерс за ноћно осматрање наочари и двоглед.

Detectors based on Microchannel Plates have variety of designs depending on the type of particles detected, throughput (counts/second), time and
position resolution, imaging area, linearity and sensitivity, signal to noise ratio and other requirements. It's a challenge to detector developer to optimize
detector design for particular application.

In general, each detector that uses MCPs consists of three parts:

1) A Converter - a mechanism to convert initial particles in photons or electrons,

2) An Assembly of MCPs - a mechanism to amplify initial single electron or photon event into electron pulse and
3) A Readout Device - a mechanism to detect the electron avalanche.

Детектори на бази Микроканална таблицама имају различите дизајна у зависности од врсте честица откривена, протока (тачке / секунди),
време и место резолуција, област снимања, линеарност и осетљивост, сигнал за ниво буке и других услова. То је изазов за детектора
програмера да оптимизују детектор дизајн за одређену апликацију.
У принципу, свака детектор који користи МЦПС састоји из три дела:
1) А конвертор - механизам за претварање почетне честице у фотона и електрона,
2) склоп МЦПС - механизам да појачају почетни ниједан електрон или фотона догађај у електронском пулс и
3) Очитавање уређаја - механизам за детекцију електрона лавину.

1. A Converter is the part responsible for conversion of initial particles into electrons or photons that in turn efficiently interact with a Microchannel
Plate. Photocathodes are used for visible and IR radiation. Open, windowless photocathodes of CsI or MgF2 deposited on MCP operate well
through the extreme UV and soft x-ray region. Specially formulated luminescent screens are used for neutrons, heavy ions and high-energy
particles. The MCP is directly sensitive to ultraviolet rays (VUV, UV), X-rays, (-rays, charged particles, and neutrons, as well as electron beams,
that's why no Converters are usually necessary for ion detection in mass-spectrometry applications and UV and VUV radiation.
2. An Assembly of MCPs consists of single, double (so-called Chevron or V-stack) or triple (Z-stack) MCPs adjacent to one another. Number of
MCPs required depends on application. For example typical image intensifier for low-level light contains single MCP, typical TOF-MS ion
detector has two MCP. A three plate (Z-stack) MCP Detector is used to detect (count) and image single particles.
3. The choice of electronic readout will depend upon requirements.
For detection and particle counting applications where position resolution is not required, single metal anode can be used as a readout device.
MCP detectors with metal anode are widely used in mass-spectrometry.
For imaging applications with low temporal resolution phosphor screen (P20, P22, P46, etc.) coupled with CCD camera can be used. Gated
versions as fast as 10 nanoseconds are also available.
For imaging applications with moderate and high temporal resolution state-of-the art anode configurations have been developed that fall into the
following classes:
Resistive anodes (one and two dimensional)
Wedge and strip designs
Delay-Line-Readout
The readout electronics should be matched to the anode configuration.
4. 1. Цонвертер је део одговоран за конверзију почетних честица у електрона или фотони који заузврат ефикасно интеракцију са
Микроканална плоча. Фотокатоде се користе за видљиву и ИР зрачења. Опен, без прозора фотокатоде оф ЦСИ или МгФ2 депонована на
МЦП раде добро преко екстремног УВ и меком Кс-зрака региону. Специјално формулисан луминисцентна екрани се користе за
неутрона, тешких јона и високо енергетских честица. МЦП је директно осетљив на ултраљубичастих зрака (вув, УВ), Кс-зраци, (-раис,
набијене честице, и неутрони, као и електрон зраке, зато постоје Цонвертер обично неопходно за детекцију јона у масовне
спектрометрија апликације и УВ и вув зрачење.
2. Скупштина МЦПС се састоји од појединачног, двоструког (тзв Цхеврон или В-Стацк) или троструке (З-Стацк) МЦПС једна до друге.
Број МЦПС потребних зависи од примене. На пример, типична слика појачивач за слабог осветљења садржи ниједну МЦП, типичан
ТОФ-МС јонска детектор има два МЦП. Три плоча (З гомила) МЦП детектор се користи за детекцију (цоунт) и слике појединачних
честица.
3. Избор електронског сатима зависиће од захтева.
За детекцију и честица апликација за бројање у којој није потребна резолуција положај, једна метална анода може да се користи као
средство очитавање. МЦП детектори са металним аноде имају широку примену у масовној-спектрометрија.
За апликације за снимање са ниским временске резолуције фосфора екрану (П20, П22, П46, итд) заједно са ЦЦД камере могу да се
користе. Гатед верзије као брз као 10 наносекунди су такође доступни.
За апликације за снимање са умереним и високим временским резолуције стате-оф-тхе арт аноде конфигурације су развијени да спадају
у следеће класе:
Отпорне аноде (један и два димензионално)
Ведге и стрип дизајн
Делаи-лине-Очитавање
У мерача електроника треба да одговара конфигурацији аноде

