Professional Documents
Culture Documents
Alpha ST
Installation Instructions
37081-IMG rev 5
Copyright Code: 37081-IMG rev 5 Date: 10/2009
Check production site at lower back end of the unit column or underneath
the unit (stereotactic devices).
This manual applies to Alpha ST units with software version 7.71a and up.
Table of Contents
1 Introduction .............................................................................................................. 1
2 Installation ................................................................................................................ 3
2.1 Preinstallation instructions..........................................................................................................................3
2.1.1 Space requirements ...................................................................................................................................... 3
2.1.2 Strength/durability requirements for the wall and floor .............................................................. 4
2.1.3 Electrical supply requirements ................................................................................................................. 4
2.1.4 Installations requirements & environment control......................................................................... 4
2.1.5 Tools needed for installation ..................................................................................................................... 4
2.1.6 Test equipment needed for installation ............................................................................................... 5
2.2 Unpacking the unit..........................................................................................................................................5
2.2.1 Unpacking the column with the carriage............................................................................................ 6
2.2.2 Unpacking the C-arm ................................................................................................................................... 6
2.2.3 Unpacking the counterweight.................................................................................................................. 7
2.2.4 Unpacking the accessories ........................................................................................................................ 7
2.2.5 Unpacking the base plate........................................................................................................................... 8
3 Installation of the counterweight ......................................................................... 9
3.1 Installation of the column to the wall..................................................................................................11
3.1.1 Attachment holes .........................................................................................................................................11
3.1.2 Attachment of the column brackets....................................................................................................12
3.1.3 Lifting the column - minimum 2 people.............................................................................................13
3.2 Installation of the column to the base plate ....................................................................................13
3.3 Localized unit label attachment ............................................................................................................15
3.4 Installation of the C-arm............................................................................................................................16
3.5 Attachment of the control panel ...........................................................................................................18
3.6 Attachment of the pedals .........................................................................................................................19
3.7 Radiation shield assembly and installation......................................................................................20
3.8 Mains plug installation ...............................................................................................................................22
3.9 Testing the control functions...................................................................................................................23
4 Alpha softstart procedure ....................................................................................27
4.1 Removing the covers...................................................................................................................................27
4.2 Preventing X-rays (for example at exhibition) .................................................................................28
4.3 Transformer voltage tapping ..................................................................................................................29
4.4 Powering up.....................................................................................................................................................29
4.5 Anode run voltage check (Pr#57 Ano) or adjustment to 50 Hz mains network ..............29
4.6 Preheat and mA-regulation check........................................................................................................31
4.7 kV- regulation check....................................................................................................................................33
1 Introduction
Term “Alpha ST” in this manual refers to Alpha ST MGF-101 unit.
NOTE!
Only engineers trained by GE Healthcare are allowed to install the unit.
parts installed by nontrained personnel carry no warranty!
CAUTION!
Always check for proper seating on all the connectors located in the areas
you have worked. equipment damage can result if power is applied or
exposures made with improperly seated connectors!
WARNING!
Always install the proper fuses into the unit. failure to do so jeopardizes
patient, operator and equipment safety.
WARNING!
This unit has floating grounds. always make all scope measurements
differentially and always be careful when servicing the unit!
CAUTION!
When handling circuit boards, always utilize ground wrist straps or
grounding kits to prevent damage to the “mos” type chips. always keep
circuit boards in a static proof bag.
WARNING!
Always disconnect the unit from line supply by removing the plug from
wall outlet before opening the covers.
WARNING!
The X-ray tube utilizes a beryllium window. do not perform any operations
on beryllium which produce dust or fumes, such as grinding, grit blasting
or acid cleaning. beryllium dust or fumes are highly toxic and breathing
them can result in serious personal injury or death.
WARNING!
This X-ray unit may be dangerous to patient and operator unless safe
exposure factors and instructions are observed!
WARNING!
Electric safety: Use fault current protection connector box between unit
power plug and mains. High voltage tank and high voltage connectors are
dangerous areas in the that require special attention as well as 220 VAC
present on Inverter Board, AMD Board and fan connectors.
NOTE!