Del Mar Photonics supply Microchannel Plates, MCP assemblies as well as custom-made systems including mounting and readout
device (s). This brochure describes Microchannel Plates, Open Microchannel Plates Detectors with Metal Anode and Open Imaging
Detectors (Image Intensifiers).

Microchannel Plates

The microchannel plate is an open multiplier intended for registration of particles and
radiations. MCPs represent 0.4-3.0 mm thick plates of round or rectangular shape. They have
a honeycomb structure and contain in one square centimeter up to one million of separate
channels of 5-15  m diameter. In addition to design simplicity, small dimensions and
absence of external voltage divider, MCPs feature high time and spatial resolution capability.

 Микроканална плоча је отворен мултипликатор намењен за регистрацију честица и зрачења. МЦПс представљају 0.4-3.0 мм дебљине плоче
округлог или правоугаоног облика. Они имају хонеицомб структуру и садрже у једном квадратном центиметру до милион засебних канала 5-15
 м пречника. Поред пројектовања једноставност, мале димензије и одсуство спољног напонског разделника, МЦПс имају високу времена и
просторни способности резолуције.

Construction and Operation

A Microchannel Plate begins as a glass tube fitted with a solid, acid-etchable core and drawn via fiberoptic techniques to form single fibers. A number of
these fibers are then stacked in a hexagonal array; the entire assembly is drawn again to form multi-fibers. The multi-fibers are then stacked together and
fused at high temperature to form a boule.

The boule is sliced on a wafer saw to the required bias angle, edged to size, and then ground and polished to an optical finish. The individual slices are
chemically processed to remove the solid core material, leaving a "honeycomb" structure of millions of tiny holes.

Through subsequent processing, this glass wafer is given its conductive and secondary emissive properties. Finally, a thin metal electrode (usually
Inconel, Nichrome or chromium) is vacuum-deposited on both input and output surfaces of the wafer to electrically connect all the channels in parallel.

Микроканална плоча почиње као стакленој цеви опремљене чврстим, језгро киселине-гравирати и извученог преко фибероптиц техника за
формирање појединачних влакана. Велики број ових влакана су затим наслагане у шестоугаоне низа; цео склоп се поново извући да се формира
мулти-влакна. Мулти-влакна се затим слажу заједно и фузионисан на високој температури да се формира боуле.
Були исечен на обланда видели на одређеним углом биас, ивица величини, а затим брушења и полирања на оптички краја. Појединачни кришке
су хемијски обрађена за уклањање чврстог материјала језгра, остављајући "хонеицомб" структуру милиона сићушних отвора.
Кроз накнадне обраде, ово стакло вафел је дала своје проводљиве и секундарне емитивних својства. Коначно, танка метална електрода (обично
Инцонел, МРЦХРОМЕ или хром) је вакуум-депонован на оба улазних и излазних површина вафла за електрично повезивање свих канала
паралелно.
Honeycomb structure of Microchannel Plates.
For additional images of Microchannel Plate surface at different spatial resolution click
here.