This unit has been tested at 60 Hz. If the unit will be installed to 50 Hz
mains network, refer to Chapter 5.2.10, Light field intensity adjustment for
adjustment of collimator light field intensity and Chapter 4.5, Anode run
voltage check (Pr#57 Ano) or adjustment to 50 Hz mains network for
adjustment of anode run voltage.
2 Installation
2.1 PREINSTALLATION INSTRUCTIONS
NOTE!
Two people are required through the mechanical installation, one person
thereafter.
If the device is mounted on a wall, the pulling force exerted on each bolt is
approximately 600 N (135 lbs). The load on the floor is approximately 7350
kg/sq.m (1500 lbs./sq.ft).
NOTE!
The crate numbers in the packing list may differ from the numbers listed
here
Crate 1:
– 2290 x 520 x 410 mm (90 3/16” x 20 1/2” x 16 3/16”)
– cardboard
– gross weight approximately 89 kg (196 lb)
– Alpha stand and column assembly
Crate 2:
– 1190 x 550 x 710 mm (46 7/8” x 21 5/8” x 28”)
– cardboard
– gross weight approximately 52 kg (115 lb)
– C-arm assembly
Crate 3:
– 1150 x 230 x 200 mm (45 5/16” x 9” x 7 7/8”)
– wooden
– gross weight: without ECS approximately 92 kg (203 lb), with ECS
approximately 104 kg (230 lb)
– counterweight (with or without additional counterweight for ECS)
Crate 5 (this will be crate 4 if ECS is not delivered with the unit):
– 460 x 320 x 370 mm (18 1/8” x 12 5/8” x 14 9/16”)
– cardboard
– gross weight approximately 8 kg (17.64 lb)
– accessories
Crate 6 (this will be crate 5 if ECS is not delivered with the unit):
– 870 x 710 x 170 mm (34 5/16” x 28” x 6 11/16”)
– wooden
– gross weight approximately 68 kg (150 lb)
– base plate
NOTE!
The crate includes the packing list. Make sure that you have received all
items.
NOTE!
The bolts holding the bracket are not meant to be used as attachment
bolts (they are too long). Attachment bolts are supplied separately in a
plastic bag.
NOTE!
The counterweight is heavy (app. weight 80 kg or 175 lbs.). Get assistance
to handle it, and be extreme careful.
WARNING!
If the carriage or counterweight moves resulting cable movement over
the pulley of the vertical motor, there is a risk to damage the pulley or
drive motor assembly (Fig. 3.1, Vertical motor and cables). Specifically the
fiber gear inside the motor assembly can be damaged. Please follow the
instructions for lifting the column.
As the column is heavy, it will require at least two people to install it.
NOTE!
Insert the assembly pin through the holes at both sides of the column, and
tape the exposed ends to prevent its sliding out.
7 Remove the block and slide the cables over the shaft making sure the
cables are not twisted or crossed.
8 If ECS or Alpha Stereo 4 is installed at same time, pay attention to
additional counterweight and extension block. See D and E in Figure
3.1, Vertical motor and cables.
9 Replace the bracket block and tighten the bolts. Make sure that the
bolts are tight.
10 Slide the counterweight until it contacts the assembly pin. The pin is
supporting the counterweight while the column is being erected.
11 Install the mechanical stop.
12 Connect the vertical motor cable.
13 Install the top column cap.
14 Slide the carriage down the column until the cables are taut.
15 Verify that the assembly pin installed above is securely fastened to
the column and cannot fall out.
Two 10 mm (3/8") attachment bolts and wall plugs are used. Each bolt
should be able to take a load of 600 N (135 lbs.).
1 Lift up the column against the wall and drill the holes like in Figure
3.2, Wall attachment holes (Long / short column version).
2 Fix the enclosed plugs in the holes.
16 5/32”
410 mm
subtract 14 mm (9/16 ”)
If no base plate, then
86 3/64” / 77 9/16”
2184 / 1969 mm
Attach the wall mount bracket and base plate to the column. See Figures
3.3, Wall mount bracket and 3.4, Base plate.
NOTE!
Make sure that the counterweight is resting against the assembly pin and
the carriage is down so that the cables are taut.
CAUTION!
The column assembly is heavy. Get assistance and be careful when
erecting the column.
NOTE!
Do not remove the rubber band, holding the cable coming out through the
axle because it helps in installing the C-arm by keeping the cables out of
the way.
CAUTION!
Damage to the surface of the base plate can occur if the column is
attempted to "walk" or slide.
NOTE!
Ensure that the counterweight is resting against the assembly pin and the
carriage is down so that the cables are taut.
6 Move the column over to the base plate and stand it up on top of
rubber bumper and the traced outline.
7 Swing one side plate A into position so that it fits into the groove of
the column (see Figure 3.5, The base plate assembly).
8 Slide the pressure plate B into the space between the back of the
column and the flange on the side plate.
NOTE!
Plate B is not symmetrical and should be installed so that its width
matches the column width. Also notice the slot for the cable.
9 Tighten all bolts and check the column with a water-level. Shift the
column if necessary and tighten slightly the four bolts for the
pressure plate.
NOTE!
DO NOT TIGHTEN THE PRESSURE PLATE TOO MUCH SINCE THE
COUNTERWEIGHT MIGHT JAM IN THE LOWER POSITION.
NOTE!
The back cover plate C will not fit if the column is not in level.
NOTE!
Do not yet remove the additional label from the release paper.
12345678
ABCDEFGH
12345678
12345678
ABCDEFGH
12345678
NOTE!
If an Alpha Stereo 4 is installed, use the installation instruction manual
code (32064-IMG).
NOTE!
If an ECS system is installed, use the installation instruction manual code
(32921-IMG).
8 Attach the upper cover after all adjustments have been completed.
NOTE!
It is strongly recommended that you check all the connectors again to
prevent power-on problems.
NOTE!
If the Alpha ID II is used, please read the separate manual.
The base of the control panel is attached on the right side of the unit. If
desired, the base can be moved to the other side of the unit or attached to
a wall. See Figure 3.8, Control panel attachment.
1 Install the control panel by pushing it against the base so that the
plastic knob mounts into the hole in the base. Use setscrews to
fasten the panel to the base.
2 Fasten the exposure button to its place with a setscrew, if required.
Attach the cable of the panel to the D-connector at the base of the
carriage.
The control panel may be installed flat to the side of the carriage, or to an
angle with the angled back piece.
NOTE!
B = vertical drive motor cable on Alpha IQ, RT, ST and III manufactured
starting from 1993.
Identify how the Alpha unit is mounted - on stationary base plate or wall
mounted. If the unit is wall mounted, the wall mount base plate is
provided with holes for the hinge support pieces (Fig. 3.10, Wall mount
base plate and hinge support plate).
Instructions:
Open the crate and compare the contents to the packing list.
If the stationary base plate is used, locate the hole provided for the hinge
in base plate. In older units the hole is under the carpet. Pierce the carpet
at the hole, and drop the lower plastic sleeve into the hole. In new units
and if wall mount is used, drop the sleeve in a hole mount.
Fig 3.10. Wall mount base plate and hinge support plate
Select suitable min 16A rated mains plugs (that complies with local
regulations) with ground (PE) connection. Loose IEC type plugs is included
in the unit delivery.
When connecting the wires, make sure that the yellow-green PE wire is
left min 10mm longer than the two other ones.
1 Blue
2 Green/Yellow
3 Brown
1
2
Fig 3.12. Patient identification window in a daylight cassette allows the use of Alpha ID II
or other identification system to print the patient data onto the film.
4 Check that all cassette tunnels, Bucky and magnification tunnel are
of the right type (Normal or Daylight).
CAUTION!
Do not press the exposure button until all necessary calibrations are
performed.
NOTE!
Press the vertical drive control down and remove the assembly pin.
10 Test the vertical drive control switches on both sides of the unit and
verify that up and down movements of the carriage starts slowly and
then speeds up as the switch is held down.
11 Test both pedals for compression adjustment.
12 Check that the breast thickness display and the compression force
changes when driving the compression down.
13 Check that the field light is switched ON when the 'down' pedal is
pushed and goes OFF automatically after approximately 40 seconds.
14 Install a Bucky and check that the grid indicator lights up. Remove
the grid locking screw before use.
15 Install a magnification table and check that the magn indicator lights
up.
16 Turn the c-arm and notice that the reading in the angle display on
the c-arm changes.
NOTE!