For normal operation, a bias of about 1000 Volts is applied across the microchannel plate, with the output at its most positive potential. The bias current
flowing through the plate resistance is what supplies the electrons necessary to continue the secondary emission process. Electron multiplying process
was described above. Below we consider most important properties of Microchannel Plates.

За нормалан рад, склоност од око 1000 волти се примењује на Микроканална плоча, са излазом на свом најпозитивнијем потенцијала.
Пристрасност струја тече кроз отпор плоче је оно што обезбеђује електроне неопходне за наставак секундарног процеса емисије. Елецтрон
Процес мултиплиинг је горе описано. У наставку сматрамо најважније особине Микроканална Платес.

Shape and Size

Microchannel plate arrays may be fabricated in a wide variety of formats. The MCPs may range in size from 6mm to 100mm or larger, and they may be
circular, rectangular, or virtually any other shape as required by the application or instrument geometry. In addition, a cylindrical or spherical radius of
curvature may be provided to conform to the focal plane of an instrument.

A border glass area surrounds an effective area of MCP where channels are arrayed. Table below shows dimensions of standard MCPs, supplied by Del
Mar Ventures.

Облик и величина
Микроканална вибро низови могу произведен у широком спектру формата. МЦПс може да се креће у величини од 6мм до 100мм или веће, и
они могу бити кружног, правоугаони, или практично било који други облик у складу са геометријом апликације или инструмент. Поред тога,
цилиндрична или сферни полупречник кривине може се обезбедити у складу са равни фокуса инструмента.
А Површина граница стакла окружује ефективну површину од МЦП где су распоређене канале. Табела испод показује димензије стандардних
МЦПС, које пружају дел Мар Вентурес.

MCP type
(part #) 25-10E* 33-10E 34-10 46-12 56-15 43*63 70*90

Diameter, mm    
24.8 32.8 34 46 56
Length, mm 63 90

Width, mm 43 70

Effective
Diameter, mm 18 25 30 40 50 38*58 65*85

Channel
Diameter,  m 10 10 10 12 15 15 15

MCP Thickness and Channel Diameter

The length of the channel of a MCP is virtually its thickness. The ratio of the channel length (L) to the channel diameter (d) L/d, as well as the inherent
secondary emission factor of the channel wall material determines the gain of the MCP. The standard MCPs are fabricated with a L/d ratio about 40 to 80.

МЦП Дебљина и канал Пречник


Дужина канала МКК је практично То је дебљина. Однос дужине канала (Л) са пречником канала (д) Л / Д, као инхерентног секундарни фактор
емисије канала зида материјала који одређује добит МЦП. Стандардни МЦПс су произведени са Л / д однос око 40 до 80.

Channel Bias Angle

The channel bias angle is an angle formed by the channel axis and the vertical axis to plate surface. Channels are tilted to prevent incident particles from
passing through the channels. The optimum angle is between 5 and 15 .
Канал Предрасуда Угао
Угао Канал пристрасност је угао формиран канала осе и вертикалне осе до плоче површину. Канали се нагиње како би се спречило инцидената
честице из пролази кроз канале. Оптимална угао између 5 и 15.

Open Area Ratio (OAR)

The OAR is the ratio of the open area to the total effective area of the MCP. For hexagonal arrays OAR=( * 3/6)*(d/P)2 where d is a channel diameter
and P is a pitch (period of the hexagonal structure, or c-c distance). For 10-12 structure (d=10 m, P=12 m) OAR=63%, for 12-15 it's 58%, for 15-18 it's
63%. OAR limits ultimate detection sensitivity of MCPs. Particles incident on the MCP between channels are not detected. In many applications it is
desired to make OAR as large as possible for more efficient input of primary electrons. For this purpose, there are custom MCPs in which the glass
channel walls on the input side have been etched to increase the OAR up to 70 to 80%.