C-arm brake can get stuck during the long storage time. Press the brake
button and release the brake by turning and vibrating the c-arm.
NOTE!
FOLLOWING VOLTAGE MEASUREMENTS, AND THE SOFT START
PROCEDURE INSTRUCTIONS ARE FOR TROUBLE SHOOTING PURPOSE.
17 Verify that the typical voltages listed are present. Refer to repair
instructions in service manual for the location of the test points.
Voltages marked with * are unregulated.
NOTE!
The purpose of this procedure is to ensure that before applying full power
to the generator, all vital signals and adjustments are correct. Use large
focus ("MAGN" lamp is off).
CAUTION!
Turn the power OFF from the main power switch.
1 Remove the side and back covers to expose the circuit boards.
2 Remove the top and bottom cover of the C-arm to expose the high
voltage unit.
Fig 4.2. Removing the top cover by opening the four screws
Fig 4.3. Removing the bottom cover by opening the four screws
Fig 4.4. Removing the back cover by opening the four screws
NOTE!
The following three items disconnects all power from the generator, but
the system can be still operated. This can be done for example when
installing for an exhibition, where X-rays are not allowed.
4.4 POWERING UP
Turn the unit ON from the mains switch.
NOTE!
The "TUBEFAIL" led H3 on the FILAMENT CONTROL Board is usually lit
before the first exposure after the unit is switched on. The activation of the
EXT signal turns the led off, and after a successful exposure the led
remains off.
1 Check that the display shows reset (r.ES in the timer display), on the
POWER SUPPLY board all LEDs are lit (3 pcs), on the FILAMENT
CONTROL board LEDs H8, H9 and H10 are lit, on the INVERTER Board
LEDs H8, H9 and H10 are lit.
2 Go into manual mode by pressing autoexp.
3 Select service mode by pressing kV- and density- buttons for 3 sec.
4 Select service mode menu by pressing Density + and Density -
buttons.
NOTE!
All testing should be performed in service mode.
1 Press kV- or kV+ buttons to scroll until the kV display shows #57 and
the mAs display shows Ano. This is the anode run voltage measuring
mode without x-rays. Press Autoexp to enter into Ano program. If
you need to adjust the anode run to 50 Hz mains network, proceed to
step 3.
2 Press exposure button to run the anode. Timer display will show
running time that is limited to 5 seconds. Check that no error
messages are displayed during that time. After this check that the
anode motor plate stops running (listen to the sound of the rotating
anode) and that no error messages are displayed on Control Panel
timer-screen. If this is not the case, perform step 3.
WARNING!
AMD BOARD HAS FLOATING GROUND OF 230 VOLTS
NOTE!
Pay attention to potentiometer setting. It is different for 50 Hz and 60 Hz.
6 During the exposure, listen that the first one of the power relays (K1)
on the INVERTER board is activated as the exposure button is
pressed, and that (K2) is activated after the preheat time, after app.
1.5 s.
7 Check also that on the INVERTER Board, the LEDs H2, H4, H5, H6, H7,
H11 and H12 and on the FILAMENT CONTROL board, the LEDs H2,
H4, H5, H6 and H7, turn on during the exposure.
3 Set the oscilloscope to 1V/div and 20ms/div. Set the trigger to app.
+1V / positive edge (storage mode).
4 Turn ON the unit. Set the unit to 27kV and 16-40 mAs (large focus).
WARNING!
X-rays !!!
5 Make an exposure and check the waveform. (see Fig. 4.7, Correct
preheat).
NOTE!
The trimmer R28 is for LARGE FOCUS. For SMALL FOCUS, insert the
magnification cassette holder to the unit and adjust the preheat from
trimmer potentiometer R96. Be careful with preheat adjustment. Turn the
potentiometer little by little and check that scope figure will look as in
Figures 4.7, Correct preheat, 4.8, Preheat rises too fast "Overshoot" and
4.9, Preheat rises too slow.
CAUTION!
Too slow a rise time will result in an E.01 error message.
NOTE!
kV is set at the factory. For calibration use Unfors, RTI, PMX-I/M or similar.
In this section the contents of the nonvolatile memory are checked and
changed if necessary. The setup information is available in two programs,
PAr and SUP. The automatic exposure control information is available in
AEC and tAr programs.
To enter the service mode press kV- and density- buttons simultaneously
until the Timer display shows Ser.
To enter the special program menu press density+ and density- buttons
simultaneously. The program number is displayed in the kV display and
the program name in the mAs display (Press DEN- first because otherwise
the system may toggle the focal spot with DEN+ key). The available
programs can be scrolled with the kV+ and kV- buttons. To enter the
desired program press the autoexp program button. To leave the
program press Autoexp buttons once more. To leave the menu press
density+ and density- simultaneously. Return to the menu from the
special programs by pressing AUTOEXP button.
NOTE!
See block diagram of software in appendix or service manual.
The effect of the 'no parity bit' and the stop bit is a total of two stop bits, so
receiving end must be configured correspondingly.
#52SUP/A The density display shows A, which means the Auto Release
Time after each exposure in seconds.
1 Select a suitable time (normally 6 sec.) using the density-
and density+ buttons. Make sure that after an exposure
the Autorelease drives the paddle up to the upper limit.
2 Press kV+
#52SUP/L The density display shows L which means the Printer Label
Type for technique printout. Select using the density- and
density+ buttons. The types are:
0 = OFF
1 = all data in one row
2 = 36.7 x 101 mm label (U.S.A. default)
3 = 26.0 x 101 mm label
Press kV+
#52SUP/H The density display shows H which indicates the Input Mode
of the text printed on the label. For each of the four rows
printed, the last 14 of 40 characters can be programmed.
See following table for how to program the label.
Sample of a label:
1234567890123456789012345678901234567890
B M 0.02s 28kV 002mAs +0 HOSPITALxNAMExx
6.5cm 200N 180^/R 0.06mGy CODExxxxxxxxxx
___/___/___ R1 12345 RADIOLOGISTxxx
_________________________ DEPARTMENTxxxxx
The first letter in row 2 is C and the character position is 15 and continues
through position 29. Row 3 contains positions 30 through 44 and row 4
positions 45 through 59. See following the ASCII table and a Label
Programming Worksheet.
Position 00 01 02 03 04 05 06 07 08 09 10 11 12 13
Text
ASCII
Position 14 15 16 17 18 19 20 21 22 23 24 25 26 27
Text
ASCII
Position 28 29 30 31 32 33 34 35 36 37 38 39 40 41
Text
ASCII
Position 42 43 44 45 46 47 48 49 50 51 52 53 54 55
Text
ASCII
Press kV+
#52SUP/i The density display shows i. In this mode the RS 232 link can
be programmed.
0 = no connections
1 = printer or Alpha ID II connected
2 = PC connected (ONLY WITH Alpha ID [Dataview])
3 = Delta 16 mode
4 = Delta 16 / Alpha ID (Dataview), requires a RS-
switcher board
Press kV+
Press kV+
Press kV+
Press kV+
Press kV+
#53PAr/b The density display shows b which means the Anode Brake
Mode. Select 0 or 1 using the Den+ and Den- buttons.
Press kV+
#53PAr/t The density display shows t. In this mode the software for
Alpha III, ALPHA ST and Alpha IQ, RT is selected.
Press kV+
Press kV+
Press kV+
CAUTION!
Perform following programming after the AEC has been adjusted and
then the density & default values have been selected and programmed.
#64AEC/c The density display shows c. This is for the AEC thickness
compensation including five (5) subparameters: P, r1, r2, S1,
S2 (in kV display). For further information on thickness
compensation refer to section AEC adjustment.
Press kV+
#64AEC/g The density display shows g. This is the software AEC gain
parameter. It is used for adjusting the basic optical density
(usually for Bucky molybdenum). refer to section AEC
adjustment.
CAUTION!
The gain affects all cassette holder/filter combinations.
Press kV+
To change kV table:
Use the initial kV table (see next page) to design new values.
Press kV+
NOTE!
If the factory default setting is changed, use the table below
to store the new values.
CAUTION!
Always drive the compression paddle from 00cm up when
reprogramming the initial kV.