Отворени простор Однос (ОАР)


Весло је однос отвореном простору на укупне ефективне површине МЦП. За шестоугаоне низове ОАР = ( *  3/6) * (г / н) 2 где је Д пречник
канала и П је корак (период шестоугаоне структуре, или ц-ц даљине). 10-12 структура (Д = 10 м, н = 12 м) ОАР = 63%, за 12-15 то је 58%, за
15-18 то је 63%. ОАР ограничава крајњи осетљивост детекције МЦПС. Честице инцидент на МЦП између канала нису откривени. У многим
апликацијама је пожељно да ОАР велики као могуће ефикасније унос примарних електрона. За ту сврху постоје прилагођене МЦПс у којима
зидови стакла канал на улазној страни су гравирани за повећање ОАР до 70 до 80%.

Metal Coating (Electrodes)

Over the input and output surfaces of a MCP, Inconel, Ni-Cr or Cr is evaporated to form electrodes. The thickness of the electrodes is controlled to have a
surface resistance of 100 to 200 between the MCP edge. In general, the electrodes are evaporated to uniformly penetrate into the channels. The
penetration depth significantly affects the angular and energy distributions of the output electrons, and usually chosen to be in the range of the channel
diameter multiplied by 0.5 to 2. In such demanding applications as image intensification where spatial resolution is of prime importance, the penetration
depth of the electrodes is controlled to be deeper in order to collimate the output electrons.

Метална оплата (електроде)


Током улазних и излазних површинама МЦП, Инцонел, Ни-Цр или Кр упари да се формира електроде. Дебљина електрода контролише да
имају отпорност површине од 100 до 200 између МЦП ивице. Уопштено, електроде су упарене до равномерно продре у канале. Дубина
продирања значајно утиче на угаоне и енергије расподеле излазних електрона, а обично изабран да буде у распону од пречника канала
помножен 0,5 до 2. У таквим захтевне апликације као слика интензивирање где просторна резолуција је од примарног значаја, пенетрација
дубина електрода контролише да дубље у циљу цоллимате излазних електроне.
Gain

The gain of an MCP, g, is given by the following equation using the length-to-diameter ratio of the channel: g= exp (G*(L/d), where G is the secondary
emission characteristics of the channel called gain factor. This gain factor is an inherent characteristic of the channel wall material and represented by a
function of the electric field intensity inside the channel. Generally, L/D is designed to be around 40, which produces a gain of 104 with an applied voltage
of 1 kV.

When an even higher gain is required, two or three MCPs are used to configure the two-stage or three-stage MCP assembly. These stacked MCP detectors
can offer higher gains up to 108-109. Multiple-stage MCP gains are not the simple multiplication of the gain of each MCP because of the gain saturation
caused by space charge effect near the output region of channels.

In these configurations the spatial resolution is degraded to some extent because a multiplying electron current spreads into several channels as it enters
the latter-stage MCP. On the other hand the saturation level increases by a factor equal to the number of those spread channels.

Добитак једног МЦП, г, даје следеће једначине применом однос дужина-пречник канала: Г = екп (Г * (л / г), где је Г секундарне карактеристике
емисије на каналу под називом добит фактор. Ово појачање фактор је присутна особина канала зида материјала у и представљени у функцији
електричног интензитета поља унутар канала. Генерално, л / Д је дизајниран да буде око 40, који производи добит од 104 са примењеном
напону од 1 кВ.
Када је потребна још већа добит, два или три МЦПс се користе за конфигурисање две фазе или три фазе МЦП монтажу. Ове наслагане МЦП
детектори може да понуди веће добитке уп то 108-109. Мултипле-стаге МЦП добици нису једноставна умножавање добити сваког МЦП због
добитак засићења изазваног ефектом пуњења простор у близини излазне региону канала.
У овим конфигурацијама просторна резолуција је деградиран у извесној мери, јер је множењем електрон струја се шири у неколико канала као
што улази у последње фазе МЦП. С друге стране повећава ниво засићења са фактором једнак броју тих ширења канала.

Pulse height distribution

When a single particle create a single electron event in MCP, the output pulse height distribution shows normally an exponential function. However, in
the region where the gain is saturated due to space charge effect, the pulse height distribution becomes peaked. This phenomenon is observed in the
MCPs operating at a high gain, for instance, stacked MCPs.