Initial kV table
0.0 23 23 25
0.5 23 23 25
1.0 23 23 25
1.5 23 23 25
2.0 23 23 25
2.5 23 23 25
3.0 23 23 25
3.5 24 24 25
4.0 24 24 26
4.5 24 24 26
5.0 25 25 26
5.5 25 25 27
6.0 26 26 27
6.5 26 26 28
7.0 27 27 28
7.5 27 27 28
8.0 28 28 29
8.5 28 28 29
9.0 28 28 30
9.5 28 28 30
1 enter program 64
2 select parameter S
3 select required step value
4 exit the service program
In this mode the target times for different thicknesses can be seen by
driving the compression paddle up and down. Insert the bucky for large
focus programming and magnification stand for small focus
programming. Also, both filters must be programmed.
Target time/sec.
(Exposure time)
4.0
3.0
2.5
T2 (2.4 sec. at 9.5 cm)
2.0
1.5
1.0
(Pr#64 AEC, parameter I) Target time LO limit
0.5
T1 (0.5 sec. at 1.5 cm)
5 10 Breast thickness/cm
NOTE!
The Target time is separately adjusted for Bucky/Magn/Cassette tunnel.
CAUTION!
The maximum and minimum target time values are limited in the #64 AEC
program (parameter I)
8 To store the new values press Den+ and Den- simultaneously until
service mode is entered.
9 Press kV- and Den- until normal mode is entered.
10 Switch OFF the power for a few seconds. Turn the power ON and
ensure that the programming has been stored.
The exposure and error counters can be checked in the special program
#58 CEC. The timer display shows the exposure counters and the error
codes: tot, cnt, E01, E02, E03, E04, E07, E08, E09, E10, E11, E12, Slo, Er1,
Er2, Er3, Er4, Er5, Eaf, SYS.
The kV and mAs displays show the number of those errors. Pressing the
mAs- resets the counter in the display.
Pressing the kV+ button selects the next counter. Reset all the counters.
When the timer display shows tot, the exposure counter is shown. It
cannot be reset. The cnt counter is a resettable counter and should be
reset.
NOTE!
REPORTS TO BE RECORDED IN U.S.A. ONLY: Record all requested data
according to the installation report enclosed. Complete the FDA form
2579(3/89) according to the DHR report guide sheet 31493-IMG enclosed.
1 Turn ON the film processor and let it warm up for at least half an
hour.
2 Run a clean-up film through the processor. This is just a piece of
regular a film, which will pick up dirt, etc., which would appear on
your film.
3 Obtain a sheet of the same film type to be used in your ALPHA. Select
the color on the sensitometer that matches the color emitted by the
intensifying screen used.
4 Ensure that the emulsion side of the film is towards the light source
on the sensitometer. Expose two edges of the film, waiting 10
seconds between exposures, and run it through the processor.
5 Measure and record the optical density of step 1 for both edges of
the film. Record the average on both the film and in Table 2.5. Step 1
is the value of GROSS FOG (BASE FOG).
NOTE!
Always run the film through the processor in the same orientation and
location.
6 Find the step, which has the density closest to 1.0. This step number
is the SPEED INDEX. Record the average density and step number on
the film and in Table 2.5.
7 Go one step above and one step below the step of Speed Index.
Measure and record both densities.
8 Subtract the density value of the step below, from the density value
of the step above the speed index step. This result is the CONTRAST
INDEX. Record this on the film and in evaluation tables on the next
page.
Date: Time:
Film type:
Water temp: Dev temp:
Gross fog: Speed index:
Contrast
index:
Date: Time:
Film type:
Water temp: Dev temp:
Gross fog: Speed index:
Contrast
index:
Date: Time:
Film type:
Water temp: Dev temp:
Gross fog: Speed index:
Contrast
index:
Table 5.5 Film processor evaluation
Check that the brake has 170 N (38 lbs) of holding force measured at the
cassette holder; if not, open the side covers of the carriage and tighten the
nyloc nuts (3).
If the brake does not release properly, open the nuts (1) and adjust the
space (4) with the screws (2). The space should be about 0.1 to 0.4 mm.
CAUTION!
Make only minor adjustments of the nyloc nuts - measured by the flats of
the nuts.
The x-ray field adjustment must be checked and aligned prior to the light
field alignment.
5 Make an exposure and develop the film. The coins outlined on the
film should not be more than 5 mm out of the exposed field and sum
of misalignments on opposite sides should not be more than 10 mm.