Pulse height distribution is usually characterized by the ratio of the half-width at peak (full width at half maximum: FWHM) to the peak value in the
pulse height distribution: FWHM/A; it is normally expressed in percentage. In general, is shows 120% or less for two-stage MCPs and 80% or less for
three-stage MCPs.
Пулс висина дистрибуција
Када један честица стварање јединственог електронски догађај у МЦП, дистрибуција висина излаз пулс показује нормално експоненцијалну
функцију. Међутим, у региону у коме је добитак засићених због ефекта пуњења простор, дистрибуција пулс висина постаје бледо. Овај
феномен се посматра у МЦПС који раде на високом добити, на пример, у низу МЦПС.
дистрибуција пулс висина се обично карактерише однос пола-ширине на врху (целе ширине на половини максимума: ФВХМ) до максималне
вредности у дистрибуцији пулс висина: ФВХМ / А; је обично изражена у процентима. У принципу, је схов емисије 120% или мање за два-
сценских МЦПС и 80% или мање за три-сценских МЦПС.

Transit time

The transit time of MCP assemblies is very small. Due to the shorter electron transit distance compared to the discrete dynode used in the conventional
PMT or SEM, transit time of the electron avalanche in MCP channels is in 100 ps range. The width of the single event peak determined mainly by
temporal characteristics of readout device and electronics. Ultimate time resolution can be achieved using anode configuration matched with 50 
connector cable.

Транзитног времена
Време транзит МЦП скупштина је веома мали. Због краћим електрона транзитног удаљености поређењу са дискретном диноде користи у
конвенционалној ПМТ или СЕМ, транзитног времена електрона лавина у МЦП каналима у 100 пс опсегу. Ширина појединачни пик догађаја
одређује углавном временских карактеристика очитавање уређаја и електронике. Крајњи резолуција време може се постићи употребом анода
конфигурацију подудара са 50  конектора кабла.

Spatial Resolution

Since each channel of the MCP serves as an independent electron multiplier, the channel diameter and center-to-center (c-c) spacing determine MCP
resolution. Channel diameters ranging from 5  m (6  m c-c) to 15  m (18  m c-c) are standard.

When the output from MCP is observed with a phosphor screen, the spatial resolution also depends on the MCP electrode depth penetrating into the
channels, the space between the MCP and the phosphor screen, and the accelerating voltage. Typical spatial resolution of a MCP composed of 10  m
diameter channels, which is observed with a phosphor screen, is about 40 l/mm. In the stacked MCP, the spatial resolution is less compared to that of a
single MCP because it spreads into many channels as it enters the latter-stage MCP, and also because the increased gain makes greater the electrostatic
repulsion in the space when the electrons are released from the MCP.

Просторна резолуција
Пошто сваки канал МЦП служи као независна електрона понављања, пречника канала и центра до центра (Ц-Ц) размак одреди МЦП
резолуцију. пречника канала у распону од 5  м (6  м ц-ц) до 15  м (18  м ц-ц) су стандардни.
Када је излаз из МЦП се посматра са екраном фосфора, просторна резолуција такође зависи од дубине МЦП електроде продире у канале,
простор између МЦП и екрана фосфора, као и напон убрзања. Типичан просторна резолуција једног МКК састоји од 10  м канале пречника,
који се посматра са екраном фосфора, износи око 40 л / мм. У табелама са МЦП, просторна резолуција је мање у поређењу са једног МЦП јер
се шири у многим каналима као што улази у последње фазе МЦП, као и због тога што је повећана добит чини већи за електростатичко
одбијање у простору када су електрони су пуштен из МЦП.

Dark Current

A typical MCP shows an exceptionally low dark current, less than 0.5pA/cm2 at an applied voltage of 1 kV. Even with a two or three-stage MCP, the
dark count rate is low, less than 3 cps/cm2 at an applied voltage of 1 kV per stage.

Тамно Тренутни
Типичан МЦП показује изузетно ниску тамне струје, мање од 0.5кпА / цм2 на примењеном напону од 1 кВ. Чак и са две или три фазе МЦП,
тамна стопа датотека је ниска, мање од 3 ЦПС / цм2 при примењеног напона од 1 кВ по фази.

Resistance

Glass composition and reduction processing conditions (time and temperature) can control the MCP resistance. Considering the output saturation, a lower
resistance is desirable; however, there is a limitation in lowering the resistance as the MCP operating temperature rises due to higher power consumption.

MCP resistance is typically in the range between 100 and 1000 M . For applications requiring high output currents, low-resistance MCPs of 20 to 30
M are available.

 отпорност
Стакло састав и смањење услови обраде (време и температура) може да контролише отпор МЦП. Имајући излазни засићеност, нижа отпорност
је пожељно; међутим, постоји ограничење у снижавању отпор као радна температура МЦП расте услед веће потрошње снаге.
МЦП отпор је типично у опсегу између 100 и 1000 М. За апликације које захтевају висок излазне струје, ниским напоном МЦПс од 20 до 30
М су доступни.

Microchannel Plate Detectors with Single Metal Anode (MCP-MA)


DEL MAR VENTURES Microchannel Plate Detectors
MCP-MA series are an open MCP detectors with one or
more microchannel plates and a single metal anode. They
are intended for time-resolved detection and make use of
high-speed response properties of the MCPs. MCP-MA
detectors are used for photons and particles detection in
vacuum chambers or in the space.

The body of assembly is a metal-


ceramic housing.

Drawing shows two matched MCPs in


V-stack (Chevron) assembly, which
are fixed in place using retainer ring
(above MCPs). Ceramic insulator
rings are shown red. Detector is spot-
welded to the support plate (available
in different sizes).
All parts of the assembly are highest quality components. Metal parts are polished to avoid electric discharges. Two MCPs are connected to each other via
thin (40 -50 m) copper or stainless steel foil ring. Direction of channel bias angle in the first MCP is opposite to one in the second MCP (chevron
assembly). Typical voltages necessary for a gain of 104, resistances and dark current densities of Microchannel Plates are shown in the table below. Each
detector is supplied with individual MCP data.

Specifications:

 
MCP-MA25 MCP-MA34 MCP-MA46

MCP detector body


metal-ceramic housing

Effective area diameter, min


18mm 25mm 40mm

MCP type 25-5, 24-10, 25-10 etc.


33-10 or 34-10 46-12

MCP Diameter, mm
24.2 or 24.8 32.8 or 34 46

MCP Thickness, mm
0.46 0.46 0.5

MCP channels
pore-pitch, m 5-6, or 10-12 10 12 12-15

Typical Gain, (one MCP)


104 - 104

(2 stack) 106 - 107


(3 stack) 108 - 109

Time resolution
< 1ns

PHD (2 stack assembly)


FWHM/A<120%

PHD (3 stack assembly)


FWHM/A<80%

Output
Single metal anode

Strip current
<20A

 
Operation conditions:

Wiring Methods

In general, MCP assemblies can be operated with any electrode (MCP-in, MCP-out or anode) at a ground potential.

1. Voltage Application

When applying a voltage, do not apply the necessary voltage to the MCP at once. Slowly increase the applied voltage, with maximum 100 V step,
until the optimum rating is reached, and verify if the MCP operates properly. In this procedure, also check the dark current by connecting an
ammeter to the readout device. If there is an increase in the dark current, which might result from a small discharge, immediately turn off the
applied voltage. After some time (depending on the situation) has passed, apply voltage to the MCP again in the same manner as described above.
Note that the applied voltage to the MCP should be increased as slowly as possible even after normal operation has been verified.
Приликом подношења захтева напон, не примењују потребне напон на МЦП одједном. Полако повећавати примењује напон, са
максималним 100 В корак, док се оптимална рејтинг је постигнут, и проверите да ли је МПЦ ради правилно. У овом поступку, такође
проверите тамне струје повезивањем амперметар за очитавање уређаја. Ако постоји повећање у мраку струје, што може довести из
малог пражњења, одмах искључите примењеног напона. Након извесног времена (зависно од ситуације) прошла, примењују напона на
МЦП поново на исти начин како је горе описано. Имајте на уму да примењује напон на МЦП треба повећати што је спорије могуће, чак
и након нормалан рад потврђена.