NOTE!
See Figure 4.6, Filament Control Board to locate the adjustment trimmers.
The anode acceleration and braking are done at full voltage, and it is not
possible to adjust them.
The anode run time voltage can be adjusted with the special program #57
Ano.
Enter the special program menu and select program #57 with the kV+
button. Press autoexp to enter the program.
The anode voltage is measured from the testpoints TP6 and TP7 located
at the edge of the AMD board (X4 pin 2 and X5 pin 2). The voltage should
be set to a value between 20...40 VAC (app. 40 VAC with TrueRMS).
NOTE!
NO radiation when pressing the exposure button inside this special
program.
The compression thickness is calibrated with the special program #56 Cth.
After entering this program the mAs display shows "to 00" with Bucky or
"to 20" with cassette tunnel.
NOTE!
The compression force is pre-adjusted at the factory and does not have to
be measured during the installation.
Strap around
the tube head
Dynamometer
User (major) adjustment is done by turning the external switch (the cable
coming to connector X5) in the front of the system. Minor adjustment can
be done by adjusting trimpot R33 (clockwise to increase).
CAUTION!
When changing the maximum force, keep external knob on maximum
position and when changing the minimum force, keep the external knob
on minimum position.
Adjust speed threshold trimpot (R43) totally clockwise (only high speed is
on). Press PADDLE pedal and adjust suitable (about 10 cm / 5 sec) high
speed from trimpot R12 (Clockwise to increase).
Use a bathroom scale to state the compression force. Use soft material on
top of the scale. Press the PADDLE pedal and adjust from trimpot R43
until the compression force is about 10 kg (22 lbs) when the speed
changes from high speed to low speed.
OR
Compress your own hand and adjust the limit for the speed change so
that you feel a strong touch and then speed changes.
OR
Turn the c-arm around, drive the compression paddle and adjust R43 so
that the paddle is just about to change the speed.
Comments
Press the REL pedal and adjust the speed from R37 to about 5 cm / 5sec.
Press the ECS pedal and adjust the C-arm speed from R59 until the
dimension between floor level and compression paddle does not change
more than ± 2 cm.
Remove the plastic plug in the back cover of the C-arm to reach the
sealed and labeled setscrew and make adjustment. Remember to seal the
screw after adjustment.
Mechanical clutch
adjustment
Sliding bar
clearance
adjustment
NOTE!
The unit has been adjusted at the factory with 60 Hz main voltage
frequency. If the unit is being installed to 50 Hz mains network, adjust
collimator lamp voltage at 11 V. Adjust and check the light intensity as
described below.
The halogen lamp lifetime increases when lamp voltage is decreased and
vice versa.
When measuring the SF, no cassette holder should be used. Set the
parameters as shown in the table below and make the exposures.
3 Record the kV, mAs and time reading after each exposure. The time
reading is to be compared with the Alpha control panel time display.
The maximum tolerance for kV, mAs and exposure time are:
For kV
For mAs
kV adjustment
kV can be adjusted from R 63 located on the Inverter Board. See
Fig. 4.10, Inverter Board.
6 Return to the normal mode by briefly pressing the kV- and density-
buttons simultaneously. The timer display shows "nor" when the
buttons are released.
The grid is aligned and the speed is adjusted at the factory and should
normally not be adjusted at the site. If adjustment and / or alignment is
necessary, however, follow the instructions below.
Remove the carbon fiber cover on the bucky. In the small bucky the grid
should be at the bottom of the grid drive block groove and fastened with
four setscrews. In the large bucky the grid should be fastened only with
two setscrews and aligned by the two rear setscrews.
Check that the grid doesn't touch the side lips of the bucky and that it is
centered. Check also that the grid is aligned parallel to the side lips.
Speed adjustment
1 Attach the bucky and enter the program by pressing autoexp. Press
the exposure button to run the grid.
The timer display shows the grid turn-around time, which cannot be
adjusted. The mAs display shows the adjustable grid travel time from side
to side.
2 Check that the fast turn-around times are equal at both ends. If
adjustment is necessary, move the slotted metal sheet traveling
through the optocoupler.