2. Applied Voltage

Recommended and maximum applied voltage to MCPs and readout devices are as follows:

 Between MCP-in and MCP-out:

Set this voltage according to the required gain, 700 -1000V per MCP typical, 1100 V maximum, MCP out at positive polarity.

 Between MCP-out and single anode:

This is normally set at about 100 - 200 V.

A system pressure better than 6.5*10-4 Pa (5 *10-6 Torr) is necessary for proper operation. The MCP detector has to be degassed before applying the
maximum voltage. Because the MCP is operated with a high voltage of about 1 kV per stage, a relatively high degree of vacuum must be required. If the
MCP is operated at a deficient vacuum, not only will the noise increase due to the ion generation in the channels, but also the lifetime may be shortened.
In the worst case, the MCP may be damaged by discharge. Therefore, it is recommended that the MCP be operated at a degree of vacuum as high as
possible. When using a new MCP, it is recommended that before applying a voltage to it, the system be evacuated at a pressure of 6.5*10-4 Pa (5*10-6
Torr) or below for more than 24 hours. If the evacuation time is short or the degree of vacuum is deficient, a discharge may occur.

Препоручује и максимално примењује напон на МЦПС и очитавање уређаја су:


• У периоду од МЦП-ин и МЦП-оут:
Сет овај напон у складу са потребном добитка, 700 -1000В по МЦП типичног, 1100 В максимум, МЦП Оут ат позитивног поларитета.
• У периоду од МЦП-оут и једним аноде:
Ово се нормално постављена на око 100 - 200 В.
Се притисак систем боље од 6.5 * 10-4 ПА (5 * 10-6 Торр) је неопходан за правилан рад. МЦП Детектор мора бити дегасира пре примене
максимални напон. Јер МЦП ради са високим напоном од око 1 кВ по фази, мора тражити релативно висок степен вакуума. Ако МЦП се ради
на мањкав вакууму, ће повећање буке захваљујући генерацији јона у каналима, али и животни век може бити скраћено само. У најгорем
случају, МЦП може бити оштећен пражњења. Стога се препоручује да МЦП радити на степеном вакуума као висок као могуће. Када користите
нову МЦП, препоручује се да пре примене напон на то, систем се евакуишу на притиску од 6,5 * 10-4 Па (5 * 10-6 Торр) или ниже за више од
24 сата. Ако је кратко време евакуација или степен вакуума је мањкава, пражњење може доћи до.

MCP Detector Mounting

MCP-MA34 detector can be mounted on the standard vacuum flange or on any other substrate. It can be either spot-welded or connected with screws.
Figure below shows MCP-MA34 mounted on the standard 6" ConFlat Flange.

Open Microchannel Plate Imaging Detectors (MCP-GPS and MCP-IFP)


Open Microchannel Plate Imaging Detectors have a design similar to
MCP detectors with Metal Anode. Instead of simple metal anode an
aluminized phosphor screen is used as a readout device. An electron cloud
is drawn across a 0.7 mm gap by a high voltage onto micro-crystalline
phosphor screen where the kinetic energy of the electrons is released as
light.
Due to a high voltage electron image transferred to the visual image
practically without distortions (this is called proximity focusing).

The phosphor screens deposited on a glass window are realized in MCP-GPS series and on a fiber-optic plate in MCP-IFP
series. Drawing shows a cross-section of the imaging detector with a fiber-optic plate.

The optical image can be viewed directly, or coupled to a camera.

MCP-GPS and MCP-IFP imaging detectors are available in the same sizes and MCP-assembly options as MCP-MA
detectors.

Open imaging detectors must be operated in pressures of less than 6.5*10-4 Pa (5 *10-6 Torr).

Recommended and maximum applied voltage to MCPs and phosphor screen are as follows:

 Between MCP-in and MCP-out:

Set this voltage according to the required gain, 700 -1000V per MCP typical, 1100 V maximum, MCP out at positive
polarity.
 Between MCP-out and phosphor screen:

A bias in the range 2.5-5 kV between MCP output and screen is required.

MCP Home - Del Mar Photonics

 
Microchannel Plate detectors optimized for imaging VUV and EUV radiation

Next generation lithography research is one of the potential applications of the new Microchannel Plate detectors optimized for imaging VUV and EUV
radiation. Current advanced lithographic equipment employs excimer lasers to produce feature sizes at 180 nm. Extreme ultraviolet (EUV) lithography
tools will use 13.5 nm light to image chips with feature sizes below 45 nm. Much development work is still required in EUV radiation sources. Del Mar
Ventures provide a complete program of stationary and gated MCP- Detectors for the registration of X-ray- and UV-radiation below 2000 Å. Each
equipped with a single Au coated MCP plate and a phosphor screen on a fiber optic plate, which also serves as a vacuum to air interface. Devending on
the detector model phosphor screen is either uniform or sectioned in several independent sectors or stripes. In multi-frame units, each individual sector or
stripe can be gated separately with a time resolusion as short as 3 ns. This allows to obtain images or spectra or both at individual times. Gated MCP
detectors supplied with optional high voltage pulse generator that provides four high voltage outputs with variable delay time and width. Versatile trigger
options, including an internal delay generator allow for an easy adaptation to the experimental requirements.

Microchannel Plate Detectors MCP-MA

DEL MAR Microchannel Plate Detectors MCP-MA series are an open MCP detectors with one or more microchannel plates and a single metal anode.
They are intended for time-resolved detection and make use of high-speed response properties of the MCPs. MCP-MA detectors are designed for photons
and particles detection in vacuum chambers or in the space.

MCP-MA detectors are used in a variety of applications including UV, VUV and EUV spectroscopy, atomic and molecular physics, TOF mass–
spectrometry of clusters and biomolecules, surface studies and space research.
MCP-MA detectors supplied as a totally assembled unit that can be easily mounted on any support substrate or directly on a vacuum flange. They also
can be supplied premounted on a standard ConFlat flanges.
 

 Model   Product Name+   Price   Buy Now 


 MCP 25-10E   Microchannel Plate MCP 25-10E   $320.00   
 MCP 33-10E   Microchannel Plate MCP 33-10E   $350.00   
 MCP 34-10   Microchannel Plate MCP 34-10   $350.00   
 MCP 43-63   Microchannel Plate MCP 43x63   $1,990.00   
 MCP 46-12   Microchannel Plate MCP 46-12   $990.00   
 MCP 56-15   Microchannel Plate MCP 56-15   $1,590.00   
 MCP 70-90   Microchannel Plate MCP 70x90   $3,690.00   
Displaying 1 to 7 (of 7 products)

 Model   Product Name+   Price   Buy Now 


 MCP 34/2 G   Microchannel plate detector MCP 34/2 G   $1,200.00   
 MCP-MA 25/2   Microchannel plate detector MCP-MA 25/2   $1,490.00   
 MCP-MA 25/2 SP37   Microchannel plate detector MCP-MA 25/2   $1,490.00   
 MCP-MA 25/2 SP70   Microchannel plate detector MCP-MA 25/2   $1,490.00   
 MCP-GPS 34/2   Microchannel plate imaging detector MCP-GPS 34/2   $2,600.00   
 MCP-IFP 25/2   Microchannel plate imaging detector MCP-IFP 25/2   $2,190.00   
 MCP-IFP 34/2   Microchannel plate imaging detector MCP-IFP 34/2   $2,800.00   