3 The grid travelling time is adjusted from R5 located on the
bucky-control board. In the CC position travel time of the grid should
be 1.6 +/- 0.1 sec. The fast travel time should be 0.17 +/- 0.1 sec.
4 Make sure that the grid moves well in lateral positions (C-arm in +/-
90°angle) as well.
1 ATTACH THE BUCKY DEVICE. Press Den+ & Den- for 3 sec. to enter
program (Pr) mode.
CAUTION!
STORE ALL DEFAULT VALUES IN THE TABLES ON THE NEXT PAGES
(MOLYBDENUM FILTER).
CAUTION!
The AEC is adjusted for and one cassette holder, with the bucky. The right
optical densities for the rest of the cassette holders are done later as
described on Density programming.
Fig 5.16. Placement of the acrylics (and phantoms!) onto the cassette holder
3 Press Autorel after the exposure and record the three values
displayed in the mAs window one after one. Each value is displayed
for 1 second. The first value represents the integration time (in ms),
the second is AEC frequency (for example 320 = 32 kHz) and the last
value is base frequency (for example 004 = 40 Hz).
4 Make exposures (do not develop the film) and adjust the trimmer R17
(CCW to increase frequency) on the AEC board so that the AEC
frequency will be appr. 32kHz (that is the second value in the mAs
window reading about 320).
Also the exposure time is shown in the timer display (for example 315 =
315 ms, 1.54 = 1.54 sec.
3 Draw a line between the two points and read the corresponding
GAIN for the wanted optical density. Notice that the object is to find
the right gain, so the initial gain values do not have to be 1.00 and
0.90. If the correct gain is for example 1.08, more proper gain values
would be 1.00 and 1.10.
4 Alter the AEC Gain parameter 'g' in the service program #64 to obtain
the desired O.D. Diminishing 'g' will increase the O.D and vice versa.
Exposure time
CAUTION!
The AEC Gain 'g' parameter will affect all exposures regardless of used
filter or cassette holder.
Rotation point P
Rotation point is the exposure time where the O.D is adjusted correctly.
Thickness compensation is achieved by altering the exposure time around
this point. Default value is 1.00.
The values of r1 and r2 can vary from 0.70 to 1.30. If the parameter is
1.30 it means that exposure times are increased by 30%. If the parameter
is 0.80 it means that the exposure times are decreased by 20% etc.
Default values are 1.0 for both.
1 Notice the exposure time, which has the correct O.D. from the
previous O.D vs. exposure time graph.
2 Notice the O.D. for the shorter exposure time (for example the 20
mm Acrylic exposure time) from the above graph. Divide the desired
O.D. with this value.
CAUTION!
Perform the thickness compensation with all cassette tunnels (Bucky/
cassette holder/magnification tunnel) separately. DO NOT change the
gain parameter anymore. Use the density correction (in Pr mode) to bring
O.D. to same level with other filter/cassette holder combinations.
NOTE!
Fine tuning is optional and may only have minor effect to compensation.
Develop the films and record the optical density of the films (=O.D.).
'S1' and 'S2' can have values 0, 50, 100, 150...up to 700. The maximum
compensation is 140% or 60% of O.D. in 'P' (rotation Point).
10 Enter the #64 AEC program. Press kV+ until the thickness
compensation parameter 'c' is reached.
11 Press autorel to scroll the 'S1' parameter in the kV display. Use Den+/
- buttons to change the value. This parameter affects the
compensation slope at exposures shorter than rotation point 'P'.
12 Press autorel to scroll the 'S2' parameter in the kV display. Use Den+/
- buttons to change the value. This parameter affects the
compensation slope at exposures longer than rotation point 'P'.
5.2.18 kV compensation
If the film/screen combination and default kV is the same for all channels
a, b and c, the density correction is the same also.
NOTE!
To enter the Program (Pr) mode, press Density+ and Density- buttons
simultaneously.
Mark the new density value in the table 5.8, Alpha ST settings for
Molybdenum filter.
World Headquarters
GE Healthcare
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
European Headquarters
GE Healthcare - Europe:
Paris, France - Fax +33-1-30-70-94-35
Asia Pacific
GE Healthcare -Asia:
Tokyo, Japan - Fax +81-425-85-5490
Hong Kong - Fax: +852-2559-3588
www.gehealthcare.com
Alpha ST